Limitations of self-inflating resuscitators.

N N Finer, K J Barrington, F Al-Fadley, K L Peters.   

Abstract

In an effort to characterize the performance of self-inflating resuscitators, three examples of three models were subjected to laboratory testing: the Ohio Hope II resuscitator (Ohio Medical Products, Madison, WI), the PMR-2 resuscitator (Puritan Medical Products, Lenexa, KS), and the Laerdal resuscitator (AS Laerdal, Stavanger, Norway). The devices were connected to a test lung and compressed at frequencies of from 10 to 60 and at greater than 60 breaths per minute at 5, 10, and 15 L/min of flow. These devices were used with and without a reservoir and were compressed at less than and more than the pop-off valve pressures. The results revealed that all devices equipped with reservoirs delivered a significantly higher oxygen concentration than those without (P less than .001), and that activation of the pop-off valve significantly reduced Fio2 under all conditions (P less than .005). The pop-off valves for each device were activated throughout a wide range of pressures, the Laerdal 41 to 72 cm H2O, PMR-2 51 to 97 cm H2O, and the Ohio Hope II 38 to 106 cm H2O, well in excess of the manufacturer's specifications. Only the Laerdal with reservoir was able to deliver an Fio2 of greater than 0.9 when compressed at more than the pop-off valve pressure at rates of up to 30 breaths per minute using flows of 10 L/min, and it was the only device to produce Fio2 values of greater than 0.9 at all rates to 60 breaths per minute when compressed at less than the pop-off valve pressure.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3456554

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  9 in total

1.  Endotracheal resuscitation of neonates using a rebreathing bag.

Authors:  C J Upton; A D Milner
Journal:  Arch Dis Child       Date:  1991-01       Impact factor: 3.791

2.  The effect of a manometer on the mean airway pressure during hand ventilation, an in vitro study.

Authors:  J Karsdon; T Stijnen; H M Berger
Journal:  Eur J Pediatr       Date:  1989-04       Impact factor: 3.183

3.  Comparison of three manual ventilation devices using an intubated mannequin.

Authors:  S G Hussey; C A Ryan; B P Murphy
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-11       Impact factor: 5.747

Review 4.  Part 13: pediatric basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Marc D Berg; Stephen M Schexnayder; Leon Chameides; Mark Terry; Aaron Donoghue; Robert W Hickey; Robert A Berg; Robert M Sutton; Mary Fran Hazinski
Journal:  Circulation       Date:  2010-11-02       Impact factor: 29.690

5.  Pediatric basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Marc D Berg; Stephen M Schexnayder; Leon Chameides; Mark Terry; Aaron Donoghue; Robert W Hickey; Robert A Berg; Robert M Sutton; Mary Fran Hazinski
Journal:  Pediatrics       Date:  2010-10-18       Impact factor: 7.124

6.  Comparison of efficacy of three devices of manual positive pressure ventilation: a mannequin-based study.

Authors:  Somashekhar M Nimbalkar; Suman Rao P N; Saudamini V Nesargi; Ashish R Dongara; Swarnarekha Bhat
Journal:  Ital J Pediatr       Date:  2015-03-31       Impact factor: 2.638

7.  Short- and intermediate-term outcomes of preterm infants receiving positive pressure ventilation in the delivery room.

Authors:  Megan O'Reilly; Po-Yin Cheung; Khalid Aziz; Georg M Schmölzer
Journal:  Crit Care Res Pract       Date:  2013-01-17

8.  Newborn self-inflating manual resuscitators: precision robotic testing of safety and reliability.

Authors:  Mark B Tracy; Robert Halliday; Sally K Tracy; Murray K Hinder
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2018-10-18       Impact factor: 5.747

9.  Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill Patients: A Randomised, Controlled, Cross-Over Simulation Study.

Authors:  Jong Won Kim; Sang O Park; Kyeong Ryong Lee; Dae Young Hong; Kwang Je Baek
Journal:  Emerg Med Int       Date:  2020-04-24       Impact factor: 1.112

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.