Quality Fade in Medical Device Manufacturing: Thinness of Airway Breathing Circuit Plastic.

Rotem Naftalovich, Marko Oydanich, Tolga Berkman, Andrew John Iskander.   

Abstract

Mechanical respirators typically use a plastic circuit apparatus to pass gases from the ventilator to the patient. Structural integrity of these circuits is crucial for maintaining oxygenation. Anesthesiologists, respiratory therapists, and other critical care professionals rely on the circuit to be free of defects. The American Society for Testing and Materials maintains standards of medical devices and had a standard (titled Standard Specification for Anesthesia Breathing Tubes) that included circuits. This standard, which was last updated in 2008, has since been withdrawn. Lack of a defined standard can invite quality fade-the phenomenon whereby manufacturers deliberately but surreptitiously reduce material quality to widen profit margins. With plastics, this is often in the form of thinner material. A minimum thickness delineated in the breathing circuit standard would help ensure product quality, maintain tolerance to mechanical insults, and avert leaks. Our impression is that over the recent years, the plastic in many of the commercially available breathing circuits has gotten thinner. We experienced a circuit leak in the middle of a laminectomy due to compromised plastic tubing in a location that evaded the safety circuit leak check that is performed prior to surgery. This compromised ventilation and oxygenation in the middle of a surgery in which the patient is positioned prone and hence with a minimally accessible airway; it could have resulted in anoxic brain injury or death. The incident led us to reflect on the degree of thinness of the circuit's plastic. © Copyright AAMI 2021. Copying, networking, and distribution prohibited.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 34727573      PMCID: PMC8641416          DOI: 10.2345/0899-8205-55.4.118

Source DB:  PubMed          Journal:  Biomed Instrum Technol        ISSN: 0899-8205


  4 in total

1.  Anesthesia respiratory circuit failure.

Authors:  Mark I Rossberg; Robert S Greenberg
Journal:  Anesthesiology       Date:  2002-09       Impact factor: 7.892

2.  An uncommon leak in the anesthesia breathing circuit.

Authors:  H R Bromley; S Tuorinsky
Journal:  Anesth Analg       Date:  1997-09       Impact factor: 5.108

3.  Leakage of disposable breathing circuits.

Authors:  J S Wang; W T Hung; C Y Lin
Journal:  J Clin Anesth       Date:  1992 Mar-Apr       Impact factor: 9.452

Review 4.  The closed circuit and the low flow systems.

Authors:  S Parthasarathy
Journal:  Indian J Anaesth       Date:  2013-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.