BACKGROUND: In this study, the tear resistance of porcine lens capsules after continuous curvilinear capsulorhexis (CCC) and femtosecond (fs)-laser-assisted capsulotomy for cataract surgery (FLC) with different laser parameters is measured with a custom-made testing setup.Entities:
Keywords: Cataract surgery; Continuous curvilinear capsulorhexis; Femtosecond (fs)-laser-assisted capsulotomy; Lens refilling; Phacoemulsification; Tear resistance
Mesh:
Year: 2021 PMID: 34713366 PMCID: PMC8971170 DOI: 10.1007/s10103-021-03448-1
Source DB: PubMed Journal: Lasers Med Sci ISSN: 0268-8921 Impact factor: 3.161
Laser parameter for capsulotomy (FLC 1, FLC 2)
| Parameter set A (FLC 1) | Parameter set B (FLC 2) | |
|---|---|---|
| Outer radius | 2.0 | 2.0 |
| Inner radius | 1.8 | 1.8 |
| Spot size | 5 | 5 |
| Spot distance lateral (μm) | 3 | 5 |
| Spot distance radial (μm) | 3 | 5 |
| Spot distance depth (μm) | 10 | 15 |
| Pulse energy (μJ) | 2 | 2 |
| Pulse length (fs) | 306 | 306 |
Fig. 1Simplified geometry (only one plane of the z-stack) of fs-laser capsulotomy pattern (parameter set B) with the most important parameters. See also Table 1
Parameters used for phacoemulsification
| Parameter | Value |
|---|---|
| Position of infusion holder | 55 cm |
| Max. vacuum | 100 mmHg |
| Max. US | 15% |
Fig. 2Porcine eye globes after phacoemulsification with CCC (a, diameter 4 ± 1 mm) and FLC with parameter set A (b, diameter 4 mm)
Fig. 3a Custom-made testing setup with scale (1), nylon thread for the connection of the left hook to the scale (2), left and right hook (3) and linear stage (4), b sclera incisions to minimize the resistance of the sclera to the stretching
Fig. 4Example plots of tear resistance measurements. Traction force as a function of strain for porcine lenses. Blue: after CCC (max. force = 135 mN); red: after FLC 1 (max force = 77 mN); green: after FLC 2 (max force = 70 mN)
Results of tear resistance measurements with calculated mean values of the three experimental series for porcine lenses (n = 15)
| Number of eye | CCC (mN) | FLC 1 (mN) | FLC 2 (mN) |
|---|---|---|---|
| 1 | 108 | 85 | 29 |
| 2 | 107 | 61 | 25 |
| 3 | 132 | 116 | 31 |
| 4 | 156 | 82 | 46 |
| 5 | 174 | 52 | 20 |
| 6 | 188 | 58 | 31 |
| 7 | 187 | 44 | 37 |
| 8 | 73 | 52 | 21 |
| 9 | 311 | 39 | 34 |
| 10 | 129 | 55 | 9 |
| 11 | 236 | 32 | 10 |
| 12 | 217 | 13 | 10 |
| 13 | 79 | 46 | 36 |
| 14 | 47 | 48 | 70 |
| 15 | 135 | 77 | 15 |
| Mean value | 150 | 60 | 30 |
| Standard deviation | 70 | 20 | 20 |
Fig. 5Box plot of maximum force before tear for the three measurement series (CCC, FLC 1, FLC 2)