Literature DB >> 34874948

Dried blood spot specimens for SARS-CoV-2 antibody testing: A multi-site, multi-assay comparison.

François Cholette1,2, Christine Mesa1, Angela Harris1, Hannah Ellis1, Karla Cachero1, Philip Lacap1, Yannick Galipeau3, Marc-André Langlois3,4, Anne-Claude Gingras5,6, Cedric P Yansouni7, Jesse Papenburg7,8, Matthew P Cheng7,8, Pranesh Chakraborty9, Derek R Stein2,10, Paul Van Caeseele2,10,11,12, Sofia Bartlett13,14, Mel Krajden13,14, David Goldfarb14, Allison McGeer12,15, Carla Osiowy1, Catherine Hankins12,16, Bruce Mazer17, Michael Drebot1, John Kim1.   

Abstract

The true severity of infection due to COVID-19 is under-represented because it is based on only those who are tested. Although nucleic acid amplifications tests (NAAT) are the gold standard for COVID-19 diagnostic testing, serological assays provide better population-level SARS-CoV-2 prevalence estimates. Implementing large sero-surveys present several logistical challenges within Canada due its unique geography including rural and remote communities. Dried blood spot (DBS) sampling is a practical solution but comparative performance data on SARS-CoV-2 serological tests using DBS is currently lacking. Here we present test performance data from a well-characterized SARS-CoV-2 DBS panel sent to laboratories across Canada representing 10 commercial and 2 in-house developed tests for SARS-CoV-2 antibodies. Three commercial assays identified all positive and negative DBS correctly corresponding to a sensitivity, specificity, positive predictive value, and negative predictive value of 100% (95% CI = 72.2, 100). Two in-house assays also performed equally well. In contrast, several commercial assays could not achieve a sensitivity greater than 40% or a negative predictive value greater than 60%. Our findings represent the foundation for future validation studies on DBS specimens that will play a central role in strengthening Canada's public health policy in response to COVID-19.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 34874948      PMCID: PMC8651133          DOI: 10.1371/journal.pone.0261003

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

According to the latest estimates from John Hopkins University (last accessed on 9 September 2021), severe acute respiratory syndrome coronavirus (SARS-CoV2), the etiological agent of coronavirus disease 2019 (COVID-19) [1], is responsible for over 220 million confirmed cases, including over 4.5 million deaths globally [2]. While nucleic acid amplification tests (NAAT) on respiratory samples remain the gold standard for COVID-19 diagnostic testing [3], they cannot provide reliable prevalence and incidence estimates at the population level given the narrow window for reliable results (<14 days post-symptom onset) [4], global supply shortages [5], and limited access due to symptom-based testing prioritization [6]. On the other hand, serological assays may provide more reliable population-level SARS-CoV-2 prevalence estimates to better inform ongoing public health responses [7]. Serological testing is also vitally important for monitoring both individual and population-level humoral immune responses to COVID-19 vaccination [8]. Several commercial serological assays rapidly became available through measures such as the US Food and Drug Administration’s (FDA) Emergency Use Authorization programme [9]. These assays have been primarily designed for the detection of SARS-CoV-2 antibodies in serum, plasma, or whole blood [10-13]. Implementing large-scale integrated biological-behavioural surveys poses a significant challenge since phlebotomy requires highly trained personnel when most health care professionals have been re-deployed to assist in the COVID-19 response [14]. Furthermore, reaching rural and remote communities adds to the complexity of providing reliable and timely testing for several reasons including lack of trained personnel to collect biological specimens, limited access to laboratory facilities, difficulties in maintaining the cold chain, and unreliable specimen transportation even within the Canadian context. The implementation of SARS-CoV-2 point-of-care (POC) testing [15-18] could alleviate some of these challenges but may not be appropriate in all settings. Rural and remote communities tend to be small and have close-knit social networks thereby making confidential or anonymous SARS-CoV-2 POC testing difficult. Thus, a practical solution is required to be able to achieve large scale sampling and circumvent these problematic issues. Dried blood spot (DBS) collection may be a practical solution owing to this method’s simplicity. DBS are prepared by placing a few drops of blood on a card made of filter paper. A finger prick is performed using standard spring-loaded lancets that do not require specialised training to utilize. Home-self collection is also a viable option that is already playing a significant role in SARS-CoV-2 sero-surveillance [19,20]. Once the DBS cards are dry, they can be stored and transported at ambient temperature to centralized laboratories through regular mail services without any adverse effect on downstream testing [21]. However, comparative performance data on SARS-CoV-2 serological assays using DBS specimens are limited. Here we present test performance data from a SARS-CoV-2 DBS panel sent to various public health and academic laboratories across Canada representing 10 commercial and 2 in-house developed assays for SARS-CoV-2 antibodies. As part of the COVID-19 Immunity Task Force (CITF, www.covid19immunitytaskforce.ca/), our goal was to provide preliminary performance data on each test to guide future large-scale validations.

Results

Assay performance

Sensitivity, specificity, positive predictive values (PPV), and negative predictive values (NPV) were assessed for 10 commercial and 2 in-house assays on 10 known negative plasma samples and 10 plasma samples from COVID-19 patients (Table 1) contrived as DBS. Nearly one third of all commercial assays (EUROIMMUN, Elecsys spike, and GSP/DELFIA) identified all positive and negative DBS correctly corresponding to a sensitivity, specificity, PPV, and NPV of 100% (95% CI = 72.2, 100) even when using a wide range of prevalence estimates (Figs 1 and S1). The Elecsys spike and GSP/DELFIA showed almost perfect agreement with EUROIMMUN as well as with one another (Tables 1 and S1, Fig 2).
Table 1

Assay performance for SARS-CoV-2 antibody testing on dried blood spot specimens (DBS).

AssayDBS punches (n =)TP (n =)FN (n =)TN(n =)FP (n =)Eq (n =)Se (95% CI [LL, UL])Sp (95% CI [LL, UL])PPV (95% CI [LL, UL])NPV (95% CI [LL, UL])Kappaa (95% CI [LL, UL])
EUROIMMUN 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)Reference
Platelia 2100901100% (72.2, 100)90% (59.6, 99.5)90.9% (62.3, 99.5)100% (70.1, 100)0.95 (0.86, 1.00)
LIASON 219100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)52.6% (31.7, 72.7)0.45 (0.20, 0.70)
LIASON 319100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)52.6% (31.7, 72.7)0.45 (0.20, 0.70)
LIASON 419100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)52.6% (31.7, 72.7)0.45 (0.20, 0.70)
COV2G 219100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)53% (32, 73)0.05 (-0.26, 0.36)
COV2G 319100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)53% (32, 73)0.05 (-0.26, 0.36)
COV2G 41982010% (0.5, 40.4)80% (49, 96.4)33.3% (1.7, 88.2)47.1% (26.2, 69)0.15 (-0.16, 0.46)
COV2T 219100010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)52.6% (31.7, 72.7)0.55 (0.32, 0.78)
COV2T 33791030% (10.8, 60.3)90% (59.6, 99.5)75% (30.1, 98.7)56.3% (33.2, 76.9)0.60 (0.36, 0.84)
COV2T 44664040% (16.8, 68.7)60% (31.3, 83.2)50% (21.5, 78.5)50% (25.4, 74.6)0.50 (0.23, 0.77)
Elecsys S 291010090% (59.6, 99.5)100% (72.2, 100)100% (70.1, 100)90.9% (62.3, 99.5)0.45 (0.203, 0.70)
Elecsys S 391010090% (59.6, 99.5)100% (72.2, 100)100% (70.1, 100)90.9% (62.3, 99.5)0.45 (0.203, 0.70)
Elecsys S 41001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
Elecsys N 264100060% (31.3, 83.2)100% (72.2, 100)100% (72.2, 100)71.4% (45.4, 88.3)0.80 (0.62, 0.98)
Elecsys N 373100070% (39.7, 89.2)100% (72.2, 100)100% (39.7, 89.2)76.9% (49.7, 91.8)0.85 (0.69, 1.00)
Elecsys N 491100090% (59.6, 99.5)100% (72.2, 100)100% (70.1, 100)90.9% (62.3, 99.5)0.95 (0.85, 1.00)
VITROS 21000100100%(72.2–100)0% (0, 27.8)50% (29.9, 70.1)N/A0.50 (0.27, 0.73)
VITROS 31000100100% (72.2, 100)0% (0, 27.8)50% (29.9, 70.1)N/A0.50 (0.27, 0.73)
VITROS 41000100100% (72.2, 100)0% (0, 27.8)50% (29.9, 70.1)N/A0.50 (0.27, 0.73)
Architect 219010010% (0.5, 40.4)100% (72.2, 100)100% (5.1, 100)52.6%(31.7, 72.7)0.05 (-0.26, 0.36)
Architect 328010020% (3.6, 51)100% (72.2, 100)100% (17.8, 100)55.6% (33.7, 75.4)0.10 (-0.21, 0.41)
Architect 428010020% (3.6, 51)100% (72.2, 100)100% (17.8, 100)55.6% (33.7, 75.4)0.10 (-0.21, 0.41)
GSP/DELFIA 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house S (U of T) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house RBD (U of T) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house N (U of T) 246100040% (16.8, 68.7)100% (72.2, 100)100% (51.0, 100)62.5% (38.6, 81.5)0.70 (0.49, 0.91)
In-house S, polyclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house RBD, polyclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house N, polyclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house S, monoclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house RBD, monoclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)
In-house N, monoclonal (U of O) 21001000100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)100% (72.2, 100)1.00 (1.00, 1.00)

DBS punches: 6 mm (1/4 inch) dried blood spot punch; TP: True positive; FN: False negative; TN: True negative; FP: False positive; Eq: Equivocal; Se: Sensitivity; Sp: Specificity; PPV: Positive predictive value; NPV: Negative predictive value; 95% CI (LL, UL): 95% confidence intervals (lower limit, upper limit); S: Spike; RBD: Receptor binding domain; N: Nucleocapsid; U of T: University of Toronto; U of O: University of Ottawa.

aKappa statistics interpreted as follows: <0 = no agreement, 0–0.20 = slight agreement, 0.21–0.40 = fair agreement, 0.41–0.60 = moderate agreement, 0.61–0.80 = substantial agreement, and 0.81–1.00 = almost perfect agreement.

The panel consisted of 10 unique SARS-CoV-2 antibody positive DBS cards contrived from patient samples and 10 unique SARS-CoV-2 negative DBS cards spotted directly from EDTA whole blood.

Fig 1

Positive predictive values (PPV) and negative predictive values (NPV) by prevalence for each commercial and in-house assay on dried blood spot specimens.

PPV and NPV is presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). The VITROS assay could not achieve a specificity greater than 0% therefore, only PPV is shown.

Fig 2

Inter-assay concordance (A) and Cohen’s kappa coefficient (B) for SARS-CoV-2 antibody testing on dried blood spot specimens. Concordance (%) and Cohen’s kappa coefficient are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In- house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa).

Positive predictive values (PPV) and negative predictive values (NPV) by prevalence for each commercial and in-house assay on dried blood spot specimens.

PPV and NPV is presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). The VITROS assay could not achieve a specificity greater than 0% therefore, only PPV is shown. Inter-assay concordance (A) and Cohen’s kappa coefficient (B) for SARS-CoV-2 antibody testing on dried blood spot specimens. Concordance (%) and Cohen’s kappa coefficient are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In- house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). DBS punches: 6 mm (1/4 inch) dried blood spot punch; TP: True positive; FN: False negative; TN: True negative; FP: False positive; Eq: Equivocal; Se: Sensitivity; Sp: Specificity; PPV: Positive predictive value; NPV: Negative predictive value; 95% CI (LL, UL): 95% confidence intervals (lower limit, upper limit); S: Spike; RBD: Receptor binding domain; N: Nucleocapsid; U of T: University of Toronto; U of O: University of Ottawa. aKappa statistics interpreted as follows: <0 = no agreement, 0–0.20 = slight agreement, 0.21–0.40 = fair agreement, 0.41–0.60 = moderate agreement, 0.61–0.80 = substantial agreement, and 0.81–1.00 = almost perfect agreement. The panel consisted of 10 unique SARS-CoV-2 antibody positive DBS cards contrived from patient samples and 10 unique SARS-CoV-2 negative DBS cards spotted directly from EDTA whole blood. The Platelia assay achieved a sensitivity, specificity, PPV, and NPV of 100% (95% CI = 72.2, 100), 90% (95% CI = 59.6, 99.5), 90.9% (95% CI = 62.3, 99.5), and 100% (95% CI = 70.1, 100) respectively due to one equivocal result from a negative DBS sample. As a result, it appears that the Platelia assay can only achieve a PPV ≥90% in high prevalence settings (Fig 1). The Platelia assay showed almost perfect agreement with EUROIMMUN (Table 1) and other high performing commercial assays (S1 Table, Fig 2). The Elecsys nucleocapsid assay achieved a sensitivity, specificity, PPV, and NPV of 90% (95% CI = 59.6, 99.5), 100% (95% CI = 72.2, 100), 100% (95% CI = 70.1, 100), and 90.9% (95% CI = 62.3, 99.5) respectively after increasing the sample input from 2 to 4 DBS punches. Increasing the sample input also significantly reduces the number of false negatives in high prevalence settings (S1 Fig) and improves agreement with other commercial assays such as EUROIMMUN, Platelia, GSP/DELFIA, and Elecsys spike (Tables 1 and S1, Fig 2). Performance varied for the remaining commercial assays. The LIASON, Architect, and COV2G assays achieved a specificity and PPV of 100% (95% CI = 72.2, 100) and 100% (95% CI = 5.1, 100) respectively but could not achieve a sensitivity greater than 20% (95% CI = 3.6, 51) or an NPV greater than 55.6% (95% CI = 33.7, 75.4) despite increasing the sample input from 2 to 4 DBS punches. The COV2T assay achieved a sensitivity, specificity, PPV, and NPV of 40% (95% CI = 16.8, 68.7), 60% (95% CI = 31.3, 83.2), 50% (95% CI = 21.5, 78.5), and 50% (95% CI = 25.4, 74.6) respectively after increasing the sample input from 2 to 4 DBS punches. Similarly, the VITROS assay achieved a sensitivity and PPV of 100% (95% CI = 72.2, 100) and 50% (95% CI = 29.9, 70.1) respectively but could not achieve a specificity greater than 0% despite increasing sample input. This would suggest that the LIASON and Architect assays would yield a high number of false negative results in high prevalence settings (≥5.0%) while the COV2G and COV2T assays would yield a high number of false negative and false positive results in low to high prevalence settings (Figs 1 and S1). The VITROS assay is likely not to be useful under any prevalence setting since all DBS samples produced positive results (Table 1). As expected, these commercial assays had no agreement to moderate agreement with EUROIMMUN (Table 1) and the higher performing assays Platelia, Elecsys spike, Elecsys nucleocapsid, and GSP/DELFIA (S1 Table, Fig 2). The in-house assays from the University of Ottawa and University of Toronto performed well and successfully identified all positive and negative DBS correctly corresponding to a sensitivity, specificity, PPV, and NPV of 100% (95% CI = 72.2, 100) even when using a wide range of prevalence estimates. However, the in-house nucleocapsid (U of T) assay only achieved a sensitivity, specificity, PPV, and NPV of 40% (95% CI = 16.8, 68.7), 100% (95% CI = 72.2, 100), 100% (95% CI = 51, 100), and 62.5% (95% CI = 38.6, 81.5) respectively suggesting that this assay would yield a high number of false negatives in high prevalence settings (Fig 1). All the in-house assays, apart from the in-house nucleocapsid (U of T) had almost perfect agreement with EUROIMMUN (Table 1) and other high performing commercial assays (S1 Table, Fig 2).

Separation between positive and negative samples

Raw data values were plotted, and receiver operating characteristic (ROC) curves were computed (Figs 3 and 4, Table 2, S2 and S3 Figs) to evaluate each assay’s ability to separate the 10 known negative plasma samples and 10 plasma samples from COVID-19 patients contrived as DBS. Among the commercial assays, the EUROIMMUN, Platelia, Elecsys spike, Elecsys nucleocapsid, and Architect showed the clearest separation of positive and negative samples by DBS testing and their overall performance was rated as excellent (Table 2). On the other hand, the LIASON, COV2G, and COV2T could not clearly distinguish between positive and negative DBS specimens and showed an overall performance ranging from bad to sufficient (Table 2). The VITROS assay showed good separation of positive and negative DBS and excellent performance according to the area under the ROC curve (AUC) suggesting that increasing the signal to cut-off (S/Co) ratio from 1.0 to 3.4 could potentially increase this assay’s sensitivity and specificity to 90% (95% CI = 60, 99) and 100% (95% CI = 72.2, 100) respectively.
Fig 3

Receiver operating characteristic (ROC) curve for each commercial and in-house assay on dried blood spot specimens.

ROC curves are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa).

Fig 4

Distribution of values obtained for each commercial and in-house assay on dried blood spot (DBS) specimens.

Each panel shows the optical density ratio (OD Ratio), arbitrary units per mL (AU/mL), index, units per mL (U/mL), cut-off index, or signal to cut-off ratio (S/Co) for SARS-CoV-2 antibody negative DBS specimens (n = 10) represented in blue and SARS-CoV-2 antibody positive DBS specimens (n = 10) represented in orange. All values are log10 transformed to aid with visualisation. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa).

Table 2

Area under the receiver operating characteristic curve (AUC) for each commercial and in-house assay on dried blood spot specimens (DBS).

AssayDBS punches (n =)AUCa (95% CI [LL, UL])P valuePerformance
EUROIMMUN 21.0 (1.0, 1.0)<0.05Excellent
Platelia 21.0 (1.0, 1.0)<0.05Excellent
LIASON 20.60 (0.35, 0.85)0.45Bad
LIASON 30.60 (0.35, 0.85)0.45Bad
LIASON 40.65 (0.40, 0.90)0.26Sufficient
COV2G 20.59 (0.33, 0.85)0.50Bad
COV2G 30.66 (0.40, 0.91)0.24Sufficient
COV2G 40.66 (0.39, 0.92)0.24Sufficient
COV2T 20.59 (0.33, 0.85)0.50Bad
COV2T 30.55 (0.29, 0.81)0.71Bad
COV2T 40.61 (0.35, 0.86)0.43Sufficient
Elecsys S 20.95 (0.84, 1.0)<0.05Excellent
Elecsys S 31.0 (1.0, 1.0)<0.05Excellent
Elecsys S 41.0 (1.0, 1.0)<0.05Excellent
Elecsys N 21.0 (1.0, 1.0)<0.05Excellent
Elecsys N 31.0 (1.0, 1.0)<0.05Excellent
Elecsys N 41.0 (1.0, 1.0)<0.05Excellent
VITROS 20.93 (0.78, 1.0)<0.05Excellent
VITROS 30.93 (0.81, 1.0)<0.05Excellent
VITROS 40.96 (0.87, 1.0)<0.05Excellent
Architect 21.0 (1.0, 1.0)<0.05Excellent
Architect 31.0 (1.0, 1.0)<0.05Excellent
Architect 41.0 (1.0, 1.0)<0.05Excellent
GSP/DELFIA 21.0 (1.0, 1.0)<0.05Excellent
In-house S (U of T) 21.0 (1.0, 1.0)<0.05Excellent
In-house RBD (U of T) 21.0 (1.0, 1.0)<0.05Excellent
In-house N (U of T) 20.98 (0.93, 1.0)<0.05Excellent
In-house S, polyclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent
In-house RBD, polyclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent
In-house N, polyclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent
In-house S, monoclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent
In-house RBD, monoclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent
In-house N, monoclonal (U of O) 21.0 (1.0, 1.0)<0.05Excellent

DBS punches: 6 mm (1/4 inch) dried blood spot punch; 95% CI (LL, UL): 95% confidence intervals (lower limit, upper limit); S: Spike; RBD: Receptor binding domain; N: Nucleocapsid; U of T: University of Toronto; U of O: University of Ottawa.

aTest performance classified according to AUC as follows: <0.5 = not useful, 0.5–0.6 = bad, 0.6–0.7 = sufficient, 0.7–0.8 = good, 0.8–0.9 = very good, and 0.9–1.0 = excellent.

The panel consisted of 10 unique SARS-CoV-2 antibody positive DBS cards and 10 unique SARS-CoV-2 negative DBS cards contrived from patient samples.

Receiver operating characteristic (ROC) curve for each commercial and in-house assay on dried blood spot specimens.

ROC curves are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa).

Distribution of values obtained for each commercial and in-house assay on dried blood spot (DBS) specimens.

Each panel shows the optical density ratio (OD Ratio), arbitrary units per mL (AU/mL), index, units per mL (U/mL), cut-off index, or signal to cut-off ratio (S/Co) for SARS-CoV-2 antibody negative DBS specimens (n = 10) represented in blue and SARS-CoV-2 antibody positive DBS specimens (n = 10) represented in orange. All values are log10 transformed to aid with visualisation. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. Data from four 6 mm (1/4 inch) punches is presented for all other assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). DBS punches: 6 mm (1/4 inch) dried blood spot punch; 95% CI (LL, UL): 95% confidence intervals (lower limit, upper limit); S: Spike; RBD: Receptor binding domain; N: Nucleocapsid; U of T: University of Toronto; U of O: University of Ottawa. aTest performance classified according to AUC as follows: <0.5 = not useful, 0.5–0.6 = bad, 0.6–0.7 = sufficient, 0.7–0.8 = good, 0.8–0.9 = very good, and 0.9–1.0 = excellent. The panel consisted of 10 unique SARS-CoV-2 antibody positive DBS cards and 10 unique SARS-CoV-2 negative DBS cards contrived from patient samples. The in-house assays from the University of Ottawa and University of Toronto (spike, RBD, and nucleocapsid) showed a clear separation between positive and negative DBS (Figs 3 and 4, Table 2, S2 and S3 Figs). However, the in-house nucleocapsid (University of Toronto) assay was unable to differentiate all positive and negative DBS successfully.

Discussion

On a small panel of DBS, this study has shown that the EUROIMMUN, Elecsys spike, and GSP/DELFIA commercial assays as well in-house assays are capable of achieving a sensitivity, specificity, PPV, and NPV of 100% (95% CI = 72.2, 100) under a broad range of prevalence scenarios following testing of DBS eluate samples collected ≥36 days post symptom onset without optimization. We interpret these findings cautiously since a recent systematic review by Lisboa Bastos et al. [22] assessing 40 serological studies reported that the maximum pooled sensitivities for enzyme linked immunosorbent assays measuring IgG or IgM (ELISAs) and chemiluminescent immunoassays (CLIAs) was 84.3% (95% CI = 75.6, 90.9) and 97.8% (95% CI = 46.2, 100) respectively. The accuracy of serological assays also varies according to sampling time frames and sensitivity typically exceeds 90% between 15–35 days post symptom onset [22]. We were unable to stratify our analysis according to days post symptom onset due to our sample size and narrow range of collection dates. Furthermore, our samples consisted of COVID-19 convalescent plasma donors who are likely to have significant antibody titers [23]. Nonetheless, we clearly show that certain assays like the LIASON, COV2G, and COV2T could not reliably identify positive and negative specimens by DBS testing and are unlikely to be useful in various prevalence settings. Based on our ROC curve analysis, assays like the Platelia, Elecsys nucleocapsid, VITROS, and Architect, which did not reach 100% for all performance characteristics, could clearly distinguish between positive and negative samples therefore could also potentially achieve a sensitivity, specificity, PPV, and NPV near 100% via threshold adjustments [24-26]. However, establishing appropriate thresholds will require a study using a large number of well characterised clinical samples representing a broad range of factors associated with developing robust immune responses to SARS-CoV-2 such as sex, age, immunodeficiencies, and disease severity [27]. This validation study has several limitations that must be considered. First, our panel was small, albeit well characterized. Current efforts are underway to validate in-house and commercial multiplex assays using a larger DBS panel that covers a broader range of days between infection and sample collection. Second, we did not take into consideration other capillary blood collection methods such as capillary tubes. While capillary tubes have been shown to be a practical alternative to venepuncture for SARS-CoV-2 antibody testing [28,29], they cannot be transported as easily or safely as DBS therefore their applicability are somewhat limited for large sero-surveys within the Canadian context. Last, DBS eluates, prepared using a well-characterized buffer used specifically by the National Laboratory for HIV Reference Services (NLHRS), were sent to the participants (with the exception of the GSP/DELFIA). The assessment of different buffers or the further optimization of DBS elution conditions was beyond the scope of this study. Additional refinement of these procedures may have yielded more optimistic results on the assays that were not reported to fare well in this evaluation. However, the purpose of this evaluation was to develop a simplified procedure based on simple well-characterized practices previously developed within the NLHRS. Implementing this standardized approach allowed for an effective and timely identification of promising platforms that could be used for DBS testing for anti-SARS-CoV-2 antibodies without extensive modification. We believe that this was accomplished despite these limitations. In conclusion, we assessed the performance of 10 commercial and 2 in-house serological assays for DBS eluate testing. Several of these assays achieved a specificity, sensitivity, PPV, and NPV adequate for sero-surveys even in low-prevalence settings. These findings suggest that the high demand for SARS-CoV-2 serology testing, mainly driven by sero-surveys within the Canadian context, could be met by the collection and testing of DBS by several different assays, thereby minimizing the risk of shortages. Furthermore, DBS collection has the potential to expand testing use and access while limiting the requirements for specimen collection on healthcare professionals already overtaxed with the COVID-19 response. While our sample size was small, this validation study, undertaken initially to determine the feasibility of a nationally representative household-based sero-survey that would be based on DBS, resulted in the selection of the in-house assays. They were chosen not only because of their test performance, but also because they are capable of distinguishing between infection- and vaccine-induced antibody responses which will be important in informing public health strategies in diverse jurisdictions across Canada. Our study represents the foundation for future validation studies on DBS specimens, that will undoubtedly play a central role in shaping Canada’s public health policy in response to COVID-19. However, larger DBS panels representing a broader range of days between symptom onset and sample collection as well as socio-demographic and clinical characteristics of COVID-19 patients will be required to substantiate our findings.

Methods

Ethics statement

All experiments were carried out in accordance with relevant guidelines and regulations. Written informed consent was obtained from all participants who provided blood samples from which we contrived SARS-CoV-2 antibody positive and negative DBS cards. All participants were 18 years of age or older. Ethical approval was obtained from the Health Canada and Public Health Agency of Canada Research Ethics Board (REB no. 2020-022P).

Patient specimens

SARS-CoV-2 antibody positive and negative plasma used to contrive DBS specimens are described in Table 3. SARS-CoV-2 antibody positive plasma was collected from COVID-19 convalescent donors at Mount Sinai Hospital (Toronto, Canada). Plasma was tested for SARS-CoV-2 antibodies using the Platelia SARS-CoV-2 Total Ab (Bio-Rad, Hercules, California) or Anti-SARS-CoV-2 ELISA IgG (EUROIMMUN, Lübeck, Germany) kits. SARS-CoV-2 negative plasma was collected from healthy donors within the National Microbiology Laboratory (Winnipeg, Canada). To prepare plasma, blood was collected in EDTA Vacutainer tubes (Beckton Dickinson, Franklin Lates, NJ) and centrifuged at 1,500 RPM for 7 minutes. Plasma was tested using the Platelia SARS-CoV-2 Total Ab (Bio-Rad) or Anti-SARS-CoV-2 ELISA IgG (EUROIMMUN, Lübeck, Germany) kits to verify donors were negative for SARS-CoV-2 antibodies. All assays were performed according to the manufacturer’s instructions.
Table 3

SARS-CoV-2 antibody positive and negative plasma used to contrive dried blood spot specimens.

Sample IDDescriptionSample collection, days post infectionEUROIMMUN Anti-SARS-CoV-2Bio-Rad Platelia SARS-CoV-2
OD ratioInterpretationaODMR4 ratioInterpretationb
J12 Healthy donorPre-COVID-190.20Negative0.08Negative
D17 Healthy donorPre-COVID-190.12Negative0.11Negative
J11 Healthy donorPre-COVID-190.16Negative0.08Negative
J1 Healthy donorPre-COVID-190.15Negative0.08Negative
D142 Healthy donorPre-COVID-190.13Negative0.09Negative
J19 Healthy donorPre-COVID-190.49Negative0.08Negative
D9 Healthy donorPre-COVID-190.18Negative0.09Negative
J30 Healthy donorPre-COVID-190.17Negative0.10Negative
J15 Healthy donorPre-COVID-190.18Negative0.07Negative
J7 Healthy donorPre-COVID-190.41Negative0.10Negative
20–1887 COVID-19 convalescent donor442.45Positive≥4.00Positive
20–1954 COVID-19 convalescent donor572.32Positive3.14Positive
20–1955 COVID-19 convalescent donor421.36Positive2.12Positive
20–1877 COVID-19 convalescent donor414.10Positive≥4.00Positive
20–1878 COVID-19 convalescent donor514.01Positive≥4.00Positive
20–1882 COVID-19 convalescent donor662.90Positive≥4.00Positive
20–1886 COVID-19 convalescent donor524.62Positive≥4.00Positive
20–1879 COVID-19 convalescent donor3611.04Positive≥4.00Positive
20–1885 COVID-19 convalescent donor385.78Positive≥4.00Positive
20–1952 COVID-19 convalescent donor424.39Positive≥4.00Positive

OD: Optical density; ODMR4: Optical density for cut-off control R4.

aResults were evaluated by calculating the ratio of the OD of the plasma specimen over the OD of the assay calibrator. Results were interpreted as follows: OD ratio <0.8 reported as positive; OD ratio ≧0.8 to <1.1 reported as borderline; OD ratio ≧1.1 reported as positive.

bResults were evaluated by calculating the ratio of the OD of the plasma specimen over the OD of the kit cut-off control R4. Results were interpreted as follows: ODMR4 ratio <0.8 reported as negative; ODMR4 ratio o ≧0.8 to <1.0 reported as equivocal; ODMR4 ratio ≧1.0 reported as positive.

OD: Optical density; ODMR4: Optical density for cut-off control R4. aResults were evaluated by calculating the ratio of the OD of the plasma specimen over the OD of the assay calibrator. Results were interpreted as follows: OD ratio <0.8 reported as positive; OD ratio ≧0.8 to <1.1 reported as borderline; OD ratio ≧1.1 reported as positive. bResults were evaluated by calculating the ratio of the OD of the plasma specimen over the OD of the kit cut-off control R4. Results were interpreted as follows: ODMR4 ratio <0.8 reported as negative; ODMR4 ratio o ≧0.8 to <1.0 reported as equivocal; ODMR4 ratio ≧1.0 reported as positive.

Contrived dried blood spot specimens

A panel consisting of 10 unique SARS-CoV-2 antibody positive DBS cards were contrived and 10 unique SARS-CoV-2 negative DBS cards were directly spotted from EDTA whole blood to assess the performance of commercial and in-house serological tests (S4 Fig). Each testing site was blinded to the status of the DBS cards. SARS-CoV-2 antibody positive plasma samples were contrived into DBS specimens by using blood collected from healthy donors within the National Microbiology Laboratory. SARS-CoV-2 antibody negative blood was centrifuged at 1,500 RPM for 7 minutes and the plasma was removed. The remaining red blood cells were re-suspended with SARS-CoV-2 antibody positive plasma using a 1:1 ratio and 75 μL was spotted onto each circle of a Whatman 903 Protein Saver card (GE Healthcare, Boston, MA). Spotted cards were allowed to air-dry for at least 2 hours in a biosafety cabinet and then packaged in a gas impermeable bag with a desiccant pack and a humidity indicator card. Packaged cards (maximum 10 per bag) were stored at -80°C until further testing. SARS-CoV-2 antibody negative blood was spotted directly from the EDTA Vacutainer tubes onto Whatman 903 Proteinsaver cards as described above.

Dried blood spot elution

DBS samples were punched using a 6 mm hole punch into a 96 deep well plate. One to four 6 mm (1/4 inch) punches were added to each well and eluted in DPBS buffer (pH 7.4) containing 0.5% BSA and 0.05% Tween-20 overnight at 4°C with agitation (400 RPM). For the Anti-SARS-CoV-2 ELISA IgG (EUROIMMUN) a single punch was eluted in 500 ul of Sample Buffer overnight without agitation and 100 μL of eluate was transferred to the assay plate. For Platelia SARS-CoV-2 Total Ab (Bio-Rad) assay, two punches were eluted in 130 μL of Sample Diluent overnight with gentle agitation (400 RPM), 75 μL of eluate was mixed with conjugate, and 100 μL of the eluate-conjugate mixture was transferred to the assay plate. Elution volumes were adjusted for all other assays according to manufacturer’s specifications or in-house developed protocols (Table 4). Afterwards, plates were incubated at room temperature for 30 minutes with agitation (400 RPM) and each DBS eluate was transferred to an individual 2.0 mL screw cap tube. Eluates were stored at -80°C until shipment on dry ice to each testing site.
Table 4

Commercial and in-house serological tests assessed for SARS-CoV-2 antibody testing on dried blood spot specimens.

AssayManufacturerSiteAntibody classTargetElution volume (μL)aInterpretation
NegativeEquivocalPositive
Anti-SARS-CoV-2 ELISAEUROIMMUNJ.C. WiltIgGS1500OD ratio <0.8OD ratio ≥0.8 <1.1OD ratio ≥1.1
LIASON SARS-CoV-2DiaSorinCadhamIgGS1, S2200<15.0 AU/mLN/A≥15.0 AU/mL
SARS-CoV-2 COV2GSiemensU of T (DID)IgGS1300Index <1.0N/AIndex ≥1.0
SARS-CoV-2 COV2TSiemensU of T (DID)TotalS1300Index <1.0N/AIndex ≥1.0
Architect SARS-CoV-2AbbottMcGillIgGN180S/Co <1.40N/AS/Co ≥1.40
GSP/DELFIA Anti-SARS-CoV-2Perkin ElmerCHEOIgGS1N/AbS/Co <1.4N/AS/Co ≥1.4
Platelia SARS-CoV-2Bio-RadJ.C. WiltTotalN130OD ratio <0.8OD ratio ≥0.8 <1.0OD ratio ≥1.1
Elecsys Anti-SARS-CoV-2RocheNMLTotalS250<0.8 U/mLN/A≥0.8 U/mL
Elecsys Anti-SARS-CoV-2RocheNMLTotalN250Cutoff index <1.0N/ACutoff index ≥1.0
VITROS Anti-SARS-CoV-2Ortho Clinical DiagnosticsBCCDCTotalS1300S/Co <1.0N/AS/Co ≥1.0
Abe KT, et al. 2020[30]In-houseU of TIgGS100S/Co <1.0N/AS/Co ≥1.0
Abe KT, et al. 2020[30]In-houseU of TIgGRBD100S/Co <1.0N/AS/Co ≥1.0
Abe KT, et al. 2020[30]In-houseU of TIgGN100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (polyclonal)S100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (monoclonal)S100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (polyclonal)RBD100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (monoclonal)RBD100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (polyclonal)N100S/Co <1.0N/AS/Co ≥1.0
Adapted from Amanat F, et al. 2020[31]In-houseU of OIgG (monoclonal)N100S/Co <1.0N/AS/Co ≥1.0

J.C. Wilt: National Microbiology Laboratory at the J.C. Wilt Infectious Diseases Research Centre, Winnipeg, Canada; U of T (DID): Division of Infectious Diseases, University of Toronto; McGill: McGill University Health Centre, McGill University; CHEO: Children’s Hospital of Eastern Ontario, Ottawa, Canada; NML: National Microbiology Laboratory at the Canadian Science Centre for Human and Animal Health, Winnipeg, Canada; BCCDC: British Columbia Centre for Disease Control, Vancouver, Canada; U of T: University of Toronto, Toronto, Canada; U of O: University of Ottawa, Ottawa, Canada; S1: Spike S1; S2: Spike S2; S: Total spike; N: Nucleocapsid; RBD: Receptor binding domain; OD: Optical density; S/Co: Signal cut-off; AU/mL: Arbitrary units/mL; N/A: Not applicable.

a The elution volume was based on the minimum volume required to perform the assay, the assay’s dead volume, and an additional 30 μL of buffer for each 6 mm (1/4 inch) DBS punch to account for liquid absorbed by the filter paper.

bDBS cards shipped directly to testing site. The GSP/DELFIA performs automated, onboard DBS punching (1 x 3.2 mm) and elution.

J.C. Wilt: National Microbiology Laboratory at the J.C. Wilt Infectious Diseases Research Centre, Winnipeg, Canada; U of T (DID): Division of Infectious Diseases, University of Toronto; McGill: McGill University Health Centre, McGill University; CHEO: Children’s Hospital of Eastern Ontario, Ottawa, Canada; NML: National Microbiology Laboratory at the Canadian Science Centre for Human and Animal Health, Winnipeg, Canada; BCCDC: British Columbia Centre for Disease Control, Vancouver, Canada; U of T: University of Toronto, Toronto, Canada; U of O: University of Ottawa, Ottawa, Canada; S1: Spike S1; S2: Spike S2; S: Total spike; N: Nucleocapsid; RBD: Receptor binding domain; OD: Optical density; S/Co: Signal cut-off; AU/mL: Arbitrary units/mL; N/A: Not applicable. a The elution volume was based on the minimum volume required to perform the assay, the assay’s dead volume, and an additional 30 μL of buffer for each 6 mm (1/4 inch) DBS punch to account for liquid absorbed by the filter paper. bDBS cards shipped directly to testing site. The GSP/DELFIA performs automated, onboard DBS punching (1 x 3.2 mm) and elution.

SARS-CoV-2 antibody testing

DBS eluates were tested for SARS-CoV-2 antibodies with 10 commercial assays and 2 in-house assays (Table 4) according to the manufacturer’s instructions or laboratory developed protocols respectively. No attempts were made at this point to optimize protocols for DBS specimens. The commercial tests consisted of 6 IgG based assays: Anti-SARS-CoV-2 ELISA (EUROIMMUN), LIASON SARS-CoV-2 (DiaSorin, Saluggia, Italy), SARS-CoV-2 COV2G (Siemens, Erlangen, Germany), Architect SARS-CoV-2 (Abbott, Mississauga, Canada), and GSP/DELFIA Anti-SARS-CoV-2 (PerkinElmer, Waltham, Massachusetts); and 5 total antibody assays: Platelia SARS-CoV-2 (Bio-Rad), SARS-CoV-2 COV2T (Siemens), Elecsys quantitative Anti-SARS-CoV-2 (Elecsys spike; Roche, Basel, Switzerland), Elecsys Anti-SARS-CoV-2 (Elecsys nucleocapsid; Roche), and VITROS Anti-SARS-CoV-2 (Ortho Clinical Diagnostics, Raritan, New Jersey). EUROIMMUN was included since it is the only assay currently approved by Health Canada for use with serum which also has a manufacturer developed protocol for use with DBS (CE marked). Both in-house tests, described in greater detail elsewhere [30] and Table 2, consist of IgG assays targeting SARS-CoV-2 spike, receptor binding domain (RBD), and nucleocapsid proteins. Each testing site was responsible for interpreting and reporting their own data.

Statistical analysis

Statistical analysis was conducted using Prism version 9.0.0 (GraphPad Software, San Diego, CA). Test performance on DBS expressed in terms of sensitivity, specificity, positive predictive values (PPV), and negative predictive values (NPV), was computed in comparison with the serum-based Anti-SARS-CoV-2 ELISA (EUROIMMUN) assay results as the reference because it is the only assay currently approved by Health Canada for use with serum which also has a manufacturer developed protocol for use with DBS (CE marked). PPV (sensitivity x prevalence/sensitivity x prevalence + [1 –specificity] x [1 –prevalence]) and NPV (specificity x [1 –prevalence]/specificity x [1 –prevalence] + [1 –sensitivity) x prevalence) was also re-calculated using a prevalence of 1.0%, 2.5%, 5.0%, 10.0%, 20.0%, and 40.0%. Equivocal results were considered positive. Agreement with the reference test was quantified using Cohen’s kappa coefficient (https://www.graphpad.com/quickcalcs/kappa1/?K=2) and interpreted as follows: <0 = no agreement, 0–0.20 = slight agreement, 0.21–0.40 = fair agreement, 0.41–0.60 = moderate agreement, 0.61–0.80 = substantial agreement, and 0.81–1.00 = almost perfect agreement [32]. Confidence intervals were computed using the hybrid Wilson/Brown method [33]. Receiver operating characteristics (ROC) curves were also computed using the Wilson/Brown method. A P-value ≤0.05 was considered statistically significant. Test performance classified according to the area under the ROC curve (AUC) was interpreted as follows: <0.5 = not useful, 0.5–0.6 = bad, 0.6–0.7 = sufficient, 0.7–0.8 = good, 0.8–0.9 = very good, and 0.9–1.0 = excellent [34,35]. Overall positive and negative percent concordance and Cohen’s kappa coefficient were used to quantify inter-assay agreement between the 10 commercial assays and 2 in-house assays. Confidence intervals were computed using the hybrid Wilson/Brown method. Cohen’s kappa coefficient was interpreted as described above. ROC curves are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In- house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). The VITROS assay could not achieve a specificity greater than 0% therefore, only PPV is shown. (TIF) Click here for additional data file.

Receiver operating characteristic curve for each commercial and in-house assay on dried blood spot specimens.

ROC curves are presented for n = 10 SARS-CoV-2 antibody negative DBS specimens and n = 10 SARS-CoV-2 antibody positive DBS specimens. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In- house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). (TIF) Click here for additional data file. Distribution of values obtained for each commercial and in-house assay on dried blood spot (DBS) specimens. Each panel shows the optical density ratio (OD Ratio), arbitrary units per mL (AU/mL), index, units per mL (U/mL), cut-off index, or signal to cut-off ratio (S/Co) for SARS-CoV-2 antibody negative DBS specimens (n = 10) represented in blue and SARS-CoV-2 antibody positive DBS specimens (n = 10) represented in orange. All values are log10 transformed to aid with visualisation. One 6 mm (1/4 inch) punch was used for the EUROIMMUN assay, and two 6 mm (1/4 inch) punches were used for the Platelia and in-house assays. EUROIMMUN = Anti-SARS-CoV-2 ELISA assay (EUROIMMUN, Lübeck, Germany). Platelia = SARS-CoV-2 assay (Bio-Rad, Hercules, California). LIASON = SARS-CoV-2 assay (DiaSorin, Saluggia, Italy). COV2G = SARS-CoV-2 COV2G assay (Siemens, Erlangen, Germany). COV2T = SARS-CoV-2 COV2T assay (Siemens). Elecsys S = Quantitative Anti-SARS-CoV-2 assay (Elecsys spike; Roche, Basel, Switzerland). Elecsys N = Anti-SARS-CoV-2 assay (Elecsys nucleocapsid; Roche). VITROS = Anti-SARS-CoV-2 assay (Ortho Clinical Diagnostics, Raritan, New Jersey). Architect = SARS-CoV-2 assay (Abbott, Mississauga, Canada). GSP/DELFIA = Anti-SARS-CoV-2 assay (PerkinElmer, Waltham, Massachusetts). In-house S (U of T) = In-house spike assay (University of Toronto). In-house RBD (U of T) = In-house RBD assay (University of Toronto). In-house N (U of T) = In-house nucleocapsid assay (University of Toronto). In-house S, mono (U of O) = In-house monoclonal spike assay (University of Ottawa). In-house RBD, mono (U of O) = In-house monoclonal RBD assay (University of Ottawa). In-house N, mono (U of O) = In-house monoclonal nucleocapsid assay (University of Ottawa). In-house S, poly (U of O) = In-house polyclonal spike assay (University of Ottawa). In-house RBD, poly (U of O) = In-house polyclonal RBD assay (University of Ottawa). In-house N, poly (U of O) = In-house polyclonal nucleocapsid assay (University of Ottawa). (TIF) Click here for additional data file.

Procedure for contriving dried blood spot specimens.

(TIF) Click here for additional data file.

Inter-assay concordance and Cohen’s kappa coefficient for SARS-CoV-2 antibody testing on dried blood spot specimens.

(DOCX) Click here for additional data file.

Raw data generated during this study for each commercial and in-house assay on dried blood spot specimens.

(DOCX) Click here for additional data file.
  30 in total

1.  Comparative Clinical Evaluation of the Roche Elecsys and Abbott Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Serology Assays for Coronavirus Disease 2019 (COVID-19).

Authors:  Shaun S Tan; Sharon Saw; Ka Lip Chew; Cindy Wang; Anastacia Pajarillaga; Candy Khoo; Weixuan Wang; Zahidah Mohamed Ali; Zhixin Yang; Yiong Huak Chan; Paul Tambyah; Roland Jureen; Sunil K Sethi
Journal:  Arch Pathol Lab Med       Date:  2021-01-01       Impact factor: 5.534

2.  A serological assay to detect SARS-CoV-2 seroconversion in humans.

Authors:  Fatima Amanat; Daniel Stadlbauer; Shirin Strohmeier; Thi H O Nguyen; Veronika Chromikova; Meagan McMahon; Kaijun Jiang; Guha Asthagiri Arunkumar; Denise Jurczyszak; Jose Polanco; Maria Bermudez-Gonzalez; Giulio Kleiner; Teresa Aydillo; Lisa Miorin; Daniel S Fierer; Luz Amarilis Lugo; Erna Milunka Kojic; Jonathan Stoever; Sean T H Liu; Charlotte Cunningham-Rundles; Philip L Felgner; Thomas Moran; Adolfo García-Sastre; Daniel Caplivski; Allen C Cheng; Katherine Kedzierska; Olli Vapalahti; Jussi M Hepojoki; Viviana Simon; Florian Krammer
Journal:  Nat Med       Date:  2020-05-12       Impact factor: 53.440

3.  Evaluation of the QIAstat-Dx Respiratory SARS-CoV-2 Panel, the First Rapid Multiplex PCR Commercial Assay for SARS-CoV-2 Detection.

Authors:  Benoit Visseaux; Quentin Le Hingrat; Gilles Collin; Donia Bouzid; Samuel Lebourgeois; Diane Le Pluart; Laurène Deconinck; François-Xavier Lescure; Jean-Christophe Lucet; Lila Bouadma; Jean-François Timsit; Diane Descamps; Yazdan Yazdanpanah; Enrique Casalino; Nadhira Houhou-Fidouh
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

4.  A new coronavirus associated with human respiratory disease in China.

Authors:  Fan Wu; Su Zhao; Bin Yu; Yan-Mei Chen; Wen Wang; Zhi-Gang Song; Yi Hu; Zhao-Wu Tao; Jun-Hua Tian; Yuan-Yuan Pei; Ming-Li Yuan; Yu-Ling Zhang; Fa-Hui Dai; Yi Liu; Qi-Min Wang; Jiao-Jiao Zheng; Lin Xu; Edward C Holmes; Yong-Zhen Zhang
Journal:  Nature       Date:  2020-02-03       Impact factor: 49.962

5.  Diagnostic Efficiency of Three Fully Automated Serology Assays and Their Correlation with a Novel Surrogate Virus Neutralization Test in Symptomatic and Asymptomatic SARS-COV-2 Individuals.

Authors:  Salma Younes; Hadeel Al-Jighefee; Farah Shurrab; Duaa W Al-Sadeq; Nadin Younes; Soha R Dargham; Nader Al-Dewik; Hamda Qotba; Mohamed Syed; Ahmed Alnuaimi; Hadi M Yassine; Patrick Tang; Laith J Abu-Raddad; Gheyath K Nasrallah
Journal:  Microorganisms       Date:  2021-01-25

6.  Estimating the efficacy of symptom-based screening for COVID-19.

Authors:  Alison Callahan; Ethan Steinberg; Jason A Fries; Saurabh Gombar; Birju Patel; Conor K Corbin; Nigam H Shah
Journal:  NPJ Digit Med       Date:  2020-07-13

7.  Multi-center evaluation of cepheid xpert® xpress SARS-CoV-2 point-of-care test during the SARS-CoV-2 pandemic.

Authors:  Femke Wolters; Jeroen van de Bovenkamp; Bart van den Bosch; Sharon van den Brink; Maaike Broeders; Ngoc Hoa Chung; Barbara Favié; Gabriel Goderski; Judith Kuijpers; Ilse Overdevest; Janette Rahamat-Langedoen; Lisa Wijsman; Willem Jg Melchers; Adam Meijer
Journal:  J Clin Virol       Date:  2020-05-11       Impact factor: 3.168

8.  A simple protein-based surrogate neutralization assay for SARS-CoV-2.

Authors:  Kento T Abe; Zhijie Li; Reuben Samson; Payman Samavarchi-Tehrani; Emelissa J Valcourt; Heidi Wood; Patrick Budylowski; Alan P Dupuis; Roxie C Girardin; Bhavisha Rathod; Jenny H Wang; Miriam Barrios-Rodiles; Karen Colwill; Allison J McGeer; Samira Mubareka; Jennifer L Gommerman; Yves Durocher; Mario Ostrowski; Kathleen A McDonough; Michael A Drebot; Steven J Drews; James M Rini; Anne-Claude Gingras
Journal:  JCI Insight       Date:  2020-10-02

9.  Clinical Evaluation of BD Veritor SARS-CoV-2 Point-of-Care Test Performance Compared to PCR-Based Testing and versus the Sofia 2 SARS Antigen Point-of-Care Test.

Authors:  Stephen Young; Stephanie N Taylor; Catherine L Cammarata; Katey G Varnado; Celine Roger-Dalbert; Amanda Montano; Christen Griego-Fullbright; Cameron Burgard; Catherine Fernandez; Karen Eckert; Jeffrey C Andrews; Huimiao Ren; Joseph Allen; Ronald Ackerman; Charles K Cooper
Journal:  J Clin Microbiol       Date:  2020-12-17       Impact factor: 5.948

10.  Evaluation of SARS-CoV-2 serology assays reveals a range of test performance.

Authors:  Jeffrey D Whitman; Joseph Hiatt; Cody T Mowery; Brian R Shy; Ruby Yu; Tori N Yamamoto; Ujjwal Rathore; Gregory M Goldgof; Caroline Whitty; Jonathan M Woo; Antonia E Gallman; Tyler E Miller; Andrew G Levine; David N Nguyen; Sagar P Bapat; Joanna Balcerek; Sophia A Bylsma; Ana M Lyons; Stacy Li; Allison Wai-Yi Wong; Eva Mae Gillis-Buck; Zachary B Steinhart; Youjin Lee; Ryan Apathy; Mitchell J Lipke; Jennifer Anne Smith; Tina Zheng; Ian C Boothby; Erin Isaza; Jackie Chan; Dante D Acenas; Jinwoo Lee; Trisha A Macrae; Than S Kyaw; David Wu; Dianna L Ng; Wei Gu; Vanessa A York; Haig Alexander Eskandarian; Perri C Callaway; Lakshmi Warrier; Mary E Moreno; Justine Levan; Leonel Torres; Lila A Farrington; Rita P Loudermilk; Kanishka Koshal; Kelsey C Zorn; Wilfredo F Garcia-Beltran; Diane Yang; Michael G Astudillo; Bradley E Bernstein; Jeffrey A Gelfand; Edward T Ryan; Richelle C Charles; A John Iafrate; Jochen K Lennerz; Steve Miller; Charles Y Chiu; Susan L Stramer; Michael R Wilson; Aashish Manglik; Chun Jimmie Ye; Nevan J Krogan; Mark S Anderson; Jason G Cyster; Joel D Ernst; Alan H B Wu; Kara L Lynch; Caryn Bern; Patrick D Hsu; Alexander Marson
Journal:  Nat Biotechnol       Date:  2020-08-27       Impact factor: 54.908

View more
  11 in total

1.  Seropositivity and risk factors for SARS-CoV-2 infection in a South Asian community in Ontario: a cross-sectional analysis of a prospective cohort study.

Authors:  Sonia S Anand; Corey Arnold; Shrikant I Bangdiwala; Shelly Bolotin; Dawn Bowdish; Rahul Chanchlani; Russell J de Souza; Dipika Desai; Sujane Kandasamy; Farah Khan; Zainab Khan; Marc-André Langlois; Jayneel Limbachia; Scott A Lear; Mark Loeb; Lawrence Loh; Baanu Manoharan; Kiran Nakka; Martin Pelchat; Zubin Punthakee; Karleen M Schulze; Natalie Williams; Gita Wahi
Journal:  CMAJ Open       Date:  2022-07-05

2.  From a recombinant key antigen to an accurate, affordable serological test: Lessons learnt from COVID-19 for future pandemics.

Authors:  Renata G F Alvim; Tulio M Lima; Danielle A S Rodrigues; Federico F Marsili; Vicente B T Bozza; Luiza M Higa; Fabio L Monteiro; Daniel P B Abreu; Isabela C Leitão; Renato S Carvalho; Rafael M Galliez; Terezinha M P P Castineiras; Leonardo H Travassos; Alberto Nobrega; Amilcar Tanuri; Orlando C Ferreira; André M Vale; Leda R Castilho
Journal:  Biochem Eng J       Date:  2022-07-16       Impact factor: 4.446

3.  A Qualitative Comparison of the Abbott SARS-CoV-2 IgG II Quant Assay against Commonly Used Canadian SARS-CoV-2 Enzyme Immunoassays in Blood Donor Retention Specimens, April 2020 to March 2021.

Authors:  Kento T Abe; Bhavisha Rathod; Karen Colwill; Anne-Claude Gingras; Ashleigh Tuite; Ninette F Robbins; Guillermo Orjuela; Craig Jenkins; Valerie Conrod; Qi-Long Yi; Sheila F O'Brien; Steven J Drews
Journal:  Microbiol Spectr       Date:  2022-06-02

4.  Evaluation of the bioMérieux VIDAS HIV Duo Quick and Anti-HCV assays for dried blood spot based serosurveillance.

Authors:  François Cholette; Braedy Farmer; Olga Balakireva; Daria Pavlova; Anna Lopatenko; Iryna Chukhalova; Svitlana Bargan; Sharmistha Mishra; Marissa Becker; Emma R Lee; John Kim; Paul Sandstrom
Journal:  Sci Rep       Date:  2022-06-17       Impact factor: 4.996

5.  Household transmission of SARS-CoV-2 from unvaccinated asymptomatic and symptomatic household members with confirmed SARS-CoV-2 infection: an antibody-surveillance study.

Authors:  Maala Bhatt; Amy C Plint; Ken Tang; Richard Malley; Anne Pham Huy; Candice McGahern; Jennifer Dawson; Martin Pelchat; Lauren Dawson; Terry Varshney; Corey Arnold; Yannick Galipeau; Michael Austin; Nisha Thampi; Fuad Alnaji; Marc-André Langlois; Roger L Zemek
Journal:  CMAJ Open       Date:  2022-04-12

6.  Monitoring of SARS-CoV-2 antibodies using dried blood spot for at-home collection.

Authors:  Peyton K Miesse; Bradley B Collier; Russell P Grant
Journal:  Sci Rep       Date:  2022-04-06       Impact factor: 4.379

7.  A scalable serology solution for profiling humoral immune responses to SARS-CoV-2 infection and vaccination.

Authors:  Karen Colwill; Yannick Galipeau; Matthew Stuible; Christian Gervais; Corey Arnold; Bhavisha Rathod; Kento T Abe; Jenny H Wang; Adrian Pasculescu; Mariam Maltseva; Lynda Rocheleau; Martin Pelchat; Mahya Fazel-Zarandi; Mariam Iskilova; Miriam Barrios-Rodiles; Linda Bennett; Kevin Yau; François Cholette; Christine Mesa; Angel X Li; Aimee Paterson; Michelle A Hladunewich; Pamela J Goodwin; Jeffrey L Wrana; Steven J Drews; Samira Mubareka; Allison J McGeer; John Kim; Marc-André Langlois; Anne-Claude Gingras; Yves Durocher
Journal:  Clin Transl Immunology       Date:  2022-03-23

8.  Performance of Immunoglobulin G Serology on Finger Prick Capillary Dried Blood Spot Samples to Detect a SARS-CoV-2 Antibody Response.

Authors:  Aidan M Nikiforuk; Brynn McMillan; Sofia R Bartlett; Ana Citlali Márquez; Tamara Pidduck; Jesse Kustra; David M Goldfarb; Vilte Barakauskas; Graham Sinclair; David M Patrick; Manish Sadarangani; Gina S Ogilvie; Mel Krajden; Muhammad Morshed; Inna Sekirov; Agatha N Jassem
Journal:  Microbiol Spectr       Date:  2022-03-10

9.  Evaluation of serological assays for SARS-CoV-2 antibody testing from dried blood spots collected from cohorts with prior SARS-CoV-2 infection.

Authors:  Beth Catlett; Mitchell Starr; Dorothy A Machalek; Thidarat Danwilai; Michael Palmer; Andrew Kelly; John Kaldor; Gregory J Dore; David Darley; Gail Matthews; Philip H Cunningham
Journal:  J Clin Virol Plus       Date:  2022-06-24

10.  Comparative performance data for multiplex SARS-CoV-2 serological assays from a large panel of dried blood spot specimens.

Authors:  François Cholette; Rissa Fabia; Angela Harris; Hannah Ellis; Karla Cachero; Lukas Schroeder; Christine Mesa; Philip Lacap; Corey Arnold; Yannick Galipeau; Marc-André Langlois; Karen Colwill; Anne-Claude Gingras; Allison McGeer; Elizabeth Giles; Jacqueline Day; Carla Osiowy; Yves Durocher; Catherine Hankins; Bruce Mazer; Michael Drebot; John Kim
Journal:  Heliyon       Date:  2022-08-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.