William P Ensminger1, Terence McIff1, Bryan Vopat1, Scott Mullen1, J Paul Schroeppel1.
Abstract
PURPOSE: To compare knot and loop characteristics of commonly used high-strength tapes and high-strength round sutures.Entities:
Year: 2021 PMID: 34712990 PMCID: PMC8527324 DOI: 10.1016/j.asmr.2021.07.014
Source DB: PubMed Journal: Arthrosc Sports Med Rehabil ISSN: 2666-061X
Fig 1Knot-tying cannula setup. Knots were tied through a standard arthroscopic cannula using a knot pusher, onto the 30-mm dowel.
Fig 2Material testing system apparatus. (A) Overview and (B) Enlarged image of the mechanical testing equipment. The upper arm (solid arrowhead) is affixed to the testing machine actuator above. The lower arm (hollow arrowhead) is attached to custom fixture within the saline bath (dashed arrow). The custom fixture and saline bath are atop an X-Y table (not pictured) to ensure inline pull during testing. The solid arrow denotes a knotted suture loop held by two stainless-steel posts, as it would appear just before testing.
Fig 3Representative image of the technique used to measure knot volume with a measurement diagram.
The 3-mm Single Load Testing (N)
| Suture/Knot | Hand-Tied | SMC Knot | Tennessee Slider | Duncan Loop | Surgeon’s Knot |
|---|---|---|---|---|---|
| BroadBand | 322.09 ± 41.36 | 266.73 ± 28.02 | 217.03 ± 49.04 | 221.73 ± 88.40 | 317.94 ± 52.78 |
| SutureTape | 294.67 ± 39.09 | 225.32 ± 40.60 | 204.91 ± 50.16 | 285.63 ± 59.02 | 293.52 ± 55.33 |
| MaxBraid | 263.27 ± 37.99 | 248.82 ± 31.04 | 200.18 ± 60.27 | 292.42 ± 52.70 | 271.79 ± 73.93 |
| FiberWire | 200.30 ± 33.34 | 203.89 ± 31.88 | 182.70 ± 49.97 | 158.66 ± 35.25 | 227.48 ± 78.51 |
| DynaCord | 331.28 ± 22.45 | 271.52 ± 42.96 | 298.86 ± 24.26 | 191.37 ± 42.57 | 283.03 ± 67.78 |
| FiberTape | 430.48 ± 89.00 | ||||
| PermaTape | 545.16 ± 95.09 |
Data are presented as mean ± standard deviation; n = 10 for all groups. Demonstrates mean maximum force (N) measured from 0 to 3 mm of displacement (or suture-knot breakage) for each suture-knot combination. P values represented are within knot comparisons between tape sutures and round sutures only.
SMC, Samsung Medical Center.
Significantly higher than MaxBraid (P < .05).
Significantly higher than FiberWire (P < .05).
Significantly higher than all narrow tapes and No. 2 sutures (P < .05).
Significantly higher than SutureTape (P < .05).
Significantly higher than BroadBand (P < .05).
Significantly higher than DynaCord (P < .05).
The 2-mm Single Load Testing (N)
| Suture/Knot | Hand-Tied | SMC Knot | Tennessee Slider | Duncan Loop | Surgeon’s Knot |
|---|---|---|---|---|---|
| BroadBand | 229.48 ± 22.90 | 189.06 ± 44.08 | 155.81 ± 56.92 | 194.43 ± 62.10 | 262.07 ± 60.31 |
| SutureTape | 234.42 ± 35.71 | 190.19 ± 39.27 | 149.18 ± 46.27 | 228.12 ± 41.11 | 234.06 ± 89.60 |
| MaxBraid | 191.64 ± 23.97 | 196.60 ± 37.57 | 133.33 ± 45.25 | 213.44 ± 41.46 | 182.16 ± 55.47 |
| FiberWire | 163.42 ± 24.92 | 168.15 ± 41.91 | 140.52 ± 40.32 | 139.27 ± 22.47 | 172.86 ± 53.27 |
| DynaCord | 239.78 ± 27.97 | 205.37 ± 24.92 | 224.35 ± 41.07 | 130.21 ± 24.82 | 194.01 ± 49.57 |
| FiberTape | 258.71 ± 46.81 | ||||
| PermaTape | 380.21 ± 67.56 |
Data are presented as mean ± standard deviation; n = 10 for all groups. Demonstrates mean maximum force (N) measured from 0 to 2 mm of displacement (or suture-knot breakage) for each suture-knot combination. P values represented are within knot comparisons between tape sutures and round sutures only.
SMC, Samsung Medical Center.
Significantly higher than MaxBraid (P < .05).
Significantly higher than FiberWire (P < .05).
Significantly higher than DynaCord (P < .05).
Significantly higher than SutureTape (P < .05).
Significantly higher than BroadBand (P < .05).
Significantly higher than all narrow tapes and No. 2 sutures (P < .05).
Cyclic Elongation (mm)
| Suture/Knot | Hand-Tied | SMC Knot | Tennessee Slider | Duncan Loop | Surgeon’s Knot |
|---|---|---|---|---|---|
| BroadBand | 0.64 ± 0.07 | 0.82 ± 0.40 | 1.63 ± 0.52 | 0.68 ± 0.30 | 0.97 ± 0.66 |
| SutureTape | 0.69 ± 0.13 | 0.84 ± 0.31 | 0.87 ± 0.39 | 0.57 ± 0.09 | 1.18 ± 0.81 |
| MaxBraid | 0.95 ± 0.10 | 1.03 ± 0.26 | 1.29 ± 0.55 | 0.80 ± 0.10 | 1.35 ± 0.47 |
| FiberWire | 0.76 ± 0.13 | 0.77 ± 0.18 | 1.16 ± 0.47 | 0.81 ± 0.26 | 0.99 ± 0.48 |
| DynaCord | 0.98 ± 0.17 | 0.90 ± 0.10 | 0.97 ± 0.22 | 1.22 ± 0.35 | 0.89 ± 0.16 |
| FiberTape | 0.88 ± 0.15 | ||||
| PermaTape | 0.59 ± 0.08 |
Data are presented as mean ± standard deviation; n = 10 for all groups. Represents the mean amount of plastic loop elongation (mm) that occurred from the initial 10 N pre-tensioning position to the end of cyclic testing (1000 cycles). P values represented are within knot comparisons between tape sutures and round sutures only.
SMC, Samsung Medical Center.
Significantly less than FiberTape (P < .05).
Significantly less than MaxBraid (P < .05).
Significantly less than DynaCord (P < .05).
Significantly less than BroadBand (P < .05).
Fig 4Load-to-failure following cyclic testing with failure defined as loop elongation of 3 mm; vertically oriented error bars represent standard deviation; n = 10 for all groups. Mean maximum force (N) was measured from 0 to 3 mm of elongation after completion of cyclic load testing. Open circles indicate values significantly higher than FiberWire (P < .05). Open circles inside squares indicate values significantly higher than all narrow tapes and no. 2 sutures (P < .05). Open diamonds indicate values significantly higher than SutureTape (P < .05). Error bars indicate values significantly higher than DynaCord (P < .05). P values represented are within knot comparisons between tape sutures and round sutures only.
Loop Circumference (mm)
| Suture/Knot | Hand-Tied | SMC Knot | Tennessee Slider | Duncan Loop | Surgeon’s Knot |
|---|---|---|---|---|---|
| BroadBand | 31.08 ± 0.22 | 30.65 ± 0.16 | 31.04 ± 0.44 | 30.47 ± 0.40 | 30.77 ± 0.52 |
| SutureTape | 31.32 ± 0.19 | 30.69 ± 0.25 | 30.92 ± 0.60 | 30.43 ± 0.18 | 30.76 ± 0.28 |
| MaxBraid | 31.43 ± 0.20 | 30.84 ± 0.19 | 30.94 ± 0.25 | 30.42 ± 0.11 | 30.93 ± 0.37 |
| FiberWire | 31.25 ± 0.21 | 30.75 ± 0.22 | 30.95 ± 0.43 | 30.52 ± 0.61 | 30.94 ± 0.42 |
| DynaCord | 30.19 ± 0.19 | 30.26 ± 0.14 | 30.20 ± 0.19 | 30.17 ± 0.24 | 30.26 ± 0.20 |
| FiberTape | 32.53 ± 0.65 | ||||
| PermaTape | 31.61 ± 0.26 |
Data are presented as mean ± standard deviation; n = 20 for all groups. Represents the mean circumference (mm) of the knotted suture loop after pre-tensioning to 10 N, with the ideal circumference being 30 mm. P values represented are within knot comparisons between tape sutures and round sutures only.
SMC, Samsung Medical Center.
Significantly less than MaxBraid (P < .05).
Significantly less than FiberTape (P < .05).
Significantly less than PermaTape (P < .05).
Significantly less than all tape sutures (P < .05).
Significantly less than BroadBand and SutureTape (P < .05).
Knot Volume (mm3)
| Suture/Knot | Hand-Tied | SMC Knot | Tennessee Slider | Duncan Loop | Surgeon’s Knot |
|---|---|---|---|---|---|
| BroadBand | 11.46 ± 1.77 | 10.05 ± 0.76 | 10.10 ± 1.00 | 14.70 ± 1.23 | 14.23 ± 2.51 |
| SutureTape | 10.50 ± 1.31 | 9.65 ± 1.39 | 9.85 ± 1.71 | 14.15 ± 1.11 | 13.82 ± 2.16 |
| MaxBraid | 12.05 ± 1.88 | 11.10 ± 0.77 | 10.79 ± 2.25 | 14.27 ± 1.32 | 14.14 ± 2.16 |
| FiberWire | 15.49 ± 2.50 | 12.93 ± 2.23 | 11.57 ± 1.73 | 17.94 ± 1.58 | 14.45 ± 2.31 |
| DynaCord | 16.82 ± 1.67 | 13.09 ± 1.27 | 14.52 ± 2.05 | 19.65 ± 1.62 | 15.54 ± 2.62 |
| FiberTape | 70.92 ± 5.99 | ||||
| PermaTape | 24.32 ± 3.40 |
Data are presented as mean ± standard deviation; n = 20 for all groups. Demonstrates the mean volume (mm3) of each suture-knot combination. P values represented are within knot comparisons between tape sutures and round sutures only.
SMC, Samsung Medical Center.
Significantly less than FiberWire (P < .05).
Significantly less than DynaCord (P < .05).
Significantly less than FiberTape and PermaTape (P < .05).
Significantly less than MaxBraid (P < .05).