Literature DB >> 34387730

Self-collected unstimulated saliva, oral swab, and nasopharyngeal swab specimens in the detection of SARS-CoV-2.

Fernanda de Paula Eduardo1, Leticia Mello Bezinelli1, Debora Heller2,3,4, Carlos Ariel Rodrigues de Araujo5, João Vitor Vanderlan Moraes6, Alexander Birbrair7,8, João Renato Rebello Pinho1, Nelson Hamerschlak1, Ibtisam Al-Hashimi9.   

Abstract

OBJECTIVES: The presence of SARS-CoV-2 virus in the saliva of patients infected with COVID-19 has been confirmed by several studies. However, the use of saliva for the diagnosis of COVID-19 remains limited, because of the discrepancies in the results, which might be due to using different saliva sampling methods. The purpose of this study was to compare the consistency of SARS-CoV-2 detection using two different saliva sampling methods (oral swab and unstimulated saliva) to that of the standard nasopharyngeal swab.
METHODS: Fifty-five subjects were recruited from a pool of COVID-19 inpatient at the Hospital Israelita Albert Einstein (HIAE), Brazil. Nasopharyngeal swab, oral swab, and self-collected unstimulated saliva samples were examined for SARS-CoV-2 using RT-PCR.
RESULTS: Self-collected unstimulated saliva demonstrated 87.3% agreement in the detection of SARS-CoV-2 virus as compared with the nasopharyngeal swab, while oral swab displayed 65.9% agreement when compared to nasopharyngeal swab and 73% when compared to self-collected unstimulated saliva.
CONCLUSION: Unstimulated self-collected saliva samples have shown a higher agreement with the nasopharyngeal swab samples for SARS-COV-2 detection than that obtained when using oral swab samples. CLINICAL RELEVANCE: This study compares the accuracy of COVID-19 test using different saliva sampling methods to that of nasopharyngeal swab. Given the need for a simple self-applied test that can be performed at home, our findings support the efficacy of self-collected unstimulated saliva samples in the diagnosis of SARS-CoV-2 infection, alleviating the demands for swab supplies, personal protective equipment, and healthcare personnel.
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Entities:  

Keywords:  Biomarkers; COVID-19; Coronavirus; Coronavirus infections; SARS-CoV-2; Saliva

Mesh:

Year:  2021        PMID: 34387730      PMCID: PMC8360804          DOI: 10.1007/s00784-021-04129-7

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


Introduction

Population-level employable rapid testing is part of a strategy, where testing should be widely available, preferably outside of healthcare or hospital settings, avoiding the overload of resources and the risk of transmission to other patients and healthcare professionals [1]. To date, the gold standard for the diagnosis of SARS-CoV-2 infection is reverse transcription-polymerase chain reaction (RT-PCR) analysis of respiratory tract swabs [2]. However, oropharyngeal and nasopharyngeal collection requires trained professionals to be performed and is associated with discomfort and inconvenience to the patients, in addition to increasing risk of virus transmission, thereby halting the possibility of employing this test in a large population setting [3-5]. Therefore, it is rather imperative to design a convenient and easily accessible test to control the spread of a highly infectious disease such as COVID-19 before vaccines or anti-viral treatment become widely available. Saliva specimen collection is one of the simplest ways to collect body fluids and represents an important source for the diagnosis and monitoring of coronavirus infection [6]. Saliva sampling has the advantage of being safe, easy to perform, and noninvasive, and may be self-collected repeatedly without discomfort to the patient. Several studies have confirmed a reliable detection of SARS-CoV-2 in the saliva of patients with COVID-19. However, the application of saliva testing for COVID-19 has been hampered due to discrepancies in the reported reliability of the test results. It is highly possible that the method of saliva sample collection could potentially be the source of the observed discrepancies in the reliability of the test results. Regarding the sampling methods, saliva samples used for COVID-19 testing have been collected in different ways, oral swab and unstimulated saliva [6]. To the best of our knowledge, there is no study that directly compared the accuracy of these two methods of sampling to that of the standard nasopharyngeal swab. Therefore, the aim of this study was to compare the consistency of SARS-CoV-2 detection using saliva samples (oral swab and self-collected unstimulated saliva) to that of standard nasopharyngeal swab.

Methods

Ethics

The study protocol was approved by the National Ethics in Human Research Committee of Brazil (CONEP, protocol #4172–20). All patients were individually informed about the nature of the proposed study and its risks and benefits, and signed informed consent forms.

Subject population and selection criteria

COVID-19 subjects aged 18–90 years were recruited from May 2020 to June 2020 from a pool of COVID-19 inpatients at the Hospital Israelita Albert Einstein (HIAE), Brazil. Inclusion criteria were patients with positive SARS-CoV-2 determined by the nasopharyngeal/oropharyngeal swab RT-PCR method and recorded in the medical record, within 3 days of hospitalization. Excluded were pediatric patients, patients with a negative RT-PCR result for SARS-CoV-2 at the time of recruitment, and patients who had other clinical illness that in the opinion of the investigators could interfere with compliance to sample collection.

Sample collection

The collection of nasopharyngeal and saliva samples was performed within 3 days of hospitalization. Senior nurse staff took nasopharyngeal samples using a flexible, mini-tip swab, which was passed through the patient’s nostril until the posterior nasopharynx was reached, left in place for several seconds to absorb secretions then slowly removed while rotating. The swab was placed in a sterile viral transport medium (total volume 3 mL) and sealed securely. Unstimulated whole saliva was self-collected by the patients in a sterile 50-mL tube as previously described [7]. Oral swabs were taken from the back of the tongue, deep groove and floor of the mouth by a trained dentist using a flexible mini-tip swab and placed in the sterile viral transport medium (total volume 3 mL). All samples were immediately transported on ice to the laboratory at the Hospital Israelita Albert Einstein and tested within 48 h of collection [8, 9].

Detection and quantification of viral RNA

Viral RNA was extracted from 200 μL of viral transport medium from the nasopharyngeal swab or from the oral swab or from unstimulated saliva specimens using a QIAsymphony™ RNA Kit (QIAGEN, Hilden, Germany) according to the manufacturer’s instructions. Following extraction, RNA was eluted with 60 µL of elution buffer. For SARS-CoV-2 RNA detection in saliva (unstimulated saliva and oral swab), RT-qPCR amplification of the Nucleocapsid (N) and Open reading frame (ORF) 1ab genes was performed using the XGen Master COVID-19 RT-PCR Kit (Mobius Life Science Comércio de Produto para Laboratórios Ltda, Pinhais, Brazil). The sensitivity of the reaction in routine diagnosis varies according to the gene analyzed, as follows: SARS-CoV-2 gene ORF1ab gene—10 copies/reaction; SARS-CoV-2 N gene—50 copies/reaction. Saliva samples were classified as positive for SARS-CoV-2 when the cycle threshold (CT) for both N and ORF1ab genes was lower than 38. For SARS-CoV-2 RNA detection in nasopharyngeal swab specimen (NPS), RT-qPCR amplification of the Envelope (E) and N genes was performed as previously described [10]. NPS samples were classified as positive for SARS-CoV-2 when the CT for both E and N genes was lower than 38 (Fig. 1).
Fig. 1

SARS-COV-2 diagnostic targets: The figure illustrates diagnostic targets for COVID-19 testing. Genes targets used for saliva and for nasopharyngeal swab specimen (NPS) are illustrated

SARS-COV-2 diagnostic targets: The figure illustrates diagnostic targets for COVID-19 testing. Genes targets used for saliva and for nasopharyngeal swab specimen (NPS) are illustrated

Statistical analysis

The results were presented by estimated mean values, 95% confidence intervals (95% CI) and p-values. The CT measurements obtained in saliva collection were compared to the CT measurements in nasopharyngeal swab collection using models of generalized estimation equations (GEE) with gamma distribution, considering the relationship between the different collections performed in the same patient. The same calculation was performed to compare the CT measurements obtained from unstimulated saliva, oral swab, and nasopharyngeal swab collections. The sensitivity of each test (self-collected unstimulated saliva and oral swab) was determined by the proportion of positive results to that of the gold standard (nasopharyngeal swab). All analyses were performed with the SPSS program (IBM Corp, IBM SPSS Statistics for Windows, Version 24.0, Armonk, NY) considering a 5% significance level.

Results

A total of 55 patients with positive COVID-19 RT-PCR using nasopharyngeal swab were included in the study. Of the 55 patients, 48 (87.3%) were also positive with self-collected unstimulated saliva. Forty-one patients also provided oral swab samples, and of those, 27 patients (65.9%) were positive for SARS-CoV-2 RNA detection (Table 1).
Table 1

SARS-CoV-2 virus detection in self-collected unstimulated saliva, oral swab, and nasopharyngeal swab specimens

SpecimenResult
Negativen (%)Positiven (%)
Unstimulated saliva (n = 55)7 (12.7)48 (87.3)
Oral swab (n = 41)14 (34.1)27 (65.9)
Nasopharyngeal swab (n = 55)0 (0.0)55 (100.0)
SARS-CoV-2 virus detection in self-collected unstimulated saliva, oral swab, and nasopharyngeal swab specimens

Unstimulated self-collected saliva and nasopharyngeal swab specimens in the detection of SARS-CoV-2

In the RT-qPCR of saliva samples, two genes from SARS-CoV-2 were amplified, ORF1ab and N genes. Amplification of N gene was observed in all saliva samples while amplification of ORF1ab, only in 72.8%. Table 2 shows the distribution of CT by the N and ORF genes in unstimulated saliva compared to the E and N genes in nasopharyngeal swab.
Table 2

Cycle threshold (CT) mean values and the 95% confidence interval (95% CI) comparisons for (A) N gene from self-collected unstimulated saliva and for the E and N genes from nasopharyngeal swab specimens (NPS) (n = 34) and (B) ORF gene from self-collected unstimulated saliva and for the E and N genes from NPS (n = 24)

Unstimulated salivaNasopharyngeal swab
AGene NGene EGene N

Mean CT

95% CI

29.9

28.4; 31.6

25.8

23.5; 28.3

28.4

26.0; 30.9

BGene ORFGene EGene N

Mean CT

95% CI

27.5

25.4; 29.7

24.1

21.5; 27.0

26.4

23.9; 29.1

Values expressed by estimated means and 95% confidence intervals (95% CI)

Cycle threshold (CT) mean values and the 95% confidence interval (95% CI) comparisons for (A) N gene from self-collected unstimulated saliva and for the E and N genes from nasopharyngeal swab specimens (NPS) (n = 34) and (B) ORF gene from self-collected unstimulated saliva and for the E and N genes from NPS (n = 24) Mean CT 95% CI 29.9 28.4; 31.6 25.8 23.5; 28.3 28.4 26.0; 30.9 Mean CT 95% CI 27.5 25.4; 29.7 24.1 21.5; 27.0 26.4 23.9; 29.1 Values expressed by estimated means and 95% confidence intervals (95% CI) We observed differences between the CT for N gene in saliva samples (CT NS) and the CT for gene E (CT E) in nasopharyngeal swab specimens, where the mean CT E was − 4.1 (95%CI: − 6.8 to − 1.5; p = 0.001). However, there was no statistical difference between CT for N gene (CT N) from self-collected unstimulated saliva specimens and nasopharyngeal swab (p = 0.183). For the nasopharyngeal swab, there was no statistical difference between the CT E (p = 0.080) and CT N (p = 0.496) values (Table 3).
Table 3

Differences between cycle thresholds obtained for the N (CT NS) and ORF (CT ORF) genes in self-collected unstimulated saliva and for the E (CT E) and N (CT NNP) genes in nasopharyngeal swab (NPS) specimens

SampleCT differenceEstimated mean difference (95% CI)p-value
SalivaCT E − CT NS − 4.1 (− 6.8; − 1.5)0.001
CT NNP − CT NS − 1.6 (− 3.9; 0.7)0.183
NPSCT E − CT ORF − 3.3 (− 7.0; 0.3)0.080
CT NNP − CT ORF − 1.1 (− 4.1; 2.0)0.496

All values are expressed by estimated means, with 95% CIs in parentheses

Differences between cycle thresholds obtained for the N (CT NS) and ORF (CT ORF) genes in self-collected unstimulated saliva and for the E (CT E) and N (CT NNP) genes in nasopharyngeal swab (NPS) specimens All values are expressed by estimated means, with 95% CIs in parentheses

Self-collected unstimulated saliva, oral swab, and nasopharyngeal swab specimens in the detection of SARS-CoV-2

Of the 55 patients, 21 had concomitantly CT values for E and N genes in nasopharyngeal swab specimens and for N gene in unstimulated saliva and oral swab specimens and 12 had concomitantly CT values for E and N genes in nasopharyngeal swab specimens and for ORF gene in unstimulated saliva and oral swab specimens. Table 4 shows the distribution of CT by the N and ORF genes in unstimulated saliva and oral swab and by the E and N (NNP) genes in nasopharyngeal swab.
Table 4

Mean values and 95% confidence intervals (95% CI) estimated for cycle threshold (CT) measurements for (A) N gene in unstimulated saliva and oral swab and by the E and N genes in nasopharyngeal swab specimens (NPS) (n = 21) and (B) ORF gene in unstimulated saliva and oral swab and by the E and N genes in NPS (n = 12)

Sample
Unstimulated salivaOral swabNasopharyngeal swabNasopharyngeal swab
AGene NGene NGene EGene N

Mean CT

95% CI

28.9

26.8; 31.1

33.3

31.5; 35.1

24.8

22.3; 27.5

27.7

25.0; 30.7

BGene ORFGene ORFGene EGene N

Mean CT

95% CI

24.3

21.8; 27.1

32.7

29.4; 36.3

23.4

20.3; 26.8

25.7

22.7; 29.2

Values expressed by estimated means and 95% confidence intervals (95% CI)

Mean values and 95% confidence intervals (95% CI) estimated for cycle threshold (CT) measurements for (A) N gene in unstimulated saliva and oral swab and by the E and N genes in nasopharyngeal swab specimens (NPS) (n = 21) and (B) ORF gene in unstimulated saliva and oral swab and by the E and N genes in NPS (n = 12) Mean CT 95% CI 28.9 26.8; 31.1 33.3 31.5; 35.1 24.8 22.3; 27.5 27.7 25.0; 30.7 Mean CT 95% CI 24.3 21.8; 27.1 32.7 29.4; 36.3 23.4 20.3; 26.8 25.7 22.7; 29.2 Values expressed by estimated means and 95% confidence intervals (95% CI) When comparing the CT values of the N gene in the oral swab to the CT values in the nasopharyngeal specimens, we observed differences in relation to the CT E (p < 0.001) and CT NNP, where the mean CT E was − 8.5 (95% CI: − 12.2 to − 4.7 and the mean CT NNP was − 5.6 (95% CI: − 9.2 to − 1.9; p = 0.001) lower than the mean CT N in oral swab. As for CT values for N gene, the mean CT in unstimulated saliva was − 4.4 (95% CI: − 6.3 to − 2.4; p < 0.001) lower than that in the oral swab. Comparing the CTs of the ORF gene in the oral swab to the CTs of the nasopharyngeal specimens, we observed evidence of differences in relation to the CT E and CT NNP, where the mean CT E was − 9.3 (95% CI: − 14.7 to − 3.9; p < 0.001) and the mean CT NNP was − 6.9 (95% CI: − 12.1 to − 1.8; p = 0.004) less than the mean of the CT ORF values in oral swab. Similarly, the mean CT value of the ORF gene in unstimulated saliva was − 8.3 (95% CI: − 11.5 to − 5.2; p < 0.001) less than the average of the value in the oral swab (Table 5).
Table 5

Differences between cycle threshold (CT) measurements obtained for the N and ORF genes in self-collected unstimulated saliva and oral swab and for the E and N genes in nasopharyngeal swab specimens (n = 21)

CT gene differences (sample source)Estimated mean difference (95% CI)p-value
CT E (NPS) − CT N (S) − 4.1 (− 7.0; − 1.2)0.003
CT N (NPS) − CT N (S) − 1.2 (− 3.8; 1.5)0.384
CT E (NPS) − CT N (OS) − 8.5 (− 12.2; − 4.7) < 0.001
CT N (NPS) − CT N (OS) − 5.6 (− 9.2; − 1.9)0.001
CT N (S) − CT N (OS) − 4.4 (− 6.3; − 2.4) < 0.001
CT E (NPS) − CT ORF (S) − 0.9 (− 3.7; 1.8)0.594
CT N (NPS) − CT ORF (S)1.4 (− 1.6; 4.4)0.594
CT E (NPS) − CT ORF (OS) − 9.3 (− 14.7; − 3.9) < 0.001
CT N2 (NPS) − CT ORF (OS) − 6.9 (− 12.1; − 1.8)0.004
CT ORF (S) − CT ORF (OS) − 8.3 (− 11.5; − 5.2) < 0.001

All values are expressed by estimated means, with 95% CIs in parentheses. CT E, CT value for E gene; CT N, CT value for N gene; CT ORF, CT value for ORF gene. Sample sources are indicated in parentheses as follows: (NPS), nasopharyngeal swab specimen; (S), saliva; (OS), oral swab

Differences between cycle threshold (CT) measurements obtained for the N and ORF genes in self-collected unstimulated saliva and oral swab and for the E and N genes in nasopharyngeal swab specimens (n = 21) All values are expressed by estimated means, with 95% CIs in parentheses. CT E, CT value for E gene; CT N, CT value for N gene; CT ORF, CT value for ORF gene. Sample sources are indicated in parentheses as follows: (NPS), nasopharyngeal swab specimen; (S), saliva; (OS), oral swab

Types of saliva sampling for SARS-CoV-2 detection

When comparing unstimulated self-collected saliva and oral swab, of the 37 patients with positive results for gene N in self-collected unstimulated saliva, 26 (70.3%) were also positive with the oral swab. Considering the positive results for the ORF gene in unstimulated saliva in 34 patients, 16 (47.1%) were also positive with the oral swab. Regardless of the gene, unstimulated saliva was positive in 37 patients, whereas the oral swab was positive only in 27 of the 37 (73.0%) samples.

Discussion

Saliva samples can be collected by a variety of methods, such as posterior oropharyngeal saliva, oral swab, and unstimulated saliva. However, the majority of studies used the term saliva failing to describe exactly which sample collection technique was used [6]. In the present study, we directly compared two saliva sampling techniques for SARS-CoV-2 detection. While self-collected unstimulated saliva showed a sensitivity of 87.3% when compared to nasopharyngeal swab, this result was decreased to 65.9% when the saliva samples were collected using oral swab. This finding is relevant in the current pandemic, as the use of a swab for saliva collection does not seem optimal. Using self-collected unstimulated saliva as the sampling method alleviates the need for swabs and healthcare workers. Thus, self-collection of saliva in simple, sterile, nuclease-free tubes, negating the high costs associated with specialized collection devices, could be one of the key obstacles to meet the mass testing demands [11]. Here, we have demonstrated that nasopharyngeal swab exhibits higher sensitivity for COVID-19 testing than unstimulated self-collected saliva. However, unstimulated self-collected saliva showed higher sensitivity than oral swab. While nasopharyngeal swab samples appear to be more suitable for COVID-19 testing, it is important to recognize that saliva-based COVID-19 testing is an important option for frequent/repeated testing [12], as being scalable, and provides a useful sustainable strategy that easily allows periodic testing over time [13]. The sensitivity of self-collected unstimulated saliva was reduced to 78.2% when using only results from ORF gene. Although RT-PCR has been used as the gold standard for diagnosing COVID-19, there are some difficulties involving its application [14]. The first RT-PCR kit made available by China’s CDC has as targets the N and ORF1ab genes of SARS-CoV-2 genome, and the infection is confirmed when both genes are amplified. However, there are reports of the amplification of only one of these gene targets [15]. Previous studies have reported a tendency for a higher viral load (or lower CTs) in nasopharyngeal swabs in comparison to saliva samples [16, 17]. However, in the present study, we found no significant differences between CT N in nasopharyngeal swab and in unstimulated saliva specimens (p = 0.183). This was not true when comparing CT N of nasopharyngeal swab to oral swab, where the latter presented a higher CT. The sensitivity of different diagnostic methods may vary according to variables such as gene target, sample processing and clinical signs and symptoms at the time of sample collection, as observed in the present study. Recent studies reported that SARS-CoV-2 could be detected in the saliva of asymptomatic individuals and outpatients [17-20], showing a higher sensitivity than nasopharyngeal swabs for diagnosis of asymptomatic and mild COVID-19 infection [21-23]. In the present study, we have only included inpatients, although we see great potential with saliva’s utility outside of the hospital setting. There are, of course, limitations in the present study. Primarily, only symptomatic inpatients testing positive for COVID-19 [24] were included in the current investigation. Furthermore, as molecular biology supplies were limited, two different COVID-19 RT-PCR kits were used. We also recognize the limited number of study participants who were able to provide all three different samples. Outside of a pandemic setting, where there are no limitations of resources, our future studies will include a larger group of patients, record the operational characteristics (sensitivity, specificity, precision, and so on) of COVID-19 diagnostic test, and comply with Standards for Reporting of Diagnostic Accuracy Studies (STARD) guidelines [25]. Finally, no biological tests have 100% sensitivity and 100% specificity, a fact which needs to be considered when diagnostic results are translated into clinical practice [26]. Future investigations should include a larger sample size and focus on the diagnostic capability of human saliva for identifying COVID-19 with new cost-effective point-of-care technologies, making diagnosis accessible for anyone and anywhere [27]. Self-collected saliva is quite comfortable for patients as well as being easy, low cost [28], and non-invasive with minimal equipment, and does not require a healthcare facility or professionals. It also minimizes the nosocomial transmission of SARS-CoV-2 to healthcare workers. In the current pandemic situation, all research centers, health agencies, and healthcare providers should explore this diagnostic opportunity and rapidly develop automated molecular point-of-care assays.

Conclusions

Unstimulated self-collected saliva samples have shown a higher agreement with the nasopharyngeal swab samples for SARS-COV-2 detection than that obtain with oral swab samples. Self-collection of unstimulated saliva samples is safe and alleviates demands for supplies of swabs, personal protective equipment, and healthcare personnel. Given the growing need for testing worldwide, our findings support the efficacy of saliva samples in the diagnosis of SARS-CoV-2 infection.
  17 in total

Review 1.  Saliva in the Diagnosis of COVID-19: A Review and New Research Directions.

Authors:  L L Fernandes; V B Pacheco; L Borges; H K Athwal; F de Paula Eduardo; L Bezinelli; L Correa; M Jimenez; N Dame-Teixeira; I M A Lombaert; D Heller
Journal:  J Dent Res       Date:  2020-09-16       Impact factor: 6.116

2.  The diagnostic potential of salivary protease activities in periodontal health and disease.

Authors:  K Thomadaki; Ja Bosch; Fg Oppenheim; Ej Helmerhorst
Journal:  Oral Dis       Date:  2013-02-04       Impact factor: 3.511

3.  COVID-19 epidemic in Switzerland: on the importance of testing, contact tracing and isolation.

Authors:  Marcel Salathé; Christian L Althaus; Richard Neher; Silvia Stringhini; Emma Hodcroft; Jacques Fellay; Marcel Zwahlen; Gabriela Senti; Manuel Battegay; Annelies Wilder-Smith; Isabella Eckerle; Matthias Egger; Nicola Low
Journal:  Swiss Med Wkly       Date:  2020-03-19       Impact factor: 2.193

4.  Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study.

Authors:  Kelvin Kai-Wang To; Owen Tak-Yin Tsang; Wai-Shing Leung; Anthony Raymond Tam; Tak-Chiu Wu; David Christopher Lung; Cyril Chik-Yan Yip; Jian-Piao Cai; Jacky Man-Chun Chan; Thomas Shiu-Hong Chik; Daphne Pui-Ling Lau; Chris Yau-Chung Choi; Lin-Lei Chen; Wan-Mui Chan; Kwok-Hung Chan; Jonathan Daniel Ip; Anthony Chin-Ki Ng; Rosana Wing-Shan Poon; Cui-Ting Luo; Vincent Chi-Chung Cheng; Jasper Fuk-Woo Chan; Ivan Fan-Ngai Hung; Zhiwei Chen; Honglin Chen; Kwok-Yung Yuen
Journal:  Lancet Infect Dis       Date:  2020-03-23       Impact factor: 25.071

5.  Diagnostic value and dynamic variance of serum antibody in coronavirus disease 2019.

Authors:  Yujiao Jin; Miaochan Wang; Zhongbao Zuo; Chaoming Fan; Fei Ye; Zhaobin Cai; Ying Wang; Huaizhong Cui; Kenu Pan; Aifang Xu
Journal:  Int J Infect Dis       Date:  2020-04-03       Impact factor: 3.623

6.  Saliva as a gold-standard sample for SARS-CoV-2 detection.

Authors:  Steph H Tan; Orchid Allicock; Mari Armstrong-Hough; Anne L Wyllie
Journal:  Lancet Respir Med       Date:  2021-04-19       Impact factor: 30.700

7.  Consistent Detection of 2019 Novel Coronavirus in Saliva.

Authors:  Kelvin Kai-Wang To; Owen Tak-Yin Tsang; Cyril Chik-Yan Yip; Kwok-Hung Chan; Tak-Chiu Wu; Jacky Man-Chun Chan; Wai-Shing Leung; Thomas Shiu-Hong Chik; Chris Yau-Chung Choi; Darshana H Kandamby; David Christopher Lung; Anthony Raymond Tam; Rosana Wing-Shan Poon; Agnes Yim-Fong Fung; Ivan Fan-Ngai Hung; Vincent Chi-Chung Cheng; Jasper Fuk-Woo Chan; Kwok-Yung Yuen
Journal:  Clin Infect Dis       Date:  2020-07-28       Impact factor: 9.079

Review 8.  Molecular immune pathogenesis and diagnosis of COVID-19.

Authors:  Xiaowei Li; Manman Geng; Yizhao Peng; Liesu Meng; Shemin Lu
Journal:  J Pharm Anal       Date:  2020-03-05

9.  Saliva is a reliable tool to detect SARS-CoV-2.

Authors:  Lorenzo Azzi; Giulio Carcano; Francesco Gianfagna; Paolo Grossi; Daniela Dalla Gasperina; Angelo Genoni; Mauro Fasano; Fausto Sessa; Lucia Tettamanti; Francesco Carinci; Vittorio Maurino; Agostino Rossi; Angelo Tagliabue; Andreina Baj
Journal:  J Infect       Date:  2020-04-14       Impact factor: 6.072

Review 10.  Unique epidemiological and clinical features of the emerging 2019 novel coronavirus pneumonia (COVID-19) implicate special control measures.

Authors:  Yixuan Wang; Yuyi Wang; Yan Chen; Qingsong Qin
Journal:  J Med Virol       Date:  2020-03-29       Impact factor: 20.693

View more
  4 in total

1.  Distinct Immunological Profiles Help in the Maintenance of Salivary Secretory IgA Production in Mild Symptoms COVID-19 Patients.

Authors:  Juliana de Melo Batista Dos Santos; Jonatas Bussador do Amaral; Carolina Nunes França; Fernanda Rodrigues Monteiro; Anuska Marcelino Alvares-Saraiva; Sandra Kalil; Edison Luiz Durigon; Danielle Bruna Leal Oliveira; Silvia Sanches Rodrigues; Debora Heller; Eliane Aparecida Rosseto Welter; João Renato Rebello Pinho; Rodolfo P Vieira; André Luis Lacerda Bachi
Journal:  Front Immunol       Date:  2022-05-24       Impact factor: 8.786

Review 2.  Saliva as a diagnostic specimen for SARS-CoV-2 detection: A scoping review.

Authors:  Yifei Wang; Akshaya Upadhyay; Sangeeth Pillai; Parisa Khayambashi; Simon D Tran
Journal:  Oral Dis       Date:  2022-04-21       Impact factor: 4.068

3.  Reverse-Transcription Loop-Mediated Isothermal Amplification Has High Accuracy for Detecting Severe Acute Respiratory Syndrome Coronavirus 2 in Saliva and Nasopharyngeal/Oropharyngeal Swabs from Asymptomatic and Symptomatic Individuals.

Authors:  Stephen P Kidd; Daniel Burns; Bryony Armson; Andrew D Beggs; Emma L A Howson; Anthony Williams; Gemma Snell; Emma L Wise; Alice Goring; Zoe Vincent-Mistiaen; Seden Grippon; Jason Sawyer; Claire Cassar; David Cross; Thomas Lewis; Scott M Reid; Samantha Rivers; Joe James; Paul Skinner; Ashley Banyard; Kerrie Davies; Anetta Ptasinska; Celina Whalley; Jack Ferguson; Claire Bryer; Charlie Poxon; Andrew Bosworth; Michael Kidd; Alex Richter; Jane Burton; Hannah Love; Sarah Fouch; Claire Tillyer; Amy Sowood; Helen Patrick; Nathan Moore; Michael Andreou; Nick Morant; Rebecca Houghton; Joe Parker; Joanne Slater-Jefferies; Ian Brown; Cosima Gretton; Zandra Deans; Deborah Porter; Nicholas J Cortes; Angela Douglas; Sue L Hill; Keith M Godfrey; Veronica L Fowler
Journal:  J Mol Diagn       Date:  2022-02-02       Impact factor: 5.341

4.  Simple and low-cost nucleic acid extraction methods for detection of SARS-CoV2 in self-collected saliva and dry oral swabs.

Authors:  J V Shwetha; Sneha K Chunchanur; T R Harsha; Silpa Mohandas; Pritik A Shah; R Ambica; Himabindhu Ks; M Sumanth
Journal:  IJID Reg       Date:  2022-09-18
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.