Literature DB >> 34511762

An unusual cause of acute mitral regurgitation in TTK Chitra heart valve prosthesis.

Muthukumar Sundaram1, Aayush Poddar1, Ancy Robinson1, Soundaravalli Balakrishnan2, Muralidharan Srinivasan1, Chandrasekar Padmanabhan1.   

Abstract

Structural failure of mechanical heart valve was a known feature when it was evolving in the 1960s and 1970s. With the advent of pyrolytic carbon and a better design, it is a rare entity with present valves. We report a case of disc fracture leading to acute mitral regurgitation in TTK Chitra heart valve prosthesis (CHVP) (TTK Healthcare Limited, India) heart valve, 6 years after its implantation in mitral position. © Indian Association of Cardiovascular-Thoracic Surgeons 2021.

Entities:  

Keywords:  Disc fracture; Structural valve dysfunction; TTK Chitra heart valve prosthesis

Year:  2021        PMID: 34511762      PMCID: PMC8387517          DOI: 10.1007/s12055-020-01134-5

Source DB:  PubMed          Journal:  Indian J Thorac Cardiovasc Surg        ISSN: 0970-9134


  7 in total

1.  Mechanical heart valves: 50 years of evolution.

Authors:  Vincent L Gott; Diane E Alejo; Duke E Cameron
Journal:  Ann Thorac Surg       Date:  2003-12       Impact factor: 4.330

2.  2017 AHA/ACC Focused Update of the 2014 AHA/ACC Guideline for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines.

Authors:  Rick A Nishimura; Catherine M Otto; Robert O Bonow; Blase A Carabello; John P Erwin; Lee A Fleisher; Hani Jneid; Michael J Mack; Christopher J McLeod; Patrick T O'Gara; Vera H Rigolin; Thoralf M Sundt; Annemarie Thompson
Journal:  J Am Coll Cardiol       Date:  2017-03-15       Impact factor: 24.094

3.  Strut fracture and disc embolization in Björk-Shiley mitral valve prostheses: diagnosis and management.

Authors:  P K Davis; J L Myers; J L Pennock; B L Thiele
Journal:  Ann Thorac Surg       Date:  1985-07       Impact factor: 4.330

4.  Complete Transversal Disc Fracture in a Björk-Shiley Delrin Mitral Valve Prosthesis 43 Years After Implantation.

Authors:  Jose María González-Santos; María Elena Arnáiz-García; María José Dalmau-Sorlí; Jose Alfonso Sastre-Rincón; Jesús Hernández-Hernández; María Elena Pérez-Losada; Víctor Sagredo-Meneses; Javier López-Rodríguez
Journal:  Ann Thorac Surg       Date:  2016-10       Impact factor: 4.330

5.  Performance characteristics of the Starr-Edwards Model 1260 aortic valve prosthesis beyond ten years.

Authors:  D C Miller; P E Oyer; R S Mitchell; E B Stinson; S W Jamieson; J C Baldwin; N E Shumway
Journal:  J Thorac Cardiovasc Surg       Date:  1984-08       Impact factor: 5.209

Review 6.  Valve-related complications after mechanical heart valve implantation.

Authors:  Yoshio Misawa
Journal:  Surg Today       Date:  2014-12-18       Impact factor: 2.549

7.  Spontaneous leaflet fracture resulting in embolization from mechanical valve prostheses.

Authors:  Gijs G J van Steenbergen; Queeny H Y Tsang; Stefan M van der Heide; Michel W A Verkroost; Wilson W L Li; Wim J Morshuis
Journal:  J Card Surg       Date:  2018-12-30       Impact factor: 1.620

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.