Merlin Behling1, Felix Wezel2, Peter P Pott1. 1. Institute of Medical Device Technology, University of Stuttgart, Stuttgart, Germany. 2. Department of Urology and Pediatric Urology, University of Ulm, Ulm, Germany.
Abstract
Detection of metastasis spread at an early stage of disease in lymph nodes can be achieved by imaging techniques, such as PET and fluoride-marked tumor cells. Intraoperative detection of small metastasis can be problematic especially in minimally invasive surgical settings. A γ-radiation sensor can be inserted in the situs to facilitate intraoperative localization of the lymph nodes. In the minimally invasive setting, the sensor must fit through the trocar and for robot-aided interventions, a small, capsule-like device is favorable. Size reduction could be achieved by using only a few simple electronic parts packed in a single-use sensor-head also leading to a low-cost device. This paper first describes the selection of an appropriate low-cost diode, which is placed in a sensor head (Ø 12 mm) and characterized in a validation experiment. Finally, the sensor and its performance during a detection experiment with nine subjects is evaluated. The subjects had to locate a 137Cs source (138 kBq activity, 612 keV) below a wooden plate seven times. Time to accomplish this task and error rate were recorded and evaluated. The time needed by the subjects to complete each run was 95 ± 68.1 s for the first trial down to 40 ± 23.9 s for the last. All subjects managed to locate the 137Cs source precisely. Further reduction in size and a sterilizable housing are prerequisites for in vitro tests on explanted human lymph nodes and finally in vivo testing.
Detection of metastasis spread at an early stage of disease in lymph nodes can be achieved by imaging techniques, such as PET and fluoride-marked tumor cells. Intraoperative detection of small metastasis can be problematic especially in minimally invasive surgical settings. A γ-radiation sensor can be inserted in the situs to facilitate intraoperative localization of the lymph nodes. In the minimally invasive setting, the sensor must fit through the trocar and for robot-aided interventions, a small, capsule-like device is favorable. Size reduction could be achieved by using only a few simple electronic parts packed in a single-use sensor-head also leading to a low-cost device. This paper first describes the selection of an appropriate low-cost diode, which is placed in a sensor head (Ø 12 mm) and characterized in a validation experiment. Finally, the sensor and its performance during a detection experiment with nine subjects is evaluated. The subjects had to locate a 137Cs source (138 kBq activity, 612 keV) below a wooden plate seven times. Time to accomplish this task and error rate were recorded and evaluated. The time needed by the subjects to complete each run was 95 ± 68.1 s for the first trial down to 40 ± 23.9 s for the last. All subjects managed to locate the 137Cs source precisely. Further reduction in size and a sterilizable housing are prerequisites for in vitro tests on explanted human lymph nodes and finally in vivo testing.
Entities:
Keywords:
Experimental evaluation; gamma detector; low-cost; miniaturization; radioguided surgery
Metastatic spread is a hallmark of malignant and deadly cancers. Under certain
circumstances, metastasis-directed therapy by, that is, surgical resection or
radiotherapy may be life prolonging or even curative in oligometastatic disease,
when resection of metastatic lesions can be achieved.
Detection of small metastasis at an early stage of disease, that is, in lymph
nodes can be achieved by modern imaging techniques, such as positron emission
tomography (PET), usually combined with computed tomography (PET/CT) or magnetic
resonance imaging (PET/MRI). For diagnostic PET, tumor cells can be marked with
fluoride 18F or other radiopharmacons attached to ligands that are
characteristic for tumor cell metabolism, such as fluorodeoxyglucose (FDG), or
ligands that specifically bind to tumor cells, that is, prostate specific membrane
antigen (PSMA) to detect prostate cancer cells.
By using such imaging modalities, even small tumor cell clusters can be
detected in the human body.However, intraoperative detection of small tumor lesions can be problematic,
especially in the case of small or atypically located lesions, which cannot be
visualized. This is further complicated in minimally invasive surgery, where direct
manual access is not possible.In recent years, salvage surgery for oligo-metastatic prostate cancer has come more
and more into focus especially due to advances in PSMA-based PET. It was shown that
intraoperative detection of cancer cells labeled with radiopharmaceutical 99mTc
attached to PSMA by using a γ-radiation sensor was feasible. Moreover, such
radioguided surgery could improve the prognosis in some patients. In early clinical
experiences, the removal of the lymph node bearing tumor cells was confirmed outside
the body.For prostate cancer, robotic-assisted laparoscopic radical prostatectomy has been
established as minimal invasive standard procedure.
In such a minimally invasive setting, a γ-radiation sensor that can be
inserted in the situs should make it easier to locate the lymph nodes
intraoperatively. In order to achieve this, the sensor must be inserted through a
conventional trocar and guided over the tissue surfaces to be examined with the aid
of rigid hand-held laparoscopic instruments or end effectors guided by robotic
manipulators. This application requires sterilizable devices as well as dimensions
that allow insertion through the trocar.The state of the art comprises handheld probes (Figure 1(a)) and gamma cameras (Figure 1(b)). Such a probe
can consist of a handle and a shaft,[6,7] which is guided through the
trocar. The sensitive sensor surface is located at the end of the shaft. The
directional sensitivity for gamma quants is achieved by using apertures, for
example, made of tungsten, which provide a relatively good shielding. In the
standard setting, the actual sensor consists of a thallium-doped sodium iodide –
NaI(Tl) scintillation crystal, whose photon emission (wavelength range 325–550 nm,
maximum emission at 415 nm) is read out by a positive intrinsic negative (PIN)
photodiode. The probe acts as a single-point detector and delivers no information on
spatial distribution of γ-radiation. According to the manufacturer’s specifications,
the sensitivity is 23 kcount/s/MBq in a gamma energy range from 60 to
511 keV. The background noise is 0.3 counts/s.
Figure 1.
(a) Schematic of a gamma quant detecting probe. (1) Scintillator crystal
converting an incoming gamma quant into a photon, (2) photo diode detecting
the photon, (3) tungsten housing. (b) Schematic of a gamma camera. The
incoming gamma quant passes the tungsten collimator layer (3) and is
converted to a photon in the scintillator crystal layer (1) and then finally
detected by one of the photo diodes in the detector layer (2) doing so, a
spatial encoding of incoming gamma quants is achieved.
(a) Schematic of a gamma quant detecting probe. (1) Scintillator crystal
converting an incoming gamma quant into a photon, (2) photo diode detecting
the photon, (3) tungsten housing. (b) Schematic of a gamma camera. The
incoming gamma quant passes the tungsten collimator layer (3) and is
converted to a photon in the scintillator crystal layer (1) and then finally
detected by one of the photo diodes in the detector layer (2) doing so, a
spatial encoding of incoming gamma quants is achieved.Gamma cameras (Figure 1(b))
consist of a flat detector array of scintillation crystals and photodiodes and a
collimator arranged in front of it. Geometry and material are matched to the isotope
to be detected. The resulting image (“scintigraphy”) encodes the local distribution
of the radionuclide, for example, in an organ. Such systems have been available for
many years as permanent installations
and as hand-held devices.[9,10] In contrast to an optical
camera there is no lens focusing an image, thus the sensing area must be brought to
the organ as close as possible.The described devices are not suitable to be used by minimally invasive robotic
surgical systems as their size does not allow robotic manipulation. It is thus
necessary to provide a small, capsule-like device that fits through the trocar and
can be navigated while being gripped by the robot’s end effector, which is important
for probe positioning.
The development of a reusable drop-in gamma probe for robotic manipulation
has been recently described by Dell’Oglio et al.
Size reduction could be achieved by using only a few simple electronic parts
compiled in a single-use sensor head.This paper first describes the selection of an appropriate diode. Requirements of a
suitable diode are high sensitivity (pulses per time interval at a given radiation
density), high signal strength for each pulse – even without the implementation of a
scintillator crystal, low dark current at 37°C, low capacity, small outlines, and
low price to allow compact and low-cost devices. Secondly, during a validation
experiment, the selected diode is characterized considering temperature, angle
between diode and radiation source, and the influence of the distance between
radiation source and diode. Thirdly, the first prototype of such a sensor and its
performance during a detection experiment is presented. Here, the hypothesis is that
the frequency of the occurrence of a sound impulse is a measure for the local
γ-radiation intensity and thus enables the surgeon to locate a lymph node labeled by
a radioactive tracer in an (in vitro mimicked) site using the presented sensor.
System design
To achieve the goal of small outlines, a detecting principle using PIN photo diodes
was chosen.
A PIN photodiode is similar to a regular (PN) photo diode, but with an
intrinsic layer between the p-doped and n-doped layers. This layer is intrinsically
conductive, and consists of pure or only very lightly doped silicon. The advantage
of this additional intermediate layer becomes apparent when a reverse voltage is
applied. The depletion region becomes significantly larger compared to a regular
diode. This has a positive effect on the detection probability and also reduces the
capacity of the diode, which limits the detection frequency. In case a gamma quant
interacts with the material in this region, a pair of charge carriers is generated,
which again result in a current pulse. This pulse can be conditioned by an amplifier
circuit. It has to be considered, that a PIN photo diode is basically detecting
light. For the intended purpose it has thus to be covered by an opaque layer. As it
is also detecting α–and β–radiation it has to be shielded against this by a layer of
steel.The amplifier circuit has two stages to improve the signal-to-noise ratio (SNR) and
at the same time to realize the smallest possible measuring head. The first stage
located directly next to the PIN diode amplifies the current pulse after gamma quant
detection to a millivolts-level voltage that can be sent to the next stage without
too much losses. This also improves SNR. The second stage then amplifies the signal
to a level that can be read out by the micro controller’s analog input (∼3 V). The
principle circuit is shown in Figure 2.
Figure 2.
Basic amplifier circuit. Primary stage of the amplifier circuit located
directly near to the sensor photo diode (left): (a) primary amplifier stage.
Main amplifying circuit located next to the electronics for evaluating the
signal (right): (b) secondary amplifier stage. The dashed lines indicate the
cable between sensor head and main amplifier circuit.
Basic amplifier circuit. Primary stage of the amplifier circuit located
directly near to the sensor photo diode (left): (a) primary amplifier stage.
Main amplifying circuit located next to the electronics for evaluating the
signal (right): (b) secondary amplifier stage. The dashed lines indicate the
cable between sensor head and main amplifier circuit.The sensor head is connected to the extracorporeally arranged control unit by means
of a tension-proof robust shielded cable. The sensor housing can be held via the
cable and manipulated by any surgical or endoscopic grasping forceps in the situs.
This is to ensure easy navigation of the sensor head while keeping the outlines as
small as possible. With the help of the cable the sensor head can be retrieved from
the situs at any time.The second stage of the amplifier circuit (Figure 2(b)) is set up from two OP amps with
high input impedance. They amplify the voltage of the first stage to be read out by
a microcontroller (Arduino™ MEGA). One single sample readout needs 120 µs on average
(8.3 kHz sampling rate, see Figure
3(3)). After detection of a pulse (Figure 3(1)) the algorithm waits 556 µs
(without sampling) for the signal to decay. If after this time the input voltage is
below the threshold (Figure
3(4)) a next pulse can be counted. If during a processing period another
pulse is measured, the signal will be above the threshold at the end. In this case a
second pulse will be counted and a new processing period is triggered. A third or
higher pulse (Figure 3(5))
would be discarded. For a pulse to be counted it has to be 5% above (see Figure 3(4)) dark current
level (see Figure 3(2)).
The processing period has a duration of 556 µs on average. Thus, a maximum number of
pulses per second can be detected by the microcontroller.
Accordingly, the sensor saturates at values above 1479 pulses per second.
Figure 3.
(a) Exemplary signal at the microcontroller input. (1) Amplified pulse after
gamma quant detection, (2) voltage derived from dark current of the photo
diode, (3) one sampling period (120 µs), (4) threshold for pulse detection,
(5) multiple pulses within one processing period (556 µs). (b) Algorithm
running on the microcontroller (“t” abbreviates threshold).
(a) Exemplary signal at the microcontroller input. (1) Amplified pulse after
gamma quant detection, (2) voltage derived from dark current of the photo
diode, (3) one sampling period (120 µs), (4) threshold for pulse detection,
(5) multiple pulses within one processing period (556 µs). (b) Algorithm
running on the microcontroller (“t” abbreviates threshold).For each detected pulse the microcontroller produces a sound of a frequency of 1 kHz
lasting approx. 25 ms, which are made audible to the operator via a loudspeaker. The
frequency of the sound appearance per time period allows a quantitative assessment
of the radiation intensity depending on the location of the measurement. The maximum
rate of pulses is
(continuous sound).
Preliminary experiment
Aim of the preliminary experiment was to identify suitable photodiodes for the
sensor. A good photodiode produces a large number of clearly detectable pulses
at a given rate of gamma quants and specific energy. Dark current and capacity
should both be low for clear and fast pulse detection. For the pre-test, the
circuit for pre-amplification and the actual sensor are accommodated in a larger
aluminum housing (see Figure
4(a)). The sensitive surface of the diodes (Table 1) is arranged perpendicular to
the axis of the housing at its distal end. In front of the sensitive surface of
the sensor, the aluminum housing (thickness 1 mm) prevents the penetration of
α–radiation and blocks large amounts of the β–radiation. It should be noted that
the measuring principle requires that radiation of a source behind the sensor
surface is also detected.
Figure 4.
Set up of the electronics for preliminary tests of diodes: (a) shows the
prototypic sensor head ((1) diode under test, (2) temperature sensor,
(3) shielded housing, (4) primary stage of the amplifier, and (5)
electric heating to 37°C). (b) Shows main amplifying circuit ((1)
printed circuit board, (2) cables to micro controller, and (3) shielded
housing).
Table 1.
Diodes under test. Values apply for capacity and dark current at 20 V
reverse voltage.
Type
Manufacturer
Cost (Δ)
Active area (mm2)
Capacity (pF)
Dark current (nA)
BPW34
OSRAM GmbH, Munich, GER
1
7.02
12
3.8
S1223-01
Hamamatsu Photonics K.K, Naka-ku, JP
10
13
20
0.2
FDS100A
Thorlabs Inc, Newton, NJ, USA
15
13
24
1
PS13-5B
First Sensor AG, Berlin, GER
35
13
72.5
0.07
Set up of the electronics for preliminary tests of diodes: (a) shows the
prototypic sensor head ((1) diode under test, (2) temperature sensor,
(3) shielded housing, (4) primary stage of the amplifier, and (5)
electric heating to 37°C). (b) Shows main amplifying circuit ((1)
printed circuit board, (2) cables to micro controller, and (3) shielded
housing).Diodes under test. Values apply for capacity and dark current at 20 V
reverse voltage.
Validation experiment
Aim of the validation experiment was to prove, that the selected diode (BPW34) is
suitable to detect γ-radiation emitted by a 137Cs source. To derive
an estimate of the sensitivity, temperature dependency of the signal, and the
dependency of the signal with regard to the distance and angular displacement of
the selected diode, the system shown in Figure 5 was used.
Figure 5.
Complete system for preliminary testing of the diodes. (1) Housing end
with diode (see Figure
4(a))), (2) shielded housing for amplifying electronics, (3)
shielded cable, (4) microcontroller for signal processing, and (5) DC
power supply (battery, 9 V). Not shown: heating control circuit and
power supply.
Complete system for preliminary testing of the diodes. (1) Housing end
with diode (see Figure
4(a))), (2) shielded housing for amplifying electronics, (3)
shielded cable, (4) microcontroller for signal processing, and (5) DC
power supply (battery, 9 V). Not shown: heating control circuit and
power supply.
Detection experiment
The pre-clinical applicable system (see Figure 6) consists of a cylindrical
stainless steel tube of 12 mm diameter and 25 mm length, which forms the sensor
head (see Figure 6),
blocks α– and β–radiation, and allows basic hygiene measures. This contains the
actual sensor diodes and the primary stage amplifier for signal
pre-amplification (see Figure
4(a)).
Figure 6.
Sensor head including primary stage of the amplifier: (a) Photo of the
device (Ø 12 × 25 mm), (b and c) internal view of the circuit. (1)
stainless steel housing of the photo diode array and the primary stage
of the amplifier circuit, (2) printed circuit of the primary stage of
the amplifier circuit, (3) shielded cable, and (4) photo diode
array.
Sensor head including primary stage of the amplifier: (a) Photo of the
device (Ø 12 × 25 mm), (b and c) internal view of the circuit. (1)
stainless steel housing of the photo diode array and the primary stage
of the amplifier circuit, (2) printed circuit of the primary stage of
the amplifier circuit, (3) shielded cable, and (4) photo diode
array.
Material and methods
To derive a test signal, a conventional incandescent gas mantle was used. This
contains Thorium(IV)-oxide and therefore through natural decay 212Pb,
which both emit γ-radiation. 212Pb is dominating the spectrum with a
gamma quant energy level of 240 keV. The gas mantle was folded to a size of
cm2 and kept in a transparent polyethylene bag.For the experiment, the plastic bag was first attached directly to the housing
leading to a minimum distance of 1 mm between gamma source and detecting diode,
and secondly mounted at a distance of 18 mm. The latter is expected to be the
practical maximum detection distance during surgery. For each diode and
distance, a measuring interval of 1 min was chosen and detection events were
counted. Ten consecutive repetitions of the measurements were performed.As the dark current of the photo diodes is strongly affected by temperature, the
primary amplifier housing temperature at the place of the diode was kept at
37°C ± 1°C by a dedicated heating circuit. This circuit was set up from a
resistive heating foil (see (5) in Figure 4(b)), a temperature sensor next
to the diode (see (2) in Figure 4(a)), and a PID controller implemented on a dedicated second
microcontroller (Arduino™ UNO).The sensor head ((1) in Figure
5), including a single BPW34 photo diode, was put in front of the
plastic bag containing the incandescent gas mantle. First, the temperature
dependency of signal strength after the main amplifying circuit and number of
pulses was measured at 22°C (room temperature) and 37°C (body temperature) at a
distance of 1 mm. Ten measurements of 60 s duration were performed each (Figure 7).
Figure 7.
Illustration of the test setup and definition of parameters. (1) PCB of
the primary stage of the amplifier, (2) two diodes, and (3) stainless
steel casing.
Illustration of the test setup and definition of parameters. (1) PCB of
the primary stage of the amplifier, (2) two diodes, and (3) stainless
steel casing.Then the dependency of the inclination angle of the diode with regard to the
plastic bag containing the incandescent gas mantle (0°, 45°, 90°) was evaluated
at a distance of 1 mm at room temperature. Ten measurements of 60 s duration
were performed each. Again, the number of pulses per minute and signal strength
were assessed.Finally, the number of detected events per minute was assessed using a
137Cs source (Amersham Buchler, activity: 370 kBq in 1977, today
approx. 138 kBq, energy: 612 keV, active area: approx. 4 mm2) at
distances of 1, 2.5, 5, 10, 20, and 40 mm. For each distance, during 20 min at
room temperature, the number of pulses was measured, and the average was
calculated.For the preclinical validation trial, the structure described below was chosen.
The experiments were carried out in the laboratory with nine volunteers (4 f,
5 m, Ø age 25 ± 2.5 years) who had no previous medical training. The subjects
had to unveil the location of the 137Cs source hidden in a wooden box
(made from medium-density fiber board, Figure 8(a))). The time taken to find
the sample and the localization accuracy were recorded.
Figure 8.
(a) Setting for the detection experiment. (1) Sensor head including
primary stage of the amplifier fixed in acrylic glass support, (2) top
plate of the box containing the 137Cs γ-radiation source, (3)
shielded cable), (4) housing for signal conditioning electronics, (5)
computer for data processing. (b) Illustration of the top plate surface.
(1) Exemplary position of the radiation source (not visible to the
subjects), (2) neighboring field, (3) any other field).
(a) Setting for the detection experiment. (1) Sensor head including
primary stage of the amplifier fixed in acrylic glass support, (2) top
plate of the box containing the 137Cs γ-radiation source, (3)
shielded cable), (4) housing for signal conditioning electronics, (5)
computer for data processing. (b) Illustration of the top plate surface.
(1) Exemplary position of the radiation source (not visible to the
subjects), (2) neighboring field, (3) any other field).A square base area was divided into 10 columns (A–J) and 10 rows (1–10) (see
Figure 8(b)). Each
square had a size of 20 × 20 mm2 and was formed by engraved lines.
The radiation source was placed underneath the plate in such a way that the
active area was always centered on one field. The effective distance between the
radioactive sample of 137Cs was thus the thickness of the plate
(5 mm). The grid and the names of the columns and rows were visible to the test
persons.For the actual experiment, the radiation source was randomly positioned under a
field of the cover plate without this being directly visible to the subjects.
The subjects held the sensor in a support (made from acrylic glass) that forces
the sensor to be placed perpendicular on the plate and the frontal surface of
the sensor head touched the cover plate. As a consequence, the same minimum
distance of 5 mm between radiation source and sensor head and thus a maximum
signal was always achieved. The subjects were then asked to locate the radiation
source as precisely as possible starting with the sensor support directly on top
of the cover plate. They were allowed start from an arbitrary starting point
deploying a self-selected search strategy. The acoustic signal triggered by each
detection of a gamma quant was always audible to the subjects. The shorter time
between the acoustic pulses, the nearer the sensor was to the radiation
source.The time between the start of the search and the statement that the sample was
found was measured by the test director. Whether the measurement was correct was
also recorded and communicated to the test persons after the last trial. The
test was repeated seven times per subject. To assess the localization quality,
it was recorded for each trial whether the sample was found exactly (“A”), a
neighboring field was determined (“B”), or a field even further away was
determined by the subjects (“C”). See Figure 8(b) for details.
Statistics
To evaluate the preliminary tests, 10 periods of 60 s were assessed. The number
of pulses per measurement period was then averaged and transferred to Figure 9 as a function of
distance.
Figure 9.
Results of the preliminary experiments for choosing the diode: (a)
detected pulses per minute at 1 and 18 mm distance to radiation source.
(b) Signal strength of the four diodes tested while detecting a gamma
quant.
Results of the preliminary experiments for choosing the diode: (a)
detected pulses per minute at 1 and 18 mm distance to radiation source.
(b) Signal strength of the four diodes tested while detecting a gamma
quant.For the validation experiment averages of the measured values were calculated.
Also, a Student’s t-test was performed to assess significance
levels. P values less than 0.05 were considered statistically
significant.To evaluate the detection experiment, a learning curve of the required time was
plotted over all test persons. The quality of the detection tests would be
recorded for each test person.
Results
The preliminary experiment allowed to choose the right diode for further studies.
Figure 9(a)) shows
the number of pulses per minute for the diodes listed in Table 1. The diode FDS100 has highest
value for pulses per minute (sensitivity) either at 1 mm distance or 18 mm.
Figure 9(b) depicts
the signal strength of the assessed diodes. Diode PS13-5B has the highest output
signal in case of detection. However, it detects less than one event per
minute.Following these results and the information collected in Table 1, the diode BPW34 was chosen as
it delivers a practical number of pulses per second and a good signal strength
at lowest cost, smallest outline, lowest capacity, and a reasonable dark
current.During the validation experiment the dependency of signal strength and the number
of detected pulses per minute with regard to temperature, angular displacement,
and distance were assessed. A visualization of the results is given in Figure 10.
Figure 10.
Results of the validation experiment with a single BPW34 photo diode: (a)
temperature dependency (note that for better visibility the display of
the values has been slightly shifted horizontally), (b) angular
dependency, and (c) distance dependency.
Results of the validation experiment with a single BPW34 photo diode: (a)
temperature dependency (note that for better visibility the display of
the values has been slightly shifted horizontally), (b) angular
dependency, and (c) distance dependency.Between room temperature and body temperature the number of detected pulses by
the photo diode is reduced by 6% (p = 0.7; not
significant).Rotating the photo diode led to a pronounced reduction in detected pulses while
the signal strength was not affected significantly. From the highest value of
140 ± 22 pulses per minute at 90°, roughly half of that is achieved at 45° and
30 ± 10 pulses were detected with the diode’s sensitive plane in parallel with
the radiation axis. While the number of detected pulses changed significantly
(p < 0.001), the signal strength did not change
significantly for 0° versus 45° (p = 0.064) and 0° versus 90°
(p = 0.25) but the difference between 45° and 90°
inclination was significant (p = 0.042).Altering the distance between the 137Cs source and the photo diode
strongly affected the number of detected pulses while the signal strength again
was not affected significantly.The 137Cs source emits 138,000 gamma quants per second evenly
distributed in all spatial directions. Given the activity of the source, the
number of events on the surface of a sphere of radius r isAt a distance of 5 mm a number of 690 counts per minute could be detected by the
selected diode. Thus, for the sensor area of 7.04 mm2 (single diode,
see Figure 4) a maximum
of 185,434 pulses per minute could be expected at 100% sensitivity. The measured
rate of 690 pulses per minute leads to a sensitivity of 0.37%.The time needed by the subjects to complete each run was averaged for each trial
and is shown in Figure
11. This shows a learning curve with a mean value of 95 ± 68.1 s for
the first trial up to 40 ± 23.9 s for the last trial. However, this appears not
to be significant.
Figure 11.
Time needed for each trial over all subjects. (1) Average value, (2)
median (3) extreme values, and (4) box depicts 25–75 percentile.
Time needed for each trial over all subjects. (1) Average value, (2)
median (3) extreme values, and (4) box depicts 25–75 percentile.Finally, the evaluation of the error rate showed that in all cases the test
persons were able to determine the position of the source correctly.
Discussion
The use of current rigid laparoscopic gamma probes can be problematic in
robot-assisted procedures because the rigid devices prove limited orientation of the
probe head. The aim of our work thus was trifold:During a dedicated preliminary experiment using a low-dose γ-radiation
source, the feasibility of setting up a low-cost gamma detector for medical
purposes from simple off-the-shelf parts and lab equipment was assessed with
the aim of selecting an appropriate and cheap diode (BPW34). The selected
diode has a bulk cost of less than 1 € allowing cost-effective realization
of such a device. Even a disposable sensor could be envisioned, which would
be economically more attractive compared to expensive reusable devices,
including costs for re-sterilization.In the validation experiment two BPW34 diodes were mounted into a stainless
steel cover to enhance sensitivity. These were then assessed with regard to
temperature dependency and the influence of angular and spatial
displacement. It became clear that at body temperature the sensitivity is
slightly reduced as the influence of dark current rises with temperature.
Still, the results show, that the threshold of the detection algorithm is
set well. Rotating the diode out of the perpendicular plane led to a
pronounced reduction of the detected pulses. This is due to the fact that
the projected area of the sensitive plane of the diode penetrated by the
gamma quants becomes smaller. Because of the plane having thickness larger
than zero, the gamma quants have to come through a thicker layer thus
increasing the probability of detection. This can explain why at 0° still a
reasonable number of pulses was detected. Finally, increasing the distance
between the source and diode led to a reduction of the number of pulses
detected. Looking at the inverse square law, an even higher reduction could
have been expected. However, this law applies only for point-sized sources.
The 137Cs source deployed in this study has a size of
4 mm2 while the active area of the photo diode has a size of
7.02 mm2. Thus, this law can be used as an estimate only.The sensitivity of the device is rather low compared to commercial devices
(e.g. Chrystal Photonics and Berlin
). This can be explained by the fact that the sensor presented does
not rely on a scintillator crystal. Also, the microcontroller is limiting
the number of counts per second to its loop execution speed – reducing
measurement dynamics. While the signal is processed, no new events can be
counted. Lastly, the circuit also needs some microseconds to recharge the
capacity of the photo diode. However, the use of a simple PIN diode greatly
reduces cost and still allows for a sufficient sensitivity and the recharge
time is far shorter than the time needed for the read-out by the
microcontroller. The acoustic display of the number of detected pulses is
limited to 40 per second. Whether this is sufficient to detect lymph nodes
in situ has to be assessed in a future pre-clinical experiment.In the third experiment the quality and time needed detecting a γ-radiation
source by human subjects was assessed. The chosen radiation source consists
of 137Cs with an activity of 138 kBq. Due to the short half-life
of 18F it was not possible to use this material in the setting.
However, the activity of the 137Cs source is comparable to human
lymph nodes during surgery after PET imaging, which is between 1 kBq and 1 MBq.
The γ-radiation source was hidden underneath a 5 mm strong
medium-density fiberboard. γ-radiation travels through this material with
almost no attenuation. This can be compared to the surgical situation where
the layer between sensor head and radiating tissue would mainly consist out
of water. It became clear that the selected special resolution of the
experimental setup (Figure
8) was too coarse – compared to the clinical use case – to derive
a meaningful result of spatial sensitivity. Future pre-clinical experiments
will take this into account to come closer to the anatomical specifications.
The subjects’ detection speed showed a slight learning curve for the time
needed to detect the position of the source. This allows the conclusion that
the acoustic user interface allows intuitive handling of the system.The experiments showed principal suitability of the BPW34 diode to detect a
137Cs γ-radiation source. The applicability of the sensor in the
medical field must be further assessed. Future developments will comprise a
reduction in size of the device, the usage of a hygienic housing of the diode and
primary stage of the amplifier to allow in vitro tests on explanted human lymph
nodes and finally in vivo testing. Also, a comparison to state-of-the-art devices
available in the operating room will be performed.
Authors: Tobias Maurer; Stephanie Robu; Margret Schottelius; Kristina Schwamborn; Isabel Rauscher; Nynke S van den Berg; Fijs W B van Leeuwen; Bernhard Haller; Thomas Horn; Matthias M Heck; Jürgen E Gschwend; Markus Schwaiger; Hans-Jürgen Wester; Matthias Eiber Journal: Eur Urol Date: 2018-04-04 Impact factor: 20.096
Authors: Matthias N van Oosterom; Hervé Simon; Laurent Mengus; Mick M Welling; Henk G van der Poel; Nynke S van den Berg; Fijs Wb van Leeuwen Journal: Am J Nucl Med Mol Imaging Date: 2016-01-28
Authors: Paolo Dell'Oglio; Philippa Meershoek; Tobias Maurer; Esther M K Wit; Pim J van Leeuwen; Henk G van der Poel; Fijs W B van Leeuwen; Matthias N van Oosterom Journal: Eur Urol Date: 2020-11-14 Impact factor: 20.096