Jennifer Wong1, Elliot Backer1, Joseph C Keenan1, Huseyin Erhan Dincer1, Roy Joseph Cho1.
Abstract
PURPOSE: Balloon-expandable stents are commonly used for the treatment of tracheobronchial strictures. We routinely perform targeted overdilation of these stents 1-2mm on initial deployment to prevent stent migration or allow foreshortening to target airway caliber; however, specific data on the effect of targeted overdilation is unknown. PATIENTS AND METHODS: We used three iCAST stents (6×22mm, 7×22mm, and 10×38mm) to perform the study. We had two sets of each size to average our results. Targeted overdilation was accomplished with Merit Elation balloons.Entities:
Keywords: airway balloon dilation; airway stenosis; airway stent; balloon expandable stent; interventional pulmonary; tracheobronchial disease
Year: 2021 PMID: 34737654 PMCID: PMC8558051 DOI: 10.2147/MDER.S336933
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Figure 1Stent break point. Red arrow points to the steel strut visualized through the PTFE.
Targeted Overdilation of 6 × 22 iCAST Stent
| Balloon Diameter (mm) | Stent OD (mm) | Stent Length (mm) |
|---|---|---|
| 6.0* | 6.2* | 20.2* |
| 8.0 | 6.2 | 18.6 |
| 9.0 | 7.1 | 18.4 |
| 10.0 | 7.9 | 17.7 |
| 11.0 | 9.2 | 16.2 |
| 12.0 | 9.4 | 14.6 |
| 13.5** | 11.2** | 13.2** |
| 15.0 | 11.4 | 11.6 |
| 20.0 | 11.4 | 5.7 |
Notes: Diameter achieved using 8-9-10mm and 12-13.5-15mm balloons. *Burst pressure dimensions per manufacturer. **Break point diameter.
Abbreviation: OD, outer diameter.
Targeted Overdilation of 10 × 38 iCAST Stent
| Balloon Diameter (mm) | Stent OD (mm) | Stent Length (mm) |
|---|---|---|
| 10.0* | 10.4* | 30.9* |
| 11.0 | 9.9 | 13.9 |
| 12.0 | 10.2 | 32.2 |
| 13.5 | 12.5 | 31.4 |
| 15.0** | 12.9** | 31.0** |
| 20.0 | 12.0 | 30.0 |
Notes: Diameter achieved using 8-9-10mm, 12-13.5-15mm, and 18-19-20mm balloons. *Burst pressure dimensions per manufacturer. **Break point diameter.
Abbreviation: OD, outer diameter.
Figure 2Over dilation of iCAST stents. Above graphically illustrates the effect of over dilation on stent outer diameter (OD) and length. The 6 and 7 × 22 stents show significant contraction at 12mm dilation. The 10 × 38 stent did not show significant contraction during over dilation. All stent appeared to have a break point at 15mm dilation.
Figure 3Change in length and outer diameter (OD) during serial targeted overdilation. Above graphically illustrates the effect of overdilation on stent outer OD (dotted line) and length (solid line). The 6 and 7 × 22 stents showed significant contraction over the series of dilations unlike the larger 10 × 38 stent. All stent appeared to have a break point at two times the original stent diameter (arrow).
Targeted Overdilation of 7 × 22 iCAST Stent
| Balloon Diameter (mm) | Stent OD (mm) | Stent Length (mm) |
|---|---|---|
| 7.0* | 7.3* | 19.4 |
| 8.0 | 7.4 | 19.0 |
| 9.0 | 7.6 | 18.6 |
| 10.0 | 7.6 | 18.3 |
| 11.0 | 9.4 | 16.2 |
| 12.0 | 9.5 | 15.7 |
| 13.5 | 10.9 | 9.9 |
| 15.0** | 11.3** | 7.0** |
| 20.0 | 11.6 | 6.0 |
Notes: Diameter achieved using 8-9-10mm and 12-13.5-15mm balloons. *Burst pressure dimensions per manufacturer. **Break point diameter.
Abbreviation: OD, outer diameter.