Literature DB >> 34470740

Use of artificial intelligence for image analysis in breast cancer screening programmes: systematic review of test accuracy.

Karoline Freeman1, Julia Geppert1, Chris Stinton1, Daniel Todkill1, Samantha Johnson1, Aileen Clarke1, Sian Taylor-Phillips2.   

Abstract

OBJECTIVE: To examine the accuracy of artificial intelligence (AI) for the detection of breast cancer in mammography screening practice.
DESIGN: Systematic review of test accuracy studies. DATA SOURCES: Medline, Embase, Web of Science, and Cochrane Database of Systematic Reviews from 1 January 2010 to 17 May 2021. ELIGIBILITY CRITERIA: Studies reporting test accuracy of AI algorithms, alone or in combination with radiologists, to detect cancer in women's digital mammograms in screening practice, or in test sets. Reference standard was biopsy with histology or follow-up (for screen negative women). Outcomes included test accuracy and cancer type detected. STUDY SELECTION AND SYNTHESIS: Two reviewers independently assessed articles for inclusion and assessed the methodological quality of included studies using the QUality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) tool. A single reviewer extracted data, which were checked by a second reviewer. Narrative data synthesis was performed.
RESULTS: Twelve studies totalling 131 822 screened women were included. No prospective studies measuring test accuracy of AI in screening practice were found. Studies were of poor methodological quality. Three retrospective studies compared AI systems with the clinical decisions of the original radiologist, including 79 910 women, of whom 1878 had screen detected cancer or interval cancer within 12 months of screening. Thirty four (94%) of 36 AI systems evaluated in these studies were less accurate than a single radiologist, and all were less accurate than consensus of two or more radiologists. Five smaller studies (1086 women, 520 cancers) at high risk of bias and low generalisability to the clinical context reported that all five evaluated AI systems (as standalone to replace radiologist or as a reader aid) were more accurate than a single radiologist reading a test set in the laboratory. In three studies, AI used for triage screened out 53%, 45%, and 50% of women at low risk but also 10%, 4%, and 0% of cancers detected by radiologists.
CONCLUSIONS: Current evidence for AI does not yet allow judgement of its accuracy in breast cancer screening programmes, and it is unclear where on the clinical pathway AI might be of most benefit. AI systems are not sufficiently specific to replace radiologist double reading in screening programmes. Promising results in smaller studies are not replicated in larger studies. Prospective studies are required to measure the effect of AI in clinical practice. Such studies will require clear stopping rules to ensure that AI does not reduce programme specificity. STUDY REGISTRATION: Protocol registered as PROSPERO CRD42020213590. © Author(s) (or their employer(s)) 2019. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Mesh:

Year:  2021        PMID: 34470740      PMCID: PMC8409323          DOI: 10.1136/bmj.n1872

Source DB:  PubMed          Journal:  BMJ        ISSN: 0959-8138


Introduction

Breast cancer is a leading cause of death among women worldwide. Approximately 2.4 million women were diagnosed with breast cancer in 2015, and 523 000 women died.1 Breast cancer is more amenable to treatment when detected early,2 so many countries have introduced screening programmes. Breast cancer screening requires one or two radiologists to examine women’s mammograms for signs of presymptomatic cancer, with the aim of reducing breast cancer related morbidity and mortality. Such screening is also associated with harms, such as overdiagnosis and overtreatment of cancer that would not have become symptomatic within the woman’s lifetime. Disagreement exists about the extent of overdiagnosis, from 1% to 54% of screen detected cancers, and about the balance of benefits and harms of screening.2 The spectrum of disease detected at screening is associated with outcomes. For example, detection of low grade ductal carcinoma in situ is more associated with overdiagnosis,3 4 whereas detection of grade 3 cancer is more likely to be associated with fewer deaths.5 Cancer is detected in between 0.6% and 0.8% of women during screening.6 7 Breast screening programmes also miss between 15% and 35% of cancers owing either to error or because the cancer is not visible or perceptible to the radiologist. Some of these missed cancers present symptomatically as interval cancers.8 Considerable interest has been shown in the use of artificial intelligence (AI) either to complement the work of humans or to replace them. In 2019, 3.8% of all peer reviewed scientific publications worldwide on Scopus related to AI.9 Claims have been made that image recognition using AI for breast screening is better than experienced radiologists and will deal with some of the limitations of current programmes.10 11 12 13 For instance, fewer cancers might be missed because an AI algorithm is unaffected by fatigue or subjective diagnosis,14 15 and AI might reduce workload or replace radiologists completely.11 12 AI might, however, also exacerbate harm from screening. For example, AI might alter the spectrum of disease detected at breast screening if it differentially detects more microcalcifications, which are associated with lower grade ductal carcinoma in situ. In such a case, AI might increase rates of overdiagnosis and overtreatment and alter the balance of benefits and harms. Autopsy studies suggest that around 4% of women die with, not because of, breast cancer,16 so there is a “reservoir” of clinically unimportant disease, including incidental in situ carcinoma, which might be detected by AI. The spectrum of disease is correlated with mammographic features (for example, ductal carcinoma in situ is often associated with microcalcifications). Therefore, the cases on which AI systems were trained, and the structures within the AI system, might considerably affect the spectrum of disease detected. These structures and algorithms within an AI system are not always transparent or explicable, making interpretation a potential problem. Unlike human interpretation, how or why an algorithm has made a decision can be difficult to understand (known as the “black box” problem).17 Unlike human decision makers, algorithms do not understand the context, mode of collection, or meaning of viewed images, which can lead to the problem of “shortcut” learning,18 whereby deep neural networks reach a conclusion to a problem through a shortcut, rather than the intended solution. Thus, for example, DeGrave et al19 have shown how some deep learning systems detect covid-19 by means of confounding factors, rather than pathology, leading to poor generalisability. Although this problem does not preclude the use of deep learning, it highlights the importance of avoiding potential confounders in training data, an understanding of algorithm decision making, and the critical role of rigorous evaluation. This review was commissioned by the UK National Screening Committee to determine whether there is sufficient evidence to use AI for mammographic image analysis in breast screening practice. Our aim was to assess the accuracy of AI to detect breast cancer when integrated into breast screening programmes, with a focus on the cancer type detected.

Methods

Data sources

Our systematic review was reported in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses of diagnostic test accuracy (PRISMA-DTA) statement.20 The review protocol is registered on PROSPERO (international prospective register of systematic reviews). We conducted literature searches for studies published in English between 1 January 2010 and 9 September 2020 and updated our searches on 17 May 2021. The search comprised four themes: breast cancer, artificial intelligence, mammography, and test accuracy or randomised controlled trials. A number of additional synonyms were identified for each theme. Databases searched were Medline (Ovid), Embase (Ovid); Web of Science, and the Cochrane Database of Systematic Reviews (CENTRAL). Details of the search strategies are shown in supplementary appendix 1. We screened the reference lists of systematic reviews and included additional relevant studies and contacted experts in the field.

Study selection

Two reviewers independently reviewed the titles and abstracts of all retrieved records against the inclusion criteria, and subsequently, all full text publications. Disagreements were resolved by consensus or discussion with a third reviewer. We applied strict inclusion/exclusion criteria to focus on the evaluation of the integration of AI into a breast cancer screening programme rather than the development of AI systems. Studies were eligible for inclusion if they reported test accuracy of AI algorithms applied to women’s digital mammograms to detect breast cancer, as part of a pathway change or a complete read (reading+decision resulting in classification). Eligible study designs were prospective test accuracy studies, randomised controlled trials, retrospective test accuracy studies using geographical validation only, comparative cohort studies, and enriched test set multiple reader multiple case laboratory studies. The enriched test set multiple reader multiple case laboratory studies included retrospective data collection of images and prospective classification by standalone AI or AI assisted radiologists. The reference standard was cancer confirmed by histological analysis of biopsy samples from women referred for further tests at screening and preferably also from symptomatic presentation during follow-up. All studies will necessarily have differential verification because not all women can or should be biopsied. In prospective test accuracy studies this will not introduce significant bias because those positive on either an index or comparator test will receive follow-up tests. In retrospective studies and enriched test set studies (with prospective readers), the decision as to whether women receive biopsy or follow-up is based on the decision of the original reader, which introduces bias because cancer, when present, is more likely to be found if the person receives follow-up tests after recall from screening. We assessed this using the QUality Assessment of Diagnostic Accuracy Studies-2 tool (QUADAS-2). When AI is used as a pre-screen to triage which mammograms need to be examined by a radiologist and which do not, we also accepted a definition of a normal mammogram as one free of screen detected cancer based on human consensus reading, as this allows estimation of accuracy in the triage. We excluded studies that reported the validation of AI systems using internal validation test sets (eg, x-fold cross validation, leave one out method), split validation test sets, and temporal validation test sets as they are prone to overfitting and insufficient to assess the generalisability of the AI system. Furthermore, studies were excluded if less than 90% of included mammograms were complete full field digital mammography screening mammograms. Additionally, studies were excluded if the AI system was used to predict future risk of cancer, if only detection of cancer subtypes was reported, if traditional computer aided detection systems without machine learning were used, or if test accuracy measures were not expressed at any clinically relevant threshold (eg, area under the curve only) or did not characterise the trade-off between false positives and false negative results (eg, sensitivity for cancer positive samples only). Finally, we excluded simulation results of the hypothetical integration of AI with radiologists’ decisions as they do not reliably estimate radiologist behaviour when AI is applied.

Data extraction and quality assessment

One reviewer extracted data on a predesigned data collection form. Data extraction sheets were checked by a second reviewer and any disagreements were resolved by discussion. Study quality was assessed independently by two reviewers using QUADAS-221 tailored to the review question (supplementary appendix 2).

Data analysis

The unit of analysis was the woman. Data were analysed according to where in the pathway AI was used (for example, standalone AI to replace one or all readers, or reader aid to support decision making by a human reader) and by outcome. The primary outcome was test accuracy. If test accuracy was not reported, we calculated measures of test accuracy where possible. Important secondary outcomes were cancer type and interval cancers. Cancer type (eg, by grade, stage, size, prognosis, nodal involvement) is important in order to estimate the effect of cancer detection on the benefits and harms of screening. Interval cancers are also important because they have worse average prognosis than screen detected cancers,22 and by definition, are not associated with overdiagnosis at screening. We synthesised studies narratively owing to their small number and extensive heterogeneity. We plotted reported sensitivity and specificity for the AI systems and any comparators in a receiver operating characteristic plot using the package “ggplot2”23 in R version 3.6.1 (Vienna, Austria).24

Patient and public involvement

The review was commissioned on behalf of the UK National Screening Committee (UKNSC), and the scope was determined by the UKNSC adult reference group, which includes lay members. The results were discussed with patient contributors.

Results

Database searches yielded 4016 unique results, of which 464 potentially eligible full texts were assessed. Four additional articles were identified: one through screening the reference lists of relevant systematic reviews, one through contact with experts, and two by hand searches. Overall, 13 articles25 26 27 28 29 30 31 32 33 34 35 36 37 reporting 12 studies were included in this review (see supplementary fig 1 for full PRISMA flow diagram). Exclusions on full text are listed in supplementary appendix 3.

Characteristics of included studies

The characteristics of the 12 included studies are presented in table 1, table 2, and table 3 and in supplementary appendix 4, comprising a total of 131 822 screened women. The AI systems in all included studies used deep learning convolutional neural networks. Four studies evaluated datasets from Sweden,26 27 35 36 three of which had largely overlapping populations,26 35 36 one from the United States and Germany,32 one from Germany,25 one from the Netherlands,33 one from Spain31 and four from the US.28 29 30 37 Four studies enrolled women consecutively or randomly,25 27 31 36 while the remaining studies selected cases and controls to enrich the dataset with patients with cancer. Three studies included all patients with cancer and a random sample of those without cancer.26 29 35 One study included all patients with cancer and controls matched by age and breast density.28 In two studies, patients and controls were sampled to meet predefined distributions and were reviewed by one radiologist to exclude images not meeting quality standards and images with obvious signs of cancer.30 32 One study used a range of rules for selection, including by perceived difficulty and mammographic features.33 Finally, one study included only false negative mammograms.37 No prospective test accuracy studies in clinical practice were included, only retrospective test accuracy studies25 26 27 29 31 35 36 and enriched test set multiple reader multiple case laboratory studies.28 30 32 33 37 Of these enriched test set laboratory studies, three reported test accuracy for a single AI read as a reader aid.30 32 37 Another nine studies reported test accuracy for a single AI read as a standalone system in a retrospective test accuracy study 25 26 27 29 31 35 36 or an enriched test set multiple reader multiple case laboratory study.28 33
Table 1

Summary of study characteristics for studies using AI as standalone system

StudyStudy designPopulationMammography vendorIndex testComparatorReference standard
Lotter 202128 Enriched test set MRMC laboratory study (accuracy of a read)285 women from 1 US health system with 4 centres (46.0% screen detected cancer); age and ethnic origin NRHologic 100%In-house AI system (DeepHealth); threshold NR (set to match readers’ sensitivity and specificity, respectively)5 MQSA certified radiologists (US), single reading; threshold of BI-RADS scores 3, 4, and 5 considered recallCancer: pathology confirmed cancer within 3 months of screening; confirmed negative: a negative examination followed by an additional BI-RADS score 1 or 2 interpretation at the next screening examination 9-39 months later
McKinney 202029 Retrospective test accuracy study (accuracy of a read)3097 women from 1 US centre (22.2% cancer within 27 months of screening); age <40, 181 (5.8%); 40-49, 1259 (40.7%); 50-59, 800 (25.8%); 60-69, 598 (19.3%); ≥70, 259 (8.4%)Hologic / Lorad branded: >99%;Siemens or GeneralElectric: <1%In-house AI system (Google Health); threshold: to achieve superiority for both sensitivity and specificity compared with original single reading using validation setOriginal single radiologist decision (US); threshold: BI-RADS scores 0, 4, 5 were treated as positiveCancer: biopsy confirmed cancer within 27 months of imaging; non-cancer: one follow-up non-cancer screen or biopsied negative (benign pathologies) after ≥21 months
Rodriguez-Ruiz 201933 Enriched test set MRMC laboratory study (accuracy of a read)199 examinations from a Dutch digital screening pilot project (39.7% cancer);age range 50-74Hologic 100%Transpara version 1.4.0 (Screenpoint Medical BV, Nijmegen, Netherlands);threshold: 8.26/10, corresponding to the average radiologist’s specificityNine Dutch radiologists, single reading, as part of a previously completed MRMC study38; no thresholdCancer: histopathology-proven cancer; non-cancer: ≥1 normal follow-up screening examination (2 year screening interval)
Salim202035 Retrospective test accuracy study (accuracy of a read)8805 women from a Swedish cohort study (8.4% cancer within 12 months of screening); median age 54.5 (IQR 47.4-63.5)Hologic 100%3 commercial AI systems (anonymised: AI-1, AI-2, and AI-3); threshold: corresponding to the specificity of the first readerOriginal radiologist decision (Sweden); (1) single reader (R1; R2), (2) consensus reading; no thresholdCancer: pathology confirmed cancer within 12 months of screening; non-cancer: ≥2 years cancer free follow-up
Schaffter 202036 Retrospective test accuracy study (accuracy of a read)68 008 consecutive women from 1 Swedish centre (1.1% cancer within 12 months of screening) mean age 53.3 (SD 9.4)NR4 in-house AI systems: 1 top performing model submitted to the DREAM challenge, 1 ensemble method of the eight best performing models (CEM), CEM combined with reader decision (single reader or consensus reading); threshold: corresponding to the sensitivity of single and consensus reading, respectivelyOriginal radiologist decision (Sweden); (1) single reader (R1; R2), (2) consensus reading; no thresholdCancer: tissue diagnosis within 12 months of screening; non-cancer: no cancer diagnosis ≥12 months after screening

AI=artificial intelligence; BI-RADS=breast imaging reporting and data system; CEM=challenge ensemble method; DREAM=Dialogue on Reverse Engineering Assessment and Methods; IQR=interquartile range; MQSA=Mammography Quality Standards Act; MRMC=multiple reader multiple case; NR=not reported; R1=first reader; R2=second reader; SD=standard deviation.

Table 2

Summary of study characteristics for studies using AI for triage

StudyStudy designPopulationMammography vendorIndex testComparatorReference standard
Balta 202025 Retrospective cohort study (accuracy of classifying into low and high risk categories)17 895 consecutively acquired screening examinations from 1 centre in Germany (0.64% screen detected cancer), age NRSiemens 70% Hologic 30%Transpara version 1.6.0 (Screenpoint Medical BV, Nijmegen, Netherlands); preselection of probably normal mammograms;Transpara risk score of 1-10, different cutoff points evaluated; optimal cutoff point ≤7: low riskNo comparator as human consensus reading decisions used as reference standard for screen negative resultsCancer: biopsy proven screen detected cancers; non-cancer: no information about follow-up for the normal examinations was available.available; for this review, a normal mammogram was defined as free of screen detected cancer based on human consensus reading
Dembrower 202026 Retrospective case-control study (accuracy of classifying into low and high risk categories)7364 women with screening examinations obtained during 2 consecutive screening rounds in 1 centre in Sweden (7.4% cancer: 347 screen detected in current round, 200 interval cancers within 30 months of previous screening round), median age 53.6 (IQR 47.6-63.0)Hologic 100%Lunit (Seoul, South Korea, version 5.5.0.16)(1) AI for preselection of mammograms probably normal, (2) AI as post-screen after negative double reading to recall women at highest risk of undetected cancer; AI risk score: decimal between 0 and 1, different cutoff points evaluatedNoneCancer: diagnosed with breast cancer at current screening round or within ≤30 months of previous screening round; non-cancer: >2 years’ follow up
Lång 202127 Retrospective cohort study (accuracy of classifying into low and high risk categories)9581 women attending screening at 1 centre in Sweden, consecutive subcohort of Malmö Breast Tomosynthesis Screening Trial (0.71% screen detected cancers), mean age 57.6 (range 40-74)Siemens 100%Transpara version 1.4.0 (Screenpoint Medical BV, Nijmegen, Netherlands); preselection of mammograms probably normal; Transpara risk score of 1–10, different cutoff points evaluated;chosen cutoff point ≤5: low riskNo comparator as human consensus reading decisions used as reference standard for screen negative resultsCancer: histology of surgical specimen or core needle biopsies with a cross reference to a regional cancer register;non-cancer: a normal mammogram was defined as free of screen detected cancer based on human consensus reading
Raya-Povedano 202131 Retrospective cohort study (accuracy of classifying into low and high risk categories)15 986 consecutive women from the Córdoba Tomosynthesis Screening Trial, 1 Spanish centre (0.7% cancer: 98 screen detected (FFDM or DBT), 15 interval cancers within 24 months of screening); mean age 58 (SD 6), range 50-69 yearsHologic (Selenia Dimensions) 100%Transpara, version 1.6.0 (ScreenPoint Medical BV, Nijmegen, Netherlands);preselection of mammograms probably normal; Transpara risk score of 1–10; cutoff point ≤7 low risk (chosen based on previous research by Balta 202035)Original radiologist decision from Córdoba Tomosynthesis Screening Trial (double reading without consensus or arbitration)Cancer: histopathologic results of biopsy, screen detected via FFDM or DBT and interval cancers within 24 months of screening; non-cancer: normal reading with 2-years’ follow-up

AI=artificial intelligence; DBT=digital breast tomosynthesis; FFDM=full field digital mammography; IQR=interquartile range; NR=not reported; SD=standard deviation.

Table 3

Summary of study characteristics for studies using AI as reader aid

StudyStudy designPopulationMammography vendorIndex testComparatorReference standard
Pacilè 202030 Enriched test set MRMC laboratory study,counterbalance design (accuracy of a read)240 women from 1 US centre (50.0% cancer), mean age 59 (range 37-85)NR14 MQSA certified radiologists (US) with AI support (MammoScreen version 1, Therapixel, Nice, France); threshold: level of suspicion (0-100) >4014 MQSA certified radiologists (US) without AI support, single reading; threshold: level of suspicion (0-100) >40Cancer: histopathology;non-cancer: negative biopsy or negative result at follow-up for ≥18 months
Rodriguez-Ruiz 201932 Enriched test set MRMC laboratory study, fully crossed (accuracy of a read)240 women (120 from 1 US centre and 120 from 1 German centre; 41.7% cancer), median age 62 (range 39-89)Hologic 50%Siemens 50%14 MQSA certified radiologists (US) with AI support (Transpara version 1.3.0, Screenpoint Medical BV, Nijmegen, the Netherlands); threshold: BI-RADS score ≥314 MQSA certified radiologists (US) without AI support, single reading;threshold: BI-RADS score ≥3Cancer: histopathology confirmed cancer; false positives: histopathologic evaluation or negative follow-up for ≥1 year;non-cancer: ≥1 year of negative follow-up
Watanabe 201937 Enriched test set MRMC laboratory study, first without AI support, then AI aided (accuracy of a read)122 women from 1 US centre (73.8% cancer, all false negative mammograms), mean age 65.4 (range 40-90)NR7 MQSA certified radiologists (US) with AI support (cmAssist, CureMetrix, Inc., La Jolla, CA); no threshold7 MQSA certified radiologists (US) without AI support, single reading; no thresholdCancer: biopsy proven cancer; non-cancer: BI-RADS 1 and 2 women with a 2 year follow-up of negative diagnosis

AI=artificial intelligence; BI-RADS=Breast Imaging-Reporting and Data System; MQSA=Mammography Quality Standards Act; MRMC=multireader multicase; NR=not reported.

Summary of study characteristics for studies using AI as standalone system AI=artificial intelligence; BI-RADS=breast imaging reporting and data system; CEM=challenge ensemble method; DREAM=Dialogue on Reverse Engineering Assessment and Methods; IQR=interquartile range; MQSA=Mammography Quality Standards Act; MRMC=multiple reader multiple case; NR=not reported; R1=first reader; R2=second reader; SD=standard deviation. Summary of study characteristics for studies using AI for triage AI=artificial intelligence; DBT=digital breast tomosynthesis; FFDM=full field digital mammography; IQR=interquartile range; NR=not reported; SD=standard deviation. Summary of study characteristics for studies using AI as reader aid AI=artificial intelligence; BI-RADS=Breast Imaging-Reporting and Data System; MQSA=Mammography Quality Standards Act; MRMC=multireader multicase; NR=not reported. In studies of standalone systems, the AI algorithms provided a cancer risk score that can be turned into a binary operating point to classify women as high risk (recall) or low risk (no recall). The in-house or commercial standalone AI systems (table 1, table 2, table 3) were evaluated in five studies as a replacement for one or all radiologists. Three studies compared the performance of the AI system with the original decision recorded in the database, based on either a single US radiologist29 or two radiologists with consensus within the Swedish screening programme.35 36 Two studies compared the performance of the AI system with the average performance of nine Dutch single radiologists33 and five US single radiologists,28 respectively, who read the images under laboratory conditions. Four commercial AI systems were evaluated as a pre-screen to remove normal cases25 26 27 31 or were used as a post-screen of negative mammograms after double reading to predict interval and next round screen detected cancers.26 In studies of assistive AI, the commercial AI systems provided the radiologist with a level of suspicion for the area clicked. All three studies compared the test accuracy of the AI assisted read with an unassisted read by the same radiologists under laboratory conditions.30 32 37 The experience of the radiologists in the reader assisted studies ranged from 3 to 25 years (median 9.5 years) in 14 radiologists,32 from 0 to 25 years (median 8.5 years) in 14 American Board of Radiology and Mammography Quality Standards Act (MQSA) certified radiologists,30 and from less than 5 to 42 years in 7 MQSA certified radiologists.37 The role of the AI system in the screening pathway in the 12 studies is summarised in figure 1.
Fig 1

Overview of published evidence in relation to proposed role in screening pathway. Purple shade=current pathway; orange shade=AI added to pathway; green shade=level of evidence for proposed AI role. AI=artificial intelligence; +/−=high/low risk of breast cancer, person icon=radiologist reading of mammograms as single, first, or second reader; MRMC=multiple reader multiple case; R1, R2=reader 1, reader 2; RCT, randomised controlled trial; sens=sensitivity; spec=specificity

Overview of published evidence in relation to proposed role in screening pathway. Purple shade=current pathway; orange shade=AI added to pathway; green shade=level of evidence for proposed AI role. AI=artificial intelligence; +/−=high/low risk of breast cancer, person icon=radiologist reading of mammograms as single, first, or second reader; MRMC=multiple reader multiple case; R1, R2=reader 1, reader 2; RCT, randomised controlled trial; sens=sensitivity; spec=specificity

Assessment of risk of bias and applicability

The evidence for the accuracy of AI to detect breast cancer was of low quality and applicability across all studies (fig 2) according to QUADAS-2 (supplementary appendix 2). Only four studies (albeit the four largest comprising 85% of all 131 822 women in the review) enrolled women consecutively or randomly, with a cancer prevalence of between 0.64% and 1.1%.25 27 31 36 The remaining studies used enrichment leading to breast cancer prevalence (ranging from 7.4%26 to 73.8%37), which is atypical of screening populations. Five studies28 30 32 33 37 used reading under “laboratory” conditions at risk of introducing bias because radiologists read mammograms differently in a retrospective laboratory experiment than in clinical practice.39 Only one of the studies used a prespecified test threshold which was internal to the AI system to classify mammographic images.31
Fig 2

Overview of concerns about risk of bias and applicability of included studies. *Low concerns about applicability for consensus reading; high concerns about applicability for single reading as comparator test. †Low concerns about risk of bias and low applicability for the previous screening round (biopsy proven cancer or at least two years’ follow-up); high concerns about risk of bias and high applicability for the current screening round (biopsy-proven cancer but no follow-up of test negatives)

Overview of concerns about risk of bias and applicability of included studies. *Low concerns about applicability for consensus reading; high concerns about applicability for single reading as comparator test. †Low concerns about risk of bias and low applicability for the previous screening round (biopsy proven cancer or at least two years’ follow-up); high concerns about risk of bias and high applicability for the current screening round (biopsy-proven cancer but no follow-up of test negatives) The reference standard was at high (n=8) or unclear (n=3) risk of bias in 11/12 studies. Follow-up of screen negative women was less than two years in seven studies,25 26 27 28 30 32 36 which might have resulted in underestimation of the number of missed cancers and overestimation of test accuracy. Furthermore, in retrospective studies of routine data the choice of patient management (biopsy or follow-up) to confirm disease status was based on the decision of the original radiologist(s) but not on the decision of the AI system. Women classified as positive by AI who did not receive biopsy based on the original radiologists’ decision only, received follow-up to confirm disease status. Therefore, cancers with a lead time from screen to symptomatic detection longer than the follow-up time in these studies will be misclassified as false positives for the AI test, and cancers which would have been overdiagnosed and overtreated after detection by AI would not be identified as such because the type of cancer that can indicate overdiagnosis, is unknown. The direction and magnitude of bias is complex and dependent on the positive and negative concordance between AI and radiologists but is more likely to be in the direction of overestimation of sensitivity and underestimation of specificity. The applicability to European or UK breast cancer screening programmes was low (fig 2). None of the studies described the accuracy of AI integrated into a clinical breast screening pathway or evaluated the accuracy of AI prospectively in clinical practice in any country. Only two studies compared AI performance with the decision from human consensus reading.35 36 The studies included only interval cancers within 12 months of screening, which is not typical for screening programmes. No direct evidence is therefore available as to how AI might affect accuracy if integrated into breast screening practice.

Analysis

AI as a standalone system to replace radiologist(s)

No prospective test accuracy studies, randomised controlled trials, or cohort studies examined AI as a standalone system to replace radiologists. Test accuracy of the standalone AI systems and the human comparators from retrospective cohort studies is summarised in table 4. All point estimates of the accuracy of AI systems were inferior to those obtained by consensus of two radiologists in screening practice, with mixed results in comparison with a single radiologist (fig 3). Three studies compared AI accuracy with that of the original radiologist in clinical practice,29 35 36 of which two were enriched with extra patients with cancer.
Table 4

Summary of test accuracy outcomes

StudyIndex test (manufacturer)/comparatorTPFPFNTN% Sensitivity(95% CI)Δ % Sensitivity,P value or (95% CI)% Specificity(95% CI)Δ % Specificity, value or (95% CI)
Standalone AI (5 studies):
 Lotter 2021,28 Index cancerAI (in-house) at reader’s specificity12651510396.2 (91.7 to 99.2)+14.2, P<0.00166.9Set to be equal
AI (in-house) at reader’s sensitivity107142414082.0Set to be equal90.9 (84.9 to 96.1)+24.0, P<0.001
Comparator: average single reader†NANANANA82.066.9
 McKinney 202029*AI (in-house)NRNRNRNR56.24+8.1, P<0.00184.29+3.46, P=0.02
Comparator: original single readerNRNRNRNR48.180.83
 Rodriguez-Ruiz 201933 AI (Transpara version 1.4.0)6325169580 (70 to 90)+3 (-6.2 to 12.6)79 (73 to 86)Set to be equal
Comparator: average single reader§NANANANA77 (70 to 83)79 (73 to 86)
 Salim 202035AI-1 (anonymised)605NRNRNR81.9 (78.9 to 84.6)See below96.6 (96.5 to 96.7)Set to be equal
AI-2 (anonymised)495NRNRNR67.0 (63.5 to 70.4)−14.9 v AI-1 (P<0.001)96.6 (96.5 to 96.7)Set to be equal
AI-3 (anonymised)498NRNRNR67.4 (63.9 to 70.8)−14.5 v AI-1 (P<0.001)96.7 (96.6 to 96.8)Set to be equal
Comparator: original reader 1572NRNRNR77.4 (74.2 to 80.4)−4.5 v AI-1 (P=0.03)96.6 (96.5 to 96.7)
Comparator: original reader 2592NRNRNR80.1 (77.0 to 82.9)−1.8 v AI-1 (P=0.40)97.2 (97.1 to 97.3)+0.6 v AI-1 (NR)
Comparator: original consensus reading628NRNRNR85.0 (82.2 to 87.5)+3.1 v AI-1 (P=0.11)98.5 (98.4 to 98.6)+1.9 v AI-1 (NR)
 Schaffter 202036Top-performing AI (in-house)NRNRNRNR77.1Set to be equal88−8.7 v reader 1 (NR)
Ensemble method (CEM; in-house)NRNRNRNR77.1Set to be equal92.5−4.2 v reader 1 (NR)
Comparator: original reader 1NRNRNRNR77.196.7 (96.6 to 96.8)
 Schaffter 202036 Top-performing AI (in-house)NRNRNRNR83.9Set to be equal81.2−17.3 v consensus (NR)
Comparator: original consensus readingNRNRNRNR83.998.5
AI for triage pre-screen (4 studies):
 Balta 202025 AI as pre-screen (Transpara version 1.6.0):
 AI score ≤2: ~15% low risk11415 02802754100.0NA15.49NA
 AI score ≤5: ~45% low risk10997915799195.61NA44.94NA
 AI score ≤7: ~65% low risk1056135911 64792.11NA65.50NA
 Lång 202027 AI as pre-screen (Transpara version 1.4.0):
 AI score ≤2: ~19% low risk68768401829100.0NA19.23NA
 AI score ≤5: ~53% low risk6144387507589.71NA53.35NA
 AI score ≤7: ~73% low risk57254111697283.82NA73.29NA
 Raya-Povedano 202131 AI as pre-screen (Transpara version 1.6.0); AI score ≤7: ~72% low risk10044501311 42488.5 (81.1 to 93.7)NA72.0 (71.3 to 72.7)NA
 Dembrower 202026§AI as pre-screen (Lunit version 5.5.0.16):
 AI score ≤0.0293: 60% low risk¶34729 787045 200100.0NA60.28NA
 AI score ≤0.0870: 80% low risk¶33814 729960 25897.41NA80.36NA
AI for triage post-screen (1 study):
 Dembrower 202026§AI as post-screen (Lunit v5.5.0.16);prediction of interval cancers:AI score ≥0.5337: ~2% high risk32141316873 92116NA98.12NA
 Dembrower 202026§AI as post-screen (Lunit version 5.5.0.16); prediction of interval and next round screen detected cancers:AI score ≥0.5337: ~2% high risk103134244473 64519NA98.21NA
AI as reader aid (3 studies):
 Pacilè 202030 AI support§ (MammoScreen version 1)NANANANA69.1 (60.0 to 78.2)+3.3, P=0.0273.5 (65.6 to 81.5)+1.0, P=0.63
Comparator: average single reader**NANANANA65.8 (57.4 to 74.3)72.5 (65.6 to 79.4)
 Rodriguez-Ruiz 201932 AI support (Transpara version 1.3.0)86291411186 (84 to 88)+3, P=0.0579 (77 to 81)+2, P=0.06
Comparator: average single reader83321710883 (81 to 85)77 (75 to 79)
 Watanabe 201937 AI support** (cmAssist)NANANANA62 (range 41 to 75)+11, P=0.0377.2−0.9 (NR)
Comparator: average single reader**NANANANA51 (range 25 to 71)78.1

AI=artificial intelligence; CEM=challenge ensemble method of eight top performing AIs from DREAM challenge; CI=confidence interval; DREAM=Dialogue on Reverse Engineering Assessment and Methods; FN=false negatives; F=false positives; NA=not applicable; NR=not reported; TN=true negatives; TP=true positives.

Inverse probability weighting: negative cases were upweighted to account for the spectrum enrichment of the study population. Patients associated with negative biopsies were downweighted by 0.64. Patients who were not biopsied were upweighted by 23.61.

Applied an inverse probability weighted bootstrapping (1000 samples) with a 14:1 ratio of healthy women to women receiving a diagnosis of cancer to simulate a study population with a cancer prevalence matching a screening cohort.

In addition, the challenge ensemble method prediction was combined with the original radiologist assessment. At the first reader’s sensitivity of 77.1%, CEM+reader 1 resulted in a specificity of 98.5% (95% confidence interval 98.4% to 98.6%), higher than the specificity of the first reader alone of 96.7% (95% confidence interval, 96.6% to 96.8%; P<0.001). At the consensus readers’ sensitivity of 83.9%, CEM+consensus did not significantly improve the consensus interpretations alone (98.1% v 98.5% specificity, respectively). These simulated results of the hypothetical integration of AI with radiologists’ decisions were excluded as they did not incorporate radiologist behaviour when AI is applied.

Applied 11 times upsampling of the 6817 healthy women, resulting in 74 987 healthy women and a total simulated screening population of 75 534.

Specificity estimates not based on exact numbers; the numbers were calculated by reviewers from reported proportions applied to 75 334 women (347 screen detected cancers and 74 987 healthy women).

In enriched test set multiple reader multiple case laboratory studies where multiple readers asses the same images, there are considerable problems in summing 2x2 test data across readers.

Fig 3

Study estimates of sensitivity and false positive rate (1−specificity) in receiver operating characteristic space by index test (artificial intelligence) and comparator (radiologist) for eight included studies. Comparators are defined as consensus of two readers and arbitration (radiologist consensus), or single reader decision/average of multiple readers (radiologist single/average). iVertical dashed lines represent specificity for screening programmes for Denmark (2% false positive rate),61 UK (3% false positive rate),62 63 and US (11% false positive rate).64 Retrospective test accuracy studies: Salim et al,35 Schaffter et al,36 and McKinney et al.29 Enriched test set multiple reader multiple case laboratory studies: Pacilè et al,30 Watanabe et al,37 Rodriguez-Ruiz et al33 (Rodriguez-Ruiz 2019a in figure), Lotter 2021,28 and Rodriguez-Ruiz et al32 (Rodriguez-Ruiz 2019b in figure)

Summary of test accuracy outcomes AI=artificial intelligence; CEM=challenge ensemble method of eight top performing AIs from DREAM challenge; CI=confidence interval; DREAM=Dialogue on Reverse Engineering Assessment and Methods; FN=false negatives; F=false positives; NA=not applicable; NR=not reported; TN=true negatives; TP=true positives. Inverse probability weighting: negative cases were upweighted to account for the spectrum enrichment of the study population. Patients associated with negative biopsies were downweighted by 0.64. Patients who were not biopsied were upweighted by 23.61. Applied an inverse probability weighted bootstrapping (1000 samples) with a 14:1 ratio of healthy women to women receiving a diagnosis of cancer to simulate a study population with a cancer prevalence matching a screening cohort. In addition, the challenge ensemble method prediction was combined with the original radiologist assessment. At the first reader’s sensitivity of 77.1%, CEM+reader 1 resulted in a specificity of 98.5% (95% confidence interval 98.4% to 98.6%), higher than the specificity of the first reader alone of 96.7% (95% confidence interval, 96.6% to 96.8%; P<0.001). At the consensus readers’ sensitivity of 83.9%, CEM+consensus did not significantly improve the consensus interpretations alone (98.1% v 98.5% specificity, respectively). These simulated results of the hypothetical integration of AI with radiologists’ decisions were excluded as they did not incorporate radiologist behaviour when AI is applied. Applied 11 times upsampling of the 6817 healthy women, resulting in 74 987 healthy women and a total simulated screening population of 75 534. Specificity estimates not based on exact numbers; the numbers were calculated by reviewers from reported proportions applied to 75 334 women (347 screen detected cancers and 74 987 healthy women). In enriched test set multiple reader multiple case laboratory studies where multiple readers asses the same images, there are considerable problems in summing 2x2 test data across readers. Study estimates of sensitivity and false positive rate (1−specificity) in receiver operating characteristic space by index test (artificial intelligence) and comparator (radiologist) for eight included studies. Comparators are defined as consensus of two readers and arbitration (radiologist consensus), or single reader decision/average of multiple readers (radiologist single/average). iVertical dashed lines represent specificity for screening programmes for Denmark (2% false positive rate),61 UK (3% false positive rate),62 63 and US (11% false positive rate).64 Retrospective test accuracy studies: Salim et al,35 Schaffter et al,36 and McKinney et al.29 Enriched test set multiple reader multiple case laboratory studies: Pacilè et al,30 Watanabe et al,37 Rodriguez-Ruiz et al33 (Rodriguez-Ruiz 2019a in figure), Lotter 2021,28 and Rodriguez-Ruiz et al32 (Rodriguez-Ruiz 2019b in figure) The DREAM challenge of 68 008 consecutive women from the Swedish screening programme found the specificity of the top performing AI system (by Therapixel in a competition between 31 AI systems evaluated in the competitive phase on the independent Swedish dataset) was inferior in comparison with the original first radiologist (88% v 96.7%) and inferior also in comparison with the original consensus decision (81% v 98.5%) when the AI threshold was set to match the first reader’s sensitivity and the consensus of readers’ sensitivity, respectively.36 The specificity of an ensemble method of the eight top performing AI systems remained inferior to that of the original first radiologist (92.5% v 96.7%, P<0.001), even in the same dataset that was used to choose the top eight. An enriched Swedish cohort study (which overlapped that of the DREAM challenge, n=8805, 8.4% cancer) used three commercially available AI systems with thresholds set to match the specificity of the original radiologists. The study found that one commercially available AI system had superior sensitivity (81.9%, P=0.03) and two had inferior sensitivity (67%, 67.4%) in comparison with the original first radiologist (77.4%).35 All had inferior sensitivity in comparison with the original consensus decision (85%, P=0.11 for best AI system v consensus). The manufacturer and identity were not reported for any of the three AI systems. An enriched retrospective cohort from the US (n=3097, 22.2% cancer) found the AI system outperformed the original single radiologist in sensitivity (56% v 48%, P<0.001) and specificity (84% v 81%, P=0.021), although absolute values for the radiologist were lower than those found in clinical practice in the US and Europe.29 Two enriched test set multiple case multiple reader laboratory studies reported that AI outperformed an average single radiologist reading in a laboratory setting, but the generalisability to clinical practice is unclear.28 33

AI as a standalone system for triage

Four studies used the Transpara versions 1.4.0 and 1.6.0 and Lunit version 5.5.0.16 AI systems, respectively, as a pre-screen to identify women at low risk whose mammograms required less or no radiological review.25 26 27 31 In this use, AI systems require high sensitivity so that few patients with cancer are excluded from radiological review, and only moderate specificity, which determines the radiology case load saved. In a retrospective consecutive German cohort (n=17 895, 0.64% cancer) the Transpara version 1.6.0 AI system achieved a sensitivity of 92% and a specificity of 66% at the Transpara score 7 to remove patients at low risk from double reading, and 96% sensitivity (45% specificity) at a Transpara score of 5.25 A Transpara version 1.4.0 score of 5 had 90% sensitivity and 53% specificity in a Swedish cohort (n=9581, 0.71% cancer).27 Both studies reported 100% sensitivity at a score of 2 (and specificities of 15% and 19%, respectively). The threshold for classification (725 and 527) was determined by exploring the full range of Transpara scores from 1 to 10 in the same dataset (fig 4A). In these studies, screen negative women were not followed up, so the sensitivity refers to detection of cancers which were detected by the original radiologists.
Fig 4

Study estimates of sensitivity and false positive rate (1−specificity) in receiver operating characteristic space for studies of artificial intelligence (AI) as a pre-screen (A) or post-screen (B). Pre-screen requires very high sensitivity, but can have modest specificity, post-screen requires very high specificity, but can have modest sensitivity. Reference standard for test negatives was double reading not follow-up. (A) Dembrower 2020a: retrospective study using AI (Lunit version 5.5.0.16) for pre-screen (point estimates not based on exact numbers). Reference standard includes only screen detected cancers. No data reported for radiologists.26 Balta 2020 (Transpara version 1.6.0),25 Raya-Povedano 2021 (Transpara version 1.6.0),31 and Lång 2020 (Transpara version 1.4.0)27: retrospective studies using AI as pre-screen. Reference standard includes only screen detected cancers. (B) Dembrower 2020b: retrospective study using AI (Lunit version 5.5.0.16) for post-screen detection of interval cancers,26 Dembrower 2020c: retrospective study using AI (Lunit version 5.5.0.16) for post-screen detection of interval cancers and next round screen detected cancers.26 Thresholds highlighted represent thresholds specified in studies. Radiologist double reading for this cohort would be 100% specificity and 0% sensitivity as this was only in a cohort of women with screen (true and false) negative mammograms

Study estimates of sensitivity and false positive rate (1−specificity) in receiver operating characteristic space for studies of artificial intelligence (AI) as a pre-screen (A) or post-screen (B). Pre-screen requires very high sensitivity, but can have modest specificity, post-screen requires very high specificity, but can have modest sensitivity. Reference standard for test negatives was double reading not follow-up. (A) Dembrower 2020a: retrospective study using AI (Lunit version 5.5.0.16) for pre-screen (point estimates not based on exact numbers). Reference standard includes only screen detected cancers. No data reported for radiologists.26 Balta 2020 (Transpara version 1.6.0),25 Raya-Povedano 2021 (Transpara version 1.6.0),31 and Lång 2020 (Transpara version 1.4.0)27: retrospective studies using AI as pre-screen. Reference standard includes only screen detected cancers. (B) Dembrower 2020b: retrospective study using AI (Lunit version 5.5.0.16) for post-screen detection of interval cancers,26 Dembrower 2020c: retrospective study using AI (Lunit version 5.5.0.16) for post-screen detection of interval cancers and next round screen detected cancers.26 Thresholds highlighted represent thresholds specified in studies. Radiologist double reading for this cohort would be 100% specificity and 0% sensitivity as this was only in a cohort of women with screen (true and false) negative mammograms One study predefined the Transpara score of 7 to identify women at low risk in a Spanish cohort (n=15 986, 0.7% cancer, including 15 interval cancers within 24 months of follow-up) and achieved 88% sensitivity and 72% specificity.31 A Swedish case-control study (n=7364, 7.4% cancer) used a range of thresholds to consider use of the Lunit version 5.5.0.16 AI system as a pre-screen to remove normal patients (fig 4A) and then as a post-screen of patients who were negative after double reading to identify additional cancers (interval cancers and next round screen detected cancers; fig 4B).26 Using 11 times upsampling of healthy women to simulate a screening population, they reported that use of AI alone with no subsequent radiologist assessment in the 50% and 90% of women with the lowest AI scores had 100% and 96% sensitivity and 50% and 90% specificity, respectively. AI assessment of negative mammograms after double reading detected 103 (19%) of 547 interval and next round screen detected cancers if the 2% women with the highest AI scores were post-screened (with a hypothetical perfect follow-up test).26 None of these studies reported any empirical data on the effect on radiologist behaviour of integrating AI into the screening pathway.

AI as a reader aid

No randomised controlled trials, test accuracy studies, or cohort studies evaluated AI as a reader aid in clinical practice. The only three studies of AI as a reader aid reported accuracy of radiologists’ reading of an enriched test set in a laboratory environment, with limited generalisability to clinical practice. Sensitivity and specificity were reported as an average of 14,30 14,32 or 737 radiologists with and without the AI reader aid. Point estimates of the average sensitivity were higher for radiologists with AI support than for unaided reading (absolute difference +3.0%, P=0.046,32 +3.3%, P=0.021,30 and +11%, P=0.03037) in all three studies of 240,30 240,32 and 12237 women. The effect of AI support on average reader specificity in a laboratory setting was small (absolute difference +2.0%, P=0.06,32 +1.0%, P=0.63,30 and −0.9%,37 no P value reported; table 4).

Cancer type

Limited data were reported on types of cancer detected, with some evidence of systematic differences between different AI systems. Of the three retrospective cohort studies investigating AI as a standalone system to replace radiologist(s), only one reported measuring whether there was a difference between AI and radiologists in the type of cancer detected. One anonymised AI system detected more invasive cancers (82.8%) than a radiologist (radiologist 1: 76.7%; radiologist 2: 79.7%, n=640) and less ductal carcinoma in situ (83.5%) than a radiologist (radiologist 1: 89.4%; radiologist 2: 89.4%, n=85), though the grades of ductal carcinoma in situ and invasive cancer were not reported.35 This same AI system detected more stage 2 or higher invasive cancers (n=204, 78.4% than radiologist 1: 68.1% and radiologist 2: 68.1%).35 The other two anonymised AI systems detected fewer stage 2 or higher invasive cancers (58.3% and 60.8%) than the radiologists. In an enriched test set multiple reader multiple case laboratory study, a standalone in-house AI model (DeepHealth Inc.) detected more invasive cancer (+12.7%, 95% confidence interval 8.5 to 16.5) and more ductal carcinoma in situ (+16.3%, 95% confidence interval 10.9 to 22.2) than the average single reader.28 This trend for higher performance of the AI model was also seen for lesion type, cancer size, and breast density. In an enriched test set multiple reader multiple case laboratory study, addition of the CureMetrix AI system to assist readers increased detection of microcalcifications (n=17,+20%) preferentially in comparison with other mammographic abnormalities such as masses (n=73,+9%).37 Microcalcifications are known to be more associated with ductal carcinoma in situ than with invasive cancer, but the spectrum of disease was not directly reported. Forty seven (87%) of 54 screen detected invasive cancers were classified as high risk using Transpara version 1.4.0 with a threshold of 5, in comparison with 14 (100%) of 14 microcalcifications.27 Using Transpara version 1.6.0 with a threshold of 7 as pre-screen, four additional cancers were classified as high risk by AI that had been missed by original double reading without consensus (two ductal carcinoma in situ, one low grade invasive ductal cancer, and one high grade invasive ductal cancer).31 No information on cancer type was reported for the two screen detected cancers that were classed by AI as low risk.

Discussion

Main findings

In this systematic review of AI mammographic systems for image analysis in routine breast screening, we identified 12 studies which evaluated commercially available or in-house convolutional neural network AI systems, of which nine included a comparison with radiologists. One of the studies reported that they followed STARD reporting guidelines.36 The six smallest studies (total 4183 women) found that AI was more accurate than single radiologists.28 29 30 32 33 37 The radiologists in five of six of these studies were examining the mammographic images of 932 women in a laboratory setting, which is not generalisable to clinical practice. In the remaining study, the comparison was with a single reading in the US with an accuracy below that expected in usual clinical practice.29 Whether this lower accuracy was due to case mix or radiologist expertise is unclear. In two of the largest retrospective cohort studies of AI to replace radiologists in Europe (n=76 813 women),35 36 all AI systems were less accurate than consensus of two radiologists, and 34 of 36 AI systems were less accurate than a single reader. One unpublished study is in line with these findings.40 This large retrospective study (n=275 900 women) reported higher sensitivity of AI in comparison with the original first reader decision but lower specificity, and the AI system was less accurate than consensus reading.40 Four retrospective studies25 26 27 31 indicated that at lower thresholds, AI can achieve high sensitivity so might be suitable for triaging which women should receive radiological review. Further research is required to determine the most appropriate threshold as the only study which prespecified the threshold for triage achieved 88.5% sensitivity.31 Evidence suggests that the accuracy and spectrum of disease detected between different AI systems is variable. Considerable heterogeneity in study methodology was found, some of which resulted in high concerns over risk of bias and applicability. Compared with consecutive sampling, case-control studies added bias by selecting cases and controls41 to achieve an enriched sample. The resulting spectrum effect could not be assessed because studies did not adequately report the distribution of original radiological findings, such as the distribution of the original BI-RADS scores. The effect was likely to be greater, however, when selection was based on image or cancer characteristics rather than if enrichment was achieved by including all available women with cancer and a random sample of those who were negative. The overlap of populations in three Swedish studies means that they represent only one rather than three separate cohorts.26 35 36 Performance of the AI system might have been overestimated if the same AI system read the same dataset more than once and, therefore, could have had the opportunity to learn. We could not confirm this as the three AI systems used by Salim et al were anonymised.35 The included studies have some variation in reference standard for the definition of normal cases, from simply consensus decision of radiologists at screening, to one to three years of follow-up. This inconsistency means accuracy estimates are comparable within, but not between, studies. Overall, the current evidence is a long way from the quality and quantity required for implementation in clinical practice.

Strengths and limitations

We followed standard methodology for conducting systematic reviews, used stringent inclusion criteria, and tailored the quality assessment tool for included studies. The stringent inclusion criteria meant that we included only geographical validation of test sets in the review—that is, at different centres in the same or different countries, which resulted in exclusion of a large number of studies that used some form of internal validation (where the same dataset is used for training and validation—for example, using cross validation or bootstrapping). Internal validation overestimates accuracy and has limited generalisability,42 and might also result in overfitting and loss of generalisability as the model fits the trained data extremely well but to the detriment of its ability to perform with new data. The split sample approach similarly does not accurately reflect a model’s generalisability.43 Temporal validation is regarded as an approach that lies midway between internal and external validation43 and has been reported by others to be sufficient in meeting the expectations of an external validation set to evaluate the effectiveness of AI.42 For screening, however, temporal validation could introduce bias because, for instance, the same women might attend repeat screens, and be screened by the same personnel using the same machines. Only geographical validation offers the benefits of external validation and generalisability.42 We also excluded computer aided detection for breast screening using systems that were categorised as traditional. The definition was based on expert opinion and the literature.14 The distinction is not clear cut and this approach might have excluded relevant studies that poorly reported the AI methods or used a combination of methods. We extracted binary classifications from AI systems, and do not know how other information on a recall to assessment form from a radiologist, such as mammographic characteristics or BI-RADS score/level of suspicion, might affect the provision of follow-up tests. In addition, AI algorithms are short lived and constantly improve. Reported assessments of AI systems might be out of date by the time of study publication, and their assessments might not be applicable to AI systems available at the time. The exclusion of non-English studies might have excluded relevant evidence. The available methodological evidence suggests that this is unlikely to have biased the results or affected the conclusions of our review.44 45 Finally, the QUADAS-2 adaptation was a first iteration and needs further refinement taking into consideration the QUADAS-2 AI version and AI reporting guides such as STARD-AI and CONSORT-AI, which are expected to be published in due course.

Strengths and limitations in comparison with previous studies

The findings from our systematic review disagree with the publicity some studies have received and opinions published in various journals, which claim that AI systems outperform humans and might soon be used instead of experienced radiologists.10 11 12 13 Our different conclusion is based on our rigorous and systematic evaluation of study quality. We did not extract the “simulation” parts of studies, which were often used as the headline numbers in the original papers, and often estimated higher accuracy for AI than the empirical data of the studies. In these simulations various assumptions were made about how radiologist arbitrators would behave in combination with AI, without any clinical data on behaviour in practice with AI. Although a great number of studies report the development and internal validation of AI systems for breast screening, our study shows that this high volume of published studies does not reflect commercially available AI systems suitable for integration into screening programmes. Our emphasis on comparisons with the accuracy of radiologists in clinical practice explains why our conclusions are more cautious than many of the included papers. A recent scoping review with a similar research question, but broader scope, reported a potential role for AI in breast screening but identified evidence gaps that showed a lack of readiness of AI for breast screening programmes.46 The 23 included studies were mainly small, retrospective, and used publicly available and institutional image datasets, which often overlapped. The evidence included only one study with a consecutive cohort, one study with a commercially available AI system, and five studies that compared AI with radiologists. We found overlap of only one study between the scoping review and our review despite the same search start date, probably because we focused on higher study quality. Our review identified nine additional recent eligible studies, which might suggest that the quality of evidence is improving, but as yet no prospective evaluations of AI have been reported in clinical practice settings.

Possible explanations and implications for clinicians and policy makers

Our systematic review should be considered in the wider context of the increasing proposed use of AI in healthcare and screening. Most of the literature focuses, understandably, on those screening programmes in which image recognition and interpretation are central components, and this is indicated by a number of reviews recently published describing studies of AI and deep learning for diabetic retinopathy screening.47 48 Beyond conventional screening programmes, the use of deep learning in medicine is increasing, and has been considered in the diagnosis of melanoma,49 ophthalmic diseases (age-related macular degeneration50 and glaucoma51), and in interpretation of histological,52 radiological,53 and electrocardiogram54 images. Evidence is insufficient on the accuracy or clinical effect of introducing AI to examine mammograms anywhere on the screening pathway. It is not yet clear where on the clinical pathway AI might be of most benefit, but its use to redesign the pathway with AI complementing rather than competing with radiologists is a potentially promising way forward. Examples of this include using AI to pre-screen easy normal mammograms for no further review, and post-screen for missed cases. Similarly, in diabetic eye screening there is growing evidence that AI can filter which images need to be viewed by a human grader, and which can be reported as normal immediately to the woman.55 56 Medical decisions made by AI independently of humans might have medicolegal implications.57 58

Implications for research

Prospective research is required to measure the effect of AI in clinical practice. Although the retrospective comparative test accuracy studies, which compared AI performance with the original decision of the radiologist, have the advantage of not being biased by the laboratory effect, the readers were “gatekeepers” for biopsy. This means that we do not know the true cancer status of women whose mammograms were AI positive and radiologist negative. Examination of follow-up to interval cancers does not fully resolve this problem of true cancer status, as lead times to symptomatic presentation are often longer than the study follow-up time. Prospective studies can answer this question by recalling for further assessment women whose mammograms test positive by AI or radiologist. Additionally, evidence is needed on the types of cancer detected by AI to allow an assessment of potential changes to the balance of benefits and harms, including potential overdiagnosis. We need evidence for specific subgroups according to age, breast density, prior breast cancer, and breast implants. Evidence is also needed on radiologist views and understanding and on how radiologist arbitrators behave in combination with AI. Finally, evidence is needed on the direct comparison of different AI systems; the effect of different mammogram machines on the accuracy of AI systems; the effect of differences in screening programmes on cancer detection with AI, or on how the AI system might work within specific breast screening IT systems; and the effect of making available additional information to AI systems for decision making. Commercially available AI systems should not be anonymised in research papers, as this makes the data useless for clinical and policy decision makers. The most applicable evidence to answer this question would come from prospective comparative studies in which the index test is the AI system integrated into the screening pathway, as it would be used in screening practice. These studies would need to report the change to the whole screening pathway when AI is added as a second reader, as the only reader, as a pre-screen, or as a reader aid. No studies of this type or prospective studies of test accuracy in clinical practice were available for this review. We did identify two ongoing randomised controlled trials, however: one investigating AI as pre-screen with the replacement of double reading for women at low risk with single reading (randomising to AI integrated mammography screening v conventional mammography screening), and one investigating AI as a post-screen (randomising women with the highest probability of having had a false negative screening mammogram to MRI or standard of care.)59 60

Conclusions

Current evidence on the use of AI systems in breast cancer screening is a long way from having the quality and quantity required for its implementation into clinical practice. Well designed comparative test accuracy studies, randomised controlled trials, and cohort studies in large screening populations are needed which evaluate commercially available AI systems in combination with radiologists. Such studies will enable an understanding of potential changes to the performance of breast screening programmes with an integrated AI system. By highlighting the shortcomings, we hope to encourage future users, commissioners, and other decision makers to press for high quality evidence on test accuracy when considering the future integration of AI into breast cancer screening programmes. A recent scoping review of 23 studies on artificial intelligence (AI) for the early detection of breast cancer highlighted evidence gaps and methodological concerns about published studies Published opinion pieces claim that the replacement of radiologists by AI is imminent Current mammography screening is repetitive work for radiologists and misses 15-35% of cancers—a prime example of the sort of -role we would expect AI to be fulfilling This systematic review of test accuracy identified 12 studies, of which only one was included in the previous review Current evidence on the use of AI systems in breast cancer screening is of insufficient quality and quantity for implementation into clinical practice In retrospective test accuracy studies, 94% of AI systems were less accurate than the original radiologist, and all were less accurate than original consensus of two radiologists; prospective evaluation is required
  49 in total

Review 1.  The effect of English-language restriction on systematic review-based meta-analyses: a systematic review of empirical studies.

Authors:  Andra Morrison; Julie Polisena; Don Husereau; Kristen Moulton; Michelle Clark; Michelle Fiander; Monika Mierzwinski-Urban; Tammy Clifford; Brian Hutton; Danielle Rabb
Journal:  Int J Technol Assess Health Care       Date:  2012-04       Impact factor: 2.188

2.  Can we open the black box of AI?

Authors:  Davide Castelvecchi
Journal:  Nature       Date:  2016-10-06       Impact factor: 49.962

3.  Preferred Reporting Items for a Systematic Review and Meta-analysis of Diagnostic Test Accuracy Studies: The PRISMA-DTA Statement.

Authors:  Matthew D F McInnes; David Moher; Brett D Thombs; Trevor A McGrath; Patrick M Bossuyt; Tammy Clifford; Jérémie F Cohen; Jonathan J Deeks; Constantine Gatsonis; Lotty Hooft; Harriet A Hunt; Christopher J Hyde; Daniël A Korevaar; Mariska M G Leeflang; Petra Macaskill; Johannes B Reitsma; Rachel Rodin; Anne W S Rutjes; Jean-Paul Salameh; Adrienne Stevens; Yemisi Takwoingi; Marcello Tonelli; Laura Weeks; Penny Whiting; Brian H Willis
Journal:  JAMA       Date:  2018-01-23       Impact factor: 56.272

4.  Locality Sensitive Deep Learning for Detection and Classification of Nuclei in Routine Colon Cancer Histology Images.

Authors:  Korsuk Sirinukunwattana; Shan E Ahmed Raza; David R J Snead; Ian A Cree; Nasir M Rajpoot
Journal:  IEEE Trans Med Imaging       Date:  2016-02-04       Impact factor: 10.048

5.  Detection of Breast Cancer with Mammography: Effect of an Artificial Intelligence Support System.

Authors:  Alejandro Rodríguez-Ruiz; Elizabeth Krupinski; Jan-Jurre Mordang; Kathy Schilling; Sylvia H Heywang-Köbrunner; Ioannis Sechopoulos; Ritse M Mann
Journal:  Radiology       Date:  2018-11-20       Impact factor: 11.105

6.  Breast screening using 2D-mammography or integrating digital breast tomosynthesis (3D-mammography) for single-reading or double-reading--evidence to guide future screening strategies.

Authors:  Nehmat Houssami; Petra Macaskill; Daniela Bernardi; Francesca Caumo; Marco Pellegrini; Silvia Brunelli; Paola Tuttobene; Paola Bricolo; Carmine Fantò; Marvi Valentini; Stefano Ciatto
Journal:  Eur J Cancer       Date:  2014-04-17       Impact factor: 9.162

7.  Predicting the Future - Big Data, Machine Learning, and Clinical Medicine.

Authors:  Ziad Obermeyer; Ezekiel J Emanuel
Journal:  N Engl J Med       Date:  2016-09-29       Impact factor: 91.245

8.  Prospective evaluation of an artificial intelligence-enabled algorithm for automated diabetic retinopathy screening of 30 000 patients.

Authors:  Peter Heydon; Catherine Egan; Louis Bolter; Ryan Chambers; John Anderson; Steve Aldington; Irene M Stratton; Peter Henry Scanlon; Laura Webster; Samantha Mann; Alain du Chemin; Christopher G Owen; Adnan Tufail; Alicja Regina Rudnicka
Journal:  Br J Ophthalmol       Date:  2020-06-30       Impact factor: 4.638

Review 9.  Prevalence of incidental breast cancer and precursor lesions in autopsy studies: a systematic review and meta-analysis.

Authors:  Elizabeth T Thomas; Chris Del Mar; Paul Glasziou; Gordon Wright; Alexandra Barratt; Katy J L Bell
Journal:  BMC Cancer       Date:  2017-12-02       Impact factor: 4.430

10.  Evaluation of Combined Artificial Intelligence and Radiologist Assessment to Interpret Screening Mammograms.

Authors:  Thomas Schaffter; Diana S M Buist; Christoph I Lee; Yaroslav Nikulin; Dezso Ribli; Yuanfang Guan; William Lotter; Zequn Jie; Hao Du; Sijia Wang; Jiashi Feng; Mengling Feng; Hyo-Eun Kim; Francisco Albiol; Alberto Albiol; Stephen Morrell; Zbigniew Wojna; Mehmet Eren Ahsen; Umar Asif; Antonio Jimeno Yepes; Shivanthan Yohanandan; Simona Rabinovici-Cohen; Darvin Yi; Bruce Hoff; Thomas Yu; Elias Chaibub Neto; Daniel L Rubin; Peter Lindholm; Laurie R Margolies; Russell Bailey McBride; Joseph H Rothstein; Weiva Sieh; Rami Ben-Ari; Stefan Harrer; Andrew Trister; Stephen Friend; Thea Norman; Berkman Sahiner; Fredrik Strand; Justin Guinney; Gustavo Stolovitzky; Lester Mackey; Joyce Cahoon; Li Shen; Jae Ho Sohn; Hari Trivedi; Yiqiu Shen; Ljubomir Buturovic; Jose Costa Pereira; Jaime S Cardoso; Eduardo Castro; Karl Trygve Kalleberg; Obioma Pelka; Imane Nedjar; Krzysztof J Geras; Felix Nensa; Ethan Goan; Sven Koitka; Luis Caballero; David D Cox; Pavitra Krishnaswamy; Gaurav Pandey; Christoph M Friedrich; Dimitri Perrin; Clinton Fookes; Bibo Shi; Gerard Cardoso Negrie; Michael Kawczynski; Kyunghyun Cho; Can Son Khoo; Joseph Y Lo; A Gregory Sorensen; Hwejin Jung
Journal:  JAMA Netw Open       Date:  2020-03-02
View more
  12 in total

1.  Impact of artificial intelligence in breast cancer screening with mammography.

Authors:  Lan-Anh Dang; Emmanuel Chazard; Edouard Poncelet; Teodora Serb; Aniela Rusu; Xavier Pauwels; Clémence Parsy; Thibault Poclet; Hugo Cauliez; Constance Engelaere; Guillaume Ramette; Charlotte Brienne; Sofiane Dujardin; Nicolas Laurent
Journal:  Breast Cancer       Date:  2022-06-28       Impact factor: 3.307

2.  Ultra-high resolution, multi-scale, context-aware approach for detection of small cancers on mammography.

Authors:  Krithika Rangarajan; Aman Gupta; Saptarshi Dasgupta; Uday Marri; Arun Kumar Gupta; Smriti Hari; Subhashis Banerjee; Chetan Arora
Journal:  Sci Rep       Date:  2022-07-08       Impact factor: 4.996

3.  Reporting guideline for the early stage clinical evaluation of decision support systems driven by artificial intelligence: DECIDE-AI.

Authors:  Baptiste Vasey; Myura Nagendran; Bruce Campbell; David A Clifton; Gary S Collins; Spiros Denaxas; Alastair K Denniston; Livia Faes; Bart Geerts; Mudathir Ibrahim; Xiaoxuan Liu; Bilal A Mateen; Piyush Mathur; Melissa D McCradden; Lauren Morgan; Johan Ordish; Campbell Rogers; Suchi Saria; Daniel S W Ting; Peter Watkinson; Wim Weber; Peter Wheatstone; Peter McCulloch
Journal:  BMJ       Date:  2022-05-18

4.  Artificial Intelligence Evaluation of 122 969 Mammography Examinations from a Population-based Screening Program.

Authors:  Marthe Larsen; Camilla F Aglen; Christoph I Lee; Solveig R Hoff; Håkon Lund-Hanssen; Kristina Lång; Jan F Nygård; Giske Ursin; Solveig Hofvind
Journal:  Radiology       Date:  2022-03-29       Impact factor: 29.146

5.  Good quality practices for artificial intelligence in genetics.

Authors:  Timothé Ménard
Journal:  Eur J Hum Genet       Date:  2022-02-07       Impact factor: 5.351

6.  Independent External Validation of Artificial Intelligence Algorithms for Automated Interpretation of Screening Mammography: A Systematic Review.

Authors:  Anna W Anderson; M Luke Marinovich; Nehmat Houssami; Kathryn P Lowry; Joann G Elmore; Diana S M Buist; Solveig Hofvind; Christoph I Lee
Journal:  J Am Coll Radiol       Date:  2022-01-20       Impact factor: 5.532

7.  Use of infrared thermography in medical diagnostics: a scoping review protocol.

Authors:  Dorothea Kesztyüs; Sabrina Brucher; Tibor Kesztyüs
Journal:  BMJ Open       Date:  2022-04-01       Impact factor: 2.692

8.  Identification of early invisible acute ischemic stroke in non-contrast computed tomography using two-stage deep-learning model.

Authors:  Jun Lu; Yiran Zhou; Wenzhi Lv; Hongquan Zhu; Tian Tian; Su Yan; Yan Xie; Di Wu; Yuanhao Li; Yufei Liu; Luyue Gao; Wei Fan; Yan Nan; Shun Zhang; Xiaolong Peng; Guiling Zhang; Wenzhen Zhu
Journal:  Theranostics       Date:  2022-07-18       Impact factor: 11.600

9.  Screen-detected and interval breast cancer after concordant and discordant interpretations in a population based screening program using independent double reading.

Authors:  Marit A Martiniussen; Silje Sagstad; Marthe Larsen; Anne Sofie F Larsen; Tone Hovda; Christoph I Lee; Solveig Hofvind
Journal:  Eur Radiol       Date:  2022-04-02       Impact factor: 7.034

Review 10.  The Role of Artificial Intelligence in Early Cancer Diagnosis.

Authors:  Benjamin Hunter; Sumeet Hindocha; Richard W Lee
Journal:  Cancers (Basel)       Date:  2022-03-16       Impact factor: 6.639

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.