Figure 1Method of data collection and parameters. The pressure and flow velocity were recorded simultaneously via a catheter for angiography and with transesophageal echocardiography. Pairs of corresponding waveforms were traced, and each parameter was measured (see also Video 1). TEE, Transesophageal echocardiography.
Figure 2Changes in aortic stiffness and cardiac workload following stent graft deployment. A, The parameter of aortic stiffness, ΔP/∫V dt, was significantly increased. B, The parameter of cardiac workload, peak P × peak V, was also increased significantly. The red lines indicate median values.