Tianao Mao1,2, Haoyang Wang3, Zheng Peng3, Taotao Ni2, Tianqi Jia2,4, Rongrong Lei2,4, Wenbin Liu1,2,4.
Abstract
A gas chromatography-mass spectrometry (GC/MS) method for the determination of hexabromocyclododecane (HBCD) in expanded polystyrene and extruded polystyrene foam (EPS/XPS) was developed. The EPS/XPS samples were ultrasonically extracted with acetone and the extracts were purified by filtration through a microporous membrane (0.22 μm) and solid-phase extraction. The samples were analyzed using a GC/MS using the selected ion monitoring mode. The ions 157, 319 and 401 were selected as the qualitative ions, while ion 239 was chosen as the quantitative ion. An HBCD standard working solution with a concentration range of 1.0-50.0 mg/L showed good linearity. The detection limit of HBCD was 0.5 mg/kg, meeting the LPC limit (<100 or 1000 mg/kg). Six laboratories were selected to verify the accuracy of the method, and 10 samples were tested. The interlaboratory relative standard deviation range was 3.68-9.80%. This method could play an important role in controlling HBCD contamination in EPS/XPS.Entities:
Keywords: EPS; HBCD; POPs; XPS; gas chromatography-mass spectrometry
Year: 2021 PMID: 34885731 PMCID: PMC8659249 DOI: 10.3390/molecules26237143
Source DB: PubMed Journal: Molecules ISSN: 1420-3049 Impact factor: 4.411
Selected monitoring ions (m/z) and allowable relative deviations.
| Monitoring of the Ion (m/z) | Ion Species | Ion Ratio (%) | Permissible Relative Deviation (%) |
|---|---|---|---|
| 157 | Qualitative ion | 100 | - |
| 239 | Quantitative ion | 89 | ±15 |
| 319 | Qualitative ion | 55 | ±20 |
| 401 | Qualitative ion | 19 | ±20 |
Figure 1Chromatograms of HBCD at different injection temperatures, (a) 5 mg/L-270 °C, (b) 5 mg/L-230 °C, (c) 5 mg/L-190 °C, (d) 10 mg/L-270 °C, (e) 10 mg/L-230 °C, and (f) 10 mg/L-190 °C.
Figure 2The calibration curve of HBCD.
Test data for the method precision and accuracy (n = 6).
| HBCD Add the Amount | 5 mg/kg | 10 mg/kg | 20 mg/kg |
|---|---|---|---|
| Measured mean | 4.9 | 9.9 | 21.3 |
| The standard deviation Si | 0.34 | 0.69 | 1.4 |
| Relative standard deviation RSD (%) | 6.90% | 6.92% | 6.64% |
| Recovery range | 87~106% | 89~108% | 96~113% |
Comparison of experimental results from different studies.
| Test Material | Testing Equipment | Concentrations | RSDs | Limit of Quantitation | The Literature |
|---|---|---|---|---|---|
| EPS/XPS | GC/MS | 4.5~33.5 | 6.64~6.92 | 0.5 | In this study |
| EPS/XPS | XRF | 5~12 | 7~16 | 50 | [ |
| EPS | LC-MS/MS | 6.849~7.105 | 0.2~0.6 | 0.005 | [ |
| EPS/XPS | XRF, LC-MS/MS, NMR | 6.7~11.1 | 44~104 | 300 | [ |
Figure 3Results from the six laboratories for EPS (A) and XPS (B).