Early Benchmarking Total Hip Arthroplasty Implants Using Data from the Michigan Arthroplasty Registry Collaborative Quality Initiative (MARCQI).

Heather A Chubb1, Eric R Cornish2, Brian R Hallstrom1, Richard E Hughes1.   

Abstract

BACKGROUND: Benchmarking arthroplasty implant revision risk is an informative way to address implant performance. National benchmarking efforts exist in the United Kingdom, Netherlands, and Australia. Recently, the International Prosthesis Benchmarking Working Group, including representatives from industry, academia, and national registries, produced a guideline describing arthroplasty benchmarking methodology. The proposal was applied to data from the Michigan Arthroplasty Registry Collaborative Quality Initiative (MARCQI) to assess its feasibility for benchmarking implants in the United States.
METHODS: Primary elective total hip arthroplasty procedures performed for osteoarthritis between 2/15/2012 and 12/31/2018 and their associated revisions were identified in the MARCQI registry. The guidelines recommend that all prostheses combinations receive an early benchmark if they have at least 250 procedures at risk and the revision rate does not exceed the pre-determined standard of 2% at 2 years and 3% at 5 years.
RESULTS: A total of 72,949 primary cases met the inclusion criteria. Of these, 1369 had revisions. Twenty-nine and six stem/cup combinations satisfied the minimum case requirement at 2 and 5 years, respectively. Three implant combinations would not receive a benchmark at 2 years: Secur-Fit/Trident, Anthology/Reflection 3, Taperloc 133/G7.
CONCLUSION: The guideline can be implemented in the United States by a regional registry. Moreover, not all hip implants currently in use would receive an early benchmark. This raises concern as these implant combinations represent a significant number of cases in Michigan, some with increasing utilization.
© 2021 Chubb et al.

Entities:  

Keywords:  arthroplasty; benchmarking; hip; implant; revision

Year:  2021        PMID: 34853539      PMCID: PMC8627892          DOI: 10.2147/ORR.S325042

Source DB:  PubMed          Journal:  Orthop Res Rev        ISSN: 1179-1462


Plain Language Summary

Some total hip replacement implants are better than others. An international group has proposed a method for “benchmarking” implants, which means identifying which ones perform well. The Michigan Arthroplasty Registry Collaborative Quality Initiative (MARCQI), which is a state-wide collaborative of hospitals and orthopaedic surgeons dedicated to improving the quality of care for hip and knee replacement patients in Michigan, has applied the proposed method to their data. They found that three implant combinations were not good enough to receive a benchmark based on their data. The results suggest health care quality can be improved by surgeons using implants that receive benchmarks.

Introduction

Elective total hip arthroplasty is a common in the United States, with over 522,00 hip replacements performed in 2014.1 There is wide variation in revision risk between total hip arthroplasty (THA) implants, with arthroplasty registry reports showing a range of 10-year revision risks for cemented implants from 1.03% to 36.2% and from 2.6% to 66.5% for uncemented fixation.2 Voluntary implant product recalls by manufacturers are rare, and the Food and Drug Administration is reluctant to recall implants. It is imperative that arthroplasty registries play a public health role in providing information for surgeons and patients. Internationally, arthroplasty registries seek to reduce the number of revisions in three ways: (1) public reporting through annual reports, (2) identifying outlier implants, (3) implant benchmarking. Benchmarking is a systematic process of determining whether an implant meets specified performance levels.3,4 There are currently three groups performing THA implant benchmarking: (1) Orthopaedic Data Evaluation Panel (ODEP) in the United Kingdom,5,6 (2) Prostheses List Advisory Committee (PLAC) in Australia, (3) Netherlands Orthopaedic Association Classification of Orthopaedic Implants (NOV).7 The International Prosthesis Benchmarking Working Group was established to review current systems and develop a global system proposal to evaluate and benchmark arthroplasty prostheses performance. The working group produced a guidance document in May of 2018 which focused on benchmarking hip and knee implants.8 The statistical subcommittee of the working group analyzed Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) data9 and determined that poor implant performance at two years is predictive of poor performance at ten years.8 Therefore, early benchmarking is extremely important, as devices with inferior performance at two years rarely recover. The purpose of this project was to assess the feasibility of applying the proposal to arthroplasty registry data collected by a regional registry in the United States.

Materials and Methods

Utilizing the methodology detailed in the International Prosthesis Benchmarking Workgroup’s proposal, benchmarking of prostheses in the MARCQI database was performed. All data collected by MARCQI is for the purposes of quality improvement. The Institutional Review Board of the University of Michigan’s Medical School (IRBMED) provided a notice of determination of “not regulated” status for this project because it does not fit the definition of human subject research according to 45 CFR 46 and 21 CFR 56. “Not regulated” status is different than “exempt,” and it reflects that the purpose of the data collection was quality or process improvement. This notice is available upon request. The details of MARCQI’s organizational structure and methodology have been previously described, but in summary, MARCQI collects data on over 97% of all elective total hip and knee arthroplasty cases performed in the state of Michigan.10–12 To qualify for inclusion in the MARCQI registry, a primary case must be elective, defined by a planned procedure, treating a non-emergent condition at the pre-planned surgical date. Hemiarthroplasty cases are excluded, as well as non-elective total arthroplasty cases such as for hip fracture. 91.9% of the cases were done for a diagnosis of osteoarthritis. All total hip and knee replacement revisions are captured and linked to the primary case. The linkage can occur across hospitals in the state of Michigan. Thus, a revision case performed at a different site than the primary is linked to the primary case. This enables MARCQI to conduct analyses based on time-to-revision for primary procedures. Each site has free access to their individual data. MARCQI performs revision risk analyses of implants and publicly reports the results in an annual report which is readily available to the public online.12–14 The dataset used to generate the most recent annual report was used in this benchmarking analysis.14 In addition to demographic and clinical data collected on each case, catalog numbers of every device implanted are captured. These catalog numbers are converted to device descriptors (stem, cup, head, liner, product name, etc.) using a device library made available by Curvo Labs (Evansville, IN). The Curvo Labs library matches 98.5% of all devices used in MARCQI cases. The cup, stem, and product name fields were utilized in the benchmarking analysis. Data for this study specifically was limited to total hip arthroplasty (THA) cases performed between 2/15/2012, and 12/31/2018. Due to the design of the MARCQI registry, all qualifying primary cases were elective, either conventional or conversion, and patients were at least 18 years old. The International Prosthesis Working Group proposed protocol describes benchmarking based only on cases performed for a diagnosis of osteoarthritis. Therefore, for this analysis, inclusion was restricted to a primary diagnosis of osteoarthritis. The clinical endpoint used was all-cause revision. The exclusion criteria were: (1) cases performed before 2/15/2012 or after 12/31/2018, (2) knee procedure, (3) resurfacing THA procedure, (4) diagnosis other than osteoarthritis, or (5) otherwise non-qualifying MARCQI case. The benchmarking guidelines proposed by the International Prosthesis Working Group focus on the 10-year time point. The philosophy of the working group was that implant combinations would receive early benchmarks by default at two and five years unless the device revision risk exceeded a predetermined standard. Thus “early” benchmark procedures at 2 and 5 years were described. MARCQI began collecting data in 2012 and does not yet have data for a 10 year benchmark analysis.8 Therefore, the analysis focused on the early benchmark time points of 2 and 5 years. The proposed methodology was based on Kaplan-Meier estimates of time-to-revision following primary procedure and the associated 95% confidence intervals. The International Prosthesis Benchmarking Working Group document proposed that all prostheses combinations will receive an early benchmark if they have at least 250 procedures at risk and the lower 95% confidence limits of the revision rate does not exceed the proposed benchmark standard of 2% at 2 years and 3% at 5 years. The early benchmarking standard would not be provided if the lower 95% confidence interval exceeds the proposed benchmark standard. The benchmarking criteria were applied to each stem/cup combination separately. Based upon working group recommendations, the criteria were then applied to stems aggregated across cups and cups aggregated across stems. The percentage of primary hip cases performed using implants that failed to receive early benchmarks was computed to provide a population-wide quality measure. This measure was computed using the total number of primary THA cases in the MARCQI database as the denominator. It was also done by year to provide a time trend. The International Prosthesis Benchmarking Working Group proposed that benchmarks be determined regardless of specific patient characteristics. However, the working group recommended that patient characteristics, such as age and gender, be summarized with revision rates. Therefore, prostheses combination revision rates were also evaluated by gender and age groups (less than 65 years and 65 years and older). All statistical analyses were performed in SAS software version 9.4 (SAS Institute Inc., Cary, NC, USA) and mathematical modeling was done using Excel (Microsoft, Redmond, WA, USA).

Results

A total of 72,949 primary cases met the inclusion criteria for benchmarking (Figure 1). Of these, 1,369 had revisions in the database. At 2 years and 5 years respectively, twenty-nine and six stem/cup combinations satisfied the requirement that there be at least 250 at-risk cases (Table 1). Twenty-six individual femoral stem components satisfied the minimum at-risk case threshold at 2 years and seven met it at 5 years (Table 2). Fifteen individual acetabular cup components met the minimum at-risk case requirement at 2 years and seven met it at 5 years (Table 3).
Figure 1

Case flow diagram.

Table 1

Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years

Stem/Cup Combinations2 Year Early Benchmark (2%)5 Year Early Benchmark (3%)
No. at RiskCPR Estimate (95% Confidence Interval)BenchmarkNo. at RiskCPR Estimate (95% Confidence Interval)Benchmark
Tri-Lock BPS/Pinnacle11540.89 (0.54, 1.45)Yes
Tri-Lock BPS/Trident2561.02 (0.33, 3.15)Yes
Fitmore/Trabecular Metal3101.26 (0.47, 3.32)Yes
Citation TMZF/Trident2941.32 (0.50, 3.47)Yes
Summit/Pinnacle35221.45 (1.14, 1.84)Yes7591.88 (1.45, 2.44)Yes
Avenir Muller/Continuum4011.49 (0.71, 3.10)Yes
Taperloc 133 Microplasty/G77341.58 (1.02, 2.45)Yes
Accolade II/Restoration ADM3161.62 (0.73, 3.58)Yes
Accolade II/Trident91861.64 (1.44, 1.86)Yes10512.69 (2.32, 3.12)Yes
Fitmore/Continuum17901.66 (1.22, 2.25)Yes
Corail/Pinnacle11321.67 (1.17, 2.38)Yes
Accolade TMZF/Trident8241.72 (1.04, 2.83)Yes
M/L Taper/Continuum46001.80 (1.49, 2.19)Yes11722.54 (2.12, 3.04)Yes
Secur-Fit Plus Max/Trident14502.04 (1.47, 2.83)Yes4792.66 (1.91, 3.70)Yes
Trabecular Metal/Continuum4722.04 (1.16, 3.56)Yes
Polarstem/Reflection 32822.05 (1.24, 3.36)Yes
M/L Taper/Trilogy10962.07 (1.41, 3.03)Yes3422.65 (1.86, 3.78)Yes
SROM/Pinnacle7002.08 (1.30, 3.33)Yes
AML/Pinnacle2612.09 (1.07, 4.03)Yes
M/L Taper/Trabecular Metal4402.14 (1.25, 3.67)Yes
M/L Taper/G74392.17 (1.42, 3.31)Yes
Taperloc 133/RingLoc+9752.33 (1.60, 3.37)Yes
Secur-Fit Max/Trident12862.42 (1.81, 3.24)Yes3073.24 (2.43, 4.30)Yes
Taperloc 133/G79372.63 (2.06, 3.36)No
Echo Bi-Metric/Regenerex RingLoc+2512.68 (1.29, 5.54)Yes
Synergy/Reflection 35312.79 (1.82, 4.25)Yes
Anthology/Reflection 312702.95 (2.27, 3.82)No
Taperloc 133/Regenerex RingLoc+3482.97 (1.74, 5.07)Yes
Secur-Fit/Trident6304.87 (3.67, 6.46)No
Table 2

Characteristics of Femoral Stem Analysis at 2 and 5 Years

Femoral Stem2 Year Early Benchmark (2%)5 Year Early Benchmark (3%)
No. at RiskCPR Estimate (95% Confidence Interval)BenchmarkNo. at RiskCPR Estimate (95% Confidence Interval)Benchmark
Tri-Lock BPS14150.92 (0.58, 1.44)Yes
Avenir Muller4941.25 (0.65, 2.39)Yes
Citation TMZF2941.32 (0.50, 3.47)Yes
Summit35381.47 (1.16, 1.86)Yes7651.93 (1.49, 2.49)Yes
Trabecular Metal6171.59 (0.94, 2.67)Yes
Natural2991.64 (0.69, 3.91)Yes
Taperloc4131.64 (0.83, 3.26)Yes
Taperloc 133 Microplasty11121.66 (1.16, 2.38)Yes
Corail11341.67 (1.17, 2.38)Yes
Accolade II96151.69 (1.50, 1.91)Yes11092.79 (2.42, 3.21)Yes
Accolade TMZF8241.72 (1.04, 2.83)Yes
Echelon2651.88 (0.85, 4.14)Yes
Fitmore23511.90 (1.49, 2.43)Yes3752.47 (1.95, 3.13)Yes
M/L Taper66651.93 (1.66, 2.24)Yes16292.62 (2.27, 3.03)Yes
Secur-Fit Plus Max14532.02 (1.46, 2.81)Yes4802.64 (1.89, 3.67)Yes
SROM7162.03 (1.27, 3.25)Yes
AML2632.08 (1.07, 4.01)Yes
Polarstem2902.14 (1.32, 3.45)Yes
M/L Taper Kinectiv4572.20 (1.25, 3.86)Yes
Secur-Fit Max12912.41 (1.80, 3.22)Yes3093.22 (2.41, 4.28)Yes
Synergy6732.41 (1.59, 3.65)Yes
Taperloc 13330872.51 (2.12, 2.97)No3253.28 (2.69, 4.00)Yes
Echo Bi-Metric4162.52 (1.64, 3.87)Yes
Anthology14662.74 (2.14, 3.52)No
Versys2613.20 (1.68, 6.06)Yes
Secur-Fit6324.70 (3.54, (6.24)No
Table 3

Characteristics of Acetabular Cup Analysis at 2 and 5 Years

Acetabular Cup2 Year Early Benchmark (2%)5 Year Early Benchmark (3%)
No. at RiskCPR Estimate (95% Confidence Interval)BenchmarkNo. at RiskCPR Estimate (95% Confidence Interval)Benchmark
Ranawat-Burstein4110.48 (0.12, 1.91)Yes
Mallory-Head3880.96 (0.36, 2.54)Yes
Pinnacle70791.41 (1.20, 1.66)Yes11452.01 (1.67, 2.42)Yes
Restoration ADM3231.56 (0.70, 3.46)Yes
Converge4471.74 (0.91, 3.31)Yes
Reflection4251.87 (0.98, 3.56)Yes
Trident14,3291.92 (1.74, 2.11)Yes22852.88 (2.60, 3.20)Yes
Continuum85851.97 (1.73, 2.25)Yes18682.82 (2.49, 3.20)Yes
Trabecular Metal10632.09 (1.43, 3.03)Yes4022.57 (1.80, 3.66)Yes
Trilogy12882.13 (1.51, 3.00)Yes4392.71 (1.96, 3.72)Yes
RingLoc+14432.2 (1.60, 3.02)Yes3072.65 (1.95, 3.60)Yes
G727982.25 (1.93, 2.63)Yes
Regenerex RingLoc+7602.38 (1.56, 3.63)Yes
RingLoc2842.51 (1.34, 4.66)Yes
Reflection 324622.74 (2.27, 3.30)No2603.80 (3.00, 4.81)No
Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years Characteristics of Femoral Stem Analysis at 2 and 5 Years Characteristics of Acetabular Cup Analysis at 2 and 5 Years Case flow diagram. The majority of stem/cup combinations and individual components achieved early benchmarks at the 2- and 5-year time points. At 2 years, twenty-six stem/cup combinations received a benchmark, while three prostheses combinations did not (Figure 2). The total number at risk at 2 years was 35,887, and the number at risk for the 3 combinations that did not receive a benchmark was 2,837. In Figure 2 the vertical dotted line denotes the 2% benchmark criteria at 2 years. Any combination where the lower confidence limit falls to the right of this line does not meet the pre-determined benchmark standard. The three combinations that do not receive an early benchmark (Secur-Fit/Trident, Anthology/Reflection 3, and Taperloc 133/G7) have lower confidence limits of 3.67%, 2.27%, and 2.06%, respectively. All other combinations had 95% confidence intervals whose lower limit was no greater than 2%. At 5 years all stem/cup combinations received a benchmark (Figure 3). The total number of risk at 5 years was 4,111. However, the three combinations that did not receive an early 2-year benchmark were not assessed, as they did not meet the minimum requirement of 250 at-risk cases in the MARCQI registry at the 5-year time point. The analysis of stem components in isolation at 5 years showed all would receive a benchmark (Figure 4), and there were 4,992 at risk at 5 years. Only six of the seven acetabular cup components aggregated across stems would receive a benchmark at 5 years (Figure 5). There were 6,706 cases at risk at 5 years for the cup analysis.
Figure 2

Benchmarking stem/cup combinations at 2-year time point.

Figure 3

Benchmarking stem/cup combinations at 5-year time point.

Figure 4

Benchmarking femoral stems at 5-year time point.

Figure 5

Benchmarking acetabular cups at 5-year time point.

Benchmarking stem/cup combinations at 2-year time point. Benchmarking stem/cup combinations at 5-year time point. Benchmarking femoral stems at 5-year time point. Benchmarking acetabular cups at 5-year time point. Specific age and gender requirements are not given for conventional hip replacement; however, benchmarking may have clinical indications following appropriate stratification. Revision rates with 95% confidence intervals, stratified by gender and age group for prostheses combinations, provide additional information about the performance of an implant. Applying the 2% pre-determined benchmark criteria at the 2-year time point, three stem/cup combinations perform better in one gender group and one combination does not perform well in males or females (Table 4). Likewise, five stem/cup combinations perform better in one age group and one does not perform well in either age group, below 65 years or 65 years and above (Table 5).
Table 4

Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years by Gender

Device CombinationsGender2 Year Time Point5 Year Time Point
No. at RiskCPR Estimate (95% Conf. Interval)No. at RiskCPR Estimate (95% Conf. Interval)
AML/PinnacleFemale1542.06 (0.84, 4.98)
Male1072.11 (0.78, 5.62)
Accolade II/Restoration ADMFemale1610.57 (0.08, 4.01)
Male1552.74 (1.14, 6.48)
Accolade II/TridentFemale50321.68 (1.41, 1.98)5862.80 (2.33, 3.37)
Male41461.59 (1.31, 1.94)4642.56 (2.01, 3.26)
Accolade TMZF/TridentFemale4562.06 (1.11, 3.79)
Male3681.29 (0.54, 3.07)
Anthology/Reflection 3Female6832.61 (1.80, 3.77)
Male5873.35 (2.32, 4.84)
Avenir Muller/ContinuumFemale2251.85 (0.77, 4.39)
Male1761.00 (0.25, 3.94)
Citation TMZF/TridentFemale1541.89 (0.61, 5.74)
Male1400.68 (0.10, 4.76)
Corail/PinnacleFemale6661.56 (0.96, 2.52)
Male4661.85 (1.10, 3.13)
Echo Bi-Metric/Regenerex RingLoc+Female1432.03 (0.66, 6.15)
Male1083.54 (1.34, 9.16)
Fitmore/ContinuumFemale8691.93 (1.26, 2.94)
Male9211.42 (0.91, 2.23)
Fitmore/Trabecular MetalFemale1781.10 (0.28, 4.31)
Male1321.48 (0.37, 5.79)
M/L Taper/ContinuumFemale24792.00 (1.56, 2.56)5952.79 (2.20, 3.53)
Male21211.58 (1.16, 2.14)5772.25 (1.69, 2.97)
M/L Taper/G7Female2522.15 (1.29, 3.57)
Male1872.26 (1.11, 4.57)
M/L Taper/Trabecular MetalFemale2362.71 (1.41, 5.16)
Male2041.45 (0.54, 3.83)
M/L Taper/TrilogyFemale5702.56 (1.60, 4.09)1853.16 (2.04, 4.87)
Male5261.52 (0.79, 2.90)1572.09 (1.13, 3.84)
Polarstem/Reflection 3Female1312.86 (1.46, 5.55)
Male1511.45 (0.69, 3.04)
SROM/PinnacleFemale3611.96 (0.98, 3.88)
Male3392.21 (1.16, 4.21)
Secur-Fit/TridentFemale3865.42 (3.82, 7.67)
Male2434.11 (2.53, 6.64)
Secur-Fit Max/TridentFemale6622.34 (1.54, 3.56)1743.55 (2.41, 5.22)
Male6242.51 (1.68, 3.73)1332.85 (1.87, 4.35)
Secur-Fit Plus Max/TridentFemale7331.52 (0.89, 2.61)2532.06 (1.18, 3.58)
Male7172.54 (1.68, 3.84)2263.24 (2.15, 4.88)
Summit/PinnacleFemale20061.75 (1.32, 2.33)4102.52 (1.83, 3.46)
Male15151.05 (0.68, 1.60)3491.05 (0.68, 1.60)
Synergy/Reflection 3Female2863.50 (2.08, 5.86)
Male2451.94 (0.93, 4.03)
Taperloc 133/G7Female5633.07 (2.26, 4.16)
Male3742.00 (1.33, 2.99)
Taperloc 133/Regenerex RingLoc+Female1893.37 (1.70, 6.63)
Male1592.50 (1.05, 5.91)
Taperloc 133/RingLoc+Female5732.54 (1.59, 4.06)
Male3962.05 (1.11, 3.78)
Taperloc 133 Microplasty/G7Female3262.03 (1.16, 3.53)
Male4061.19 (0.58, 2.44)
Trabecular Metal/ContinuumFemale2673.32 (1.85, 5.91)
Male2050.38 (0.05, 2.70)
Tri-Lock BPS/PinnacleFemale6621.17 (0.66, 2.06)
Male4910.53 (0.20, 1.42)
Tri-Lock BPS/TridentFemale2190.80 (0.20, 3.19)
Male372.50 (0.36, 16.45)

Note: Bold items do not meet pre-detemined standards.

Table 5

Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years by Age Group.

Device CombinationsAge Group2 Year Time Point5 Year Time Point
No. at RiskCPR Estimate (95% Conf Interval)No. at RiskCPR Estimate (95% Conf Interval)
AML/PinnacleBelow 65 years1080.49 (0.07, 3.40)
65 years or Above1533.29 (1.63, 6.57)
Accolade II/Restoration ADMBelow 65 years1532.19 (0.82, 5.73)
65 years or Above1631.07 (0.26, 4.24)
Accolade II/TridentBelow 65 years45471.45 (1.19, 1.77)5212.58 (2.08, 3.19)
65 years or Above46391.81 (1.53, 2.13)5302.79 (2.27, 3.42)
Accolade TMZF/TridentBelow 65 years4401.08 (0.45, 2.58)
65 years or Above3842.43 (1.31, 4.47)
Anthology/Reflection 3Below 65 years6022.65 (1.76, 3.98)
65 years or Above6683.20 (2.28, 4.49)
Avenir Muller/ContinuumBelow 65 years1771.97 (0.74, 5.17)
65 years or Above2241.12 (0.36, 3.42)
Citation TMZF/TridentBelow 65 years1272.27 (0.74, 6.86)
65 years or Above1670.58 (0.08, 4.05)
Corail/PinnacleBelow 65 years4991.54 (0.87, 2.70)
65 years or Above6331.78 (1.13, 2.80)
Echo Bi-Metric/Regenerex RingLoc+Below 65 years1092.61 (0.85, 7.87)
65 years or Above1422.74 (1.04, 7.13)
Fitmore/ContinuumBelow 65 years9381.58 (1.02, 2.44)
65 years or Above8521.74 (1.13, 2.69)
Fitmore/Trabecular MetalBelow 65 years1160.83 (0.12, 5.77)
65 years or Above1941.51 (0.49, 4.61)
M/L Taper/ContinuumBelow 65 years22021.72 (1.28, 2.29)5512.31 (1.76, 3.04)
65 years or Above23981.88 (1.45, 2.43)6212.74 (2.15, 3.48)
M/L Taper/G7Below 65 years1792.23 (1.17, 4.24)
65 years or Above2602.11 (1.21, 3.67)
M/L Taper/Trabecular MetalBelow 65 years1951.90 (0.86, 4.18)
65 years or Above2452.34 (1.12, 4.85)
M/L Taper/TrilogyBelow 65 years4232.11 (1.14, 3.89)1362.54 (1.38, 4.64)
65 years or Above6732.04 (1.25, 3.31)2062.70 (1.74, 4.17)
Polarstem/Reflection 3Below 65 years1410.76 (0.25, 2.37)
65 years or Above1413.29 (1.89, 5.69)
SROM/PinnacleBelow 65 years4071.89 (0.99, 3.61)
65 years or Above2932.32 (1.17, 4.60)
Secur-Fit/TridentBelow 65 years2954.88 (3.20, 7.41)
65 years or Above3354.86 (3.31, 7.11)
Secur-Fit Max/TridentBelow 65 years5711.89 (1.16, 3.07)1382.75 (1.68, 4.48)
65 years or Above7152.86 (2.00, 4.09)1693.64 (2.56, 5.16)
Secur-Fit Plus Max/TridentBelow 65 years7731.60 (0.95, 2.68)2701.89 (1.16, 3.08)
65 years or Above6772.50 (1.64, 3.81)2093.53 (2.25, 5.51)
Summit/PinnacleBelow 65 years15671.36 (0.94, 1.97)3372.02 (1.31, 3.12)
65 years or Above19551.52 (1.12, 2.07)4221.77 (1.30, 2.42)
Synergy/Reflection 3Below 65 years2002.14 (0.96, 4.73)
65 years or Above3313.18 (1.92, 5.22)
Taperloc 133/G7Below 65 years4892.15 (1.46, 3.16)
65 years or Above4483.11 (2.26, 4.27)
Taperloc 133/Regenerex RingLoc+Below 65 years1883.95 (2.07, 7.45)
65 years or Above1601.88 (0.71, 4.92)
Taperloc 133/RingLoc+Below 65 years4271.19 (0.53, 2.62)
65 years or Above5483.19 (2.09, 4.86)
Taperloc 133 Microplasty/G7Below 65 years4291.28 (0.67, 2.41)
65 years or Above3052.02 (1.10, 3.71)
Trabecular Metal/ContinuumBelow 65 years1962.53 (1.14, 5.55)
65 years or Above2761.71 (0.77, 3.77)
Tri-Lock BPS/PinnacleBelow 65 years5350.39 (0.12, 1.24)
65 years or Above6191.29 (0.75, 2.22)
Tri-Lock BPS/TridentBelow 65 years850.00 (0.00, 0.00)
65 years or Above1711.50 (0.48, 4.61)

Note: Bold items do not meet pre-detemined standards.

Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years by Gender Note: Bold items do not meet pre-detemined standards. Characteristics of Stem/Cup Combination Analysis at 2 and 5 Years by Age Group. Note: Bold items do not meet pre-detemined standards. The proportion of cases in Michigan utilizing implant combinations which did not receive a 2-year benchmark was 8.6% of primary THA cases from 2/15/2012 through 12/31/2018. Moreover, some combinations show an increasing utilization trend over time (Figure 6).
Figure 6

Percent of MARCQI total hip arthroplasty cases using implant combinations that would not receive an early (2 year) benchmark over time.

Percent of MARCQI total hip arthroplasty cases using implant combinations that would not receive an early (2 year) benchmark over time.

Discussion

The purpose of this project was to assess the feasibility of applying the implant benchmarking methodology developed by the International Prosthesis Benchmarking Working Group to a regional arthroplasty registry in the United States. The result was that there were sufficient numbers of implants in the MARCQI registry to conduct benchmarking at the early time points (2 and 5 years), but the registry has not been in existence long enough to conduct a later assessment (10 year). While the majority of implants received a benchmark, some did not. MARCQI’s application of the proposed benchmarking methodology revealed that 8.6% of primary THA cases captured by MARCQI across the state of Michigan were done with an implant combination that would not receive an early benchmark. The rising use of these non-benchmarked implants may increase the risk of revision among patients and merits continued surveillance. It is important to note that one limitation of benchmarking is the difficulty to detect early impact of small changes in a prosthesis until a sufficient number of cases (250) are performed. There is ongoing debate to “lump” similar prostheses together for larger numbers and statistical significance, or “split” prostheses with minor changes into smaller groups for analysis which spreads out the time to achieve statistical significance. At this time, there are no established guidelines to categorize a new change as significantly different to “lump” or “split.” Splitting may have some benefit in the interest of promoting innovation. An additional limitation of the early benchmarking methodology proposed by the International Prosthesis Benchmarking Working Group is that benchmarks are based on a non-inferiority analytical framework rather than superiority. In simplistic terms, a superiority analysis requires that the upper end of a 95% confidence interval be less than a pre-specified threshold. In a non-inferiority analysis, a margin is added to the threshold to obtain a new non-inferiority threshold. Non-inferiority is determined if the upper end of the 95% confidence interval is no greater than the non-inferiority threshold. Applying a clinically accepted non-inferiority margin of 20% to the pre-determined criteria of 2% at 2 years sets the non-inferiority threshold at 2.4%. A non-inferiority analysis finds that of the three combinations that would not receive an early benchmark, Secur-Fit/Trident is classified as inferior, but the evidence against Anthology/Reflection 3 and Taperloc 133/G7 is inconclusive (Figure 7).
Figure 7

Non-inferiority analysis at 2-year time point.

Non-inferiority analysis at 2-year time point. The working group proposed a superiority approach at 10 years, which is a more definitive statement that an implant performs well. In contrast, the group’s proposal for earlier benchmarks gives a benchmark by default, and it is only withheld if the implant proves to be inferior with respect to the 2- and 5-year pre-determined criteria of 2% and 3%, respectively. This approach may allow a mediocre product to initially be portrayed as an acceptable product. Differences between the two approaches at the early time points and the 10 year time point appear to be a compromise between the competing interests of innovation and public health. Another obvious limitation of this work arises from the structure of MARCQI, which is limited to the state of Michigan. MARCQI does receive full abstraction on over 97% of all primary and revision total hips in the state and performs audits to ensure that all primary and revision surgeries are captured at each site. While MARCQI identifies revision surgeries that occur in the state, it has no mechanism for finding revision cases performed outside Michigan. However, Etkin et al15 reported that only 4.1% of patients having primary THA or TKA migrate out of Michigan within 5 years based on Medicare claims data between 2004 and 2016. While this only represents the over 65 year-old population, it suggests a low fraction of patients would be lost due to the inability to follow-up outside the state. Despite its limitations, the International Prosthesis Benchmarking Working Group proposal has major strengths. Among these strengths is the belief that the preferred data source for benchmarking is accurate and complete registry data. The combination of data from multiple sites in a registry environment allows benchmarking to be based on statistically significant numbers. This is an advantage over analysis from the scientific literature where studies generally have small numbers, and those from the developers of an implant have better outcomes than demonstrated in national registry data.16,17 An additional strength of the proposal is that it was developed by a broad group of stakeholders around the globe. The adoption of a global methodology for benchmarking would serve to make benchmarks more transparent to payers, hospitals, surgeon, regulators, and patients. A single accepted methodology would also benefit implant manufacturers by reducing the cost of preparing and submitting data for benchmarking organizations and regulatory bodies. Such efficiencies would be advanced if additional methodology were developed to aggregate data from multiple registries into a single world-wide benchmark. However, accomplishing this would require adapting the benchmarking proposal to include sound meta-analysis methods for analyzing data from multiple sources. While sponsors, medical device manufacturers, registries and organizations currently involved in benchmarking were the intended audience, the group recognized their proposal would receive interest from additional stakeholders with the potential to be broadened for consideration in other joints as well. MARCQI’s application of the benchmarking proposal reflects community use performance in real-world settings and hopes to strengthen the arthroplasty and scientific communities in registry involvement. It is important to differentiate benchmarking from implant outlier detection. The two processes use different analytics, thresholds, and metrics.18 The most important difference may possibly lie in whether outlier surgeons and hospitals are analyzed. The benchmarking process does not control for confounding at the site or surgeon level. It is based on the analysis of the cumulative percent revision and number at risk at each benchmarking time point. The outlier detection model is based on component time incidence rate and allows registries to develop a standardized process in which to identify outliers and determine possible reasons for any difference, including device and non-device concerns. This opens the possibility that poor performance indicated by not receiving a benchmark at two or five years could be due to the implant performing poorly in the hands of only a few surgeons or a few sites. An additional difference between early benchmarking and outlier detection is that early benchmarking uses a non-inferiority analysis and outlier detection seeks to determine inferiority. Investigating the gap between benchmarking and outlier detection might prove useful for future implant performance detection.

Conclusion

The International Prosthesis Benchmarking Working Group protocol for benchmarking THA implants was found to be applicable to a regional arthroplasty registry in the United States. We found three implant combinations that did not perform sufficiently well to receive a benchmark at 2 years. Due to the fact that MARCQI is a young registry, we did not have sufficient numbers at risk at 5 years to conduct a benchmark assessment of these combinations. 8.6% of MARCQI cases were done with implant combinations that did not receive a 2-year benchmark. Moreover, the number of cases done with these non-benchmarked implant combinations is increasing over time in the state of Michigan. This presents a significant opportunity for quality improvement.
  9 in total

1.  What Are the Migration Patterns for U.S. Primary Total Joint Arthroplasty Patients?

Authors:  Caryn D Etkin; Edmund C Lau; Heather N Watson; Steven M Kurtz; Terrence J Gioe; Bryan D Springer; David G Lewallen; Kevin J Bozic
Journal:  Clin Orthop Relat Res       Date:  2019-06       Impact factor: 4.176

Review 2.  Arthroplasty registries around the world: valuable sources of hip implant revision risk data.

Authors:  Richard E Hughes; Aditi Batra; Brian R Hallstrom
Journal:  Curr Rev Musculoskelet Med       Date:  2017-06

3.  The Michigan Arthroplasty Registry Collaborative Quality Initiative Experience: Improving the Quality of Care in Michigan.

Authors:  Richard E Hughes; Huiyong Zheng; Rochelle M Igrisan; Mark E Cowen; David C Markel; Brian R Hallstrom
Journal:  J Bone Joint Surg Am       Date:  2018-11-21       Impact factor: 5.284

4.  Finding the right hip implant for patient and surgeon: the Dutch strategy--empowering patients.

Authors:  Rudolf W Poolman; Jan A N Verhaar; B Willem Schreurs; L Paul A Bom; Rob G H H Nelissen; Henk W J Koot; Jon H M Goosen; Cees C P M Verheyen
Journal:  Hip Int       Date:  2015-02-28       Impact factor: 2.135

Review 5.  Outcome and reproducibility of data concerning the Oxford unicompartmental knee arthroplasty: a structured literature review including arthroplasty registry data.

Authors:  Gerold Labek; Kathrin Sekyra; Wolfram Pawelka; Wolfgang Janda; Bernd Stöckl
Journal:  Acta Orthop       Date:  2011-03-24       Impact factor: 3.717

Review 6.  Outcome of the cementless Taperloc stem: a comprehensive literature review including arthroplasty register data.

Authors:  Gerold Labek; Stefan Frischhut; Rainer Schlichtherle; Alexandra Williams; Martin Thaler
Journal:  Acta Orthop       Date:  2011-04       Impact factor: 3.717

7.  Joint registry approach for identification of outlier prostheses.

Authors:  Richard N de Steiger; Lisa N Miller; David C Davidson; Philip Ryan; Stephen E Graves
Journal:  Acta Orthop       Date:  2013-08       Impact factor: 3.717

8.  Assessing the non-inferiority of prosthesis constructs used in total and unicondylar knee replacements using data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man: a benchmarking study.

Authors:  Kevin C Deere; Michael R Whitehouse; Martyn Porter; Ashley W Blom; Adrian Sayers
Journal:  BMJ Open       Date:  2019-04-29       Impact factor: 2.692

9.  Determining the sample size required to establish whether a medical device is non-inferior to an external benchmark.

Authors:  Adrian Sayers; Michael J Crowther; Andrew Judge; Michael R Whitehouse; Ashley W Blom
Journal:  BMJ Open       Date:  2017-08-28       Impact factor: 2.692

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.