| Literature DB >> 34654870 |
Debajit Bhowmick1, Frank van Diepen2, Anita Pfauth2, Renaud Tissier3, Michelle L Ratliff4.
Abstract
In conventional flowcytometry one detector (primary) is dedicated for one fluorochrome. However, photons usually end up in other detectors too (fluorescence spillover). 'Compensation' is a process that corrects the spillover signal from all detectors except the primary detector. Post 'compensation', the photon counting error of spillover signals become evident as spreading of the data. The spreading induced by spillover impairs the ability to resolve stained cell population from the unstained one, potentially reducing or completely losing cell populations. For successful multi-color panel design, it is important to know the expected spillover to maximize the data resolution. The Spillover Spreading Matrix (SSM) can be used to estimate the spread, but the outcome is dependent on detector sensitivity. Simply, the same single stained sample produces different spillover spread values when detector(s) sensitivity is altered. Many researchers mistakenly use this artifact to "reduce" the spread by decreasing detector sensitivity. This can result in diminished capacity to resolve dimly expressing cell populations. Here, we introduce SQI (Spread Quantification Index), that can quantify the spillover spread independent of detector sensitivity and independent of dynamic range. This allows users to compare spillover spread between instruments having different types of detectors, which is not possible using SSM.Entities:
Mesh:
Year: 2021 PMID: 34654870 PMCID: PMC8520008 DOI: 10.1038/s41598-021-99831-7
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Concept of the SQI: The figure shows the concept of the SQI value. Spillover fluorescence from primary fluorochrome X results in spread in detector B, and vice versa. The spread is quantified as the difference of the 99th percentile and 50th percentile in dimension B. Subsequently, this value is multiplied by a normalization factor to make the SQI value independent of detector gain settings. The 99th percentile is marked by a Green line.
Figure 2Effect of PMT voltage on ISS: In (A), the primary detector (V605) voltage was changed from 600 to 400 V in steps of 100 V. This did not change the rSD and SQI values measuring the spread in the secondary detector (V677). However, the ISS values changed noticeably. In (A–C), voltages of both detectors were changed simultaneously to 600 V, 500 V or 400 V respectively. The SQI values remained the same for both detectors in all cases, but the ISS values decreased considerably. In (D) only voltage of V677 was lowered which resulted in a huge increase of ISS but SQI remain same.
Comparison of spread in four different flow cytometers: ten different single stained bead sets were run on four different instruments.
SQI was calculated for every combination. For BV711 there was no data available for the Cytoflex S. The SQI values are categorized as follows: 1–120 (Green); 121–199 (Yellow); 200–299 (Orange); and 300 + (Red).
Comparison of all four Fusion setup: the SQI values for 10 fluorochromes run on three different Fusions are shown.
For Fusion 2, “2_High” denotes higher laser power as for the other two instruments. The SQI values are categorized as follows: 1–120 (Green); 121–199 (Yellow); 200–299 (Orange); and 300 + (Red).
PMT voltage independent measure of spread: two different sets of PMT voltages were applied to set the MFI of the positive population to 110,000 (110 K) or 50,000 (50 K).
ISS values are shown in Italics.
Six single stained bead samples were run using these two voltage sets, and the ISS and SQI values were calculated. The SQI values were similar at the different voltages, whereas the ISS values were voltage different. ISS values are shown in Italics.
Figure 3Effect of detector settings on SQI and ISS: In (A, B) we have only changed the PMT voltage for the primary detector. SQI remains same but ISS dropped with increase of voltage. In (C, D), we only changed PMT voltage for secondary detector. SQI again remain same but ISS increased with voltage.