| Literature DB >> 34820748 |
Roman Krumpholz1, Jonas Fuchtmann2, Maximilian Berlet2,3, Annika Hangleiter2, Daniel Ostler2, Hubertus Feussner2,3, Dirk Wilhelm2,3.
Abstract
PURPOSE: While demand for telemedicine is increasing, patients are currently restricted to tele-consultation for the most part. Fundamental diagnostics like the percussion still require the in person expertize of a physician. To meet today's challenges, a transformation of the manual percussion into a standardized, digital version, ready for telemedical execution is required.Entities:
Keywords: Auscultation; Percussion; Robotic examination; Robotic percussion; Tele-diagnostic; Telemedicine
Mesh:
Year: 2021 PMID: 34820748 PMCID: PMC8612625 DOI: 10.1007/s11548-021-02520-z
Source DB: PubMed Journal: Int J Comput Assist Radiol Surg ISSN: 1861-6410 Impact factor: 2.924
Fig. 1Left: Ivory pelximeter with handles used to transfer the strikes during a percussion onto the patient’s tissue [8]. Right: Sibson’s percussor-pleximeter, a percussion device with an integrated pleximeter, allowing for higher standardized diagnostics [8]
Fig. 2Left: Photograph of the examination cabin. White circles indicate the distinct examination modules. Right: Detail shot, showing the revolver-like fixture at the robotic arm of module 3 with the attached palpation and auscultation device, as well as the magnetic percussion end-effector during evaluation
Fig. 3Inner workings of the three different percussion instruments. Picture a shows the spring-loaded motorized version, picture b depicts the transistor controlled solenoid and picture c displays the pneumatic percussion instrument, driven by an external supply of compressed air
Overview of the key parameters of the developed percussion instruments
| Percussion instruments | Frequency (Hz) | Force (N) | Stroke (mm) |
|---|---|---|---|
| Motorized percussion (MOT) | 3.6 | 17.85 | 116.8 |
| Magnetic percussion (MAGN) | 4 | 19 | 7.5 |
| Pneumatic Percussion (PNEU) | 4 | 20 | 10 |
In order to create a comparable setup for all instrument versions, same percussion rate and impact force were chosen
Fig. 4Percussion points on the ventral (left) and dorsal (right) patient side. The lung region is highlighted. Auscultation was always carried out dosal and in close proximity to the percussion end-effector
Baseline characteristics of the participants
| In-ear headphones | Classic headphones | Device speakers | |||
|---|---|---|---|---|---|
| % | 23.8 | 28.5 | 47.6 | ||
| 5 | 6 | 10 | |||
| Medical student | 33.3 | 7 | 2 | 1 | 4 |
| Medical doctor (MD) | 33.3 | 7 | 1 | 2 | 4 |
| Other | 33.3 | 7 | 2 | 3 | 2 |
n, absolute frequency; %, relative frequency
Fig. 5Device abbreviations—MAN manual percussion, MOT-D mechanic device with dorsal percussion and auscultation, MOT-V mechanic device with ventral percussion and dorsal auscultation, MAGN-D magnetic device with dorsal percussion and auscultation, MAGN-V magnetic device with ventral percussion and dorsal auscultation, PNEU-D pneumatic device with dorsal percussion and auscultation, PNEU-V pneumatic device with ventral percussion and dorsal auscultation, the color scheme from A and B is also used in C, in B the symbols mark the mean and the whiskers the standard deviation on the respective axis/**p value < 0.01, *p value < 0.05, n.s. p value 0.05 in Friedman test with Wilcoxon rank sum test as post-hoc analysis, p values are adjusted by Bonferroni correction
Correct answers by qualification and device
| Group | MAN | MOT-D | MOT-V | MAGN-D | MAGN-V | PNEU-D | PNEU-V |
|---|---|---|---|---|---|---|---|
| Mean (SD) | 47.1 | 52.9 | 65.7 | 80.0 | 62.9 | 50.0 | 54.3 |
| (11.1) | (13.8) | (28.2) | (10.0) | (22.9) | (18.3) | (9.8) | |
| Median (IQR) | 50.0 | 50.0 | 60.0 | 80.0 | 50.0 | 40.0 | 50.0 |
| (15.0) | (10.0) | (30.0) | (5.0) | (25.0) | (20.0) | (10.0) | |
| Mean (SD) | 50.0 | 55.7 | 74.3 | 51.4 | 52.9 | 45.7 | 40.0 |
| (5.8) | (9.8) | (11.3) | (35.3) | (24.3) | (27.0) | (27.1) | |
| Median (IQR) | 50.0 | 60.0 | 80.0 | 50.0 | 60.0 | 40.0 | 50.0 |
| (0.0) | (10.0) | (15.0) | (65.0) | (35.0) | (30.0) | (40.0) | |
| Mean (SD) | 55.0 | 48.3 | 75.0 | 51.7 | 65.0 | 46.7 | 46.7 |
| (13.8) | (11.7) | (18.7) | (34.9) | (19.7) | (18.6) | (13.7) | |
| Median (IQR) | 55.0 | 50.0 | 75.0 | 50.0 | 70.0 | 45.0 | 45.0 |
| (25.0) | (15.0) | (25.0) | (60.0) | (20.0) | (17.5) | (10.0) | |
| Mean (SD) | 50.5* | 53.3* | 62.4 | 61.4 | 47.1* | 47.6* | |
| (10.2)* | (12.0)* | (30.2) | (22.4) | (20.3)* | (18.4)* | ||
| Median (IQR) | 50.0* | 50.0* | 80.0 | 60.0 | 40.0* | 50.0* | |
| (20.0)* | (10.0)* | (60.0) | (30.0) | (20.0)* | (20.0)* | ||
Numbers represent percentages (%) of correct answers
Device abbreviations—MAN manual percussion, MOT-D mechanic device with dorsal percussion and auscultation, MOT-V mechanic device with ventral percussion and dorsal auscultation, MAGN-D magnetic device with dorsal percussion and auscultation, MAGN-V magnetic device with ventral percussion and dorsal auscultation, PNEU-D pneumatic device with dorsal percussion and auscultation, PNEU-V pneumatic device with ventral percussion and dorsal auscultation, SD standard deviation, IQR interquartile range
The bold printed results differ statistically significantly from the others marked with an asterisk (*)