Marco Mura1,2,3, Wendy Martin4, K Keven Williams4, Dina Joy K Abulon5.
Abstract
PURPOSE: Novel 3D digital display systems, such as NGENUITY 3D digitally assisted visualization system (DAVS), can provide enhanced illumination, depth of field, and digital filtering. This study compared vitreous removal using NGENUITY 3D DAVS with a standard surgical microscope.Entities:
Keywords: 3D digital display systems; NGENUITY; vitrectomy
Year: 2021 PMID: 34824525 PMCID: PMC8610758 DOI: 10.2147/OPTH.S327570
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Equipment Settings
| Arm 1 | Arm 2 | |
|---|---|---|
| Vitrectomy platform | CONSTELLATION Vision System | CONSTELLATION Vision System |
| Visualization system | LuxOR LX3 with NGENUITY 3D DAVS | LuxOR LX3 with standard binocular |
| Vitrectomy optics | Oculus BIOM 200 mm | Oculus BIOM 200 mm |
| Optimal magnification | × 5 | × 5 |
Abbreviation: DAVS, digitally assisted visualization system.
NGENUITY 3D DAVS Settings Applied During Surgery
| Parameters | NGENUITY 3D DAVS Settings |
|---|---|
| Camera | Standard |
| Light profile | Xenon |
| Gain | 1 |
| Channel | Off |
| RGB | 100, 100, 100 |
| Brightness | 49.00 |
| Contrast | 52.50 |
| Gamma | 1.00 |
| Hue | 32 |
| Saturation | 87 |
Abbreviations: DAVS, digitally assisted visualization system; RGB, red green blue.
Figure 1Residual vitreous assessment. Posterior segment of the eye was removed (A), the eye was bisected (B), and residual vitreous was stripped (C and D).
Figure 2Comparison of NGENUITY 3D DAVS and a standard visualization system for the removal of vitreous. Mean residual vitreous (A) and mean percentage of vitreous removed (B) are shown (n = 15). *P < 0.0001.
Comparison of OD and OS
| Mean ± SD Vitreous Weight, g | Mean | Difference Between OD and OS | ||
|---|---|---|---|---|
| Mean ± SD Difference in Vitreous Weight, g | Percentage Difference in Vitreous Weight, % | |||
| OD | 2.47 ± 0.15 | 17.48 ± 0.19 | 0.046 ± 0.035 | 1.87 |
| OS | 2.49 ± 0.15 | 17.54 ± 0.20 | ||
Abbreviations: OD, right eye; OS, left eye.
Vitreous Weight and Axial Length for Contralateral Eyes in Comparison Analysis
| Animal | Vitreous Weight, g | Axial Length, mm | ||||
|---|---|---|---|---|---|---|
| OD | OS | Difference Between OD and OS, g | OD | OS | Difference Between OD and OS, mm | |
| 1 | 2.1517 | 2.2378 | 0.0861 | 17.46 | 17.18 | 0.28 |
| 2 | 2.4288 | 2.3884 | 0.0404 | 17.19 | 17.28 | 0.09 |
| 3 | 2.5095 | 2.5925 | 0.0830 | 17.82 | 17.71 | 0.11 |
| 4 | 2.7157 | 2.7145 | 0.0012 | 17.69 | 17.72 | 0.03 |
| 5 | 2.5450 | 2.5368 | 0.0082 | 17.42 | 17.53 | 0.11 |
| 6 | 2.3515 | 2.3408 | 0.0107 | 17.60 | 17.80 | 0.20 |
| 7 | 2.4919 | 2.5024 | 0.0105 | 17.55 | 17.60 | 0.05 |
| 8 | 2.3787 | 2.4473 | 0.0686 | 17.61 | 17.77 | 0.16 |
| 9 | 2.4122 | 2.5300 | 0.1178 | 17.52 | 17.55 | 0.03 |
| 10 | 2.4118 | 2.3767 | 0.0351 | 17.27 | 17.36 | 0.09 |
| 11 | 2.3791 | 2.3124 | 0.0667 | 17.21 | 17.34 | 0.13 |
| 12 | 2.6853 | 2.7471 | 0.0618 | 17.66 | 17.66 | 0 |
| 13 | 2.5239 | 2.4871 | 0.0368 | 17.45 | 17.68 | 0.23 |
| 14 | 2.6228 | 2.6008 | 0.0220 | 17.32 | 17.34 | 0.02 |
| Mean (SD) | 2.47 (0.15) | 2.49 (0.15) | 0.0464 (0.0352) | 17.48 (0.18) | 17.54 (0.20) | 0.11 (0.08) |
| 0.35 | 0.15 | |||||
Note: P values were calculated using paired t-test.
Abbreviations: OD, right eye; OS, left eye.
Figure 3Vitreous visualization using NGENUITY 3D digitally assisted visualization system alternated between right eye and left eye to reduce surgical bias; vitreous was highlighted in a blue color.