Abstract
BACKGROUND: Atrial fibrillation (AF) is the most prevalent cardiac dysrhythmia with high morbidity and mortality rate. Evidence shows that in every three patients with AF, one is asymptomatic. The asymptomatic and paroxysmal nature of AF is the reason for unsatisfactory and delayed detection using traditional instruments. Research indicates that wearing a dynamic electrocardiogram (ECG) recorder can guide accurate and safe analysis, interpretation, and distinction of AF from normal sinus rhythm. This is also achievable in an upright position and after exercises, assisted by an artificial intelligence (AI) algorithm.Entities:
Keywords: Accuracy; Atrial fibrillation; Detection; Dynamic ECG recorder; One-lead ECG
Mesh:
Year: 2021 PMID: 34800984 PMCID: PMC8606080 DOI: 10.1186/s12872-021-02363-1
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Fig. 1A flowchart of the study. AF atrial fibrillation, ECG electrocardiogram, SR sinus rhythm
Fig. 2A prototype for AF detection using wearable dynamic ECG recorder and 12-lead ECG. A A patient is simultaneously tested with a wearable dynamic ECG recorder and 12-lead ECG. B A screenshot of the Amazfit CardiDoc app installed on the mobile phone. C Representative pulse waveform recording a participant with normal sinus rhythm. D Representative pulse waveform recording a participant with AF. AF atrial fibrillation, ECG electrocardiogram
Baseline demographic of in Non-AF and AF group (N = 114)
| Variable | Non-AF (n = 61) | AF (n = 53) | |
|---|---|---|---|
| Demographics | |||
| Age (years) | 55.15 ± 11.01 | 64.00 ± 9.38 | < 0.001 |
| Females (n, %) | 25 (41.0) | 24 (45.3) | 0.644 |
| Height (m) | 1.66 ± 0.08 | 1.66 ± 0.08 | 0.619 |
| Weight (kg) | 66.16 ± 11.89 | 66.07 ± 12.15 | 0.965 |
| BMI (kg/m2) | 23.97 ± 3.21 | 23.74 ± 3.38 | 0.711 |
| Medical history (n, %) | |||
| Hypertension | 21 (34.4) | 24 (45.3) | 0.237 |
| Diabetes mellitus | 6 (9.8) | 9 (17.0) | 0.260 |
| Heart failure | 14 (23.0) | 13 (24.5) | 0.843 |
| Stroke | 2 (3.3) | 1 (1.9) | 0.643 |
| Coronary artery disease | 0 (0.0) | 5 (9.4) | 0.014 |
| Vascular disease | 3 (4.9) | 5 (9.4) | 0.346 |
| Pulmonary embolism | 0 (0.0) | 1 (1.9) | 0.285 |
| Hyperthyroidism | 0 (0.0) | 1 (1.9) | 0.285 |
| Renal dysfunction | 0 (0.0) | 1 (1.9) | 0.285 |
| Hepatic dysfunction | 1 (1.8) | 5 (9.4) | 0.076 |
| Current drinking | 17 (27.9) | 13 (24.5) | 0.686 |
| Current smoking | 1 (1.6) | 2 (3.8) | 0.090 |
| CHA2DS2-VASc score | 1 (1–2.5) | 2 (1–3) | 0.001 |
| Medication, n (%) | |||
| Oral anticoagulant | 1 (1.6) | 21 (39.6) | < 0.001 |
| Antiplatelet drug | 1 (1.6) | 11 (20.8) | 0.001 |
| ACEI/ARBs | 14 (23.0) | 14 (26.4) | 0.668 |
| CCB | 7 (11.7) | 18 (34.0) | 0.004 |
| Diuretic | 0 (0.0) | 8 (15.1) | 0.002 |
| Digoxin | 0 (0.0) | 10 (19.2) | < 0.001 |
| Beta-blocker | 3 (4.9) | 28 (52.8) | < 0.001 |
| Amiodarone | 1 (1.6) | 5 (9.4) | 0.063 |
AF, atrial fibrillation; BMI, body mass index; CHA2DS2-VASc, congestive heart failure, hypertension, age ≥ 75, diabetes, stroke, vascular disease, age 65–74 and sex category(female); ACEI/ARBs, angiotensin-converting enzyme (ACE) inhibitors, angiotensin-receptor blockers; CCB, calcium channel blocker
Testing results of AF detection in participants and overall diagnostic performance of wearable dynamic ECG recorder in 60 s in a supine position
| Wearable dynamic ECG recorder | Standard physician read ECG | |||
|---|---|---|---|---|
| AF | Sinus rhythm | Unclassified | N | |
| AF | 47 | 0 | 0 | 47 |
| Non-AF | 4 | 61 | 0 | 65 |
| Unclassified | 2 | 0 | 0 | 2 |
| n | 53 | 61 | 0 | 114 |
ECG electrocardiogram, AF atrial fibrillation
Detailed diagnostic performance of the wearable dynamic ECG recorder for AF screening in a supine position
| Unable to unclassified as TPTN | Unable to unclassified as FPFN | |
|---|---|---|
| Accuracy, %(95% CI) | 96.49% (91.03–98.92%) | 94.74% (88.76–97.80%) |
| Sensitivity, %(95% CI) | 92.45% (81.64–97.52%) | 88.68% (77.06–95.07%) |
| Specificity, %(95% CI) | 100% (92.91–100%) | 100% (92.91–100%) |
| PPV, %(95% CI) | 100% (91.32–100%) | 100% (90.98–100%) |
| NPV, %(95% CI) | 93.85% (84.78–98.02%) | 91.04% (81.48–96.16%) |
ECG electrocardiogram, AF atrial fibrillation, TP true positives, TN true negatives, FP false positives, FN false negative, PPV positive predictive value, NPV negative predictive value
Testing results of AF detection in participants and overall diagnostic performance of wearable dynamic ECG recorder in 60 s in an upright position
| Wearable dynamic ECG recorder | Standard physician read ECG | |||
|---|---|---|---|---|
| AF | Sinus rhythm | Unclassified | N | |
| AF | 50 | 0 | 0 | 50 |
| Non-AF | 2 | 61 | 0 | 63 |
| Unclassified | 1 | 0 | 0 | 1 |
| n | 53 | 61 | 0 | 114 |
ECG electrocardiogram, AF atrial fibrillation
Detailed diagnostic performance of the wearable dynamic ECG recorder for AF screening in an upright position
| Unable to unclassified as TPTN | Unable to unclassified as FPFN | |
|---|---|---|
| Accuracy, %(95% CI) | 98.25% (93.43–99.91%) | 97.37% (92.21%, 99.44%) |
| Sensitivity, %(95% CI) | 96.23% (86.51–99.69%) | 94.34% (84.03–98.65%) |
| Specificity, %(95% CI) | 100% (92.91–100%) | 100% (92.91–100%) |
| PPV, %(95% CI) | 100% (91.63–100%) | 100% (91.48–100%) |
| NPV, %(95% CI) | 96.83% (88.50–99.77%) | 95.31% (86.57–98.92%) |
ECG electrocardiogram, AF atrial fibrillation, TP true positives, TN true negatives, FP false positives, FN false negative, PPV positive predictive value, NPV negative predictive value
Testing results of AF detection in participants and overall diagnostic performance of wearable dynamic ECG recorder in 60 s after climbed the third floor in an upright position
| Wearable dynamic ECG recorder | Standard physician read ECG | |||
|---|---|---|---|---|
| AF | Sinus rhythm | Unclassified | N | |
| AF | 50 | 0 | 0 | 50 |
| Non-AF | 2 | 61 | 0 | 63 |
| Unclassified | 1 | 0 | 0 | 1 |
| n | 53 | 61 | 0 | 114 |
ECG electrocardiogram, AF atrial fibrillation
Detailed diagnostic performance of the wearable dynamic ECG recorder for AF screening after climbed the third floor in an upright position
| Unable to unclassified as TPTN | Unable to unclassified as FPFN | |
|---|---|---|
| Accuracy, %(95% CI) | 98.25% (93.43–99.91%) | 97.37%(92.21%, 99.44%) |
| Sensitivity, %(95% CI) | 96.23% (86.51–99.69%) | 94.34% (84.03–98.65%) |
| Specificity, %(95% CI) | 100% (92.91–100%) | 100% (92.91–100%) |
| PPV, %(95% CI) | 100% (91.63–100%) | 100% (91.48–100%) |
| NPV, %(95% CI) | 96.83% (88.50–99.77%) | 95.31% (86.57–98.92%) |
ECG electrocardiogram, AF atrial fibrillation, TP true positives, TN true negatives, FP false positives, FN false negative, PPV positive predictive value, NPV negative predictive value