| Literature DB >> 34707908 |
XiaoHao Du1,2, Jia Zhang1,3, Meng Su1, WenJia Cao1, Shuang Zeng1, QinMei Wang1,3, Ioannis M Aslanides4, ShiHao Chen1,3,5.
Abstract
PURPOSE: To compare the clinical outcomes of aberration-free all surface laser ablation (ASLA) with and without the use of smart pulse technology (SPT) in high myopia.Entities:
Year: 2021 PMID: 34707908 PMCID: PMC8545587 DOI: 10.1155/2021/2588765
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Baseline preoperative valuesa.
| Baseline preoperative values | SPT group | Non-SPT group |
|
|---|---|---|---|
| Number of eyes | 53 | 85 | |
| Mean age (years) | 26.25 ± 5.64 | 24.62 ± 4.91 | 0.082 |
| Mean preoperative logMAR UDVA | 1.294 ± 0.245 | 1.280 ± 0.269 | 0.725 |
| Mean preoperative logMAR CDVA | −0.007 ± 0.017 | −0.024 ± 0.034 | 0.02b |
| Mean preoperative refractive errors (D) | |||
| Sphere | −7.60 ± 1.31 | −7.60 ± 1.06 | 0.925 |
| Cylinder | −0.83 ± 0.64 | −1.00 ± 0.71 | 0.206 |
| Spherical equivalent | −8.02 ± 1.32 | −8.10 ± 1.05 | 0.507 |
| Optical zone diameter (mm) | 6.12 ± 0.26 | 6.18 ± 0.26 | 0.494 |
| Treatment diameter (mm) | 7.91 ± 0.33 | 8.09 ± 0.33 | 0.001b |
| Ablation depth ( | 160.85 ± 13.05 | 171.49 ± 11.25 | <0.001b |
| Residual stromal bed thickness ( | 345.83 ± 21.44 | 330.11 ± 18.71 | <0.001b |
SPT = smart pulse technology; UDVA = uncorrected distance visual acuity; CDVA = corrected distance visual acuity. aValues are presented as mean ± standard deviation. bSignificant difference between the SPT and non-SPT groups (Student's t-test/Mann–Whitney U-test).
Figure 1Postoperative healing time of corneal epithelium in the SPT and the non-SPT groups. SPT = smart pulse technology.
Visual acuity and refraction outcomesa.
| Parameter | Group | Preoperative | 12 months |
|
|---|---|---|---|---|
| LogMAR UDVA | SPT | 1.294 ± 0.245 | −0.040 ± 0.054 | 0.533 |
| Non-SPT | 1.280 ± 0.269 | −0.009 ± 0.080 | ||
|
| ||||
| LogMAR CDVA | SPT | −0.007 ± 0.017 | −0.046 ± 0.045 | 0.004b |
| Non-SPT | −0.024 ± 0.034 | −0.038 ± 0.041 | ||
|
| ||||
| Sphere | SPT | −7.60 ± 1.31 | 0.24 ± 0.50 | 0.770 |
| Non-SPT | −7.60 ± 1.06 | 0.26 ± 0.51 | ||
|
| ||||
| Cylinder | SPT | −0.83 ± 0.64 | −0.45 ± 0.34 | 0.169 |
| Non-SPT | −1.00 ± 0.71 | −0.42 ± 0.28 | ||
|
| ||||
| Spherical equivalent | SPT | −8.02 ± 1.32 | 0.01 ± 0.47 | 0.493 |
| Non-SPT | −8.10 ± 1.05 | 0.05 ± 0.50 | ||
SPT = smart pulse technology; UDVA = uncorrected distance visual acuity; CDVA = corrected distance visual acuity. aValues are presented as mean ± standard deviation. bSignificant difference between the SPT and non-SPT groups (Student's t-test/Mann–Whitney U-test).
Figure 2(a) Efficacy of all surface laser ablation (ASLA) assisted by smart pulse technology (SPT group) versus ASLA without SPT (non-STP group) 12 months postoperatively. (b) Change in corrected distance visual acuity. (c) Attempted vs. achieved spherical equivalent (SE) refraction in the SPT group and the non-SPT group. (d) SE refraction accuracy. (e) Preoperative, 1 week, and 1, 3, 6, and 12 months SE values in the SPT group and the non-SPT group.
Changes of safety and efficacy index after 12 months in SPT group and non-SPT groupa.
| Time | Safety index | Efficacy index | ||||
|---|---|---|---|---|---|---|
| SPT group | Non-SPT group |
| SPT group | Non-SPT group |
| |
| 7 days | 0.81 ± 0.17 | 0.71 ± 0.17 | <0.001b | 0.79 ± 0.22 | 0.65 ± 0.17 | 0.001b |
| 1 month | 0.99 ± 0.08 | 0.91 ± 0.14 | <0.001b | 0.98 ± 0.13 | 0.84 ± 0.24 | <0.001b |
| 3 months | 1.07 ± 0.11 | 1.02 ± 0.13 | 0.033b | 1.09 ± 0.18 | 0.99 ± 0.19 | 0.005b |
| 6 months | 1.08 ± 0.11 | 1.07 ± 0.12 | 0.477 | 1.08 ± 0.15 | 0.99 ± 0.18 | 0.009b |
| 12 months | 1.10 ± 0.12 | 1.04 ± 0.11 | 0.004b | 1.09 ± 0.14 | 0.98 ± 0.17 | <0.001b |
SPT = smart pulse technology. Safety index = postoperative mean CDVA/preoperative mean CDVA. Efficacy index = postoperative mean UDVA/preoperative mean CDVA. aValues are presented as mean ± standard deviation. bSignificant difference between the SPT and non-SPT groups (Student's t-test/Mann–Whitney U-test).
Figure 3Variable occurrence rate of haze after SPT or non-SPT. SPT = smart pulse technology.
Anterior corneal aberration valuesa.
| Parameter | Group | Preoperative | After 12 months |
| △(12-month postoperative vs. preoperative) |
|---|---|---|---|---|---|
| RMS HOA | SPT group | 0.35 ± 0.09 | 0.89 ± 0.24 | <0.001c | 0.54 ± 0.25 |
| Non-SPT group | 0.38 ± 0.10 | 1.03 ± 0.36 | <0.001c | 0.65 ± 0.39 | |
|
| 0.069 | 0.059 | 0.238 | ||
|
| |||||
| Spherical aberration | SPT group | 0.25 ± 0.09 | 0.73 ± 0.19 | <0.001c | 0.48 ± 0.20 |
| Non-SPT group | 0.22 ± 0.08 | 0.77 ± 0.21 | <0.001c | 0.56 ± 0.22 | |
|
| 0.024b | 0.461 | 0.091 | ||
|
| |||||
| Coma 0° | SPT group | −0.05 ± 0.10 | −0.29 ± 0.26 | <0.001c | −0.24 ± 0.28 |
| Non-SPT group | −0.11 ± 0.10 | −0.31 ± 0.39 | <0.001c | −0.21 ± 0.39 | |
|
| 0.001b | 0.840 | 0.231 | ||
|
| |||||
| Coma 90° | SPT group | −0.00 ± 0.15 | −0.15 ± 0.42 | 0.012c | −0.15 ± 0.39 |
| Non-SPT group | −0.04 ± 0.20 | −0.40 ± 0.51 | <0.001c | −0.36 ± 0.47 | |
|
| 0.234 | 0.001b | 0.003b | ||
RMS HOA = root mean square higher order aberrations; SPT = smart pulse technology. Compared with the SPT group and non-SPT group. aValues are presented as mean ± standard deviation. bSignificant difference between the SPT and non-SPT groups (Student's t-test/Mann–Whitney U-test). cSignificant difference between preoperative and postoperative (Student's t-test/Wilcoxon test).
Figure 4In the mode of non-SPT, the closer to the corneal periphery area is, the greater the energy loss is. While in the SPT mode, the energy loss is relatively reduced. SPT = smart pulse technology.