Literature DB >> 34428090

Contrast-Enhanced Ultrasound and Shear Wave Elastography: Novel Methods for the Evaluation of Urethral Stricture Disease.

Paul H Chung1, Joon Yau Leong1, Priscilla Machado2, Corinne E Wessner2, Edouard J Trabulsi1, Ethan J Halpern2, John R Eisenbrey2, Flemming Forsberg2.   

Abstract

PURPOSE: Urologists will benefit from an imaging modality which can assess intra and extraluminal characteristics of urethral strictures. We conducted a prospective pilot study evaluating the utility of contrast-enhanced ultrasound and shear wave elastography for the evaluation of bulbar urethral stricture disease.
MATERIALS AND METHODS: Patients with a single, bulbar urethral stricture were prospectively recruited. Contrast-enhanced ultrasound and shear wave elastography were performed at the time of surgical repair and at 4 months' followup using an Aplio i800 scanner (Canon Medical Systems, Tustin, California) with an i8CX1 transducer. Sulfur hexafluoride lipid-type A microsphere ultrasound contrast (Lumason®, Bracco Imaging, Princeton, New Jersey) was injected retrograde through the urethra. Stiffness of the corpus spongiosum was measured at and adjacent to the stricture site. Stricture lengths based on retrograde urethrogram, grayscale ultrasound and contrast-enhanced ultrasound were correlated with measured intraoperative stricture length.
RESULTS: Thirty men were enrolled. Contrast-enhanced ultrasound (R2=0.709) showed the best correlation with intraoperative measured stricture length compared to retrograde urethrogram (R2=0.016) or grayscale ultrasound (R2=0.471). Stiffness of the spongiosum was greater at the site of the stricture (32.6±5.4 vs 27.3±5.8 kPa, p=0.044) and in narrower caliber strictures (p=0.044) but did not differ by stricture length (p=0.182). At followup (4.3±1.1 months) contrast-enhanced ultrasound detected stricture recurrence with 80% sensitivity, 100% specificity, and 93% accuracy compared to cystoscopy.
CONCLUSIONS: This pilot study demonstrates the ability of contrast-enhanced ultrasound and shear wave elastography to become safe, accurate, and potentially efficacious modalities for assessing bulbar urethral strictures and spongiofibrosis.

Entities:  

Keywords:  contrast media; elasticity imaging techniques; microbubbles; ultrasonography; urethral stricture

Mesh:

Substances:

Year:  2021        PMID: 34428090     DOI: 10.1097/JU.0000000000002146

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.600


  2 in total

Review 1.  The evolving role of contrast-enhanced ultrasound in urology: a review.

Authors:  Andrew Salib; Ethan Halpern; John Eisenbrey; Thenappan Chandrasekar; Paul H Chung; Flemming Forsberg; Edouard J Trabulsi
Journal:  World J Urol       Date:  2022-08-15       Impact factor: 3.661

Review 2.  Elastography in the Urological Practice: Urinary and Male Genital Tract, Prostate Excluded-Review.

Authors:  Vasile Simon; Sorin Marian Dudea; Nicolae Crisan; Vasile Dan Stanca; Marina Dudea-Simon; Iulia Andras; Zoltan Attila Mihaly; Ioan Coman
Journal:  Diagnostics (Basel)       Date:  2022-07-16
  2 in total