| Literature DB >> 34427246 |
Vanita Pathak Ray1, Smitha B Ramesh2, Varsha Rathi3.
Abstract
PURPOSE: There is a considerable lack of awareness of slit-lamp measurement of anterior chamber depth (ACD) by the Redmond Smith method (SACD) in present day-to-day clinical practice, which may provide rapid assessment in pseudoexfoliation (PXF) when assessing for angle closure and planning for cataract surgery. This assumes importance not only in outreach clinics but also in the ongoing pandemic caused by the highly contagious novel coronavirus, where social distancing is advocated to contain the spread. We aimed to compare the axial ACD in PXF and normal patients by SACD, and its agreement with the anterior segment optical coherence tomography (ASOCT) and LenstarLS-900.Entities:
Keywords: ACD; ASOCT; Lenstar LS 900; Redmond Smith; noncontact technique; pseudoexfoliation; slit-lamp anterior chamber depth measurement
Mesh:
Year: 2021 PMID: 34427246 PMCID: PMC8544031 DOI: 10.4103/ijo.IJO_3738_20
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1(a) Diagrammatic representation of Positions A and B for clinical assessment of anterior chamber depth at the slit lamp. Courtesy Smith.[15] (Reproduced from Smith, R.J., A new method of estimating the depth of the anterior chamber. Br J Ophthalmol. 1979;63 (4):215-20. Permission Licence no. 4973651427126, dated December 21, 2020). (b) Slit-lamp view of Positions A and B for clinical assessment of anterior chamber depth at the slit lamp
Anterior chamber depth (ACD) measurement by three methodologies in pseudoexfoliation group (PXF) and in normal controls
| SACD† in mm | AACD‡ in mm | LACD¶ in mm | |
|---|---|---|---|
| Mean (SD) | 2.63 (0.46) | 2.48 (0.38) | 2.40 (0.37) |
| Mean (SD) | 2.95 (0.35) | 2.78 (0.32) | 2.72 (0.31) |
| ANOVA |
SD=Standard deviation, CI=Confidence interval; ANOVA=Analysis of variance. †SACD - ACD measurement by Redmond Smith method ‡AACD - ACD obtained by anterior segment OCT ¶LACD - ACD obtained by Lenstar
Figure 2Anterior chamber depth (y-axis) via 3 methods in pseudoexfoliation group (cases) and normal controls
Figure 3ABC: Bland–Altman plots in pseudoexfoliation (PXF) group. Dashed lines represent the 95% confidence intervals. Solid line represents the mean difference
Figure 4ABC: Bland–Altman plots in normal control group. Dashed lines represent the 95% confidence intervals. Solid line represents the mean difference