| Literature DB >> 34510073 |
Qinjian Cao1, Jun Tan, Yue Ren, Wanchun Xiong, Xiaofeng Wei, Wei Cheng.
Abstract
ABSTRACT: In order to provide safe operating conditions for radiation workers and the public at large, the radiation shielding for radiotherapy treatment rooms will have to be determined by an expert physicist according to the applicable radiation control regulations. A new radiosurgery system with integrated self-shielding, called the Zap-X, obviates the need for costly and complex radiation bunkers. Radiation levels in the vicinity of the ZAP-X radiosurgery system were acquired for a number of different operating conditions, and a 3D radiation dose distribution was measured for better visualization of hot spots and the general dose distribution. The radiation shielding requirements were evaluated according to the International and Chinese standards, respectively. While the integral self-shielding of the Zap-X was designed in accordance with international standards, it was found to be insufficient for Chinese standards. In order to meet the needs of the growing new generation of radiotherapy equipment, several suggestions for improvement of Chinese standards are proposed.Entities:
Mesh:
Year: 2021 PMID: 34510073 PMCID: PMC8505132 DOI: 10.1097/HP.0000000000001465
Source DB: PubMed Journal: Health Phys ISSN: 0017-9078 Impact factor: 1.316
Fig. 1The Zap-X system installed at PLAGH 301.
Fig. 2Scheme of Zap-X system (Weidlich et al. 2017): (a) cross-sectional view, (b) room’s eye-view.
The different gantry positions for dose rate measurement.
| No. | Gantry position | FSa | Phantom |
|---|---|---|---|
| A | Home position (appr. 45 deg off vertical) | 0 mm | no phantom |
| B | Home position | 25 mm | head phantom |
| C | Frontal position (pointing straight down) | 25 mm | head phantom |
| D | Gantry 90° (pointing right) | 25 mm | head phantom |
| E | Gantry 270° (pointing left) | 25 mm | head phantom |
| F | Gantry angle 180° (pointing up) | 25 mm | head phantom |
aFS: Field size, the diameter of collimator.
Fig. 3Scheme of gantry position for Zap-X system.
Fig. 4Measurement points around the perimeter of the system.
Fig. 5Reference station and dose rate meter used for radiation field measurement: (a) reference station, (b) radiation dose meter, (c) computer.
Measured dose rate (μSv h−1) at each station for different situation (A–F).
| No. | Survey stations | A | B | C | D | E | F | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 m | 1.5 m | 1 m | 1.5 m | 1 m | 1.5 m | 1 m | 1.5 m | 1 m | 1.5 m | 1 m | 1.5 m | ||
| 1 | Foot End Table Shield | 3.6 | 10.8 | 6.1 | 25.3 | 2.18 | 5.7 | 3.5 | 14 | 3.5 | 12.4 | 1.42 | 1.8 |
| 2 | Right Side Main Gantry | 1.48 | 1.31 | 1.89 | 1.55 | 1.12 | 1.38 | 1.03 | 1.01 | 2.65 | 2.38 | 0.92 | 1.28 |
| 3 | Left Side Main Gantry | 1.34 | 1.07 | 1.31 | 1.19 | 1.41 | 1.2 | 1.11 | 2.03 | 1.08 | 1.3 | 1.47 | 1.21 |
| 4 | Head End of System | 17.7 | 6.8 | 17.3 | 12.8 | 7.2 | 6.1 | 14.4 | 5.4 | 1.67 | 2.07 | 1.27 | 2.7 |
| 5 | Table Right | 1.65 | 1.65 | 1.7 | 1.83 | 1.32 | 1.94 | 3.6 | 5.6 | 1.4 | 1.38 | 1.85 | 1.46 |
| 6 | Table-Orbit Right | 1.4 | 1.47 | 1.68 | 1.88 | 1.89 | 3.2 | 6.6 | 5.8 | 1.71 | 1.52 | 1.56 | 3.8 |
| 7 | Right Gantry | 0.76 | 1.15 | 0.98 | 1.11 | 2.69 | 4.5 | 5.2 | 4.3 | 2.6 | 1.37 | 1.78 | 6.4 |
| 8 | Right Gantry | 7.2 | 8.4 | 8.2 | 9.4 | 2.26 | 1.68 | 2.03 | 1.57 | 7.9 | 13.4 | 2.21 | 2.5 |
| 9 | Right – Head | 11.9 | 14.8 | 12.2 | 14.5 | 6.2 | 5.6 | 1.87 | 1.53 | 14.3 | 14.2 | 2.11 | 2.5 |
| 10 | Left – Head | 16.4 | 18.1 | 18.3 | 18.5 | 2.41 | 2.39 | 13.8 | 11.6 | 1.15 | 1.16 | 2.8 | 7.3 |
| 11 | Left Gantry | 9.3 | 9.8 | 9.6 | 10.3 | 2.8 | 2.1 | 7.0 | 14.2 | 1.79 | 1.53 | 1.25 | 2.13 |
| 12 | Left Gantry | 0.9 | 1.33 | 1.36 | 1.34 | 10 | 8.9 | 1.19 | 1.27 | 3.4 | 4.1 | 13.7 | 6.7 |
| 13 | Table – Orbit Left | 1.54 | 1.21 | 1.66 | 1.87 | 8.6 | 6.7 | 1.07 | 1.34 | 3.1 | 5.4 | 10.6 | 6.8 |
| 14 | Table Left | 1.23 | 1.55 | 1.8 | 2.06 | 1.92 | 1.81 | 0.88 | 1.38 | 2.82 | 4.5 | 2.81 | 1.67 |
| 15 | Control Console | 0.55 | 0.92 | 0.3 | 1.97 | 1.89 | 0.28 | ||||||
Fig. 6Radiation field distribution of Zap-X system with gantry at home position: (a) 3D radiation field distribution, (b) 2D projection of the radiation field on a plane at the height of isocenter.
Comparison between Chinese standards and American standards.
| Chinese (GBZ 121-2020) | US (NCRP 151) | |
|---|---|---|
| Dose limit for public | Same (1 mSv y−1) | |
| Max. allowable dose rate | Instantaneous dose rate | Time averaged dose rate |
| Occupancy factor | Same (<1/2) | |
| Use factor | 0.25 | 0.20 |
| Duty cycle | Not accounted for | 0.097 |
| Method of calculation | Same (workload based including scaling factors such as | |