Sulaiman Al-Alawi1, Hala Al-Hinai2, Nawal Al-Kindi2, Mohammed Al-Rashidi1, Hanan Al-Kindi3, Intisar Al-Shukri3, Azza Al-Rashdi3, Sachin Jose4, Amina Al-Jardani3.
Abstract
OBJECTIVES: Considering the increasing, significant burden that coronavirus disease 2019 (COVID-19) imposes on the healthcare system, the need for simple, rapid, and affordable diagnostic tests to support the existing costly and demanding polymerase chain reaction (PCR) assay becomes required. This prospective diagnostic test accuracy study aims to evaluate the performance of four different COVID-19 rapid antigen tests compared to real-time reverse transcription PCR (rRT-PCR) between June and July 2020 to determine the feasibility of integrating these tests into the diagnostic algorithm in clinical settings.Entities:
Keywords: Reverse Transcriptase Polymerase Chain Reaction; COVID-19; Diagnostic Tests, Routine
Year: 2021 PMID: 34631156 PMCID: PMC8491111 DOI: 10.5001/omj.2021.106
Source DB: PubMed Journal: Oman Med J ISSN: 1999-768X
Diagnostic test characteristics of the four antigen detection assays and comparison of sensitivity, PPV, and NPV of rapid antigen tests in relation to Ct values.
| Antigen detection kit | PCR Ct value | Sensitivity | Specificity | PPV | NPV | Correlation with |
|---|---|---|---|---|---|---|
| STANDARDTM Q COVID-19 Ag test | Overall | 65.8 | 100 | 100 | 96.7 | rpb = -0.549, |
| Ct ≤ 35 | 71.0 | 100 | 97.1 | |||
| Ct ≤ 30 | 85.0 | 100 | 98.5 | |||
| Ct ≤ 25 | 92.3 | 100 | 99.2 | |||
| PCL COVID-19 Ag Rapid FIA test | Overall | 69.8 | 94.1 | 54.6 | 96.9 | rpb = -0.744, |
| Ct ≤ 35 | 75.5 | 56.5 | 97.4 | |||
| Ct ≤ 30 | 91.9 | 61.3 | 99.1 | |||
| Ct ≤ 25 | 100 (85.75–100) | 63.3 | 100 | |||
| BIOCREDIT COVID-19 Ag test | Overall | 64.0 | 100 | 100 | 96.5 | rpb = -0.645, |
| Ct ≤ 35 | 71.1 | 100 | 97.2 | |||
| Ct ≤ 30 | 85.7 | 100 | 98.6 | |||
| Ct ≤ 25 | 92.3 | 100 | 99.2 | |||
| Sofia SARS-CoV-2 antigen FIA test | Overall | 64.3 | 100 | 100 | 96.5 | rpb = -0.820, |
| Ct ≤ 35 | 66.0 | 100 | 96.7 | |||
| Ct ≤ 20 | 100 (89.11–100) | 100 | 100 |
PPV: positive predictive value; NPV: negative predictive value; PCR: polymerase chain reaction; rpb: point-biserial correlation coefficient; FIA: fluorescent immunoassay.
Data given as percentage (95% confidence interval).
Figure 1Correlation of rapid antigen tests (STANDARD™ Q COVID-19, PCL COVID-19 Ag Rapid FIA, BIOCREDIT COVID-19 Ag, and Sofia FIA) results with rRT-PCR Ct values: 0, Negative; 1, Positive.