Repeatability and agreement of wavefront aberrations of a new hybrid topographer and aberrometer in healthy eyes.

Rohit Shetty1, Divya Trivedi, Reshma Ranade, Spriha Arun, Pooja Khamar, Gairik Kundu.   

Abstract

PURPOSE: To assess repeatability and agreement of the Pentacam AXL Wave with that of the iTrace aberrometer.
SETTING: Narayana Nethralaya, Bangalore, India.
DESIGN: Prospective, cross-sectional study.
METHODS: All eyes of patients underwent 3 sets of measurements at a single visit, using Pentacam AXL Wave and iTrace. Lower-order aberrations (LOAs), higher-order aberrations (HOAs), and coma and spherical aberrations (SAs) were analyzed. Repeatability was assessed using within-subject (Sw) SD, test-retest variability (TRT), Sw coefficient of variation (CoV), and intraclass correlation coefficient (ICC). The Bland-Altman analysis was used to assess 95% limits of agreement (LoA) and the correlation coefficient (r) to determine strength of relationship between measurements.
RESULTS: 100 eyes of 100 patients underwent measurements. Pentacam AXL Wave had repeatability with an ICC for total ocular aberrations ranging from 0.84 for HOAs to 0.92 for LOAs and for corneal from 0.76 for HOAs to 0.86 for LOAs. The Sw, TRT, and CoV of all aberrations were significantly lower (better) than those of iTrace (P < .001). The mean differences between 2 devices were insignificant for the comparison of all parameters, except for spherical aberration (0.011 [0.002571 to 0.02008]; P = .01). Both devices showed excellent correlations for ocular aberrations (total HOAs [P = .12, LoA = -0.52 to 0.025; r = 0.89]; LOAs [P = .14, LoA = -2.71 to 2.61; r = 0.96]; coma [P = .27, LoA = -0.11 to 0.12; r = 0.89]; and SA [P = .01, LoA = -0.09 to 0.07; r = 0.97]).
CONCLUSIONS: The repeatability estimates for wavefront aberrations using Pentacam AXL Wave were significantly better than those using iTrace. Both devices showed excellent correlations for total ocular aberrations.
Copyright © 2021 Published by Wolters Kluwer on behalf of ASCRS and ESCRS.

Entities:  

Mesh:

Year:  2022        PMID: 34393184     DOI: 10.1097/j.jcrs.0000000000000775

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  1 in total

1.  The Pentacam® AXL Wave provides a reliable wavefront-based objective refraction when compared to manifest subjective refraction: A prospective study.

Authors:  Kepa Balparda; Andrea Acevedo-Urrego; Laura Andrea Silva-Quintero; Tatiana Herrera-Chalarca
Journal:  Indian J Ophthalmol       Date:  2022-05       Impact factor: 2.969

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.