Fig. 1Intact Lumbar Model. a AP view; b LAT view
Lumbar finite element model: unit type and material properties
| Structure | Unit Type | Thickness(mm) | Elastic Modulus(MPa) | Poisson’s ratio |
|---|---|---|---|---|
| Vertebra cortical bone | S3R | 0.4 | 12,000 | 0.3 |
| Vertebra cancellous bone | C3D4 | N/A | 100 | 0.2 |
| Vertebra endplate cartilage | S3R | 0.25 | 1000 | 0.2 |
| Vertebra posterior structure | C3D4 | N/A | 3500 | 0.25 |
| Fibrous annulus(Normal) | C3D4 | N/A | 2.6 | 0.4 |
| Nucleus pulposus(Normal) | C3D4 | N/A | 1.0 | 0.49 |
| Fibrous annulus(Degenerative) | C3D4 | N/A | 12.3 | 0.35 |
| Nucleus pulposus(Degenerative) | C3D4 | N/A | 1.7 | 0.4 |
Lumbar finite element model main ligament material properties
| Ligament | Elastic Modulus(MPa) | Cross-sectional area (mm2) | Average Length(mm) | Stiffness(kg•m−2•s− 2) |
|---|---|---|---|---|
| ALL | 7.8 | 22.4 | 20 | 8.74 |
| PLL | 10 | 7.0 | 12 | 5.83 |
| LF | 17 | 14.1 | 15 | 15.83 |
| TIL | 10 | 0.6 | 32 | 0.19 |
| JCL | 7.5 | 10.5 | 5 | 15.75 |
| IVL | 10 | 14.1 | 13 | 10.85 |
| SVL | 8.0 | 10.5 | 22 | 2.39 |
ALL anterior longitudinal ligament, PLL posterior longitudinal ligament, LF ligamentum Flavum TIL Transverse intersegmental ligament, JCL Joint capsule ligament, IVL Intervertebra ligament, SVL Supravertebra ligament
Fig. 2PLIF Model. a AP View; b LAT View
Fig. 3PLF model. a AP view; b LAT view
Fig. 4MFF model. a AP view; b LAT view
Material properties of different parts of the internal fixation system
| Internal Fixation System | Elastic Modulus(MPa) | Poisson’s ratio |
|---|---|---|
| Legacy Pedicle Screw | 110,000 | 0.3 |
| Legacy Connecting Rod | 110,000 | 0.3 |
| Johnson-johnson Cage(PEEK) | 3600 | 0.3 |
L4–5 Postoperative Range of Movement(Unit:°)
| Model | Flexion | Extension | Lateral Bending | Rotation |
|---|---|---|---|---|
| Intact Model | 5.6 ± 1.3 | 4.6 ± 1.1 | 3.4 ± 1.2 | 2.3 ± 0.8 |
| PLF Model | 1.1 ± 0.2 | 1.6 ± 0.1 | 0.4 ± 0.2 | 0.2 ± 0.1 |
| PLIF Model | 1.0 ± 0.1 | 1.3 ± 0.1 | 0.3 ± 0.2 | 0.3 ± 0.8 |
| Modified articular fusion model | 1.2 ± 0.3 | 1.1 ± 0.1 | 0.5 ± 0.2 | 0.3 ± 0.1 |
Fig. 5L4–5 Posterior Range of Motion. *P < 0.05
Stiffness and stress distribution of intact model and different postoperative fusion models
| Model | Stiffness (N/mm) | Maximum Fusion Site Stress (MPa) | Maximum Internal Stress (MPa) |
|---|---|---|---|
| Intact Model | 29.2 ± 2.1 | – | – |
| PLF Model | 56.9 ± 5.2 | 136.8 ± 4.5 | 210.1 ± 8.3 |
| PLIF Model | 63.3 ± 5.5 | 180.7 ± 3.7 | 83.3 ± 7.5 |
| Modified articular fusion model | 53.1 ± 4.6 | 186.1 ± 3.1 | 194.8 ± 10.2 |
Fig. 6Lumbar Intact Model
Fig. 7Stress distribution of PLF Model
Fig. 8Stress distribution of PLIF Model
Fig. 9Stress distribution of MFF Model
Fig. 10Stiffness of different models. *P < 0.05