Misbah Tariq1, Jian Hur2, Jun-Won Seo1, Da Young Kim1, Na Ra Yun1, You Mi Lee1, Mi-Seon Bang1, Seong Yeon Hwang1, Choon-Mee Kim3, Ju-Hyung Lee4, Kyoung-Ho Song5, Hyunju Lee6, Jongtak Jung5, Ji Young Park6, Hong Bin Kim5, Eu Suk Kim5, Sangmin Lee7, Dong-Min Kim1.
Abstract
Here, we aimed to investigate the diagnostic value of a serological assay using the nucleocapsid protein developed for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) detection and evaluated its performance using three commercial enzyme-linked immunosorbent assays (ELISAs), namely, Standard E 2019 novel coronavirus disease (COVID-19) total antibody (Ab) ELISA (SD Biosensor), and EDI novel coronavirus COVID-19 IgG and IgM ELISA. A recombinant nucleocapsid protein (rNP) was expressed from plants and Escherichia coli for the detection of serum total Ab. We prospectively collected 141 serum samples from 32 patients with reverse transcription-PCR (RT-PCR)-confirmed COVID-19 and determined the sensitivity and dynamics of their total Ab response. Specificity was evaluated using 158 prepandemic samples. To validate the assays, we evaluated the performance using two different cutoff values. The sensitivity and specificity for each assay were as follows: 92.91% and 94.30% (plant-rNP), 83.69% and 98.73% (SD Biosensor), 75.89% and 98.10% (E. coli-rNP), 76.47% and 100% (EDI-IgG), and 80.39% and 80% (EDI-IgM). The plant-based rNP showed the highest sensitivity and area under the receiver operating characteristic (ROC) curve (0.980) among all the assays (P < 0.05). The seroconversion rate for total Ab increased sequentially with disease progression, with a sensitivity of 100% after 10 to 12 days of post-symptom onset (PSO) for both rNP-plant-based and SD Biosensor ELISAs. After 2 weeks of PSO, the seroconversion rates were >80% and 100% for EDI-IgM and EDI-IgG ELISA, respectively. Seroconversion occurred earlier with rNP plant-based ELISA (5 days PSO) compared with E. coli-based (7 days PSO) and SD Biosensor (8 days PSO) ELISA. We determined that rNP produced in plants enables the robust detection of SARS-CoV-2 total Abs. The assay can be used for serosurvey and complementary diagnosis of COVID-19. IMPORTANCE At present, the principal diagnostic methods for COVID-19 comprise the identification of viral nucleic acid by genetic approaches, including PCR-based techniques or next-generation sequencing. However, there is an urgent need for validated serological assays which are crucial for the understanding of immune responses against SARS-CoV-2. In this study, a highly sensitive and specific serological antibody assay was developed for the detection of SARS-CoV-2 with an overall accuracy of 93.56% using a recombinant nucleoprotein expressed from plants.Entities:
Keywords: COVID-19; E. coli; ELISA; SARS-CoV-2; nucleoprotein; plant; total antibody
Mesh:
Substances:
Year: 2021 PMID: 34817278 PMCID: PMC8612163 DOI: 10.1128/Spectrum.00672-21
Source DB: PubMed Journal: Microbiol Spectr ISSN: 2165-0497
Performance of in-house rNP plant- and E. coli-based, SD Biosensor total Ab, and EDI novel coronavirus COVID-19 IgG and IgM ELISAs
| Method | Data by ELISA | |||||
|---|---|---|---|---|---|---|
| rNP (plant based) | rNP ( | SD Biosensor total Ab | EDI-IgG | EDI-IgM | ||
| ROC | Cutoff | 0.5 | 1.8 | 0.18 | 0.28 | 0.09 |
| True positive ( | 131/141 | 107/141 | 118/141 | 39/51 | 41/51 | |
| True negative ( | 149/158 | 155/158 | 156/158 | 20/20 | 16/20 | |
| Intermediate ( | N/A | N/A | N/A | N/A | N/A | |
| Sensitivity (% [95% CI | 92.91 (87.34–96.55) | 75.89 (67.97–82.69) | 83.69 (76.54–89.37) | 76.47 (62.51–87.21) | 80.39 (66.88–90.18) | |
| Specificity (% [95% CI]) | 94.30 (89.46–97.36) | 98.10 (94.55–99.61) | 98.73 (95.50–99.85) | 100 (83.16–100) | 80 (56.34–94.27) | |
| Accuracy (% [95% CI]) | 93.65 (90.25–96.13) | 87.63 (83.35–91.14) | 91.64 (87.90–94.52) | 83.10 (72.34–90.95) | 80.28 (69.14–88.78) | |
| AUC | 0.980 (0.957–0.993) | 0.929 (0.893–0.955) | 0.923 (0.887–0.951) | 0.894 (0.798–0.955) | 0.805 (0.694–0.890) | |
| Mean + 3 SD or range/value | Cutoff | 0.7 | 2.2 | 0.36 | 0.32–0.39 | 0.18–0.22 |
| True positive ( | 118/141 | 90/141 | 115/141 | 37/51 | 24/51 | |
| True negative ( | 158/158 | 158/158 | 157/158 | 20/20 | 19/20 | |
| Intermediate ( | N/A | N/A | N/A | 2/71 | 1/71 | |
| Sensitivity (% [95% CI]) | 83.69 (76.54–89.37) | 63.83 (55.32–71.75) | 81.56 (74.16–87.59) | 72.55 (58.26–84.11) | 45.28 (31.56–59.55) | |
| Specificity (% [95% CI]) | 100 (97.69–100) | 100 (97.69–100) | 99.37 (96.52–99.98) | 100 (83.16–100) | 95 (75.13–99.87) | |
| Accuracy (% [95% CI]) | 92.31 (88.68–95.06) | 82.94 (78.19–87.03) | 90.97 (87.13–93.96) | 80.28 (69.14–88.78) | 58.90 (46.77–70.29) | |
The data set incorporates all 299 sample results for rNP plant-based, E. coli-based, and SD Biosensor-Total Ab ELISA, obtained from 32 (symptomatic and asymptomatic) patients with COVID-19 (141 samples) and 158 samples from negative controls, while 71 samples were included for EDI novel coronavirus COVID-19 IgG and IgM ELISA containing 51 samples from 21 patients (symptomatic and asymptomatic) with COVID-19 and 20 samples from negative controls.
ROC, receiver operating characteristic.
95% CI, 95% confidence interval.
N/A, not applicable.
P < 0.05 vs all.
P < 0.05 vs all except SD Biosensor total Ab.
P < 0.05 vs versus EDI-IgM.
FIG 1Optical density at 450 nm (OD450) for antibody detection by days after symptom onset. Plant-based rNP (A), E. coli-based rNP (B), SD Biosensor total Ab (C), EDI-IgG (D), and EDI-IgM (E) ELISAs. The gray line shows the calculated cutoff values by mean + 3 SD or range recommended by commercial assay. The red line shows the cutoff values recommended by the ROC curve. Green lines indicate the median with interquartile ranges. The data set contained 286 samples for rNP plant-based, rNP E. coli-based, and SD Biosensor total Ab ELISAs, including 128 samples from 30 symptomatic COVID-19 patients and 158 samples from negative controls. For the EDI-IgG and -IgM ELISAs, 64 samples were analyzed, including 44 samples from 19 symptomatic COVID-19 patients and 20 samples from negative controls. mean + 3 SD, mean optical density plus 3-fold of standard deviation. ROC, receiver operating characteristic.
Sensitivity of in-house rNP plant-based, E. coli-based, SD Biosensor total Ab, and EDI novel coronavirus COVID-19 IgG and IgM ELISAs with the duration of illness
| Method by wk | Sensitivity (% [ | ||||
|---|---|---|---|---|---|
| rNP (plant based) | rNP ( | SD Biosensor total Ab | EDI-IgG | EDI-IgM | |
| ROC curve | |||||
| 1 | 67.86 (19/28, 47.65–84.12) | 50.00 (14/28, 30.65–69.35) | 67.86 (19/28, 47.65–84.12) | 36.36 (4/11, 10.93–69.21) | 72.73 (8/11, 39.03–93.98) |
| 2 | 96.15 (25/26, 80.36–99.90) | 69.23 (18/26, 48.21–85.67) | 84.62 (22/26, 65.13–95.64) | 80.00 (8/10, 44.39–97.48) | 80.00 (8/10, 44.39–97.48) |
| 3 | 100 (21/21, 83.89–100) | 95.24 (20/21, 76.18–99.88) | 100 (21/21, 83.89–100) | 100 (11/11, 71.51–100) | 90.91 (10/11, 58.72–99.77) |
| >3 | 100 (53/53, 93.40–100) | 94.34 (50/53, 84.34–98.82) | 100 (53/53, 93.40–100) | 100 (12/12, 73.54–100) | 83.33 (10/12, 51.59–97.91) |
| Mean + 3 SD or range/value | |||||
| 1 | 50.00 (14/28, 30.65–69.35) | 35.71 (10/28, 18.64–55.93) | 32.14 (9/28, 15.88–52.35) | 27.27 (3/11, 6.02–60.97) | 18.18 (2/11, 2.28–51.78) |
| 2 | 88.77 (21/26, 60.65–93.45) | 53.85 (14/26, 33.37–73.41) | 73.08 (19/26, 52.21–88.43) | 70.00 (7/10, 34.75–93.33) | 20.00 (2/10, 2.52–55.61) |
| 3 | 100 (21/21, 83.89–100) | 95.24 (20/21, 76.18–99.88) | 100 (21/21, 83.89–100) | 100 (11/11, 71.51–100) | 81.82 (9/11, 48.22–97.72) |
| >3 | 100 (53/53, 93.28–100) | 83.02 (44/53, 70.20–91.93) | 100 (53/53, 93.28–100) | 100 (12/12, 73.54–100) | 83.33 (10/12, 51.59–97.91) |
The data set contained 128 samples for rNP plant-based, rNP E. coli-based, and SD Biosensor total Ab assays from 30 symptomatic COVID-19 patients.
Wk 1, 0–6 days PSO; wk 2, 7–13 days PSO; wk 3, 14–20 days PSO; >wk 3, ≥21 to 91 days PSO.
95% CI, 95% confidence interval.
For the EDI-IgG and -IgM ELISA assays, 44 samples were analyzed from 19 symptomatic COVID-19 patients.
ROC, receiver operating characteristic.
Mean + 3 SD, mean optical density plus 3-fold standard deviation.
FIG 2Graph of the positive rate of SARS-CoV-2-specific total antibodies, IgG, and IgM by days post-symptom onset. (A) Positive rate obtained from recommended cutoff values by ROC curve. (B) Positive rate obtained from calculated cutoff values by mean + 3 SD or range recommended by commercial assay. The data set contains 128 samples for rNP plant-based and rNP E. coli-based assays from 30 symptomatic COVID-19 patients. For EDI-IgG ELISA, 44 samples were analyzed from 19 symptomatic COVID-19 patients. The cutoffs recommended by ROC curve were 0.5, 1.8, 0.18, 0.28, and 0.09 for rNP plant-based, rNP E. coli-based, SD Biosensor total Ab, EDI-IgG, and EDI-IgM ELISAs, respectively. The calculated cutoff values by mean + 3 SD or range/value recommended by commercial assay were 0.7, 2.2, 0.36, 0.32 to 0.39, and 0.18 to 0.22 for rNP plant-based, rNP E. coli- based, SD Biosensor test Ab, EDI-IgG, and EDI-IgM ELISAs, respectively. The borderline results obtained from the EDI commercial assay were considered positive.
FIG 3Receiver operating characteristic (ROC) curves for the evaluation and comparison of diagnostic accuracy. (A) Plant rNP-based, E. coli rNP-based, and SD Biosensor total Ab ELISAs. (B) Plant rNP-based, E. coli rNP-based, SD Biosensor total Ab, EDI-IgG, and EDI-IgM ELISAs. The data set contain all 299 samples for rNP plant-based, rNP E. coli-based, and SD Biosensor total Ab ELISAs, including 141 samples from 32 patients with COVID-19 and 158 samples from negative controls. For comparison with EDI-IgG and -IgM ELISA, the data set contains 71 samples, including 51 samples from 21 patients with COVID-19 and 20 samples from negative controls.
Correlation analysis of the OD values between in-house rNP plant-based, E. coli-based, SD Biosensor total Ab, and EDI novel coronavirus COVID-19 IgG and IgM ELISAs
| ELISA | Parameter | Data by ELISA | ||||
|---|---|---|---|---|---|---|
| rNP (plant based) | rNP ( | SD Biosensor total Ab | EDI-IgG | EDI-IgM | ||
| rNP (plant based) | Correlation coefficient | 0.869 | 0.887 | 0.908 | 0.523 | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | |||
|
| 299 | 299 | 71 | 71 | ||
| rNP ( | Correlation coefficient | 0.869 | 0.782 | 0.821 | 0.556 | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | |||
|
| 299 | 299 | 71 | 71 | ||
| SD Biosensor total Ab | Correlation coefficient | 0.887 | 0.782 | 0.745 | 0.531 | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | |||
|
| 299 | 299 | 71 | 71 | ||
| EDI-IgG | Correlation coefficient | 0.908 | 0.821 | 0.745 | 0.512 | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | |||
|
| 71 | 71 | 71 | 71 | ||
| EDI-IgM | Correlation coefficient | 0.523 | 0.523 | 0.531 | 0.512 | |
| <0.0001 | <0.0001 | <0.0001 | <0.0001 | |||
|
| 71 | 71 | 71 | 71 | ||
The analysis was performed with Pearson's correlation coefficient.
The interpretations of correlation coefficient (r) were characterized as follows: 0.00–0.19, very weak; 0.20–0.39, weak; 0.40–0.59, moderate; 0.60–0.799, strong; and 0.80–1.00, very strong.
n, no. of samples.