Literature DB >> 34606573

Unexpected Anesthetic Circuit Leak Attributed to Improper Use of a Tube Holder: A Case Report.

Naotaka Kishimoto1, Akiko Otsuka1, Tatsuru Tsurumaki1, Kenji Seo1.   

Abstract

Leaks involving the anesthesia circuit can cause significant complications including hypoxia and hypoventilation. We present a case of a circuit leak caused by damage to the corrugated tubing attributed to improper use of the tube holder. A 58-year-old male was scheduled for resection of a palatal tumor under an intubated general anesthetic. After successful nasotracheal intubation, the anesthesiologist inserted the corrugated tubing of the anesthetic circuit into the tube holder. A leaking sound was heard and a tear in the corrugated tubing was promptly discovered. The corrugated tubing of the anesthetic circuit presumably tore because it was inserted into the groove of the tube holder at an inappropriate angle with excessive force. Anesthesiologists should be aware of potential leaks if the anesthesia circuit is damaged, which may be caused by improper use of tube holders.
© 2021 by the American Dental Society of Anesthesiology.

Entities:  

Keywords:  Anesthesia circuit; Corrugated tubing; Leak; Tube holder

Mesh:

Substances:

Year:  2021        PMID: 34606573      PMCID: PMC8500317          DOI: 10.2344/anpr-68-02-02

Source DB:  PubMed          Journal:  Anesth Prog        ISSN: 0003-3006


  15 in total

1.  An unusual anesthesia machine leak.

Authors:  D Mark
Journal:  J Clin Anesth       Date:  2000-02       Impact factor: 9.452

2.  Circuit leak from capnograph sampling line lodged under adjustable pressure limiting valve.

Authors:  Stephen Hennenfent; Barry Suslowicz
Journal:  Anesth Analg       Date:  2010-08       Impact factor: 5.108

3.  Unusual cause of leak in Datex Aisys.

Authors:  Prashant Kummar; Grace Korula; Sathish Kumar; P A Saravanan
Journal:  Anesth Analg       Date:  2009-10       Impact factor: 5.108

4.  A leak of concern.

Authors:  T Jack
Journal:  Br J Anaesth       Date:  1998-06       Impact factor: 9.166

5.  Gas leak due to a damaged GE Disposable Multi Absorber Canister used with an EZchange Module following its reinstallation during anesthesia.

Authors:  Hitomi Sano; Takashi Suzuki; Yuri Kaneda
Journal:  Can J Anaesth       Date:  2014-10-18       Impact factor: 5.063

6.  Another cause for a leak in a disposable breathing circuit.

Authors:  M B Sosis; M N Payne
Journal:  Anesthesiology       Date:  1989-11       Impact factor: 7.892

7.  Unusual source of air leak in a pediatric anesthesia breathing circuit.

Authors:  S Needleman; R F Kaplan
Journal:  Anesth Analg       Date:  1995-09       Impact factor: 5.108

8.  Leak in circuit: An unusual cause!

Authors:  Jyotsna Punj; Meenu Batra; V Darlong; R Pandey
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2012-01

9.  Circuit leak with D-vapor.

Authors:  Jeetinder Kaur Makkar; Kajal Jain; Amit Jain
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2015 Apr-Jun

10.  An unusual site of leak in anaesthesia circuit.

Authors:  P M Singh; Manpreet Kaur; Vimi Rewari
Journal:  Anesth Essays Res       Date:  2012 Jan-Jun
View more