Jacob S Parzen1, Weili Zheng1, Xiaoqiang Li1, Xuanfeng Ding1, Peyman Kabolizadeh1.
Abstract
PURPOSE: Preoperative chemoradiation represents the standard of care in patients with locally advanced rectal cancer. Robustness is often compromised in the setting of proton beam therapy owing to the sensitivity of proton particles to tissue heterogeneity, such as with intestinal gas. The ideal beam arrangement to mitigate the anatomic uncertainty caused by intestinal gas is not well defined. METHODS AND MATERIALS: We developed pencil beam scanning plans using (1) 1-beam posteroanterior (PA) plans, (2) 2-beam with right and left posterior oblique (RPO and LPO) plans, (3) 3-beam with PA and opposed lateral plans, and (4) 5-beam with PA, RPO, LPO, and opposed lateral plans. We created 12 plans with robustness optimization and ran a total of 60 plan evaluations for varying degrees of intestinal gas distension to evaluate which plans would maintain clinical goals to the greatest degree.Entities:
Year: 2021 PMID: 34646968 PMCID: PMC8498733 DOI: 10.1016/j.adro.2021.100749
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
Institutional clinical goals and dose constraints based on RTOG 0822
| Target/Organ at risk | Dose constraint | Clinical goal |
|---|---|---|
| CTV45 | V100% | >95-97% |
| CTV50 | V100 | >95-97% |
| CTV50 | Dmax | <110% |
| Small bowel | V40Gy | <70 cm3 |
| V35Gy | <300 cm3 | |
| V30Gy | <350 cm3 | |
| Dmax | <52 Gy | |
| Bladder | V40Gy | <50% |
| V30Gy | <60% | |
| Genitalia | V30Gy | <35% |
| V20Gy | <50% | |
| Femoral heads | V45Gy | <5% |
| V40Gy | <30% | |
| V30Gy | <35% |
Abbreviations: CTV45 = clinical target volume prescribed 45 GyE; CTV50 = clinical target volume prescribed 50 GyE; Dmax = maximum dose to 0.03 cubic centimeters; RTOG = Radiation Therapy Oncology Group; V100% = volume receiving 100% of the prescribed dose; V40Gy = percentage of volume receiving 40 Gy.
Dosimetric parameters for the 4 beam arrangements optimized to the planning CT and iterated across rectal diameter sizes
| Dosimetric parameter | Beam arrangement | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 beam (PA) | 3 beam (PA, opposed lateral) | ||||||||||||
| Plan CT | 1 cm | 2 cm | 3 cm | 4 cm | 5 cm | Plan CT | 1 cm | 2 cm | 3 cm | 4 cm | 5 cm | ||
| CTV45 | V100% (%) | 99.0 | 98.8 | 99.1 | 99.1 | 99.0 | 99.0 | 99.9 | 99.9 | 99.9 | 99.9 | 99.9 | 99.9 |
| CTV50 | V100% (%) | 98.5 | 96.0 | 93.0 | 88.8 | 83.0 | 78.4 | 100.0 | 99.9 | 99.9 | 99.8 | 99.7 | 98.8 |
| CTV50 | Dmax (%) | 52.6 | 53.5 | 53.4 | 53.6 | 53.3 | 53.8 | 52.4 | 52.8 | 52.8 | 53.0 | 53.1 | 52.9 |
| Small bowel | V40Gy (%) | 68.2 | 68.2 | 71.0 | 75.7 | 81.8 | 88.8 | 75.3 | 75.5 | 76.0 | 76.8 | 77.5 | 78.0 |
| V35Gy (%) | 88.7 | 89.1 | 92.3 | 97.6 | 104.1 | 112.4 | 98.6 | 98.8 | 99.7 | 100.9 | 102.0 | 103.0 | |
| V30Gy (%) | 105.6 | 106.0 | 109.6 | 115.1 | 122.2 | 130.1 | 116.5 | 116.8 | 118.1 | 119.7 | 121.0 | 122.4 | |
| Dmax (Gy) | 49.3 | 49.3 | 49.5 | 49.5 | 49.2 | 49.5 | 47.7 | 47.7 | 47.6 | 47.6 | 47.6 | 47.8 | |
| Bladder | V40Gy (%) | 23.0 | 26.9 | 36.7 | 50.4 | 65.9 | 81.6 | 24.8 | 25.5 | 26.5 | 27.5 | 28.7 | 29.4 |
| V30Gy (%) | 31.0 | 36.3 | 47.3 | 61.4 | 76.7 | 88.9 | 34.4 | 36.0 | 38.2 | 40.7 | 42.7 | 44.2 | |
| Internal genitalia | V30Gy (%) | 14.7 | 25.2 | 35.1 | 40.8 | 41.1 | 41.2 | 15.2 | 17.3 | 18.1 | 18.3 | 18.1 | 18.3 |
| V20Gy (%) | 19.2 | 31.6 | 41.9 | 46.1 | 46.5 | 46.3 | 21.6 | 25.6 | 27.6 | 28.2 | 28.2 | 28.1 | |
Abbreviations: CT = computed tomography; CTV45 = clinical target volume prescribed 45 GyE; CTV50 = clinical target volume prescribed 50 GyE; Dmax = maximum dose to 0.03 cubic centimeters; LPO = left posterior oblique; PA = posteroanterior; RPO = right posterior oblique; V100% = volume receiving 100% of the prescribed dose; V40Gy = percentage of volume receiving 40 Gy.
Values did not meet institutional dose constraints.
Fig. 1Plan comparison for (A) posteroanterior (PA) field arrangement; (B) right posterior oblique (RPO) / left posterior oblique (LPO); (C) PA/opposed lateral; and (D) RPO/LPO/PA/opposed lateral on plan CT (above) and with 5-cm rectal distension (below). With PA field arrangement, there was a high-dose region extending anteriorly to the bladder with maximal rectal distension. In contrast, with maximal rectal distension, target coverage perturbations in dose distributions were much more modest for the other beam arrangements.
Dosimetric parameters for the 4 beam arrangements optimized to the 5-cm rectal diameter and iterated across rectal diameter sizes including plan CT
| Dosimetric parameter | Beam arrangement | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 beam (PA) | 3 beam (PA, opposed lateral) | ||||||||||||
| 5 cm | 4 cm | 3 cm | 2 cm | 1 cm | Plan CT | 5 cm | 4 cm | 3 cm | 2 cm | 1 cm | Plan CT | ||
| CTV45 | V100 | 99.9 | 98.1 | 94.2 | 89.0 | 84.1 | 81.5 | 100.0 | 99.7 | 97.9 | 90.1 | 79.0 | 72.5 |
| CTV50 | V100 | 97.6 | 94.7 | 87.5 | 77.1 | 67.0 | 63.4 | 99.8 | 98.2 | 84.8 | 61.2 | 41.4 | 34.7 |
| CTV50 | Dmax (%) | 55.6 | 56.5 | 58.9 | 60.9 | 58.9 | 56.6 | 53.9 | 53.8 | 54.4 | 55.5 | 56.6 | 54.7 |
| Small bowel | V40Gy (%) | 68.6 | 64.7 | 62.6 | 61.9 | 61.9 | 61.8 | 77.2 | 75.5 | 73.8 | 72.9 | 72.6 | 72.4 |
| V35Gy (%) | 85.9 | 81.4 | 78.6 | 77.5 | 77.2 | 77.1 | 100.3 | 98.1 | 96.4 | 95.3 | 94.7 | 94.6 | |
| V30Gy (%) | 102.0 | 96.4 | 93.0 | 91.7 | 91.5 | 91.3 | 119.5 | 117.1 | 115.1 | 114.1 | 113.6 | 113.6 | |
| Dmax (Gy) | 50.6 | 50.4 | 51.0 | 50.6 | 50.7 | 50.3 | 48.1 | 48.0 | 48.1 | 48.1 | 48.1 | 48.1 | |
| Bladder | V40Gy (%) | 24.0 | 8.4 | 2.4 | 1.8 | 1.8 | 1.8 | 26.0 | 19.9 | 14.4 | 7.9 | 6.6 | 6.6 |
| V30Gy (%) | 36.0 | 17.0 | 6.5 | 4.7 | 4.7 | 4.7 | 36.0 | 31.3 | 27.5 | 25.6 | 25.0 | 25.0 | |
| Internal genitalia | V30Gy (%) | 23.6 | 23.3 | 20.0 | 13.0 | 5.7 | 2.1 | 17.8 | 17.7 | 17.5 | 16.4 | 14.6 | 12.6 |
| V20Gy (%) | 33.2 | 32.7 | 29.6 | 21.8 | 13.1 | 6.3 | 26.4 | 26.3 | 26.1 | 23.7 | 21.8 | 20.3 | |
Abbreviations: CT = computed tomography; CTV45 = clinical target volume prescribed 45 GyE; CTV50 = clinical target volume prescribed 50 GyE; Dmax = maximum dose to 0.03 cubic centimeters; LPO = left posterior oblique; PA = posteroanterior; RPO = right posterior oblique; V100% = volume receiving 100% of the prescribed dose; V40Gy = percentage of volume receiving 40 Gy.
Values did not meet institutional dose constraints.
Dosimetric parameters for the 2 and 3 beam arrangements optimized to the 4-cm rectal diameter and 3-cm rectal diameter iterated across rectal diameter sizes including plan CT
| Dosimetric parameter | Beam arrangement | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 2 beam (RPO, LPO), plan 4 cm | 2 beam (RPO, LPO), plan 3 cm | ||||||||||||
| 5 cm | 4 cm | 3 cm | 2 cm | 1 cm | Plan CT | 5 cm | 4 cm | 3 cm | 2 cm | 1 cm | Plan CT | ||
| CTV45 | V100% (%) | 100.0 | 99.9 | 99.9 | 99.0 | 94.5 | 88.9 | 100.0 | 100.0 | 100.0 | 100.0 | 99.4 | 95.5 |
| CTV50 | V100% (%) | 99.9 | 99.9 | 99.7 | 92.4 | 78.8 | 69.0 | 99.9 | 100.0 | 100.0 | 99.7 | 93.1 | 82.6 |
| CTV50 | Dmax (%) | 53.9 | 54.3 | 54.5 | 54.6 | 54.8 | 55.5 | 54.3 | 54.3 | 54.3 | 54.2 | 54.5 | 55.5 |
| Small bowel | V40Gy (%) | 69.5 | 68.5 | 67.8 | 67.7 | 67.7 | 67.4 | 70.4 | 70.0 | 69.6 | 69.6 | 69.3 | 69.3 |
| V35Gy (%) | 85.3 | 84.4 | 83.6 | 83.6 | 83.2 | 83.4 | 86.3 | 85.8 | 85.3 | 85.2 | 85.0 | 85.0 | |
| V30Gy (%) | 98.8 | 98.0 | 97.6 | 97.3 | 97.1 | 97.1 | 100.1 | 99.3 | 98.7 | 98.5 | 98.6 | 98.5 | |
| Dmax (Gy) | 48.4 | 48.3 | 48.4 | 49.8 | 50.0 | 50.0 | 48.5 | 48.4 | 48.4 | 48.4 | 48.5 | 48.5 | |
| Bladder | V40Gy (%) | 21.4 | 20.8 | 20.6 | 20.5 | 20.5 | 20.4 | 21.6 | 21.3 | 21.1 | 21.1 | 21.0 | 21.0 |
| V30Gy (%) | 27.1 | 26.3 | 25.9 | 25.7 | 25.7 | 25.6 | 26.9 | 26.4 | 26.1 | 26.0 | 26.0 | 26.0 | |
| Internal genitalia | V30Gy (%) | 15.2 | 15.2 | 14.5 | 12.7 | 11.3 | 11.1 | 16.9 | 16.9 | 16.5 | 15.1 | 14.4 | 14.1 |
| V20Gy (%) | 22.5 | 22.5 | 22.1 | 20.6 | 20.2 | 20.1 | 24.4 | 24.4 | 23.9 | 22.9 | 22.6 | 22.4 | |
Abbreviations: CT = computed tomography; CTV45 = clinical target volume prescribed 45 GyE; CTV50 = clinical target volume prescribed 50 GyE; Dmax = maximum dose to 0.03 cubic centimeters; LPO = left posterior oblique; PA = posteroanterior; RPO = right posterior oblique; V100% = volume receiving 100% of the prescribed dose; V40Gy = percentage of volume receiving 40 Gy.
Values did not meet institutional dose constraints.