Zanzhe Yu1, Zhuqing Wang1, Qin Wang1, Minfang Zhang1, Haijiao Jin1, Li Ding1, Hao Yan1, Jiaying Huang1, Yan Jin1, Simon Davies2, Wei Fang1, Zhaohui Ni3.
Abstract
BACKGROUND: It has been noticed for years that ultrafiltration (UF) is important for survival in peritoneal dialysis. On the other hand, precise and convenient UF measurement suitable for patient daily practice is not as straight forward as it is to measure UF in the lab. Both overfill and flush before fill used to be source of measurement error for clinical practice. However, controversy finding around UF in peritoneal dialysis still exists in some situation. The current study was to understand the difference between clinical measured UF and real UF. The effect of evaporation and specific gravity in clinical UF measurement were tested in the study.Entities:
Mesh:
Substances:
Year: 2021 PMID: 34781890 PMCID: PMC8594243 DOI: 10.1186/s12882-021-02589-3
Source DB: PubMed Journal: BMC Nephrol ISSN: 1471-2369 Impact factor: 2.388
Features of the four brands bags used in the study
| Glucose (%) | Buffer agent | Ca++ (mmol/L) | PVC/non-PVC | Days after manufacture | |
|---|---|---|---|---|---|
| Brand A (Baxter, Guangzhou) | 2.27% | Lactate | 1.77 | PVC | 43 |
| Brand B (Shuanghe, Beijing) | 2.27% | Lactate | 1.77 | PVC | 50 |
| Brand C (Huaren, Shanghai) | 2.27% | Lactate | 1.77 | non-PVC | 62 |
| Brand D (Fresenius, Jiangsu) | 2.27% | Lactate | 1.24 | non-PVC | 105 |
Dialysate weight at baseline in different brand
| whole bag | outer package | bag without outer package | |
|---|---|---|---|
| A( | 2250 ± 3.6 | 19.7 ± 0.1 | 2230.3 ± 3.5 |
| B( | 2221.9 ± 1.9 | 21.6 ± 0.3 | 2200.3 ± 1.8 |
| C( | 2261.2 ± 3.7 | 20.4 ± 0.2 | 2240.8 ± 3.6 |
| D( | 2229.8 ± 2.1 | 19.1 ± 0.4 | 2210.7 ± 2.1 |
| < 0.01 | < 0.01 | < 0.01 |
Dialysate weight at baseline, 6 months and 12 months in different conditions
| baseline | 6 months | 12 months | ||
|---|---|---|---|---|
| condition 1( | 2238.9 ± 16.5 | 2238.6 ± 16.5 | 2238.3 ± 16.6 | < 0.01 |
| condition 2 ( | 2238.6 ± 14.9 | 2230.2 ± 15.2 | 2224.8 ± 15.8 | < 0.01 |
| condition 3 ( | 2239.1 ± 13.7 | 2225.6 ± 14.2 | 2217.2 ± 15.3 | < 0.01 |
| condition 4 ( | 2240 ± 16 | 2202.9 ± 21.3 | 2178.3 ± 26.7 | < 0.01 |
Fig. 1Weight loss in each brand over 12 months storage in different condition. Over the 12 month’s storage, dialysate bags lost significant weight. The weight loss was greater in higher temperature and lower humidity. PVC package was related to more significant weight loss over 12 month’s storage. Brand A and brand B were in PVC package. Brand C and brand D were in non-PVC package
generalized linear model of dialysate weight at 12 months of storage
| Multivariate | |||
|---|---|---|---|
| Coeff. | 95% CI | ||
| baseline weight | .871 | 0.75,0.99 | < 0.01 |
| condition 1 | 60.975 | 56.21,65.74 | < 0.01 |
| condition 2 | 48.131 | 43.3,52.97 | < 0.01 |
| condition 3 | 39.756 | 34.85,44.66 | < 0.01 |
| condition 4(ref) | 0 | ||
| non-PVC | 16.296 | 12.69,19.91 | < 0.01 |
| PVC(ref) | 0 | ||
| intercept | 220.28 | − 49.2489.8 | |
correlation between specific gravity and other solute concentration
| specific gravity | potassium | sodium | chlorine | glucose | protein | |
|---|---|---|---|---|---|---|
| Mean ± SD | 1.0136 ± 0.009 (g/l) | 3.35 ± 0.67 (mmol/L) | 136.14 ± 4.5 (mmol/L) | 99.65 ± 5.22 (mmol/L) | 30.54 ± 9.84 (mmol/L) | 664.74 ± 280.03 (mg/mL) |
| specific gravity | 1 | .064 | .113 | −.037 | −.011 | .139* |
| potassium | .064 | 1 | .157* | .261** | −.344** | .254** |
| sodium | .113 | .157* | 1 | .680** | −.328** | .270** |
| chlorine | −.037 | .261** | .680** | 1 | −.450** | .346** |
| glucose | −.011 | −.344** | −.328** | −.450** | 1 | −.368** |
| protein | .139* | .254** | .270** | .346** | −.368** | 1 |
* P < 0.05, **P < 0.01, SD, standard deviation