The toolset of mass spectrometry (MS) is still expanding, and the number of metal ion complexes researched this way is growing. The Cu(II) ion forms particularly strong peptide complexes of biological interest which are frequent objects of MS studies, but quantitative aspects of some reported results are at odds with those of experiments performed in solution. Cu(II) complexes are usually characterized by fast ligand exchange rates, despite their high affinity, and we speculated that such kinetic lability could be responsible for the observed discrepancies. In order to resolve this issue, we selected peptides belonging to the ATCUN family characterized with high and thoroughly determined Cu(II) binding constants and re-estimated them using two ESI-MS techniques: standard conditions in combination with serial dilution experiments and very mild conditions for competition experiments. The sample acidification, which accompanies the electrospray formation, was simulated with the pH-jump stopped-flow technique. Our results indicate that ESI-MS should not be used for quantitative studies of Cu(II)-peptide complexes because the electrospray formation process compromises the entropic contribution to the complex stability, yielding underestimations of complex stability constants.
The toolset of mass spectrometry (MS) is still expanding, and the number of metal ion complexes researched this way is growing. The Cu(II) ion forms particularly strong peptide complexes of biological interest which are frequent objects of MS studies, but quantitative aspects of some reported results are at odds with those of experiments performed in solution. Cu(II) complexes are usually characterized by fast ligand exchange rates, despite their high affinity, and we speculated that such kinetic lability could be responsible for the observed discrepancies. In order to resolve this issue, we selected peptides belonging to the ATCUN family characterized with high and thoroughly determined Cu(II) binding constants and re-estimated them using two ESI-MS techniques: standard conditions in combination with serial dilution experiments and very mild conditions for competition experiments. The sample acidification, which accompanies the electrospray formation, was simulated with the pH-jump stopped-flow technique. Our results indicate that ESI-MS should not be used for quantitative studies of Cu(II)-peptide complexes because the electrospray formation process compromises the entropic contribution to the complex stability, yielding underestimations of complex stability constants.
Complexation
to peptides is proposed to play important roles in
Cu(II) physiology and toxicology. Cu(II) activates GHK, a wound healing
factor[1,2] and α-factor, a yeast pheromone,[3,4] and is likely transported to neurons by neurokinin B.[5] It also elicits toxicity and probably gets detoxified
by some variants of Aβ peptides[6,7] and protamine
HP2[8−10] and likely participates in the antifungal action of histatins, salivary
antimicrobial peptides.[11] A recent study
indicated that more peptides with such properties remain to be identified
in human proteome.[12] Peptides have also
been used extensively to model Cu(II) binding to its transport proteins,
such as albumin and hCtr1 membrane transporter,[13−15] and synaptic
proteins, such as prions, APP, and α-synuclein.[16,17] Two N-terminal sequence motifs, Xaa-His and Xaa-Zaa-His (where Xaa
is any α-amino acid except of Cys, and Zaa is any α-amino
acid except of Cys or Pro), provide the highest Cu(II) complex affinities
by virtue of synergistic formation of chelate rings involving peptide
nitrogen atoms (Figure ).[18−20] The logarithmic conditional stability constants at
physiological pH 7.4, log CK, for Xaa-His complexes are in the range of 12.5–13,
while those of Xaa-Zaa-His complexes range from 12.3 to ca. 15.[21,22] The latter are also known as ATCUN or NTS complexes.[18]
Figure 1
Structures of Cu(II) complexes at the N-terminal site
of peptides
having His in the 2nd (B) or 3rd position (A)
and (C). R(n) mark side chains and the remainder of the
peptide chain.
Structures of Cu(II) complexes at the N-terminal site
of peptides
having His in the 2nd (B) or 3rd position (A)
and (C). R(n) mark side chains and the remainder of the
peptide chain.Mass spectrometry (MS) in its
many variants is one of the most
versatile techniques of peptide and protein research, providing information
on their composition, sequence, and post-translational modifications
and via hyphenated techniques also on protein structure. It has also
been widely used to determine affinity constants of molecular complexes,
using a number of experimental approaches,[23−27] including studies of metal complexes.[28] A significant work was devoted to developing
best conditions for quantitative determinations, including optimization
of spray formation.[29,30]Under certain conditions,
MS can also be used to obtain quantitative
information on metal ion complexes, e.g., Zn(II), Cu(I) or As(III).[31−34] It has also been used to determine the binding constants for Cu(II)
complexes.[34−36]One issue to consider is the rate at which
the charged molecular
ions are generated with respect to parallel changes in their environment.
During the electrospray formation the time from injection of the sample
from the capillary to the droplet fission takes a few milliseconds.[37−39] Many metal ion complexes remain intact in this time scale due to
their sufficient inertness in ligand exchange reactions.[40] However, as recently demonstrated by stopped-flow
and freeze-quench techniques, the formation of Cu(II) complexes with
ATCUN peptides Gly-Gly-His and Aβ4–16 proceeds
via a reactive intermediate step whose lifetime is in the hundred
milliseconds range.[20,41] The Cu(II) ion in this species
is partially coordinated to the amine and imidazole nitrogens only
(Figure A) and is
considered to be kinetically labile. Such lability may in turn affect
the apparent binding constant determination by exposing partially
folded/coordinated complexes to the gas phase. Moreover, alternative
complex stoichiometries can occur. For example, Matsumoto et al. discovered
Cu(II) complexes that emerged rapidly in the gas phase but were not
present in solution.[42]In addition,
the presence of stable, e.g., “physiological”,
conditions in the gas phase is very debatable. Protein and peptide
structures are maintained through covalent bonds, hydrogen bonds,
and hydrophobic interactions with a strong contribution from the solvent
entropy.[43,44] The pH and ionic strength cannot be defined
anymore for the molecule after its transition to gas phase. Protonation
states are altered[45] and hydrophobic interactions
get lost altogether, along with the evaporation of solvating water
molecules.[46] In contrast, electrostatic
interactions, hydrogen bonds and van der Waals interactions are enhanced
in the gas phase, prompting the loss of native structure.[46] The MS measurement time scale is generally considered
as too short for the protein or peptide unfolding, because it requires
a concerted breaking of multiple noncovalent bonds.[47,48] Some proteins may actually unfold in less than 1 ms, however.[49] Metal binding to a peptide contributes only
a few coordination bonds, typically two to six, and perhaps several
weaker interactions in the second coordination sphere. The loss of
even one of them due to gas phase conditions will have a profound
effect on the complex stability. This is particularly important for
peptidic Cu(II) complexes in which the entropic contribution to stability
is very significant.[50,51]Next, the equilibrium could
be shifted because the pH of electrospray
droplets might be different from that set in the sample prior to the
analysis. Finding an appropriate buffer for ESI-MS studies is a difficult
task.[52] Most buffers that maintain physiological
pH do not qualify because their ionic character suppresses the signal.
Ammonium acetate is commonly used as a “buffer” for
so-called native ESI-MS experiments due to volatility of neutral forms
of its components, acetic acid and ammonia, which form upon the spray
evaporation and do not contribute to ionic noise. However, as pointed
out by Konermann, ammonium acetate is not really a buffer at pH 7.[52] It has two buffering areas around the pK values of its components, at 4.75 ± 1 for acetate
and 9.25 ± 1 for ammonia. Hence, the rising number of H+ ions produced by electric field in shrinking solvent droplets in
the positive ion measurement mode rapidly decreases the droplet pH,
down to the acetate buffering range of 4.75 ± 1. Under these
circumstances there is no straightforward way of determining how exactly
the pH has changed during the measurement. This poses a serious problem
for quantitative analysis because the binding constants may change
drastically with pH (see Table ). The situation is further aggravated when mixtures of water
and organic solvents are used, making the pH even more difficult to
ascertain.[53] The discrepant effects of
droplet acidification and other effects mentioned above on the characterization
of metal ion complexes have been noted, e.g., for relatively weak
complexes of alkaline earth metal ions with EDTA,[54] or lanthanide complexes with acetate,[55] but also for very tightly bound species, e.g., the Bi3+ complex of transferrin.[56]
Table 1
Comparison of Conditional Cu(II) Binding
Constants of the Studied Peptides at pH = 7.4 (CK7.4), Compared to These Constants Determined
Previously by Other Methods and Publisheda
peptide sequence (name)
published log CK7.4b
log Kd ± SD
from serial dilution ESI-MS experiment in this study
log ratio log Kd (ESI-MS)
– log CK7.4 (published)
range of concentrations in this study (μM)
DTHFPI-NH2(hepc6)
14.7c
7.6 ± 0.7
–7.1
100–0.01
MNH-NH2
14.5d
6.3 ± 0.4
–8.2
100–2
FRHDSG (Aβ4–9)
14.2e
6.9 ± 0.9
–5.3
100–1
FRHDSGYEVHHQK-NH2 (Aβ4–16)
13.5f
7.0 ± 0.2
–6.5
100–0.5
MDH-NH2
13.1d
7.1 ± 0.6
–6.0
100–1
GGH
12.2g
6.0 ± 0.5
–6.2
100–10
Standard
deviation of ESI-MS
experimental values are given in parentheses. Logarithmic values are
given for better clarity. The published log CK7.4 values were determined by potentiometry and corroborated
by spectroscopic methods.
The SD values were considerably
less than 0.1 log unit in all cases.
Reference (63).
Reference (64).
Reference (65).
Reference (6).
Reference (21).
Standard
deviation of ESI-MS
experimental values are given in parentheses. Logarithmic values are
given for better clarity. The published log CK7.4 values were determined by potentiometry and corroborated
by spectroscopic methods.The SD values were considerably
less than 0.1 log unit in all cases.Reference (63).Reference (64).Reference (65).Reference (6).Reference (21).Other available
buffers offer little alternative. Ammonium bicarbonate
has pKa = 6.4 but is unstable and prone
to CO2 evolution, while common biologically friendly buffers,
like Tris or HEPES, contain nonvolatile cations and anions, resulting
in a poor MS data quality.Furthermore, redox-prone metal complexes,
including the Cu(II)
species, may undergo reduction in the gas phase. This effect is independent
of the method of ionization, as such behavior was reported in both
electrospray and plasma desorption studies.[57,58] It is difficult to monitor and can influence quantitative studies,
as the copper oxidation state is crucial for the complex formation
and stability. For example, the ESI-MS experiments on the Cu(II)–GHK
peptide complex yielded a mixture of Cu(II) and Cu(I) species in a
proportion depending on the electrostatic potential in the ion source.[59] Tsybizova et al. listed three possible mechanisms
for Cu(II) reduction: reduction on the capillary walls, desolvation
with electron transfer and reduction while still in the solution.[60] Reduction was more likely to occur when the
Cu(II) ion was coordinated by no more than two atoms.The unpredictability
of issues with reduction and pH shift is aggravated
by an uncertainty of the measurement time scale which depends on the
droplet size. As a general rule, the bigger the droplet is, the longer
time it takes to dissipate, extending the opportunity for unwanted
processes to occur.[61]In our laboratory
practice, we frequently recorded ESI-MS spectra
of Cu(II) complexes of peptides. Nearly always strong signals of unbound
peptide were observed, although they were not present in equilibrium
(see Figure for examples).
This strongly indicated that complex dissociation occurred in the
mass spectrometer, yielding artifacts. On the other hand, ESI-MS is
often treated, e.g. by less experienced researchers as if it faithfully
reflected the solution equilibria. The goal of our study was therefore
to systematically explore the issue of a systematic error in attempts
to study Cu(II)/peptide complexes quantitatively by ESI-MS. In order
to achieve it, we performed ESI-MS titration experiments on six high-affinity
Cu(II)/peptide systems which we previously characterized in a comprehensive
fashion.
Figure 2
Theoretical dissociation curves of Cu(II)/peptide 1:1 complexes
at pH 7.4.
Theoretical dissociation curves of Cu(II)/peptide 1:1 complexes
at pH 7.4.We started with serial dilution
experiments on a standard ESI-MS
instrument in order to obtain apparent affinity constants serving
for quantitative illustration of the scale of the systematic error
generated by ESI-MS. This was followed by pH–jump kinetic experiments
aimed to model events occurring in the evaporating spray droplets
and by additional competition-based affinity studies performed on
an ESI-MS instrument dedicated for studies of noncovalent interactions.
All these experiments yielded a clear and unequivocal view that ESI-MS
should not be used for quantitation of Cu(II)/peptide complexes.
Experimental
Section
l-Histidine and ammonium acetate were purchased
from Sigma.
All peptides, except for GGH purchased from Sigma, were synthesized
in-house with standard Fmoc solid phase synthesis, as described before.[62] Crude synthesis products were purified with
RP-HPLC on ACE C18-300 column 250 × 8 mm with a rising gradient
of acetonitrile in water with 0.1% TFA.ESI-MS spectra were
recorded on a Premier ESI-QToF spectrometer
(Waters). All measurements were performed in the positive ion mode.
The source temperature 80 °C was used for a complete desolvation
of the peptide ions. The cone voltage was 10 V for shorter peptides
and 30 V for Aβ4–16. The transmission of the
ions was optimized on the quadrupole for the required mass range (m/z 200 to 1000 for shorter peptides, m/z 300 to 1500 for Aβ4–16). Mass spectra were accumulated over 2 to 3 min to improve the signal-to-noise
ratio. The sample flow was 20 μL/min. The 100 μM Cu(II)-peptide
solutions in 20 mM ammonium acetate at pH 7.4 were used for serial
dilutions of the whole complex with the same ammonium acetate. For
quantitation, all peaks corresponding to each charged state were integrated
for each peptide. Ratios of free peptide to Cu(II)/peptide complex
were extracted from the integration of MS intensities with the assumption
of similar ionization efficacy. For each dilution the Kd was calculated according to eq , followed by the averaging as required (see
the Results for details).The free Cu2+ ions could not be
detected directly, due to low m/z and instead were calculated from the mass balance (free Cu2+ assumed to be equal to free peptide).Isotopic distributions
were calculated using built-in MassLynx
Isotope model software (Waters).Competition experiments with
histidine were carried out on Q Exactive
UHMR Hybrid Quadrupole-Orbitrap mass spectrometer (ThermoFisher Scientific).
Samples were prepared by dissolving peptides in 50 mM ammonium acetate
pH 7.4 in 100 nM concentration. To each sample a mixture of 100 nM
CuCl2 and differing concentrations of histidine were added.
Measurements started with an 1 h delay to allow for reaching equilibrium.
Samples were introduced into the mass spectrometer with a syringe
pump using 10 μL/min flow rate, by electrospray ionization using
positive mode in HESI source. MS measurements were conducted under
the following settings: desolvation voltage: 20 V, capillary temperature
320 °C, detector m/z optimization:
low m/z; ion transfer optimization
to low m/z. The RF applied throughout
the instrument were set to 150 Vp-p for injection flatapole, 300 Vp-p
for bent flatapole, 250 for transfer multipole and HCD cell, and 2300
for C-trap. The ions transfer optics was to 5 V for injection flatapole,
4 V for intel flatapole, 2 V for bent flatapole and 0 V for transfer
multipole. Integration of the resulting peaks was achieved with built-in
software FreeStyle 1.4 (Thermo Scientific).The stopped-flow
SFM-300 (BioLogic) instrument was used to measure
the rates of dissociation of Cu(II)–GGH and Cu(II)–hepc6
complexes upon acidification. The kinetic runs were observed with
a diode-array detector (TIDAS S 500 K, J&M Analitik AG), with
the spectra recorded in the 400–900 nm wavelength range at
1.5 ms intervals. Reactions were performed in a 1 cm path length cuvette
at 25 °C. The dead time of the instrument was 2 ms at total flow
rate 15 mL/min.The pH–jump experiments were carried
out for solutions of
2 mM peptide and 1.8 mM CuCl2 in 50 mM ammonium acetate
at pH = 7.4 mixed with equal volumes of 45 mM, 0.25 M, or 2.5 M acetic
acid, which resulted in sample acidification to pH 5, 4, or 3, respectively.
The dilution during the mixing yielded final cuvette concentrations
of 1 mM peptide, 0.9 mM Cu(II), and 22.5 mM, 125 mM and 1.25 M acetic
acid, respectively. The solutions were freshly prepared and degassed
before each series of reactions. The final pH value was measured in
the samples collected after each experiment. To show the differences
in absorbance signal before the first recorded time point, the dilution
of the Cu(II)–peptide complex was measured as a control (shown
as t = 0 s). In all cases, the peptides were in a
slight excess over Cu(II) to avoid Cu(OH)2 precipitation.
Results
ESI-MS
Attempts at Determining CK
To test
the pseudophysiological ESI-MS conditions for the binding constant
determination we employed two different mass spectrometers and two
different techniques. Serial dilution studies of Cu–peptide
complexes were performed on a standard ESI-MS instrument, while histidine
competition experiments were performed on a sensitive instrument dedicated
for studies of noncovalent complexes.In the first approach,
we performed measurements on several synthetic peptides with well-established
Cu(II) binding properties. Equimolar Cu(II) peptide mixtures (100
μM) were dissolved in 20 mM ammonium acetate at pH 7.4. The
dissociation of complexes was monitored by serial dilutions of both
reagents. Table provides
the Kd values obtained according to eq for each data point and
averaged. These values were compared to the literature CK7.4 values.[6,21,63−65] Very large discrepancies
were observed, ranging from five to eight log units.The ESI-MS
spectra are presented in Figures S1–S6. As seen in these figures, substantial amounts
of unbound peptides were observed in all cases at all tested concentrations.
For example, ca. 20% free peptide was detected throughout the dilution
experiment for hepc6. This degree of Cu(hepc6) complex dissociation
at pH 7.4 should not occur above a complex concentration of 9 fM (Figure ).[63] Further experiments were undertaken to explain the sources
of this discrepancy.For the peptides forming the weakest (GGH)
and the strongest complexes
(hepc6), different Cu(II)/peptide ratios were analyzed (Figure S7 and S8, respectively). The increasing
Cu(II) amounts should eventually saturate the peptide, eliminating
the apopeptide signals. This did not happen, even at the highest Cu(II)
excess that could be achieved due to solubility, more than hundred-fold
for GGH (Figure S7). As presented in Figure S9, increasing the Cu/peptide ratio lowered
the apparent Kd value, contrary to what
should be expected according to eq .We also analyzed the spectra for the signs
of Cu(II) reduction
to Cu(I) as a possible source of Kd value
deviation. This effect would manifest itself qualitatively in the
ESI-MS spectra as a m/z shift of
+1 for the molecular ions containing Cu(I), because the positive charge
decrease on the copper ion would have to be compensated by an additional
H+ ion attached elsewhere. This, in turn, would affect
the peak proportions in the ionic manifold. This effect was analyzed
by comparing the peak maximum intensities with respect to the monoisotopic
(highest) peak. Indeed, very slight increases of the second and fourth
peaks were observed in most peptides, consistent with the +1 mass
shift, but the effect was small, not exceeding a few percent of the
monoisotopic peak (Figure S10). Thus, some
Cu(II) reduction to Cu(I) could have occurred in our experiments,
but its contribution to Kd was minimal
to negligible.
Competition Experiments with Histidine
Having demonstrated
that serial dilution experiments on a standard ESI-MS instrument inadvertently
yield artifacts, we checked whether competition experiments could
provide a better result. For that we used a state-of-the-art Q-Exactive
Ultra-High Mass Range mass spectrometer designed specially to preserve
native noncovalent bonds. The initial experiment in 50 mM ammonium
acetate, pH 7.4, proved that no artifactual complex dissociation occurred
for the CuAβ4–16 complex at the 100 pM concentration,
near the instrument’s detection limit (Figure S11). According to Figure , this dilution was still 2 orders of magnitude
away from the range at which serial dilution could be applied but
gave promise for competition experiments. In this approach, 100 nM
peptides Aβ4–16 and hepc6 mixed with 100 nM Cu(II) were
titrated with histidine, which served as a weak competitor. With full
knowledge of the protonation states and stability constants for each
complex species in solution (Tables S1 and S2) we were able to calculate the theoretical binding isotherms indicating
a feasibility of such approach. The experimental curves were unfortunately
different (Figure ).
Figure 3
Experimental and theoretical binding isotherms for Cu(II)/Aβ4-16
(A) and Cu(II)/hepc6 (B) 100 nM 1:1 complexes titrated with histidine
as a weak competitor. Dotted lines mark concentration at which 50%
of the peptide is occupied with Cu(II) ion.
Experimental and theoretical binding isotherms for Cu(II)/Aβ4-16
(A) and Cu(II)/hepc6 (B) 100 nM 1:1 complexes titrated with histidine
as a weak competitor. Dotted lines mark concentration at which 50%
of the peptide is occupied with Cu(II) ion.Calculating binding constants from competition experiments where
the competitor forms both 1:1 and 2:1 complexes is not an easy task
in general, and competition may produce errors that are difficult
to account for,[66] but we were able to circumvent
the calculation problem by extracting the histidine concentration
at which 50% of peptides were bound to Cu(II) from Figure . These values were 24.9 μM
His for Aβ4–16 and 93.3 μM His for hepc6. With
these values and the knowledge of all protonation constants of histidine
and stability constants for Cu(II)/histidine complexes[67] we could calculate the binding constants using
the competitivity index (CI) approach.[68,69] The CI for
a binary metal/ligand system is defined as the logarithm of the conditional
stability constant of MZ (the metal complex of a theoretical molecule
Z), such that Σ([MHL]) = [MZ], at given overall component concentrations, where
M is a metal ion, H is hydrogen, and L is the metal binding ligand.
In other words, the apparent stability constant of MZ, in units of
M–1, reflects the overall metal ion binding ability
of all other molecules in equilibrium. CI is equivalent to log CK7.4 when there is only one complex
species under given conditions but has a broader relevance if more
than one species coexist in solution. In such case CI represents the
metal binding properties of the whole ensemble of complex species.
In the studied system L is histidine and Z is the peptide. The data
obtained through these calculations are gathered in Table , and raw spectra are shown
in Figures S12 and S13.
Table 2
Comparison of Conditional Cu(II) Binding
Constants Obtained with Competition Experiments with Histidine of
the Studied Peptides at pH = 7.4 (CK7.4), Compared to These Constants Determined Previously by
Other Methods and Publisheda
peptide sequence (name)
published log CK7.4
log Kd from His
competition ESI-MS experiments in this study
log ratio log Kd (ESI-MS)
- log CK7.4 (published)
range of concentrations of histidine in this
study (μM)
DTHFPI-NH2 (hepc6)
14.66b
13.98
–0.68
1000-10
FRHDSGYEVHHQK-NH2 (Aβ4–16)
13.53c
12.84
–0.69
500-1
Logarithmic
values are given
for better clarity. The published log CK7.4 values were determined by potentiometry and corroborated
by spectroscopic methods.
Reference (63).
Reference (6).
Logarithmic
values are given
for better clarity. The published log CK7.4 values were determined by potentiometry and corroborated
by spectroscopic methods.Reference (63).Reference (6).
Modeling the pH Drop in Electrospray Droplets Using pH–jump
Stopped Flow
The formation of Cu(II)–peptide complexes
in general and ATCUN complexes in particular is very strongly pH-dependent.[18] This is illustrated in Figure for hepc6 and GGH and in Figure S14 for other studied peptides. These diagrams indicate
that the significant coexistence of Cu(II)-bound and unbound peptides
occurs typically in the pH range between 3 and 5. The presence of
unbound peptides in all serial dilution ESI-MS experiments could thus
be a simple consequence of acidification of sample droplets containing
ammonium acetate during their evolution in electrospray.[52]
Figure 4
pH-dependent species distribution of Cu(II) complexes
of hepc6
(top) and GGH (bottom) calculated for 1 mM peptides and 0.8 mM Cu(II)
based on potentiometric data from ref (65) and ref (21), respectively.
pH-dependent species distribution of Cu(II) complexes
of hepc6
(top) and GGH (bottom) calculated for 1 mM peptides and 0.8 mM Cu(II)
based on potentiometric data from ref (65) and ref (21), respectively.Figure provides
the correlation diagram indicating what pH values in electrospray
droplets would explain the apparent loss of Cu(II) affinity if droplet
acidification were the only source of the observed log Kd decrease in serial dilution experiments. Although all
values are within the buffering range of acetate, their spread is
rather random and covers about one pH unit, despite using the same
experimental conditions in all respective experiments. Therefore,
the acidification, while plausible, is not the only source of the
studied effect and the actual Kd values
cannot be simply recovered from ESI-MS data by adjusting them for
a hypothetical electrospray pH value.
Figure 5
Prediction of pH values in electrospray
stipulated by the literature
log CK values compared with the values
calculated from serial dilution ESI-MS (data from Table ). Solid curves represent the
pH dependence of literature log CK values
for individual peptides. Circles mark the ESI-MS-derived log Kd values ± SD placed along these curves
to indicate the pH of spray droplets expected if acidification were
the sole source of the observed Kd decrease.
Gray area marks the buffering zone of ammonium acetate (pH 4.75 ±
1).
Prediction of pH values in electrospray
stipulated by the literature
log CK values compared with the values
calculated from serial dilution ESI-MS (data from Table ). Solid curves represent the
pH dependence of literature log CK values
for individual peptides. Circles mark the ESI-MS-derived log Kd values ± SD placed along these curves
to indicate the pH of spray droplets expected if acidification were
the sole source of the observed Kd decrease.
Gray area marks the buffering zone of ammonium acetate (pH 4.75 ±
1).Cu(II) complexes are usually considered
to equilibrate rapidly,
but our recent study demonstrated that Cu(II) complexes of ATCUN peptides
are formed in a stepwise manner, with the lifetime of most stable
intermediate species around several hundred milliseconds.[20] A typical time of electrospray formation is
≤10 ms.[37,38] These facts prompted us to investigate
whether the complexes had enough time to reach the new pH-dependent
equilibrium within the time of droplet evolution preceding the gas
phase transition. To answer this question, we performed stopped-flow
pH–jump experiments. The samples containing Cu(II)–hepc6
and Cu(II)–GGH complexes in ammonium acetate at pH 7.4 were
mixed with acetic acid solutions of concentrations adjusted to reach
pH 5, 4, or 3 after the sample mixing. The evolution of the systems
was monitored using the visible absorption spectra, as shown in Figure for Cu(II)–hepc6.
The data for Cu(II)–GGH are provided in Figure S15.
Figure 6
Stopped-flow spectra of Cu(II)-hepc6 in 20 mM ammonium
acetate
initially at pH 7.4, subjected to rapid pH decrease by mixing with
concentrated acetic acid. Color-coded spectra were recorded every
1.5 ms with the instrument dead time of 2 ms. (A) First 11 ms of the
reaction and (B) all spectra recorded until equilibrium was reached.
The band at 525 nm represents the fully formed (4N) ATCUN complex,
stable at pH 7.4, and the band at 750 nm represents the intermediate
(2N) complex.[20]
Stopped-flow spectra of Cu(II)-hepc6 in 20 mM ammonium
acetate
initially at pH 7.4, subjected to rapid pH decrease by mixing with
concentrated acetic acid. Color-coded spectra were recorded every
1.5 ms with the instrument dead time of 2 ms. (A) First 11 ms of the
reaction and (B) all spectra recorded until equilibrium was reached.
The band at 525 nm represents the fully formed (4N) ATCUN complex,
stable at pH 7.4, and the band at 750 nm represents the intermediate
(2N) complex.[20]The traces at 525 nm, corresponding to the decomposition of the
4N ATCUN complex, are presented in Figure for Cu(II)–hepc6 and in Figure S16 for Cu(II)–GGH. The extent
of dissociation of 4N complexes after 10 ms at pH 5, 4, and 3 was
measured by following the decay of their absorption peaks at 525 nm.
Due to significant noise of the absorption signal the averaged signal
intensities at 0 and 10 ms were obtained by fitting a first order
kinetic function to the absorption data. The overall fits and expanded
regions of interest (0 to 30 ms) are presented for clarity. As seen
in these figures, the inertness of ATCUN complexes to acid-catalyzed
dissociation in the 10 ms time window is significant (except for Cu(II)–GGH
at pH 3) and must be taken into account. Table S3 presents the results of a simulation of relative abundances
of apopeptides and Cu(II) complexes that should occur if the ESI-MS
signals depended only on droplet acidification and the resulting complex
dissociation after 10 ms. This analysis indicates that the abundance
of hepc6 apopeptide in ESI-MS spectra is consistent with pH the droplet
drop to slightly above 3, while the abundance of GGH apopeptide appears
to correspond with the final droplet pH clearly above 5. Both cannot
be true for experiments performed under identical conditions. Hence,
acidification and complex decay kinetics do not explain the pattern
of free and complexed peptides observed in ESI-MS.
Figure 7
Kinetic traces at 525
nm for pH–jump experiments on Cu(II)-hepc6
presented in Figure for pH 3 (A), 4 (B), and 5 (C). Left: full traces recorded over
1.5 s and 1st order kinetic fits (red lines). Right: the first 30
ms of traces. Vertical lines mark the 0 ms (reaction start) and the
10 ms time points.
Kinetic traces at 525
nm for pH–jump experiments on Cu(II)-hepc6
presented in Figure for pH 3 (A), 4 (B), and 5 (C). Left: full traces recorded over
1.5 s and 1st order kinetic fits (red lines). Right: the first 30
ms of traces. Vertical lines mark the 0 ms (reaction start) and the
10 ms time points.The application of the
above methodology to decipher the mechanism
responsible for the deviation in competition experiments is not possible
because of the overlap of the pH and desolvation effects exerted simultaneously
on the peptide and histidine complexes. The droplet acidification
and/or complex unfolding processes are less significant in the mild
desolvation conditions of the respective instrument, as evidenced
by the control experiment. The theoretical speciation for Cu(Aβ4–16)
at the 100 pM concentration suggests that the pH drop may be actually
very low or none (Figure S11). In accord,
the attempt to assign a pH drop to competition experiments proved
futile, as presented in Figure . Therefore, the systematic error embedded in these experiments
is likely due to complex desolvation phenomena, as discussed below.
Figure 8
Experimental
and theoretical binding isotherms of Cu/Aβ4-16
(A) and Cu/hepc6 (B) 1:1 complexes titrated with histidine as weak
competitor.
Experimental
and theoretical binding isotherms of Cu/Aβ4-16
(A) and Cu/hepc6 (B) 1:1 complexes titrated with histidine as weak
competitor.
Discussion
The
experimental results presented in Figure and Figures S1–S6 and S12–S13 showed that the proportions of free ATCUN
peptides and their complexes in ammonium acetate solutions at pH 7.4,
obtained by ESI-MS, do not match the expectations based on CK7.4 values obtained by other techniques
(Tables and 2). This cannot be assigned to a nonproportionate
efficacy of ionization of respective species, a well-known phenomenon
in ESI-MS,[30] because the discrepancy is
far too high.In the case of histidine competition experiments,
a disproportional
ionization would shift the binding curve but should not yield a steeper
transition from free to complexed peptide. Doing so would indicate
that ionization efficiency difference reverses in changing concentrations
of the competitor. Interestingly, at low histidine concentrations
the apparent Cu(II) binding by the peptides is slightly enhanced beyond
the expected value. This could be caused by the fact that during shrinking
of the droplet, when concentrations rise, His escapes the droplets
earlier. This would shift the equilibrium in favor of the peptide.
No such effect is observed, however, at higher histidine concentrations.In the case of serial dilutions, the artifactual surplus of free
peptides was not alleviated even at a high Cu2+ excess
(Figures S7–S9). The analysis of
relative peak intensities excluded the Cu(II) reduction to Cu(I) as
a significant source of this discrepancy (Figure S10). The significant amounts of apopeptides in the spectra
resembled the Cu2+/peptide equilibria under the acidic
conditions (Figures and S14), which appeared to correlate
with the noted acidification of ammonium acetate solutions upon the
electrospray droplet formation.[52] Quantitative
analysis demonstrated, however, that the apparent pH values derived
from ESI-MS data are spread randomly over nearly one pH unit (5.0–5.7, Figure ), effectively precluding
a quantitative use of such data. Additionally, this apparent pH shift
was less than expected for droplet acidification according to Konermann
(around 4.75).[52] Looking for a reason,
we turned to pH–jump kinetic experiments, guided by our recent
discovery of intermediate steps in the process of ATCUN complex formation,
which could overlap with the lifetime of electrospray droplets.[20,41] These experiments, presented in Figures , 7, S15, and S16, revealed that the studied complexes are far
too inert to undergo a significant decomposition within the 10 ms
time window facing the relevant extent of acidification. Therefore,
the main reason for the complex dissociation was in this case an inadequate
ESI source or its parameters. Tuning of voltage, gas flow, or injection
flow did not yield considerable success. However, changing the spectrometer
to the one dedicated to intact noncovalent bonds yielded much more
accurate results. The apparent binding constants were much closer
to expected values derived from solution studies but were still significantly
underestimated.We propose that the clue for the observed effect
is provided by
thermodynamics of Cu(II) complex formation. The ITC investigation
of DAHK, an ATCUN peptide model of human serum albumin, demonstrated
that the Cu(DAHK) complex at neutral pH is stabilized solely by entropic
contribution.[50] At this pH the enthalpic
contributions such as Coulombic attraction are canceled out by the
cost of deprotonations of peptide nitrogens some 8 orders of magnitude
below its Ka.[70] The main contributor to complex stability is believed to be a release
of water molecules from hydration shells of Cu(II) and the peptide,
as well as deprotonation. Stable 4-nitrogen–Cu(II) complex
leaves much less peptide and metal exposed to the solvent; thus, the
entropic factor comes from the combination of folding and desolvation.
A similar feature was observed for Cu(II) binding to metalloproteins.[51,71] Therefore, in accordance with the data presented above, the removal
of water from the second coordination sphere of a Cu(II)–ATCUN
complex will likely lead to the equilibrium shift and possible subsequent
complex decomposition upon entry of the molecules into the gas phase.
It was previously established that thermodynamic changes occur during
transition to the gas phase and entropically favored forces become
destabilized.[28] Additionally, acid-dissociated
Cu2+ ions may recombine with the peptide upon its transition
to the gas phase, additionally confounding the observations.[72] From this perspective, one can state that neither
ESI-MS nor other mass spectrometry techniques should be applied in
quantitative studies of Cu(II)–peptide complexes. If such experiments
are performed, the resulting binding constant will always be lower
than the real one, but to an extent impossible to predict without
independent data obtained by a different method. Therefore, mass spectrometry
data for Cu(II)/peptide binding constants should be regarded as “greater
than x”, always with an assumption that the
in-solution binding constant is higher than that measured. On the
other hand, enthalpic contributions are significant or decisive for
the stability of other metal chelates in water,[73,74] and hence, metal complex formation can be studied by MS upon maintaining
the utmost scrutiny and validating the results by independent techniques
whenever possible.
Authors: Mariusz Mital; Nina E Wezynfeld; Tomasz Frączyk; Magdalena Z Wiloch; Urszula E Wawrzyniak; Arkadiusz Bonna; Carolin Tumpach; Kevin J Barnham; Cathryn L Haigh; Wojciech Bal; Simon C Drew Journal: Angew Chem Int Ed Engl Date: 2015-07-14 Impact factor: 15.336