| Literature DB >> 34786927 |
Eduardo Mazzucchi1, Giovanni Scala Marchini1, Fernanda Christina Gabrigna Berto2, John Denstedt2, Alexandre Danilovic1, Fabio Carvalho Vicentini1, Fabio Cesar Miranda Torricelli1, Carlos Alfredo Battagello1, Miguel Srougi1, William Carlos Nahas1.
Abstract
Flexible ureteroscopy is a well-established method for treatment of urinary stones but flexible ureteroscopes are expensive and fragile devices with a very limited lifetime. Since 2006 with the advent of digital flexible ureteroscopes a great evolution has occurred. The first single-use flexible ureteroscope was launched in 2011 and new models are coming to the market. The aim of this article is to review the characteristics of these devices, compare their results with the reusable devices and evaluate the cost-benefits of adopting single-use flexible ureteroscopes in developing countries.Entities:
Keywords: Meta-Analysis [Publication Type]; Surgical Procedures, Operative; Ureteroscopes; Kidney Calculi
Mesh:
Year: 2022 PMID: 34786927 PMCID: PMC9060176 DOI: 10.1590/S1677-5538.IBJU.2021.0475
Source DB: PubMed Journal: Int Braz J Urol ISSN: 1677-5538 Impact factor: 3.050
Figure 1Flow diagram of evidence acquisition in a review on single-use and reusable flexible ureteroscopy.
Figure 2shows the different types of single-use flexible ureteroscopes.
Shows the comparison of the physical (weight, shaft length, working channel) and functional(deflection, irrigation flow) characteristics of the single-use flexible ureteroscopes currently disponible in the market (modified from Scotland et al. (7), Proietti et al. (28), Whelan et al. (30) and Dragos et al. (31).
| LithoVue™ | Uscope UE3022™ | Neoflex™ | Dornier - Axis™ | WiScope™ | |
|---|---|---|---|---|---|
| Shaft length (cm) | 68 | 65 | 68 | 66 | 67 |
| Tip outer diameter (Fr) | 9.5 | 9.0 | 9.0 | 8.5 | 7.4 |
| Deflection up/down (degree) | 270 | 270 | 280 | 275 | 275 |
| Working channel (Fr) | 3.6 | 3.6 | 3.6 | 3.6 | 3.6 |
| Irrigation flow empty channel (mL/min) | 42 | 52 | 40 | NA | 49 |
| Illumination | LED | Fiberoptic fiber | LED | LED | LED |
| Imager technology | CMOS | CMOS | CMOS | CMOS | CMOS |
| Weight (g) | 277 | 147 | 119 | 160 | 200 |
| Connector type | Round and 8 pins | Flat | USB | USB | Round |
shows the comparison of results (stone-free rates, complications and operative time) of renal stone treatment with single-use and reusable flexible ureteroscopes.
| Stone-free rates (%) | ||||
|---|---|---|---|---|
| Single-use | Reusable | p | ||
| Usawachintachit et al., 2017 ( | 60 | 44.7 | 0.36 | |
| Mager et al., 2018 ( | 82 | 85 | 0.8 | |
| Salvadó et al., 2019 ( | 95 | 88.2 | 0.1 | |
| Kam et al., 2019 | 87 | 90 | ns | |
|
| ||||
| Usawachintachit et al., 2017 ( | 5.4 | 18 | < 0.05 | |
| Mager et al., 2018 ( | 17 | 7 | 0.06 | |
| Kam et al., 2019 ( | 29 | 19 | ns | |
|
| ||||
| Usawachintachit et al., 2017 ( | 54.1 ± 25.7 | 64.5 ± 37.0 | < 0.05 | |
| Mager et al.,2018 ( | 76.2 | 76.8 | 0.9 | |
| Salvadó et al., 2019 ( | 56.1 | 77 | <0.01 | |
| Kam et al., 2019 ( | 86.1 | 72.3 | ns | |
Stone-free rate calculated based on the need for a second look pyeloscopy.