Literature DB >> 34651705

Comparison of a contrast-to-noise ratio-driven exposure control and a regular detector dose-driven exposure control in abdominal imaging in a clinical angiography system.

Thomas Werncke1, Marius Kemling1, Stanislav Tashenov2, Jan B Hinrichs1, Timo C Meine1, Sabine K Maschke1, Yiannis Kyriakou3, Frank K Wacker1, Bernhard C Meyer1.   

Abstract

PURPOSE: The first purpose of this phantom study was to verify whether a contrast-to-noise ratio (CNR)-driven exposure control (CEC) can maintain target CNR in angiography more precisely compared to a conventional detector dose-driven exposure control (DEC). The second purpose was to estimate the difference between incident air kerma produced by CEC and DEC when both exposure controls reach the same CNR.
METHODS: A standardized 3D-printed phantom with an iron foil and a cavity, filled with iodinated contrast material, was developed to measure CNR using different image acquisition settings. This phantom was placed into a stack of polymethylmethacrylate and aluminum plates, simulating a patient equivalent thickness (PET) of 2.5-40 cm. Images were acquired using fluoroscopy and digital radiography modes with CEC using one image quality level and four image quality gradients and DEC having three different detector dose levels. The spatial frequency weighted CNR and incident air kerma were determined. The differences in incident air kerma between DEC and CEC were estimated.
RESULTS: When using DEC, CNR decreased continuously with increasing attenuation, while CEC within physical limits maintained a predefined CNR level. Furthermore, CEC could be parameterized to deliver the CNR as a predefined function of PET. To provide a given CNR level, CEC used equal or lower air kerma than DEC. The mean estimated incident air kerma of CEC compared to DEC was between 3% (PET 20 cm) and 40% (PET 27.5 cm) lower in fluoroscopy and between 1% (PET 20 cm) and 55% (PET 2.5 cm) lower in digital radiography while maintaining CNR.
CONCLUSION: Within physical and legislative limits, the CEC allows for a flexible adjustment of the CNR as a function of PET. Thus, the CEC enables task-dependent examination protocols with predefined image quality in order to easier achieve the as low as reasonably achievable principle. CEC required equal or lower incident air kerma than DEC to provide similar CNR, which allows for a substantial reduction of skin radiation dose in these situations.
© 2021 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  Image quality; air kerma; angiography; contrast-to-noise ratio; fluoroscopy; radiation dose reduction; radiography; skin-dose

Mesh:

Year:  2021        PMID: 34651705     DOI: 10.1002/mp.15288

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Tantalum-specific contrast-to-noise ratio or conventional detector dose-driven exposure control in angiography: radiation dose and image quality evaluation in a porcine model.

Authors:  Thomas Werncke; Timo Christian Meine; Jan B Hinrichs; Sabine K Maschke; Lena Sophie Becker; Inga Brüsch; Regina Rumpel; Frank K Wacker; Bernhard C Meyer
Journal:  Eur Radiol Exp       Date:  2022-05-17

Review 2.  Review and investigation of automatic brightness/dose rate control logic of fluoroscopic imaging systems in cardiovascular interventional angiography.

Authors:  Pei-Jan Paul Lin; Allen R Goode; Frank D Corwin
Journal:  Radiol Phys Technol       Date:  2022-01-20