| Literature DB >> 34717062 |
M Konarik1, M Sramko, Z Dorazilova, M Blaha, I Netuka, P Ivak, J Maly, O Szarszoi.
Abstract
Mechanical circulatory support (MCS) with an implantable left ventricular assist device (LVAD) is an established therapeutic option for advanced heart failure. Most of the currently used LVADs generate a continuous stream of blood that decreases arterial pulse pressure. This study investigated whether a change of the pulse pressure during different pump speed settings would affect cerebral autoregulation and thereby affect cerebral blood flow (CBF). The study included 21 haemodynamically stable outpatients with a continuous-flow LVAD (HeartMate II, Abbott, USA) implanted a median of 6 months before the study (interquartile range 3 to 14 months). Arterial blood pressure (measured by finger plethysmography) was recorded simultaneously with CBF (measured by transcranial Doppler ultrasound) during baseline pump speed (8900 rpm [IQR 8800; 9200]) and during minimum and maximum tolerated pump speeds (8000 rpm [IQR 8000; 8200] and 9800 rpm [IQR 9800; 10 000]). An increase in LVAD pump speed by 800 rpm [IQR 800; 1000] from the baseline lead to a significant decrease in arterial pulse pressure and cerebral blood flow pulsatility (relative change ?24% and ?32%, both p < 0.01), but it did not affect mean arterial pressure and mean CBF velocity (relative change 1% and ?1.7%, p = 0.1 and 0.7). In stable patients with a continuous-flow LVAD, changes of pump speed settings within a clinically used range did not impair static cerebral autoregulation and cerebral blood flow.Entities:
Mesh:
Year: 2021 PMID: 34717062 PMCID: PMC8815470 DOI: 10.33549/physiolres.934738
Source DB: PubMed Journal: Physiol Res ISSN: 0862-8408 Impact factor: 1.881