| Literature DB >> 34778080 |
Gustavo R Sarria1, Zuzanna Smalec1, Thomas Muedder1, Jasmin A Holz1, Davide Scafa1, David Koch1, Stephan Garbe1, Matthias Schneider2, Motaz Hamed2, Hartmut Vatter2, Ulrich Herrlinger3, Frank A Giordano1, Leonard Christopher Schmeel1.
Abstract
PURPOSE: To simulate and analyze the dosimetric differences of intraoperative radiotherapy (IORT) or pre-operative single-fraction stereotactic radiosurgery (SRS) in addition to post-operative external beam radiotherapy (EBRT) in Glioblastoma (GB).Entities:
Keywords: IORT; SRS; dose escalation; glioblastoma; kilovoltage
Year: 2021 PMID: 34778080 PMCID: PMC8581360 DOI: 10.3389/fonc.2021.759873
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Figure 1(A, B) Generated contours for intraoperative radiotherapy (IORT) and stereotactic radiosurgery (SRS) boosting. GTV: gross target volume, PTV: planning target volume, mGTV: modified gross target volume. mGTV was adjusted to fit the immediate larger apposite applicator. (C, D) Simulated path for the applicator placement. The healthy brain contour was removed from the hypothetical surgical trajectory.
Planning baseline features.
| Structure volume (cm3) | ||
|---|---|---|
| Mean | SD | |
| GTV | 21.72 | 19.36 |
| PTV | 36.2 | 31.48 |
| mGTV | 29.64 | 25.64 |
| EBRT | 254.09 | 80 |
| Boost delivered doses (Gy) | ||
| 20 | 15 | |
| SRS (%) | 52.94 | 47.06 |
| IORT (%) | 100 | |
| Applicator diameter (mm) | ||
| Median | Range | |
| Applicator size | 35 | 15 - 50 |
GTV, gross target volume; PTV, planning target volume; mGTV, modified gross target volume; SRS, stereotactic radiosurgery; IORT, intraoperative radiotherapy.
Organs at risk dosimetric differences.
| Dose exposure | |||||
|---|---|---|---|---|---|
| Organs at risk | EBRT+SRS | EBRT+IORT | p | ||
| Mean | SD | Mean | SD | ||
| Brain V12 (no EBRT) | 49.68 | 26.7 | 16.94 | 13.33 |
|
| Brainstem Dmax | 44.06 | 17.75 | 42.91 | 16.84 | 0.228 |
| Chiasma Dmax | 28.96 | 19.09 | 27.72 | 18.5 |
|
| Left optic nerve Dmax | 15.71 | 16.23 | 14.6 | 14.66 | 0.023 |
| Right optic nerve Dmax | 19.51 | 18.69 | 19.48 | 19.48 | 0.977 |
| Left retina Dmean | 6.24 | 5.66 | 5.61 | 5.57 |
|
| Right retina Dmean | 7.28 | 6.37 | 6.5 | 5.71 |
|
| Left cochlea Dmean | 9.43 | 10.75 | 9.72 | 10.96 | 0.645 |
| Right cochlea Dmean | 10.74 | 14.9 | 11.85 | 15.35 | 0.295 |
EBRT, external beam radiotherapy; SRS, stereotactic radiosurgery; IORT, intraoperative radiotherapy; V12, volume receiving 12 Gy. The brain V12 was not added to EBRT doses due to practical reasons. Statistically significant dosimetric differences are displayed and highlighted.
Figure 2Exemplary case with three-dimensional views of isodose line distribution for a 2.5-cm IORT applicator. Doses reaching the reconstruction plaque should be disregarded to resemble a real-world surgical situation.
Figure 3Exemplary case with three-dimensional views of isodose line distribution for upfront SRS boosting, homogeneously delivered to a 2-mm PTV.