| Literature DB >> 34668071 |
Nastasja Burgardt1, Melanie Weissenberger2.
Abstract
The newly released Spectrum Compact CE System by Promega is a capillary electrophoresis instrument developed for DNA-fragment separation and sequencing. In this study, its compatibility to 8 commercial short tandem repeat (STR) kits from 4 different manufacturers, reproducibility (sizing precision, accuracy and concordance) and robustness (sensitivity and mixture resolution) were tested and compared to the ABI PRISM® 310 Genetic Analyzer. The instrument was able to successfully analyse amplicons of all tested kits, proved to be as precise as claimed by manufacturer specifications and was shown to be more robust than the ABI PRISM® 310 Genetic Analyzer in some aspects. Analyses on the Spectrum Compact CE System were able to resolve peaks with length differences of 1-basepair in the short and long fragment range and mixtures of mixture ratios down to 1:30. We describe the advantages and limitations we have observed so far working with this instrument in our forensic genetics laboratory.Entities:
Keywords: DNA; Forensic genetics; Promega; STR; Spectrum Compact CE System
Mesh:
Substances:
Year: 2021 PMID: 34668071 PMCID: PMC8813819 DOI: 10.1007/s00414-021-02673-1
Source DB: PubMed Journal: Int J Legal Med ISSN: 0937-9827 Impact factor: 2.686
Fig. 1Pie chart of percentages of electrophoresis methods/instruments used in forensic genetics publications released going back 1 year from 18.01.2021 (n = 86) on https://pubmed.ncbi.nlm.nih.gov/, accessed under the search terms “forensic genetics DNA”, G.A., Genetic Analyzer
Instrument features of the ABI PRISM® 310 Genetic Analyzer and the Spectrum Compact CE System in comparison
| ABI PRISM® 310 Genetic Analyzer | Spectrum Compact CE System | |
|---|---|---|
| Capillary number/array | 1 | 4 |
| Capillary length | 47 cm | 36 cm |
| Possible dye channel colours | Up to 5 | Up to 6 |
| Polymer for fragment analysis | POP4 (in this study) | Spectrum Compact Polymer4 |
| Laser/wavelength | 10-mW argon ion class 1 laser (when safety interlocks are enabled, otherwise class 3b), 488 nm and 514 nm | 20-mW class 1 laser, 505 nm |
| Detection unit | CCD camera | CCD camera |
| Sample tray | 48 single sample tubes (in this study) | 4 strips × 8 sample wells = 32 samples |
| Control system | External computer required | All system functions can be controlled at the instrument touchscreen itself, external computer required for data analysis |
| Output data format for fragment analysis | .fsa-files | .fsa-files |
| Data connection | Automatic transfer of.fsa-files to connected computer | Instrument has its own network connection, download of.fsa-files by remote software from any device that is connected to the network possible, as well as directly from the instrument (USB port at the front of the instrument) |
| Quality indicators calculated by instrument | None | Spectral calibration: Quality value and condition number Fragment analysis: Sizing quality and electrophoresis quality |
| Runtime (including washing and filling capillary steps) | 35 min/sample | 42 min/4 samples |
| Size and weight | 61 cm × 55.9 cm × 86.4 cm, 90 kg | 40 cm × 60 cm × 60 cm, 45 kg |
| Power input | 240 Vac; 50/60 Hz; 2.4 kVA | 100–240Vac; 50/60 Hz; 260VA |
| Operating temperature | 15 to 30 °C | 15 to 30 °C |
| Soft stops | Warning when there is not enough polymer left for the planned run setup Warning when the capillary exceeded its expiry (after 100 injections) | Warning anode buffer (number of usage/expiry) Warning cathode buffer (number of usage/expiry) Warning capillary array (number of usage/expiry) Warning polymer (expiry date) |
| Hard stops | None (except when polymer is empty, if the warning was ignored, the instrument stops in the middle of the run when polymer is empty) | When polymer is empty or is less than required for the planned run set up, the run cannot be started |
STR-kits with corresponding pre-installed filter assay colour sets/run-assays that were used on the Spectrum Compact CE System
| Kit | Assay | Kit dyes | Filter assay colours |
|---|---|---|---|
| MPX5ESSv5 | Filter6_5Dye_O600(60–600)_36_P4 | FAM, JOE-C6, YEL, RED & ORN | 6-FAM™, VIC®, NED™, PET™ & LIZ™ |
| NGM Detect™ | T_6Dye_O600(60–600)_36_P4 | 6-FAM™, VIC™, TED™, TAZ™, SID™ & LIZ™ | 6-FAM™, VIC™, NED™, TAZ™, SID™, LIZ™ |
| PowerPlex® ESX17 & PowerPlex® Y23 | Promega_5Dye_WENILS_36_P4 | Fluorescein, JOE, TMR-ET, CXR-ET & WEN | Fluorescein, JOE, TMR-ET, CXR-ET & WEN |
| Investigator® Argus X-12 QS, Investigator® ESSplex SE QS & Investigator® IDPlex Plus | Q_5Dye_BTO550_36_P4 | 6-FAM™, BTG, BTY, BTR & BTO | 6-FAM™, BTG, BTY, BTR & BTO |
| Investigator® 24plex QS | Q_6Dye_BTO550_36_P4 | 6-FAM™, BTG, BTY, BTR2, BTP & BTO | 6-FAM™, BTG, BTY, BTR2, BTP & BTO |
Run conditions used on the ABI PRISM® 310 Genetic Analyzer and Spectrum Compact CE System
| Instrument | Injection time | Injection voltage | Run voltage | Run temperature | Run time |
|---|---|---|---|---|---|
| ABI PRISM® 310 Genetic Analyzer | 5 s | 15 kV | 15 kV | 60 °C | 28 min |
| Spectrum Compact CE System | 9 s | 1.6 kV | 13 kV | 60 °C | 32.17 min |
Spectral calibration parameter pass conditions and measured values for all four capillaries of the capillary-array of the Spectrum Compact CE System
| Min. quality value | Max. condition number | Capillary number | Quality value measured | Condition number measured | |
|---|---|---|---|---|---|
| Promega 5-DYE | ≥ 0.95 | ≤ 8.5 | 1 | 0.996 | 5.66 |
| 2 | 0.996 | 5.81 | |||
| 3 | 0.995 | 5.67 | |||
| 4 | 0.997 | 5.54 | |||
| T6dye | ≥ 0.95 | ≤ 8.0 | 1 | 0.995 | 5.91 |
| 2 | 0.991 | 6.30 | |||
| 3 | 0.994 | 6.05 | |||
| 4 | 0.998 | 5.90 | |||
| Q5dye | ≥ 0.95 | ≤ 20 | 1 | 0.994 | 6.84 |
| 2 | 0.994 | 6.78 | |||
| 3 | 0.996 | 6.75 | |||
| 4 | 0.994 | 6.73 | |||
| Q6dye | ≥ 0.95 | ≤ 13.5 | 1 | 0.993 | 11.19 |
| 2 | 0.993 | 11.23 | |||
| 3 | 0.991 | 11.16 | |||
| 4 | 0.996 | 11.08 | |||
| Filter6 5dye | ≥ 0.95 | ≤ 13.5 | 1 | 0.979 | 6.13 |
| 2 | 0.997 | 6.83 | |||
| 3 | 0.998 | 6.71 | |||
| 4 | 0.997 | 6.62 |
Limits of detection (LOD) and limits of quantitation (LOQ) on the Spectrum Compact CE System (n = 40) for the kits PowerPlex® ESX17, PowerPlex® Y23, Investigator® Argus X-12 QS and NGM Detect™, rounded to the closest 5-value (rfu)
| LOD (rfu) Spectrum Compact CE System | LOQ (rfu) Spectrum Compact CE System | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Colour channel | Blue | Green | Yellow | Red | Violet | Blue | Green | Yellow | Red | Violet |
| PowerPlex® ESX17 | 25 | 20 | 20 | 35 | n/a | 60 | 40 | 50 | 75 | n/a |
| PowerPlex® Y23 | 25 | 25 | 20 | 50 | n/a | 50 | 55 | 55 | 105 | n/a |
| Investigator® Argus X-12 QS | 35 | 20 | 20 | 55 | n/a | 75 | 40 | 45 | 120 | n/a |
| NGM Detect™ | 30 | 30 | 35 | 40 | 30 | 50 | 50 | 75 | 65 | 55 |
Fig. 2Dynamic ranges of the Spectrum Compact (S.C.) CE System (continuous arrows) and the ABI Prism® 310 (ABI310) Genetic Analyzer (dashed arrows) with the limit of detection (LOD), limit of quantitation (LOQ) and limit of linearity (LOL) thresholds for each instrument for each dye of the PowerPlex® ESX17 kit in relation to the maximum on-scale peak signals per DNA template input amount, (n = 76 for each colour on the Spectrum Compact CE System, n = 19 for each colour on the ABI Prism® 310 Genetic Analyzer); a blue colour channel; b green colour channel; c yellow colour channel; d red colour channel
Fig. 3Electropherograms of Gednap 47 trace 4 sample in the locus D1S1656 with indicated valley heights; a PowerPlex® ESX17 on the Spectrum Compact CE System, amplicon 1; b PowerPlex® ESX17 on the ABI PRISM® 310 Genetic Analyzer; c NGM Detect™ on the Spectrum Compact CE System, amplicon 2
Fig. 4Electropherograms of a mixture of the two DNA amplification positive controls 2800 M and DNA Control 007 in 5 mixture ratios at the locus SE33 (minor alleles 17/25.2 and major alleles 15/16), amplified with the PowerPlex® ESX17 kit and analysed on the Spectrum Compact CE System, only true alleles are labelled; a mixture ratio 1:1; b mixture ratio 5:1; c mixture ratio 10:1; d mixture ratio 20:1; e mixture ratio 30:1
Fig. 5Electropherograms of a mixture of the two DNA amplification positive controls 2800 M and DNA Control 007 in 5 mixture ratios at the locus D16S539 (minor alleles 9/10 and major alleles 9/13) amplified with the PowerPlex® ESX17 kit and analysed on the Spectrum Compact CE System, only true alleles are labelled; a mixture ratio 1:1; b mixture ratio 5:1; c mixture ratio 10:1; d mixture ratio 20:1; e mixture ratio 30:1
Fig. 6Electropherograms of a mixture of the two DNA amplification positive controls 2800 M and DNA Control 007 in 4 mixture ratios at the locus D22S1045 (minor alleles 11/16 and major alleles 16/16), amplified with the PowerPlex® ESX17 kit and analysed on the Spectrum Compact CE System, only true alleles are labelled; a mixture ratio 1:1; b mixture ratio 5:1; c mixture ratio 10:1; d mixture ratio 20:1, only true alleles are labelled, unlabelled peaks that are not in stutter-positions correspond to pull-ups from other systems
Fig. 7Colour channel balances (%) on the Spectrum Compact CE System and the ABI PRISM® 310 Genetic Analyzer based on n = 17 samples per instrument per kit with the ideal percentage share indicated by a thick line; a PowerPlex® ESX17; b PowerPlex® Y23; c Investigator® Argus X-12 QS; d NGM Detect™