| Literature DB >> 34376969 |
Sheetal Brar1, Sri Ganesh1, Nikhil Rp1, Roopashree Cr1.
Abstract
PURPOSE: To evaluate the clinical outcomes and patient satisfaction after implantation of Optiflex Trio, a new trifocal intraocular lens (IOL) following cataract surgery.Entities:
Keywords: IOL; Optiflex; intraocular lens; spectacle independence; trifocal
Year: 2021 PMID: 34376969 PMCID: PMC8349194 DOI: 10.2147/OPTH.S320202
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Description of Optiflex Trio IOL Specifications
| Material | Hydrophobic Acrylic Containing Natural Chromophore |
|---|---|
| Optic type | Single piece, 360° square edge with aspheric optic, Diffractive- Refractive, Negative aspheric |
| Optic size | 6.00 mm |
| Haptic design | Optimized C haptic design |
| Overall size | 13.00 mm |
| Angulation | 0° |
| Diffractive surface | Anterior |
| Zone | 4mm Diffractive, Peripheral – Refractive Zone |
| Near Addition | +3.5D |
| Intermediate Addition | +1.85D |
| ACD | 5.28 mm |
| Refractive index | 1.48 |
| Dioptre range | +7.00 to +30.00 D in 0.5 D steps |
| A constant (Optical) | 118.5 |
| Light distribution | 45% for Far, 27% for Intermediate and 28% for near distances |
| Abbe Number | 49 |
| Injector | Optiject BES28 Injector system |
| Implantation site | Capsular bag |
| Sterilization | Irradiation |
Figure 1Representative image showing the design of the Optiflex Trio trifocal IOLs.
Demographics and Baseline Pre-Operative Parameters of All Eyes (n = 54) Included in the Study
| Parameter | Mean± SD |
|---|---|
| Age (years) | 66.30±7.48 |
| Male: Female | 12 Males, 15 Females |
| K1 (D) | 43.51±1.39 |
| K2(D) | 44.26±1.47 |
| Astigmatism (D) | 0.74±0.38 |
| Axial length (mm) | 23.35±0.75 |
| ECD (cells/mm2) | 2618.59±186.19 |
| IOL Power (D) | 21.17±1.611 |
Abbreviations: K, keratometry; D, dioptre; ECD, endothelial cell density; IOL, intraocular lens; SD, standard deviation.
Visual and Refractive Outcomes at 1 Week, 1 Month, 3 Months, 6 Months, and 12 Months Post-Op
| Parameter (Mean± SD) | 1-Week | 1-Month | 3-Months | 6-Months | 12 -Months | p-value |
|---|---|---|---|---|---|---|
| UDVA (LogMAR) | 0.06±0.09 | 0.06±0.09 | 0.04±0.09 | 0.04±0.08 | 0.04±0.08 | 0.54 |
| CDVA (LogMAR) | −0.00±0.04 | −0.01±0.04 | −0.02±0.04 | −0.02±0.04 | −0.02±0.04 | 0.56 |
| Sph (D) | −0.07±0.28 | −0.07±0.22 | −0.07±0.19 | −0.07±0.19 | −0.07±0.19 | 0.99 |
| Cyl (D) | −0.26±0.32 | −0.24±0.28 | −0.21±0.27 | −0.19±0.24 | −0.18±0.23 | 0.52 |
| SE (D) | −0.20±0.31 | −0.19±0.26 | −0.17±0.23 | −0.17±0.22 | −0.16±0.21 | 0.95 |
| Binocular UNVA (LogMAR) | 0.02±0.05 | 0.01±0.05 | 0.01±0.05 | 0.01±0.05 | 0.01±0.05 | 0.97 |
| Binocular DCNVA (LogMAR) | −0.00±0.03 | −0.01*±0.03 | −0.01±0.03 | −0.01±0.03 | −0.01±0.03 | 0.88 |
| Binocular UDVA (LogMAR) | 0.00±0.07 | 0.00±0.07 | 0±0.07 | 0±0.07 | −0.00±0.07 | 0.95 |
| Binocular CDVA (LogMAR) | −0.05±0.05 | −0.05±0.05 | −0.06±0.05 | −0.06±0.04 | −0.06±0.04 | 0.96 |
Abbreviations: UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; Sph, sphere; Cyl, cylinder; SE, spherical equivalent; IOL, intraocular lens; UNVA, uncorrected near visual acuity; DCNVA, distance corrected near visual acuity; D, dioptre; SD, standard deviation.
Figure 2Histogram showing binocular results for UDVA and CDVA obtained following implantation of Optiflex Trio IOL at 12 months post-operatively.
Figure 3Binocular uncorrected near vision results at 12 months post-op.
Figure 4Histogram showing the accuracy to the intended spherical equivalent refraction (A) and refractive astigmatism (B) at 12 months post-op.
Binocular Near and Intermediate Visual Acuity with ETDRS Charts and Reading Speeds at 40, 60, and 80 Cm and CS with CSV-1000 at Post-Operative Visits from 1-Week Through 12 Months
| Mean±SD | 1 Week | 1 Month | Months | 6 Months | 1 Year | p-value* | ||
|---|---|---|---|---|---|---|---|---|
| 0.02±0.05 | 0.01±0.05 | 0.01±0.05 | 0.01±0.05 | 0.01±0.05 | 0.97 | |||
| −0.00±0.03 | −0.01±0.03 | −0.01±0.03 | −0.01±0.03 | −0.01±0.03 | 0.88 | |||
| 0.08 | 0.06 | 0.06 | 0.06 | 0.06 | ||||
| 0.08±0.07 | 0.08±0.07 | 0.07±0.06 | 0.07±0.06 | 0.07±0.06 | 0.93 | |||
| 0.07±0.06 | 0.07±0.06 | 0.06±0.05 | 0.06±0.05 | 0.06±0.05 | 0.85 | |||
| 0.54 | 0.54 | 0.49 | 0.49 | 0.49 | ||||
| 0.04±0.05 | 0.04±0.05 | 0.03±0.05 | 0.03±0.05 | 0.03±0.05 | 0.94 | |||
| 0.03±0.05 | 0.03±0.05 | 0.02±0.04 | 0.02±0.04 | 0.02±0.04 | 0.93 | |||
| 0.58 | 0.58 | 0.55 | 0.55 | 0.55 | ||||
| 162±40 | 163±40 | 164±40 | 164±40 | 165±40 | 0.99 | |||
| 163±40 | 165±40 | 166±40 | 166±40 | 167±40 | 0.99 | |||
| 0.86 | 0.85 | 0.85 | 0.85 | 0.86 | ||||
| 167±31 | 168±33 | 169±33 | 169±33 | 170±33 | 0.98 | |||
| 169±31 | 170±33 | 171±33 | 171±33 | 172±33 | 0.98 | |||
| 0.76 | 0.76 | 0.77 | 0.77 | 0.76 | ||||
| 170±31 | 171±33 | 172±34 | 172±34 | 173±34 | 0.99 | |||
| 172±31 | 173±34 | 174±34 | 174±34 | 175±34 | 0.98 | |||
| 0.78 | 0.78 | 0.82 | 0.82 | 0.85 | ||||
| 1.55±0.13 | 1.56±0.13 | 1.57±0.13 | 1.57±0.13 | 1.58±0.14 | 0.94 | |||
| 1.90±0.08 | 1.91±0.08 | 1.92±0.08 | 1.93±0.07 | 1.94±0.07 | 0.15 | |||
| 1.54±0.08 | 1.54±0.08 | 1.55±0.09 | 1.56±0.08 | 1.57±0.06 | 0.43 | |||
| 1.05±0.06 | 1.06±0.06 | 1.07±0.05 | 1.07±0.05 | 1.08±0.05 | 0.09 | |||
Notes: P value** calculated using Independent t-test. P value* calculated using Anova single factor test.
Abbreviations: SRD, Salzburg Reading Desk; wpm, words per minute; CSV, contrast sensitivity (Vector Vision); SD, standard deviation.
Figure 5Photopic CS evaluated binocularly (with correction) at 12 months.
Figure 6Binocular distance corrected defocus curve evaluated from +2 to −4 D defocus at 12 months.
Comparison of the Design and Clinical Outcomes Between Optiflex Trio and Published Results of Currently Available Trifocal Intraocular Lenses
| Optiflex Trio (Biotech Healthcare Holding GmbH, Luzren, Switzerland) | Zeiss AT LISA Tri 839MP (Carl Zeiss Meditec, Germany) | AcrySof IQ PanOptix (Alcon Laboratories, Switzerland) | RayOne Trifocal (Rayner Intraocular Lenses Ltd, England, UK) | Fine Vision POD F/ Micro F (PhysIOL, BVI, Belgium) | ||
|---|---|---|---|---|---|---|
| Optical Design | Diffractive- Refractive | Diffractive | Diffractive- Refractive | Diffractive- Refractive | Diffractive | |
| Optic material | Hydrophobic acrylic containing natural chromophore | Hydrophilic acrylic with hydrophobic surface properties | Hydrophobic acrylic | Hydrophilic acrylic | Hydrophilic acrylic | |
| Abbe number | 49 | 58 | 37 | 56 | 58 | |
| Spherical aberrations | −0.20µm | −0.18 µm | −0.10 µm | Neutral SA | −0.11 µm | |
| Number of diffractive rings | 12 rings on anterior surface. Monofocal beyond 4mm | 29 rings anteriorly on entire optic surface. Bifocal beyond 4.34 mm | 15 rings on 4.5mm anterior optic surface. Monofocal beyond 4.5mm | 16 rings on 4.5mm anterior optic surface. Monofocal beyond 4.5mm | 26 rings on entire optic surface | |
| Addition | +1.85D I(72cm)/+3.5D N(38 cm) | +1.66D I(80cm)/ +3.33D N (40cm) | +2.17D I(62cm)/+3.25D N(42cm) | +1.75D I (75cm)/+3.50D N (37.5cm) | +1.75D I (75cm) /+3.5D N (37.5 cm) | |
| Light Utilization | 88.3% | 85.7% | 88% | 89% | 86% | |
| Light distribution | 45% D/27% I/28% N | 50% D/20%I/30% N | 42% D/ 24% I/ 22% N | 52% D/22% I /26% N | 49% D/ 17% I/ 34% N | |
| Binocular UDVA (LogMAR) | Present study: −0.00±0.07 | Ganesh et al | Ribeiro et al | Ribeiro et al | Ribeiro et al | |
| Mencucci et al | Mencucci et al | Ferreira et al | Ferreira et al | |||
| Doroodgar et al | Doroodgar et al | Marques et al | ||||
| Binocular CDVA (LogMAR) | Present study: −0.06±0.04 | Ganesh et al | Ribeiro et al | Ribeiro et al | Ribeiro et al | |
| Mencucci et al | Mencucci et al | Ferreira et al | Ferreira et al | |||
| Doroodgar et al | Doroodgar et al | Marques et al | ||||
| Binocular UIVA (LogMAR) | Present study: 1) 60 cm: 0.07±0.06 | Doroodgar et al | Ribeiro et al | Ribeiro et al | Ribeiro et al | |
| 2) 80cm: 0.03±0.05 | Kim et al | Mencucci et al | Ferreira et al | Ferreira et al | ||
| Asena et al | Doroodgar et al | Marques et al | ||||
| Binocular UNVA (LogMAR) | Present study: 40 cm 0.01±0.05 | Ganesh et al | Ribeiro et al | Ribeiro et al | Ribeiro et al | |
| Doroodgar et al | Doroodgar et al | Ferreira et al | Ferreira et al | |||
| Kim et al | Asena et al | Marques et al | ||||
| Reading speed (wpm, uncorrected) | 40cm | Present study: 165±40 | - | - | - | - |
| 60cm | Present study: 170±33 | Mencucci et al: | Mencucci et al: | - | - | |
| 80cm | Present study: 173±34 | Ganesh et al: | - | - | - | |
Abbreviations: D, distance; I, intermediate distance; N, near distance; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; UIVA, uncorrected intermediate visual acuity; UNVA, uncorrected near visual acuity; wpm, words per minute.