| Literature DB >> 34345606 |
Ishtiaque Ahmad1, Zabih Ullah2, Mohd Ibrahim Khan3, Abdullah Khaloofa Alahmari4, Mohd Faiyaz Khan4.
Abstract
A wide-range, specific, and precise liquid chromatography tandem mass spectrometric (LC-MS/MS)technique for quantifying fluoxetine (FLX) in human plasma was developed using the RapidTrace® automated solid-phase extraction (SPE) method; the analyte and internal standard (IS) were extricated on Oasis MCX SPE cartridges. Acetonitrile and 5 mM ammonium formate buffer (90:10 v/v) were used as mobile phase to achieve chromatographic separation on the reverse phase (C18 column). The analyte and IS were ionized using +ve electrospray ionization approach which was further traced by multiple-reaction monitoring on a tandem mass spectrometer. To quantify the FLX and FLX-d5, the parent-to-daughter ion transition of m/z of 310.0/44.1 and 315.0/44.0 was used, respectively. The method demonstrated a linear active limit of 0.20-30 ng/ml with recoveries ranging from 63.04% to 79.39% for quality control samples and 61.25% for IS samples. The concentrations over the calibration range demonstrated acceptable precision and accuracy. Due to the high inconsistency of the FLX concentration data, the minimum threshold of the assay was kept at 0.20 ng/ml. The flow rate was maintained at 500 μL/min, and the time for sample analysis for each injection was 3.5 min. The method was found to be specific, sensitive, and faster with minimum utilization of organic solvents and was utilized further for metabolic and pharmacokinetic studies. Copyright:Entities:
Keywords: Automation; fluoxetine; human plasma; mass spectrometry; solid-phase extraction
Year: 2021 PMID: 34345606 PMCID: PMC8300324 DOI: 10.4103/japtr.JAPTR_308_20
Source DB: PubMed Journal: J Adv Pharm Technol Res ISSN: 0976-2094
Figure 1Chemical structure of fluoxetine
Extraction procedure (RapidTrace®)
| Steps | Sources | Destination | Volumes (mL) | Flow rate (ml/min) |
|---|---|---|---|---|
| Conditioning | Methanol | Organic waste | 1 | 15 |
| Conditioning | Water | Aqueous waste | 1 | 15 |
| Loading | Sample | Biowaste | 0.65 | 1 |
| Purge cannula | Water | Cannula waste | 2 | 30 |
| Rinsing | Washing solution | Biowaste | 1 | 2 |
| Rinsing | Washing solution | Biowaste | 1 | 2 |
| Rinsing | Methanol | Biowaste | 1 | 2 |
| Rinsing | Methanol | Biowaste | 1 | 2 |
| Drying | Time (5 min) | |||
| Collection | Elution solution | Fraction-1 | 1 | 30 |
| Collection | Elution solution | Fraction-1 | 1 | 30 |
| Purge cannula | Methanol | Cannula waste | 2 | 30 |
| Purge cannula | Water | Cannula waste | 2 | 30 |
Summary of chromatographic and mass spectrometric conditions
| Parameter | Optimized Condition |
|---|---|
| Column name | Ascentis Express C18 column (75 mm×4.6 mm×2.7 µm) |
| Mobile phase | Acetonitrile: 5.0 mM ammonium formate (90:10 v/v) |
| Oven temperature | 40 |
| Injected volume (µl) | 10 |
| Flow rate (mL/min) | 0.5 |
| Detector | MS/MS |
| Gas-1 (nebulizer) | 50 |
| Gas-2 (heater gas) | 50 |
| Curtain gas | 18 |
| Collision gas | 6 |
| Ion source | Turbo-ion spray (positive ion mode) |
| Temperature | 450 |
| IS voltage | 4500 V |
| DE clustering potential | 18 |
| Entrance potential | 10 |
| Collision energy | 10 |
| collision cells exit potential | 9 |
| Dwell time (ms) | 200 |
| M/z ion fluoxetine | 310.0/44.1 |
| M/z fluoxetine-d5 | 315.0/44.0 |
| Sample cooler temperature | 10±2 |
| Rinsing solution | Methanol: Water (50:50 v/v) |
| Retention time (fluoxetine), min | 1.5-1.8 |
| Right (fluoxetine d5), min | 1.5-1.8 |
| Run time (min) | 3.5 |
IS: Internal standard, MS: Mass Spectrometry
Blank screening and selectivity
| Blank ID | Interference fluoxetine | LOQ area-fluoxetine (%) | Interference fluoxetine-d5 | IS area (%) |
|---|---|---|---|---|
| B-01 | 83 | 16,283 (0.50) | 80 | 1,462,938 (0.005) |
| B-02 | 62 | 17,995 (0.30) | 90 | 1,398,671 (0.006) |
| B-03 | 2 | 15,388 (0.00) | 63 | 1,351,752 (0.005) |
| B-04 | 31 | 14,124 (0.20) | 23 | 1,230,928 (0.002) |
| B-05 | 20 | 14,850 (0.10) | 50 | 1,434,412 (0.003) |
| B-06 | 109 | 14,036 (0.80) | 211 | 1,132,306 (0.019) |
| B-07 | 12 | - | 300 | |
| B-08 | 6 | - | 230 | |
| B-09 | 78 | - | 150 | |
| B-10 | 45 | - | 186 | |
| Mean | 15,446 | 1,335,168 | ||
| SD | 1502.820 | 128,453 | ||
| Percentage CV | 9.730 | 9.6 |
IS: Internal standard, SD: Standard deviation, LOQ: Limit of quantitation, CV: Coefficient of variation
Figure 2Calibration curve of fluoxetine concentration versus area ratio of drug/internal standard
Intraday and interday precision and accuracy of fluoxetine
| Nominal concentration (ng/mL) | Intraday ( | Inter day ( | ||||
|---|---|---|---|---|---|---|
| Mean concentration±SD | Precision (%) | Accuracy (%) | Mean concentration±SD | Precision (%) | Accuracy (%) | |
| 0.203 (LOQQC) | 0.203±0.01 | 7.26 | 99.92 | 0.192±0.01 | 7.7 | 94.69 |
| 0.583 (LQC) | 0.603±0.01 | 1.92 | 108.06 | 0.629±0.02 | 2.57 | 107.8 |
| 12.677 (MQC) | 12.405±0.32 | 2.58 | 97.93 | 12.543±0.38 | 3.02 | 99.02 |
| 25.353 (HQC) | 25.218±0.51 | 2.02 | 99.47 | 25.018±0.61 | 2.42 | 98.68 |
SD: Standard deviation, LOQQC: Lowest limit of quantification quality control, LQC: Low quality control, MQC: Medium quality control, HQC: High quality control
Figure 3Representative chromatograms of blank (left) and blank internal standard (right) in plasma samples
Figure 6Representative multiple reaction monitoring chromatograms of upper level of quantification samples in plasma: fluoxetine (left) and fluoxetine-d5 (right)
Matrix effect and recovery for fluoxetine and fluoxetine d5
| QC | A (CV %) | B (CV %) | C (CV %) | Absolute ME (%) | RE (%) |
|---|---|---|---|---|---|
| LQC | 45142.20 (4.6) | 44423.70 (8.66) | 35268.20 (6.51) | 98.41 | 79.39 |
| MQC | 974522.10 (4.64) | 960072.67 (2.48) | 609186.83 (6.52) | 98.52 | 63.45 |
| HQC | 1915432.40 (2.99) | 1905052.00 (2.90) | 1200998.20 (4.10) | 99.46 | 63.04 |
| ISD | 1234584.60 (7.39) | 1229850.70 (0.70) | 753278.20 (4.25) | 99.62 | 61.25 |
QC: Quality control, ME: Matrix effect, LQC: Low quality control, MQC: Medium quality control, HQC: High quality control, RE: Relative recovery, ISD: Internal standard dilution, CV: Coefficient of variation
Stability data for fluoxetine and fluoxetine-d5
| Storage condition | QC (ng/mL) | Precision (%) | Accuracy (%) | Stability (%) |
|---|---|---|---|---|
| Benchtop stability (14.50 h) | LQC | 4.35 | 96.10 | 101.44 |
| HQC | 2.52 | 100.71 | 98.14 | |
| Benchtop extraction stability (14.53 h) | LQC | 2.38 | 98.54 | 98.37 |
| HQC | 1.94 | 102.96 | 102.52 | |
| Long-term stability (15 days) | LQC | 3.93 | 95.31 | 96.52 |
| HQC | 1.34 | 102.50 | 98.97 | |
| Freeze-thaw stability (3 cycles) | LQC | 2.81 | 102.8 | 95.45 |
| HQC | 2.70 | 99.30 | 101.27 | |
| In-injector stability (52.50 h) | LQC | 5.29 | 94.16 | 96.35 |
| HQC | 2.74 | 95.21 | 93.05 | |
| Reinjection reproducibility (52.50 h) | LQC | 1.21 | 91.00 | 98.61 |
| HQC | 0.82 | 98.27 | 101.46 | |
| Short-term stability of reference standard (9.50 h) | 0.87 | 100.24 | ||
| Stock solution stability of fluoxetine (7 day) | 2.32 | 94.80 | ||
| Stock solution stability of fluoxetine-d5 (24 h) | 0.44 | 100.15 |
QC: Quality control, LQC: Low quality control, HQC: High quality control
Precision and accuracy batch for the ruggedness
| Nominal concentration (ng/mL) | Mean±SD | Precision (%) | Accuracy (%) |
|---|---|---|---|
| 0.203 (LOQQC) | 0.192±0.01 | 2.82 | 94.33 |
| 0.583 (LQC) | 0.615±0.02 | 3.95 | 105.55 |
| 12.677 (MQC) | 12.847±0.12 | 0.93 | 101.34 |
| 25.353 (HQC) | 25.480±0.04 | 0.16 | 100.51 |
SD: Standard deviation, LOQQC: Lowest limit of quantification quality control, LQC: Low quality control, MQC: Medium quality control, HQC: High quality control