Validation of Central Pressure Estimation in Patients with an Aortic Aneurysm Before and After Endovascular Repair.

S Holewijn1, J J M Vermeulen2,3, M van Helvert2,4, L van de Velde2,4, M M P J Reijnen2,4.   

Abstract

PURPOSE: The aim of this study was to investigate if non-invasive central pressure estimations are accurate in patients with an abdominal aortic aneurysm, before and after endovascular repair. Secondary evaluation was if measurement-accuracy was dependent on anatomical characteristics.
METHODS: Procedural invasive and non-invasive pressure-measurements were performed simultaneously both before and after endovascular repair in 20 patients with an infrarenal abdominal aortic aneurysm. Invasive catheter measurements were performed in the abdominal aorta. A tonometric device was used to perform non-invasive pressure-wave-analysis at the radial artery. A generalized transfer-function was used to generate an ascending aortic waveform for both measurements, allowing for direct comparison.
RESULTS: Pre-treatment the mean differences between methods were - 5.5 mmHg (p = .904), - 11.8 (p < .001), and - 7.2 mmHg (p = .124) for central systolic, diastolic, and mean pressure, respectively. The accuracy was dependent of aneurysm sac volume and intraluminal thrombus volume. Post-treatment limits of agreement were smaller for all pressure parameters compared to pre-treatment. The mean differences were 6.5 mmHg (p = .007), - 6.4 (p < .020), and 1.6 mmHg (p = .370) for central systolic, diastolic, and mean pressure, respectively.
CONCLUSION: In untreated AAA's the accuracy of non-invasive central pressure estimation was acceptable (mean difference between 5 and 10 mmHg) when compared to invasive pressures, but dependent of AAA characteristics. After EVAR the accuracy of central pressure estimation improved (reduction of 75% of the mean difference between pre and post measurements) TRIAL REGISTRATION NUMBER: NCT03469388; 3-5-2018.
© 2021. Biomedical Engineering Society.

Entities:  

Keywords:  Abdominal aortic aneurysm; Augmentation index; Central blood pressure; Endovascular aneurysm repair; Pressure wave analysis; Subendocardial viability ratio

Mesh:

Year:  2021        PMID: 34585343     DOI: 10.1007/s13239-021-00574-3

Source DB:  PubMed          Journal:  Cardiovasc Eng Technol        ISSN: 1869-408X            Impact factor:   2.305