Literature DB >> 34799784

Corneal power measurements by ray tracing in eyes after small incision lenticule extraction for myopia with a combined Scheimpflug Camera-Placido disk topographer.

Mª Victoria de Rojas Silva1,2, Adrián Tobío Ruibal3, Jorge Suanzes Hernández4,5.   

Abstract

PURPOSE: To evaluate the accuracy of the measurements of corneal power obtained by ray tracing with a combined Scheimpflug camera-Placido disk corneal topographer (Sirius) in eyes after small incision lenticule extraction for myopia (SMILE).
METHODS: Retrospective cases study includes 50 eyes of 50 patients who underwent myopic SMILE. Mean value of simulated keratometry (Kpost), mean pupil power (MPP) (ray tracing, diameter of the entrance pupil range 3-6 mm), anterior and posterior corneal radius, and corneal thickness were obtained with Sirius topographer preoperatively and three months postoperatively, as well as cycloplegic refraction. True net power, equivalent keratometry readings, and Haigis equivalent power formula were calculated, and these measurements, MPP and Kpost, were compared with the corneal power calculated with the clinical history method (CHM).
RESULTS: Corneal power measurements obtained with all methods were significantly different from CHM (P < 0.001), except the value of MPP obtained at 5.5 mm (P = 0.927). A good direct correlation was found between CHM and all measurements. The distribution of differences as compared with the CHM showed that the lowest difference corresponded to the value of MMP at 5.5 mm (- 0.002 ± 0.6). The Bland-Altman plots for the MPP at 5.5 mm showed 95% limits of agreement between - 1.1787 D and 1.1741 D.
CONCLUSIONS: MPP obtained by ray tracing within a diameter of entrance pupil of 5.5 mm could predict corrected corneal power derived from the CHM in eyes following SMILE surgery.
© 2021. The Author(s), under exclusive licence to Springer Nature B.V.

Entities:  

Keywords:  Corneal power measurement; Placido disk; Ray tracing; SMILE; Scheimpflug

Mesh:

Year:  2021        PMID: 34799784     DOI: 10.1007/s10792-021-02073-9

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  2 in total

1.  Predictability of intraocular lens power calculation after small-incision lenticule extraction for myopia.

Authors:  Apostolos Lazaridis; Florian Schraml; Paul-Rolf Preußner; Walter Sekundo
Journal:  J Cataract Refract Surg       Date:  2021-03-01       Impact factor: 3.351

2.  Evaluation of Equivalent Keratometry Readings Obtained by Pentacam HR (High Resolution).

Authors:  Yanjun Hua; Xiaolan Zhang; Tor Paaske Utheim; Jinhai Huang; Chao Pan; Weina Tan; Qinmei Wang
Journal:  PLoS One       Date:  2016-03-07       Impact factor: 3.240

  2 in total