| Literature DB >> 34672690 |
Shao Jin Ong1, Gopinathan Anil1,2, Koon Liang Chia1, Deborah Khoo3, Joseph Kt Lee1,4, Priscilla Xh Chen1, Teddy M Nares1, Calvin J Koh5,6, Peijing Su7, Cunli Yang1, Pavel Singh1, Prapul C Rajendran1, Timothy Fotheringham8, Swee T Quek1, Ian Renfrew8.
Abstract
OBJECTIVE: To evaluate the efficacy of a barrier shield in reducing droplet transmission and its effect on image quality and radiation dose in an interventional suite.Entities:
Mesh:
Year: 2021 PMID: 34672690 PMCID: PMC8722246 DOI: 10.1259/bjr.20210835
Source DB: PubMed Journal: Br J Radiol ISSN: 0007-1285 Impact factor: 3.039
Figure 1.SIR AGP Shield. (a) Demonstration of use within a bi-planar Interventional radiology theatre. (b) Side-holes with overlapping silicone flaps for assistant access in simulated intubation–extubation procedures and a cranial access port for intubation bougie and endotracheal tube access. AGP, Aerosol Generating Procedure; SIR, Safety in Interventional Radiology.
Figure 2.Fluorescent cough droplets. Edible fluorescent fluid was gargled and ingested by healthy human volunteers prior to coughing supine in a dark room. Cough droplet splatter was visualised with ultraviolet-A light.
Figure 3.Optical clarity pre- and post-200 cycles of cleaning with hospital disinfectants. Optical clarity of 5 mm polycarbonate was measured pre- and post-200 cycles of cleaning with Ammonia-based Mikrozid Sensitive wipes, Virusolve+® Sporicidal Wipes and 70% isopropyl alcohol wipes. No significant difference was observed between the optical clarity pre- and post-200 cycles of cleaning with either of the disinfectants.
Line pairs per mm visible on fluoroscopy, 2 FPS DSA, 3FPS DSA, 6FPS DSA and single shot acquisition with and without the shield
| Line pairs per mm | Fluoroscopy | 2FPS DSA | 3FPS DSA | 6FPS DSA | Single shot acquisition | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Reader | No shield | AGP shield | No shield | AGP shield | No shield | AGP shield | No shield | AGP shield | No shield | AGP shield |
| 1 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 |
| 2 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 |
| 3 | 1.6 | 1.6 | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 |
| 4 | 1.6 | 1.6 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 5 | 1 | 1 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 6 | 1.25 | 1.25 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 7 |
|
| 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 8 | 1.25 | 1.25 | 2 | 2 | 2 | 2 | 2 | 2 |
|
|
| 9 | 2 | 2 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 |
| 10 | 1.6 | 1.6 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
| 11 | 1.25 | 1.25 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 | 1.6 |
| 12 | 1.6 | 1.6 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 |
AGP, Aerosol Generating Procedure; DSA, digital subtraction angiography; FPS, frames per second.
Apart from one outlier reading each for reader 7 and 8, no difference in spatial resolution was observed with or without the shield by other readers.
Figure 4.Contrast to noise ratios with and without the Shield on image quality test phantom. The CNR of positive contrast dots on an image quality test phantom was plotted for fluoroscopy (A), digital subtraction angiogram (DSA) at two frames per second (FPS) (B), DSA at 3 FPS (C), DSA at 6 FPS (D) and single shot acquisition (E). No statistical significance was demonstrated between the CNR with and without use of the Shield. DSA, digital subtraction angiography; FPS, frames per second.
Measured and reported radiation dosage with and without the Shield
| Positioning and exposure protocols | kV | Time (s) | Ka,I (mGy) | Ka,r (mGy) | PKA (mGycm2) | |
|---|---|---|---|---|---|---|
| Fluoroscopy | Native | 74 | 59.9 | 1.86 | - | - |
| Elevated | 68 | 59.5 | 4.10 | - | - | |
| Elevated with Shield | NR | 59.4 | 4.49 | - | - | |
| DSA 2FPS | Native | 69 | 20.3 | 7.23 | 13.3 | 5.88 |
| Elevated | 67 | 20.3 | 22.9 | 25.9 | 7.99 | |
| Elevated with Shield | 68 | 20.3 | 28.9 | 32.3 | 9.98 | |
| DSA 3FPS | Native | 69 | 20.0 | 11.01 | 20.3 | 8.95 |
| Elevated | 68 | 20.0 | 40.0 | 44.5 | 13.7 | |
| Elevated with Shield | 68 | 20.1 | 43.8 | 48.8 | 15.1 | |
| DSA 6FPS | Native | 69 | 20.0 | 22.2 | 40.9 | 18.0 |
| Elevated | 71 | 20.0 | 95.3 | 103 | 31.9 | |
| Elevated with Shield | 71 | 20.0 | 107 | 117 | 36.0 | |
| Single Shot Acquisition | Native | 72 | 0.0168 | 0.40 | 0.70 | 0.312 |
| Elevated | 67 | 0.0641 | 0.59 | 0.70 | 0.207 | |
| Elevated with Shield | 68 | 0.0520 | 0.77 | 0.90 | 0.266 | |
AGP, Aerosol Generating Procedure; DSA, digital subtraction angiography; FPS, frames per second.
Incident Air Kerma (Ka,I) readings were obtained from the dosemeter while Air–Kerma at the Interventional Reference Point (Ka,r) and Kerma–Air Product (PKA) were recorded from the Arits System.