Literature DB >> 34372464

Investigation and Optimization of a Line-Locked Quartz Enhanced Spectrophone for Rapid Carbon Dioxide Measurement.

Hui Zhang1,2, Wenling Jin3, Mengpeng Hu1,2, Mai Hu1, Jingqiu Liang1, Qiang Wang1,2.   

Abstract

We have developed a rapid quartz enhanced spectrophone for carbon dioxide (CO2) measurement, in which the laser wavelength was tightly locked to a CO2 absorption line and a custom quartz tuning fork (QTF) operating at 12.5 kHz was employed. The intrinsic QTF oscillation-limited response time, as well as the optimal feedback interval, was experimentally investigated. By tightly locking the laser to the R(16) transition of CO2, we obtained a stable laser operation with its center wavelength variation kept within 0.0002 cm-1, merely three times the laser linewidth. The reported CO2 sensor achieved a detection limit of 7 ppm, corresponding to a normalized noise equivalent absorption coefficient (NNEA) of 4.7 × 10-9 W·cm-1·Hz-1/2, at a response time of 0.5 s. The detection limit can be further improved to 0.45 ppm at an integration time of 270 s, illustrating a good system stability. This spectrophone enables the realization of compact and fast-response gas sensors for many scenarios, where CO2 concentration from sub-ppm to hundreds of thousands of ppm is expected.

Entities:  

Keywords:  carbon dioxide; custom quartz tuning fork; laser spectroscopy; quartz enhanced photoacoustic spectroscopy; wavelength locking

Year:  2021        PMID: 34372464     DOI: 10.3390/s21155225

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


  1 in total

1.  Frequency-Domain Detection for Frequency-Division Multiplexing QEPAS.

Authors:  Xiang Chen; Hao Liu; Mai Hu; Lu Yao; Zhenyu Xu; Hao Deng; Ruifeng Kan
Journal:  Sensors (Basel)       Date:  2022-05-26       Impact factor: 3.847

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.