Load applied on osseointegrated implant by transfemoral bone-anchored prostheses fitted with state-of-the-art prosthetic components.

Laurent Frossard1, Stefan Laux2, Marta Geada2, Peter Paul Heym3, Knut Lechler4.   

Abstract

BACKGROUND: This study presented the load profile applied on transfemoral osseointegrated implants by bone-anchored prostheses fitted with state-of-the-art ÖSSUR microprocessor-controlled Rheo Knee XC and energy-storing-and-returning Pro-Flex XC or LP feet during five standardized daily activities.
METHODS: This cross-sectional cohort study included 13 participants fitted with a press-fit transfemoral osseointegrated implant. Loading data were directly measured with the tri-axial transducer of an iPecsLab (RTC Electronics, USA) fitted between the implant and knee unit. The loading profile was characterized by spatio-temporal gaits variables, magnitude of loading boundaries as well as onset and magnitude of loading extrema during walking, ascending and descending ramp and stairs.
FINDINGS: A total of 2127 steps was analysed. The cadence ranged between 36 ± 7 and 47 ± 6 strides/min. The absolute maximum force and moments applied across all activities was 1322 N, 388 N and 133 N as well as 22 Nm, 52 Nm and 88 Nm on and around the long, anteroposterior and mediolateral axes of the implant, respectively.
INTERPRETATION: This study provided new benchmark loading data applied by transfemoral bone-anchored prostheses fitted with selected ÖSSUR state-of-the-art components. Outcomes suggested that such prostheses can generate relevant loads at the interface with the osseointegrated implant to restore ambulation effectively. This study is a worthwhile contribution toward a systematic recording, analysis, and reporting of ecological prosthetic loading profiles as well as closing the evidence gaps between prescription and biomechanical benefits of state-of-the-art components. Hopefully, this will contribute to improve outcomes for growing number of individuals with limb loss opting for bionic solutions.
Copyright © 2021 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Amputation; Artificial limbs; Bionics; Bone-anchored prosthesis; Kinetics; Loading; Prosthesis

Mesh:

Year:  2021        PMID: 34454327     DOI: 10.1016/j.clinbiomech.2021.105457

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

Review 1.  EMG-driven control in lower limb prostheses: a topic-based systematic review.

Authors:  Andrea Cimolato; Josephus J M Driessen; Leonardo S Mattos; Elena De Momi; Matteo Laffranchi; Lorenzo De Michieli
Journal:  J Neuroeng Rehabil       Date:  2022-05-07       Impact factor: 5.208

2.  Loading characteristics data applied on osseointegrated implant by transfemoral bone-anchored prostheses fitted with state-of-the-art components during daily activities.

Authors:  Laurent Frossard; Stefan Laux; Marta Geada; Peter Paul Heym; Knut Lechler
Journal:  Data Brief       Date:  2022-02-10

3.  Editorial: Bionics limb prostheses: Advances in clinical and prosthetic care.

Authors:  Laurent Frossard; Silvia Conforto; Oskar C Aszmann
Journal:  Front Rehabil Sci       Date:  2022-08-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.