Literature DB >> 34498748

In vivo biophysical characterization of very high power, short duration, temperature-controlled lesions.

Giuseppe Stabile1,2,3, Vincenzo Schillaci1, Teresa Strisciuglio1,4, Alberto Arestia1, Alessia Agresta1, Gergana Shopova1, Antonio De Simone2, Francesco Solimene1.   

Abstract

INTRODUCTION: A very high-power short-duration (vHPSD) strategy of radiofrequency (RF) ablation aims to minimize conductive heating and increase resistive heating. This study aimed to clarify the contribution of contact force (CF) and temperature and their interrelationship in making an adequate lesion with the vHPSD catheter.
METHODS: We enrolled 46 consecutive patients undergoing first catheter ablation for atrial fibrillation (AF). The vHPSD ablation was performed applying 90 W, for 4 s, with an irrigation of 8 ml/min. During an application, an impedance drop (ID) ≥10 Ω was regarded as an adequate lesion formation.
RESULTS: The mean procedural time was 95 ± 15 min. First-pass isolation was reached in 89% of patients and at the end of the procedure all pulmonary veins were isolated. No steam pop nor procedural complication occurred. A total of 3829 qualified RF points were analyzed and the median values of ID, CF and maximum temperature were respectively 10.6 (IQR 8.6-13.1) Ohm, 9 (5.8-13.8) g, 46.8 (44.1-49.8) °C. The mean ID significantly increased in parallel with the increasing CF as well as with the increasing maximum temperature. In the multivariable analysis only the CF and the maximum temperature were independent predictors of ID. From receiver operating characteristic curve analysis, a CF of 8 g and a maximum temperature of 47°C are the optimal cutoff discriminatory value for adequate lesion formation.
CONCLUSIONS: The vHPSD ablation is highly effective and safe. The CF and the maximum temperature are independent predictors of adequate lesion formation assessed by means of ID.
© 2021 Wiley Periodicals LLC.

Entities:  

Keywords:  atrial fibrillation; catheter ablation; high power short duration

Mesh:

Year:  2021        PMID: 34498748     DOI: 10.1111/pace.14358

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  In-Silico Modeling to Compare Radiofrequency-Induced Thermal Lesions Created on Myocardium and Thigh Muscle.

Authors:  Juan J Pérez; Enrique Berjano; Ana González-Suárez
Journal:  Bioengineering (Basel)       Date:  2022-07-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.