Giovanni Luca Bartolomeo1, Yao Zhang2,3, Naresh Kumar1, Renato Zenobi1. 1. Department of Chemistry and Applied Biosciences, ETH Zürich, 8093 Zürich, Switzerland. 2. Hefei National Laboratory for Physical Sciences at the Microscale and Synergetic Innovation Center of Quantum Information and Quantum Physics, University of Science and Technology of China, Hefei 230026, Anhui, China. 3. State Key Laboratory of Quantum Optics and Quantum Optics Devices, Shanxi University, Taiyuan 030006, China.
Abstract
Tip-enhanced Raman spectroscopy (TERS) is a powerful tool for nondestructive and label-free surface chemical characterization at nanometer length scales. However, despite being considered nondestructive, the interaction of the TERS probe used in the analysis can alter the molecular organization of the sample. In this study, we investigate the role of the atomic force microscopy (AFM) feedback (contact mode and tapping mode) on molecular perturbation in TERS analysis of soft samples using a self-assembled monolayer (SAM) of 2-chloro-4-nitrobenzene-1-thiol (Cl-NBT) as a test sample. Surprisingly, the tapping mode shows a consistently higher TERS signal resulting from a minimal perturbation of the Cl-NBT SAM compared to the contact mode. This study provides novel insights into the choice of the correct AFM-TERS operation mode for nanoscale chemical analysis of soft and delicate samples and is expected to expedite the growing application of TERS in this area.
Tip-enhanced Raman spectroscopy (TERS) is a powerful tool for nondestructive and label-free surface chemical characterization at nanometer length scales. However, despite being considered nondestructive, the interaction of the TERS probe used in the analysis can alter the molecular organization of the sample. In this study, we investigate the role of the atomic force microscopy (AFM) feedback (contact mode and tapping mode) on molecular perturbation in TERS analysis of soft samples using a self-assembled monolayer (SAM) of 2-chloro-4-nitrobenzene-1-thiol (Cl-NBT) as a test sample. Surprisingly, the tapping mode shows a consistently higher TERS signal resulting from a minimal perturbation of the Cl-NBT SAM compared to the contact mode. This study provides novel insights into the choice of the correct AFM-TERS operation mode for nanoscale chemical analysis of soft and delicate samples and is expected to expedite the growing application of TERS in this area.
The interest of the
scientific community in nanotechnologies has
increased substantially in the last few decades, leading to the development
of powerful tools to characterize nanostructured materials with higher
sensitivity and spatial resolution than ever before. Today, several
nano-analytical techniques are available, each having its own strengths
and limitations. However, nondestructive and label-free nanoscale
molecular imaging under ambient conditions still remains challenging.
For example, nanoscale secondary ion mass spectrometry (nanoSIMS)
can provide elemental and isotopic information with a spatial resolution
of down to tens of nanometers but requires the sample to be compatible
with ultrahigh vacuum conditions and easily ionizable, and is destructive.[1,2] On the other hand, super-resolution fluorescence microscopy provides
a spatial resolution down to ca. 25 nm nondestructively
but requires fluorophore labeling that can alter the native state
of the sample.[3,4] When noninvasiveness is the highest
priority, especially for the study of delicate systems in their native
state, label-free spectroscopies, such as nanoscale infrared (nanoIR)
spectroscopy and tip-enhanced Raman spectroscopy (TERS), are the techniques
of choice. Nano-IR combines scanning probe microscopy (SPM) with IR
spectroscopy and provides nondestructive chemical and topographical
information.[5] However, single-molecule
resolution is not possible and application in liquid environments
is limited.[6]TERS overcomes most
of these limitations and provides nondestructive
and label-free molecular imaging at the nanoscale level in both air
and liquid environments.[7,8] In TERS, a metallic
SPM tip is used to dramatically enhance the electromagnetic (EM) field
in the focal spot of an excitation laser and confine the EM field
within a nanoscopic volume via a combination of
localized surface plasmon resonance and lightning rod effect.[9] The spatially confined optical field enhancement
in TERS can provide single-molecule sensitivity with angstrom-scale
resolution.[10] TERS can be performed in
both atomic force microscopy (AFM) and scanning tunneling microscopy
(STM) SPM modes.[11] However, since AFM feedback
does not necessitate an electrical contact with the sample, AFM-TERS
is more versatile allowing a much wider range of applications including
organic solar cells,[12] biological cells,[13] DNA strands,[14] lipid
membranes, two-dimensional (2D) nanomaterials,[15,16] and catalytic materials.[17−19] AFM-TERS is most commonly operated
using either the contact or tapping mode feedback. In contact mode,[7] the AFM probe simply scans the sample and the
resulting cantilever deflection is measured; whereas in tapping mode,[20,21] which is a milder interaction regime, the sample is imaged by measuring
the reduction in amplitude of an oscillating probe put in close proximity
to a surface.[22] Although both contact and
tapping modes have been successfully used to perform TERS measurements,[14,16,23] to the best of our knowledge,
the influence of the AFM feedback mode on molecular perturbation in
chemical analysis using TERS has never been investigated.In
this work, we perform a comparative study of TERS measurements
using contact and tapping mode AFM. Both AFM modes are applied on
a self-assembled monolayer (SAM) of 2-chloro-4-nitrobenzene-1-thiol
(Cl-NBT) using TERS probes with a similar Au coating. Intuitively,
we expected a higher TERS signal in contact mode AFM-TERS because
of the closer proximity of the probe to the sample. Nevertheless,
interestingly, a consistently higher TERS signal is observed in tapping
mode compared to contact mode. After a detailed analysis of Cl-NBT
vibrational modes supported by density functional theory (DFT) calculations,
we attribute the higher TERS signal in tapping mode to the lower molecular
perturbation of the Cl-NBT SAM. Our results highlight a subtle but
important role of the AFM feedback in choosing the correct operational
mode for chemical analysis of soft and delicate molecular samples
using TERS.
Experimental Section
TERS System
TERS experiments were
performed on an NTEGRA
Spectra II TERS system (NT-MDT Spectrum Instruments, Russia) equipped
with an AFM and a Raman spectrometer in top illumination geometry.
A schematic representation of the setup is depicted in Figure . A 632.8 nm He–Ne laser
(LASOS, Germany) was focused onto the sample using a 100×, 0.9
NA objective (Olympus, Japan). Unless otherwise specified, a laser
power of 10 μW was utilized on the sample surface for TERS measurements
with a spectrum acquisition time of 2 s. In contact mode, a vertical
force of ∼9 nN was applied to the sample (see Table S1 for details). In tapping mode, a free oscillation
amplitude of ∼100 nm was imposed and the oscillation amplitude
of the probe in contact with the sample was set to 70% of the free
oscillation. After bringing the probe in contact with the sample,
the oscillation amplitude was rechecked through amplitude–distance
curves. In tapping mode, a maximum vertical force of ∼1570
nN was applied to the sample (see Table S2 for details). Note that although the maximum vertical force in the
tapping mode is nearly 2 orders of magnitude higher than the contact
mode, the lateral force is expected to be significantly higher in
contact mode,[24] potentially resulting in
a substantial disruption of analyte molecules.
Figure 1
Schematic representation
of the TERS setup used to study molecular
perturbation effects in chemical analysis using TERS.
Schematic representation
of the TERS setup used to study molecular
perturbation effects in chemical analysis using TERS.
TERS Probes
ATEC-NC and ATEC-CONT Si AFM cantilevers
(NanoAndMore, Germany) were used to prepare contact mode and tapping
mode TERS probes. To increase the refractive index of the surface,[25] Si cantilevers were oxidized in a furnace (Carbolite
Gero, U.K.) at 1000 °C for 22.5 h to obtain a SiO2 layer of ∼300 nm. Before metal coating, oxidized cantilevers
were cleaned inside a UV–ozone cleaner (Ossila, United Kingdom)
for 1 h. Probes were then coated with Au (99.99%, Acros Organics)
to a nominal thickness of 100 nm at 0.02 nm/s and a pressure of <10–6 mbar inside a thermal evaporation system (MBraun,
Germany). To prepare contamination-free TERS probes with high plasmonic
sensitivity, the entire thermal evaporation system was placed inside
a nitrogen glovebox (MBraun, Germany) with <0.1 ppm of oxygen and
moisture.
Preparation of Cl-NBT SAM on Au
Ultraflat Au substrates
were prepared using the template-stripping method.[26] Cl-NBT SAM was constructed by soaking the Au substrates
within a 5 mM solution of Cl-NBT in ethanol for 12 h. Cl-NBT SAM samples
were rinsed with a copious amount of ethanol and dried under nitrogen
before use.
Scanning Electron Microscopy (SEM)
SEM imaging was
performed using a Hitachi SU5000 instrument (Hitachi, Japan) with
an acceleration voltage of 5 kV and a current intensity of 0.1 nA.
SEM imaging of the probes was carried out after TERS measurements.
Density Functional Theory (DFT) Calculations
Raman
vibrational modes of Cl-NBT were simulated with Gaussian 9.0 and
GaussView 5.0 softwares (Gaussian Inc.) using B3LYP hybrid functional
and the 6-31G(+) basis set. Orientation-dependent DFT simulations
of the Raman spectra of Cl-NBT and Cl-NBT-Au were performed using
in-house-developed scripts in MATLAB R2021a.
Data Analysis
AFM and SEM images were processed using
Gwyddion software (gwyddion.net). Spectral data were analyzed and
plotted using a custom-made Python routine. TERS spectra were background-subtracted
using an asymmetric-least-squared algorithm prior to Gaussian fitting
of peaks as illustrated in Figure S1. TERS
signal intensity was evaluated from the fitted peak height.
Results
and Discussion
The Cl-NBT SAM used in this study produces
a strong TERS signal[27] however, it is also
deformable[28] making it a good model system
for the investigation of
the probe-sample interaction. An AFM topography image of the Cl-NBT
SAM is shown in Figure a, which shows a smooth topography with RMS roughness of 0.6 nm.
To rule out the possibility of any surface-enhanced Raman spectroscopy
(SERS) contribution from the Au substrate, far-field Raman imaging
of the sample was first performed. However, no Raman bands were observed
as shown in the far-field spectrum in Figure S2, confirming the SERS inactivity of the Au substrate. The Au coating
of the probes was examined using SEM, as shown in Figures b,c, S3, and S4 (top panels). All TERS probes showed a similarly homogeneous
Au coating. Moreover, the contact mode and tapping mode TERS probes
exhibited a similar apex diameter as listed in Table S3.
Figure 2
(a) AFM topography image of the Cl-NBT SAM on a Au substrate.
SEM
images of representative Au-coated (b) contact mode and (c) tapping
mode TERS probes. (d) TERS image (20 × 20 pixels) of the 1336
cm–1 signal intensity measured using tapping mode
AFM feedback. (e) Histogram of signal intensity in the TERS image
shown in (d). Averaged spectra of the TERS images of the Cl-NBT SAM
on Au measured using five different probes in (f) contact mode and
five different probes in (g) tapping mode. Each spectrum represents
an average of 400 spectra measured within a TERS image. Far-field
(FF) confocal Raman spectra measured with the TERS probes retracted
from the sample and confocal Raman spectrum of bulk Cl-NBT are also
plotted for comparison. Both (f) and (g) share the same y-axis; the spectra are vertically shifted for easier visualization.
(a) AFM topography image of the Cl-NBT SAM on a Au substrate.
SEM
images of representative Au-coated (b) contact mode and (c) tapping
mode TERS probes. (d) TERS image (20 × 20 pixels) of the 1336
cm–1 signal intensity measured using tapping mode
AFM feedback. (e) Histogram of signal intensity in the TERS image
shown in (d). Averaged spectra of the TERS images of the Cl-NBT SAM
on Au measured using five different probes in (f) contact mode and
five different probes in (g) tapping mode. Each spectrum represents
an average of 400 spectra measured within a TERS image. Far-field
(FF) confocal Raman spectra measured with the TERS probes retracted
from the sample and confocal Raman spectrum of bulk Cl-NBT are also
plotted for comparison. Both (f) and (g) share the same y-axis; the spectra are vertically shifted for easier visualization.To compare TERS signals in contact and tapping
modes, we carried
out TERS imaging of the Cl-NBT SAM on Au in each mode with five different
probes in 2 μm × 2 μm areas with 20 × 20 pixels
(400 spectra). A representative TERS image of the 1336 cm–1 signal (−NO2 stretching mode) intensity measured
in the tapping mode is shown in Figure d. Rather consistent TERS spectra of the sample were
recorded throughout the image, demonstrating the stability of the
TERS system. Nonetheless, a spatial variation in the signal intensity
is observed across the TERS image, possibly coming from the variation
in the plasmonic resonance of the Au-coated tip-Au substrate geometry.
A similar behavior is observed in the TERS images measured with other
tapping mode and contact mode probes as shown in Figures S3 and S4 (middle panels), respectively. Note that
spurious signals, possibly from organic contaminants, were also observed
in some spectra in the TERS images measured using both tapping and
contact modes, as shown in Figure S5. However,
since such spurious signals are observed randomly, their contribution
to the average spectrum is minimal. A histogram of the TERS signal
intensity measured in Figure d is shown in Figure e, which shows only a small intensity variation indicating
good probe stability and consistency of the tip-substrate plasmon
resonance. Intensity histograms of other TERS images measured in tapping
and contact modes are shown in Figures S3 and S4 (bottom panels).Average spectra of the TERS images
measured in tapping and contact
modes are plotted in Figure f,g, respectively, together with the far-field Raman spectra
of Cl-NBT SAM and Raman spectrum of bulk Cl-NBT. All TERS spectra
show a very good correlation of the band positions with the Raman
spectrum of bulk Cl-NBT. A summary of the TERS results is shown in Figure . Figure a shows a comparison of the
average intensity of the 1336 cm–1 signal measured
in TERS images (400 spectra in total) using five different probes
in tapping and contact modes, respectively. Small error bars in the
TERS signal of most probes indicate the stability of plasmonic signal
enhancement of the Cl-NBT SAM on Au. Interestingly, in tapping mode,
a higher TERS signal is consistently observed for all five probes
compared to the contact mode. This trend is even more clear in Figure b, where in the comparison
of average TERS spectra measured in all TERS images (2000 spectra),
TERS signal in tapping mode is found to be 2.5 times higher than
contact mode. Despite the difference in signal intensity, the average
TERS spectra measured in contact and tapping modes showed no significant
difference from the far-field Raman spectrum in terms of peak positions
and relative peak intensity, as shown in Figure S6.
Figure 3
(a) Comparison of average TERS signal intensity measured with contact
and tapping mode feedback. Each TERS intensity value represents an
average intensity of the 1336 cm–1 signal in 400
spectra measured in a TERS image. Error bars represent the standard
deviation of the signal over the TERS image. (b) Plot showing an average
of 2000 TERS spectra measured using five different TERS probes in
contact and tapping modes, respectively. A simulated Raman spectrum
of Cl-NBT is also plotted for comparison. (c) Same as plot (a), with
the values for tapping mode TERS measurements normalized for the time
the TERS probe spends in close proximity to the sample, as discussed
in Figure S7. (d) Comparison of the average
TERS signal in 2000 TERS spectra measured using contact and tapping
modes. Error bars represent the standard deviation of the signal in
all TERS spectra measured with a particular feedback mode. “Tapping
(normalized)” represents the contact-time-normalized TERS signal
in tapping mode.
(a) Comparison of average TERS signal intensity measured with contact
and tapping mode feedback. Each TERS intensity value represents an
average intensity of the 1336 cm–1 signal in 400
spectra measured in a TERS image. Error bars represent the standard
deviation of the signal over the TERS image. (b) Plot showing an average
of 2000 TERS spectra measured using five different TERS probes in
contact and tapping modes, respectively. A simulated Raman spectrum
of Cl-NBT is also plotted for comparison. (c) Same as plot (a), with
the values for tapping mode TERS measurements normalized for the time
the TERS probe spends in close proximity to the sample, as discussed
in Figure S7. (d) Comparison of the average
TERS signal in 2000 TERS spectra measured using contact and tapping
modes. Error bars represent the standard deviation of the signal in
all TERS spectra measured with a particular feedback mode. “Tapping
(normalized)” represents the contact-time-normalized TERS signal
in tapping mode.In TERS geometry of a
metallic tip in contact with a metal substrate,
also known as “gap mode”, the plasmonically enhanced
optical field is confined to a nanoscopic volume. It has been well
established from a number of experimental and theoretical TERS studies
that the gap-mode, TERS enhancement decreases dramatically over a
distance of ca. 3 nm[29] and becomes negligible above 5 nm,[30] as
summarized in Table S4. Tapping mode TERS
probes typically oscillate with a frequency of 210 490 kHz spending
only a fraction of time in contact with the Cl-NBT SAM in contrast
to contact mode, in which TERS probes remain virtually stationary
and in contact with the sample at all times.[31] Therefore, the observation of a higher TERS signal in tapping mode
shown in Figure a,b
is counterintuitive and surprising. In our TERS experiments, tapping
mode probes typically oscillate with an amplitude of ∼70 nm.
Assuming a harmonic oscillator model, the time spent by tapping mode
probes in close contact (i.e., within less than 5 nm) with the sample
is ∼17.2% of the total oscillation period, as illustrated in Figure S7. This implies that for the same spectrum
integration time, the actual time contributing to the near-field signal
in tapping mode is ∼6 times shorter compared to contact mode.
Normalizing the tapping mode TERS signal intensity with this factor
reveals an even higher signal compared to the contact mode as shown
in Figure c. A comparison
of the average signal in 2000 TERS spectra measured in tapping and
contact modes is shown in Figure d, where the contact-time-normalized TERS signal intensity
in tapping mode is found to be 13 times higher than contact mode.This result is confounding since for the same integration time
a higher TERS signal is expected in contact mode as the continuous
contact of the probe with the sample will allow a greater sampling
of plasmonically enhanced signal. However, contrary to this expectation,
a significantly higher TERS signal is observed in tapping mode under
identical measurement conditions as shown in Figure . To understand this, we need to consider
the probe-sample interaction in contact and tapping mode AFM-TERS.
Despite careful control of the measurement parameters, soft samples
have been shown to alter during AFM imaging in contact mode.[28] We tested this hypothesis in our own setup by
performing the AFM imaging of a 1,2-dioleoyl-sn-glycero-3-phosphocholine
(DOPC) bilayer lipid grating sample in contact and tapping modes,
which is shown in Figure S8. A severe disruption
of the surface features is observed in the contact mode (Figure S8a), whereas the lipid grating remains
intact in the tapping mode AFM image (Figure S8b) indicating a minimal perturbation of molecular arrangement. On
the other hand, the same phenomenon is not observed during the imaging
of a silicon grating in contact and tapping modes, which is a relatively
hard sample as shown in Figure S8c,d.The AFM imaging results of a lipid grating may not be directly
comparable to a thiol SAM; however, perturbation of the molecular
arrangement of thiol molecules in the contact mode AFM imaging cannot
be ruled out. In fact, the thiol SAM does not necessarily have to
be permanently removed or modified to produce a relatively lower TERS
signal. Our group has previously shown that molecular orientation
can have significant influence on the intensity of bands in a TERS
spectrum.[32] To confirm this, we performed
DFT simulation of the Cl-NBT Raman spectrum, which is shown in Figures b and S9. The simulated Raman spectrum corroborates
the measured TERS spectra very well both in terms of the peak position
and relative peak intensity. The assignment of the TERS and simulated
Raman bands is listed in Table . Analysis of the simulated vibrational modes reveals that
all major bands of the spectrum, i.e., ring-breathing mode at 1135
cm–1, −NO2 stretching mode at
1336 cm–1, and C=C stretching mode at 1567
cm–1, represent in-plane modes as shown in Figure a–c and the Supporting MP4 Files. In a gap-mode TERS configuration,
since the plasmonically enhanced optical field has a vertical polarization,
these in-plane Raman modes will experience a strong enhancement when
the plane of vibration is vertically aligned.[33,34] We tested this hypothesis with DFT calculation of the orientation-dependent
Raman spectra of Cl-NBT molecules under irradiation of vertically
polarized light. These results are presented in Figure d,e, which clearly show a rapid decrease
in the intensities of all major Raman bands as the molecule bends
toward the substrate surface. Furthermore, the result did not change
even when the thiol molecule was bound to a Au atom, as shown in Figure S10.
Table 1
Assignment of the Main Bands in the
TERS and Simulated Raman Spectra of Cl-NBT; See Figure S9 for Further Details
TERS signal
(cm–1)
simulated
signal (cm–1)
assignment
1118
1100
C–N stretching
1135
1116
ring breathing
1242
1249
C–H rocking
1336
1336
–NO2 stretching
1567
1574
C=C stretching
Figure 4
Representation of the in-plane (a) ring-breathing
mode at 1135
cm–1, (b) −NO2 stretching mode
at 1336 cm–1, and (c) in-plane C=C stretching
mode at 1567 cm–1 calculated with DFT simulations.
DFT simulations of Cl-NBT Raman spectra at different orientation angles
with respect to the substrate surface, for molecules lying (d) sideways
or (e) flat at 0°. The polarization of incident light is aligned
with z-axis, which is indicated by a red arrow. In
both cases, the intensity of all Raman bands decreases rapidly as
the molecule bends toward the substrate surface.
Representation of the in-plane (a) ring-breathing
mode at 1135
cm–1, (b) −NO2 stretching mode
at 1336 cm–1, and (c) in-plane C=C stretching
mode at 1567 cm–1 calculated with DFT simulations.
DFT simulations of Cl-NBT Raman spectra at different orientation angles
with respect to the substrate surface, for molecules lying (d) sideways
or (e) flat at 0°. The polarization of incident light is aligned
with z-axis, which is indicated by a red arrow. In
both cases, the intensity of all Raman bands decreases rapidly as
the molecule bends toward the substrate surface.The experimental and
theoretical results strongly indicate that
in the tapping mode TERS imaging, the Cl-NBT SAM remains largely unperturbed
in which the molecules lie at an angle of 60–70°[35] and a large component of the polarizability
tensor of in-plane vibrational modes stands parallel to the vertical
polarization of the near-field as schematically illustrated in Figure a. However, in contact
mode TERS imaging of the Cl-NBT SAM, direct continuous contact of
the probe with the sample likely perturbs the molecular arrangement
by flattening the molecules, thereby reducing the vertical component
of their polarizability tensor as well as the number of molecules
in the near-field as illustrated in Figure b. The TERS intensity of out-of-plane vibrational
modes, for example the out-of-plane wagging at 807 cm–1 shown in Figure S9 and animated files
available in the Supporting Information, should increase when the molecules are bent by large angles or
flattened. However, these vibrational modes are not detectable in
the experimental spectrum because of a very low cross section as shown
by DFT simulations.
Figure 5
(a) In tapping mode, a minimal molecular disruption
results in
a higher number of molecules in the near-field as well as a greater
plasmonic enhancement of the vertical component of in-plane vibrational
modes. (b) In contact mode, the direct contact of the probe with the
sample can laterally disrupt the thiol SAM, decreasing the number
of molecules inside the near-field as well as changing their orientation.
(a) In tapping mode, a minimal molecular disruption
results in
a higher number of molecules in the near-field as well as a greater
plasmonic enhancement of the vertical component of in-plane vibrational
modes. (b) In contact mode, the direct contact of the probe with the
sample can laterally disrupt the thiol SAM, decreasing the number
of molecules inside the near-field as well as changing their orientation.The noninvasiveness of the tapping mode TERS measurements
observed
in our study corroborates similar results obtained by Wang et al. on a different molecular system, where intermittent
contact mode was found to be relatively more invasive for molecular
analysis.[36] However, in our contact mode
TERS measurements (Figures f and S6), we did not consistently
observe any new Raman bands or the Stark-shifting or broadening of
the existing Raman bands, which have been previously attributed to
molecular charging,[37] chemical transformation[36] or optical rectification effects.[38] Furthermore, in the last decade, a number of
experimental and theoretical studies have demonstrated that the nanogap
plasmons within a metallic tip-sample junction are governed by quantum
mechanical effects including electron tunneling and nonlocal screening.[39] Particularly, when the tip-sample nanogap becomes
≤1 nm, electron tunneling could lead to a decrease in the intensity
of EM field enhancement.[40−42] For example, Zhu et al. observed quenching of SERS enhancements of thiophenol molecules
adsorbed on Au nanoparticles below an interparticle gap distance of
0.67 nm due to the electron tunneling effect.[40] In our study, the length of the molecular reporter, Cl-NBT, is ∼0.74
nm. Although precise determination of the tip-sample nanogap distance
in contact mode is extremely challenging, we expect it to be between
0.74 and 1 nm in our experiments. Therefore, in contact mode TERS
measurements, the possibility of quenching of electromagnetic field
enhancement due to electron tunneling cannot be completely ruled out.We propose that in contact mode TERS imaging, the molecular disruption
of the Cl-NBT SAM and the possible electron tunneling effect cause
a lower signal compared to the tapping mode. Therefore, for nanoscale
chemical analysis of “soft samples” such as SAMs of
organic molecules or lipid monolayer/bilayer systems in their native
state using TERS, tapping mode should be the method of choice because
of a lower degree of molecular perturbation as well as lower quantum-mechanical
EM field quenching, and, consequently, higher signal intensity.
Conclusions
In this work, we have performed a comparative study of the influence
of AFM feedback on molecular perturbation in chemical analysis using
TERS. Despite the conventional belief that “closer is better”
in terms of TERS signal enhancement, the data presented herein show
that this approach could be counterproductive for soft samples. A
consistently higher TERS signal is obtained in tapping mode despite
the probe spending 6-fold less time in close proximity to the sample
than contact mode. Comparative TERS imaging of soft and hard grating
samples indicates that a larger perturbation of the organization and
orientation of the probed molecules in contact mode, possibly in
combination with electron tunneling induced EM field quenching, diminishes
the TERS signal of the probed species. Permanent damage has not been
proven but appears to be a likely scenario when working in contact
mode, particularly for soft samples that are not chemically bound
to the substrate surface. Note that, caution should be exercised in
directly applying the results of this study to other “soft”
samples as the precise TERS signal difference between contact and
tapping modes may be affected by several factors including sample
thickness, sample adhesion to the substrate, tip-sample distance,
etc. Nevertheless, this study clearly demonstrates that although both
tapping and contact modes are capable of producing a TERS signal with
a high signal-to-noise ratio, the degree of molecular perturbation
could be significantly different. Therefore, for the TERS measurements
of soft samples, tapping mode should be the preferred method of choice.
The novel insights gained in this study are expected to accelerate
the growing application of TERS to nondestructive and label-free nanoscale
chemical analysis of delicate samples.