Michel Hesse1, Philipe d'Abadie1, Renaud Lhommel1, Francois Jamar1, Stephan Walrand1.
Abstract
PURPOSE: Explaining why 90Y TOF-PET based equivalent uniform dose (EUD) using recommended manufacturer FDG reconstruction parameters has been shown to predict response.Entities:
Keywords: EUD; TOF-PET; dose-response; radioembolization; reconstruction
Year: 2021 PMID: 34676157 PMCID: PMC8523947 DOI: 10.3389/fonc.2021.592529
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Figure 1(A) Jaszczak deluxe phantom set in a vertical position on a paper bloc modelling the patient attenuation. Only a part of the hot rod insert was filled with an active solution to reach a typical clinical tumour absorbed dose using a 3GBq total 90Y activity. (B) Actual hot rods map. (C) Autoradiography of a normal liver (NL) tissue resected 9 days post 90Y resin spheres liver radioembolization delivering 52Gy to the NL tissue [reprinted from (35) with permission of EJNMMI]. (B, C) are represented at the same scale.
Figure 2Hot rod slice. (A, C, D, F) 550ps TOF-PET. B,E,H: 320ps TOF-PET. (A, B) 4mm-thick slice in 50Gy-90Y-setup. (D, E) 4mm-thick slice in 100Gy-90Y-setup. (C) 4mm-thick slice in 5Gy-90Y-setup with a 10 fold longer acquisition time. Note that the acquisition time increasing (C) did not counterbalance the count rate reduction and result in lower image quality than (A, D). (F, G) 4mm-thick slice in clinical 18FDG WB setup. As the intent is to show the capability of each modality to assess the heterogeneity pattern, the maximum voxel intensity of each image was scaled to 255.
True (T) + scatter (S) coincidences numbers acquired on the 550ps TOF-PET for the three 90Y phantom setups.
| Sector Dose[Gy] | Filled Height[mm] | Duration/Bed[min] | P[Mcts] | R[Mcts] | S[Mcts] | T[Mcts] | R/T | SNR |
|---|---|---|---|---|---|---|---|---|
| 100 | 18 | 20 | 6.49 | 5.50 | 0.35 | 0.64 | 8.6 | 0.85 |
| 50 | 36 | 20 | 6.56 | 5.55 | 0.35 | 0.66 | 8.5 | 0.85 |
| 5 | 36 | 200 | 36.31 | 35.25 | 0.37 | 0.69 | 51.2 | 0.60 |
Prompts (P) and randoms (R) correspond to the counts recorded in the prompt and delayed coincidence window. S was estimated as 0.35 (P-R). (Mcts: megacounts). See the for an SNR curve estimation.
Mean and variance of the counts per pixel normalized to 100 for the largest hot rod sector of each setup for the two different TOF resolution (TOFr) systems.
| TOFr | Phantom | 12.7 mm | 11.1 mm | 9.5 mm | 7.9 mm | 6.4 mm |
|---|---|---|---|---|---|---|
| 550 ps | D: 90Y 100Gy | 100.0 ± 15.9 | 96.2 ± 10.9 | 80.8 ± 11.9 | 68.2 ± 13.0 | 52.2 ± 8.9 |
| G: 18F DG WB | 100.0 ± 5.5 | 89.0 ± 5.4 | 77.9 ± 6.1 | 63.1 ± 5.1 | 48.7 ± 5.3 | |
| 320 ps | E: 90Y 100Gy | 100.0 ± 7.7 | 109.5 ± 11.5 | 84.6 ± 7.7 | 68.7 ± 10.8 | 62.2 ± 11.6 |
| H: 18F DG WB | 100.0 ± 7.1 | 89.5 ± 10.1 | 82.2 ± 10.5 | 68.0 ± 6.1 | 51.4 ± 5.4 |
Figure 3(A) Overall survival fraction (OSF) in HCC liver radioembolizations with resin spheres (solid line) and glass spheres (dashed line) using the same 40Gy-dose threshold on the 550ps TOF-PET based EUD:α=0.035Gy-1 (reprinted from (34) with permission of IOP publishing). (B) comparison of the 550ps TOF-PET based sectors EUD:α=0.035Gy-1 (triangles, diamonds are without and with 6mm-FWHM filtering, respectively) with the true sectors EUD:α=0.021Gy-1(circles) for a mean sector dose D of 50Gy (blue) and 100Gy (brown).