Ian R Macdonald1, Elena A Cora1, Ian Grant2, David Volders1. 1. Division of Neuroradiology, Department of Diagnostic Radiology, 3688Dalhousie University, Canada. 2. Division of Neurology, Department of Medicine, Dalhousie University, Canada.
Abstract
Endovascular thrombectomy (EVT) is part of first-line intervention for acute ischemic stroke management. Recent technological advances have demonstrated that large-bore catheters are an attractive approach for EVT. A multitude of approaches such as A Direct Aspiration first Pass Technique (ADAPT) or in conjunction with stent retrieval (Solumbra technique) have been developed with increasingly large-bore catheters, demonstrating safety and efficacy. Furthermore, these techniques have demonstrated promise for the intervention of cerebral venous thrombosis as well as posterior circulation ischemic events. Recently, advances in neurointerventional catheters have focused on improved maneuverability to navigate the neurovasculature, as well as larger inner diameters for improved procedural versatility, including aspiration. We describe a case report highlighting our early institutional experience with the recently developed large-bore catheter, the BENCHMARK™ BMX™ 96. The case report entails near complete occlusion of the internal carotid artery from acute thrombus and the utility of the BMX™ 96 catheter for treatment of such extensive clot burden. The applicability of large-bore aspiration catheters, with an emphasis on recent advances, for mechanical thrombectomy in arterial as well as venous systems is discussed. To our knowledge, this is the first reported case of use of the BENCHMARK™ BMX™ 96 access system for EVT in acute ischemic stroke. Such new-generation large-bore catheters are a promising advance in neurointervention, and our early institution experience highlights the ease of use and versatility for neurointerventional procedures such as EVT.
Endovascular thrombectomy (EVT) is part of first-line intervention for acute ischemic stroke management. Recent technological advances have demonstrated that large-bore catheters are an attractive approach for EVT. A multitude of approaches such as A Direct Aspiration first Pass Technique (ADAPT) or in conjunction with stent retrieval (Solumbra technique) have been developed with increasingly large-bore catheters, demonstrating safety and efficacy. Furthermore, these techniques have demonstrated promise for the intervention of cerebral venous thrombosis as well as posterior circulation ischemic events. Recently, advances in neurointerventional catheters have focused on improved maneuverability to navigate the neurovasculature, as well as larger inner diameters for improved procedural versatility, including aspiration. We describe a case report highlighting our early institutional experience with the recently developed large-bore catheter, the BENCHMARK™ BMX™ 96. The case report entails near complete occlusion of the internal carotid artery from acute thrombus and the utility of the BMX™ 96 catheter for treatment of such extensive clot burden. The applicability of large-bore aspiration catheters, with an emphasis on recent advances, for mechanical thrombectomy in arterial as well as venous systems is discussed. To our knowledge, this is the first reported case of use of the BENCHMARK™ BMX™ 96 access system for EVT in acute ischemic stroke. Such new-generation large-bore catheters are a promising advance in neurointervention, and our early institution experience highlights the ease of use and versatility for neurointerventional procedures such as EVT.
Entities:
Keywords:
A Direct Aspiration first Pass Technique (ADAPT); BENCHMARK™ BMX™ 96; Interventional neuroradiology; endovascular thrombectomy
Endovascular thrombectomy (EVT) is a mainstay of acute ischemic stroke management and is
considered part of first-line intervention as supported by recent guidelines,
with a diverse range of techniques described.
Continued recent technological advances have positioned increasingly large-bore
catheters as an attractive approach for EVT in ischemic stroke. These catheters are
typically used either in the setting of A Direct Aspiration first Pass Technique (ADAPT)
or in conjunction with stent retrieval (Solumbra technique).
Stand-alone large-bore catheter aspiration has also shown efficacy for cerebral
venous thrombosis
as well as posterior circulation EVT.
Recent advances in neurointerventional catheters have been focused on improved system
maneuverability and conformability to optimize navigation of the neurovasculature.
Furthermore, larger inner diameters (ID) improve procedural versatility, such as aspiration,
which has translated into safe and efficacious ADAPT approaches for acute ischemic
stroke.[7,8] Recent clinical trials (COMPASS
and ASTER
) have positioned ADAPT as an attractive first-line approach for EVT. We describe a
case report highlighting our early institutional experience with the recently developed
large-bore aspiration catheter, the BENCHMARK™ BMX™ 96 (Penumbra, Inc., Alameda, CA).
Case
A 66-year-old female presented to an outside hospital three hours after acute stroke
symptom onset with right hemiplegia and aphasia, with a National Institutes of Health Stroke
Scale (NIHSS) of 12. A non-contrast computed tomography (NCCT) scan of the head confirmed
preserved gray–white differentiation, with no signs of early infarction or hemorrhage
(Alberta Stroke Program Early CT Score (ASPECTS) 10). A CT angiogram (CTA) of the arch to
the vertex revealed absent contrast filling from the mid-portion of the left common carotid
artery (CCA) to the supraclinoid segment of the left internal carotid artery (ICA). The left
middle cerebral artery (MCA) circulation was preserved secondary to collateral flow via the
anterior communicating artery. Intravenous thrombolytic therapy was contraindicated, as the
patient was taking warfarin for atrial fibrillation (International Normalized Ratio 2.3).
Given the patient’s excellent ASPECTS score, good collaterals, and moderate stroke symptoms
(a NIHSS score of 12), the patient was transported to our comprehensive stroke center for
emergency EVT consideration. After arrival, approximately four hours later, a repeat NCCT
scan of the head and CTA of the arch to the vertex showed no interval change, with no signs
of early infarction (ASPECTS 10) and persistent left ICA occlusion (Figure 1). Additional CT perfusion (CTP) of the brain
showed decreased perfusion to the left MCA territory (Figure 1). Given the persistent moderate stroke
symptoms (a NIHSS score of 12) and the significant penumbra on imaging, the patient was
urgently transferred to the angiography suite for an emergency EVT.
Figure 1.
Computed tomography (CT) scan of the head, time-to-peak perfusion map, and CT angiogram
study. (a) Select unenhanced axial CT head slice, demonstrating no evidence of infarct
(ASPECTS 10). (b) Corresponding time-to-peak perfusion map, demonstrating diffuse
abnormality in the left middle cerebral artery territory. (c) Coronal maximum intensity
projection (MIP), demonstrating occlusion of the left internal carotid artery (ICA;
white arrow). (d) Axial MIP, demonstrating persistent filling of the left middle
cerebral artery (black arrow) secondary to collateral flow. Select axial (e) and coronal
(f) CT angiogram slices, demonstrating slow flow at the level of the mid-left common
carotid artery (CCA; white arrows). Note that the patient clinically demonstrated right
hemiparesis as well as aphasia.
Computed tomography (CT) scan of the head, time-to-peak perfusion map, and CT angiogram
study. (a) Select unenhanced axial CT head slice, demonstrating no evidence of infarct
(ASPECTS 10). (b) Corresponding time-to-peak perfusion map, demonstrating diffuse
abnormality in the left middle cerebral artery territory. (c) Coronal maximum intensity
projection (MIP), demonstrating occlusion of the left internal carotid artery (ICA;
white arrow). (d) Axial MIP, demonstrating persistent filling of the left middle
cerebral artery (black arrow) secondary to collateral flow. Select axial (e) and coronal
(f) CT angiogram slices, demonstrating slow flow at the level of the mid-left common
carotid artery (CCA; white arrows). Note that the patient clinically demonstrated right
hemiparesis as well as aphasia.General anesthesia was induced, and the patient was prepped in standard fashion. After
right common femoral artery access, a co-axial system of a 90 cm BENCHMARK™ BMX™ 96 access
catheter and 125 cm 6F Bern select catheter (Penumbra, Inc.) was easily maneuvered into the
mid left CCA over a Terumo Advantage guidewire (Terumo Medical Canada, Inc., Vaughan,
Canada). Angiographic runs demonstrated near occlusion of the carotid bifurcation with a
string sign and large amount of thrombus throughout the left ICA (Figure 2(a)). The left ICA was successfully selected,
and the BMX™ 96 catheter was positioned in the mid-cervical ICA, with no issues navigating
the severe bifurcation stenosis. No spontaneous back flow was noted from the BMX™ 96 after
removal of the hemostatic valve. Aspiration tubing paired with a ENGINE aspiration pump
(Penumbra, Inc.) was connected to the BMX™ 96, and aspiration was applied for two minutes.
Subsequently, 13 cm of thrombus was spontaneously aspirated through the BMX™ 96 and its
connecting tube to the pump without manipulating the catheter (Figure 3). A subsequent angiogram showed complete
recanalization of the left ICA, with no evidence of distal emboli (TICI 3; Figure 2(b)). The time from groin
puncture to first recanalization was 15 minutes. The BMX™ 96 was subsequently pulled back to
the proximal cervical ICA (Figure
2(c)), and aspiration at this level yielded additional small pieces of thrombus.
The patient was loaded with 325 mg aspirin and 300 mg clopidogrel through an OG tube.
Carotid bifurcation stenting using a tapered 8–6 mm × 40 mm EV3 Protégé RX stent (Medtronic,
Minneapolis, MN) and balloon angioplasty using an Armada 18 4 mm × 20 mm balloon (Abbott,
Plymouth, MN) was successfully performed. A control left CCA angiogram showed satisfactory
result with no residual stenosis (Figure
2(d)). Hemostasis of the right groin was achieved through manual compression for 30
minutes. The patient was extubated and transported to our recovery room, with subsequent
transfer to home hospital several hours later. Twenty-four hours post procedure, the patient
had neurologically returned to normal, with no residual weakness or aphasia (a NIHSS score
of 0).
Figure 2.
Endovascular thrombectomy (EVT) and stent deployment. (a) Anterior-posterior digital
subtraction angiogram with BMX™ 96 catheter positioned in the mid-left CCA,
demonstrating severe stenosis at the left CCA bifurcation (black arrow), with extensive
near occlusive thrombus throughout the left ICA (black arrowhead). (b) Post aspiration
thrombectomy. demonstrating widely patent left ICA with no distal intracranial emboli
(TICI 3; white arrow). (c) Persistent severe left carotid bifurcation stenosis post EVT
(black arrow). (d) Angiogram post stent and angioplasty, demonstrating patency of the
left carotid bifurcation (white arrowhead).
Figure 3.
Thrombectomy specimen. Photo of the aspirated left common carotid/ICA thrombus
measuring 13 cm in length, obtained after one pass of the BECNHMARK™ BMX™ 96 aspiration
catheter.
Endovascular thrombectomy (EVT) and stent deployment. (a) Anterior-posterior digital
subtraction angiogram with BMX™ 96 catheter positioned in the mid-left CCA,
demonstrating severe stenosis at the left CCA bifurcation (black arrow), with extensive
near occlusive thrombus throughout the left ICA (black arrowhead). (b) Post aspiration
thrombectomy. demonstrating widely patent left ICA with no distal intracranial emboli
(TICI 3; white arrow). (c) Persistent severe left carotid bifurcation stenosis post EVT
(black arrow). (d) Angiogram post stent and angioplasty, demonstrating patency of the
left carotid bifurcation (white arrowhead).Thrombectomy specimen. Photo of the aspirated left common carotid/ICA thrombus
measuring 13 cm in length, obtained after one pass of the BECNHMARK™ BMX™ 96 aspiration
catheter.
Discussion
This case report highlights the use of a BMX™ 96 catheter in the setting of extensive
arterial clot burden. The BMX™ 96 has hydrophilic coating along the distal 11 cm, as well as
progressive laser-cut patterns along the entire length, contributing to the distal ease of
maneuverability at the level of the aortic arch and cervical ICA while balancing proximal
support. This can reduce the possible need to perform an exchange maneuver and therefore
translates into reduced procedure times. The BMX™ 96 has an 8 Fr outer diameter (OD; 0.109
inch) and a 0.096 inch ID. To our knowledge, this ID is the largest currently available for
aspiration catheters used in EVT treatment and has favorable dimensions compared to other
widely used aspiration catheters such as Neuron Max (Penumbra, Inc.; OD of 0.112 inch and ID
of 0.088 inch). The larger ID in the setting of ADAPT conceptually provides capacity to
aspirate a greater thrombus burden from larger vessels such as the ICA with minimal
repositioning and potentially minimal passes, as recently supported in a meta-analysis.
When paired with a suitable vacuum pump, even small increases in catheter ID
translate into exponentially greater force applied to the thrombus for removal. Considering
Poiseuille’s law, the rate of flow through a pipe for steady flow relates to the pressure
difference across the two ends of the pipe, the radius of the tube, as well as the fluid
viscosity and the length of the tube. In particular, the radius has a direct relationship to
the 4th power to the flow (Figure
4). When confronted with a large ICA clot burden, the BMX™ 96 can provide an
opportunity to aspirate large amounts of thrombus, without the need to pass first with a
microcatheter/microwire combination, thereby likely reducing the risk of dislodging distal
emboli. Furthermore, techniques using the combination of EVT with a proximal balloon guide
catheter (BCG) have emerged
to minimize the risk of potential distal emboli. However, much of this evidence
predates recent large-bore aspiration catheters. More recent BCG studies using Sofia
catheter systems (MicroVention, Aliso Viejo, C; up to 0.070 inch ID) demonstrate efficacy
and shorter procedure times in conjunction with large-bore aspiration catheters.[12,13] Additionally, the aspiration catheter
size positively correlates with the rate of first-pass effect when used in conjunction with
BCG and stent retrieval.
However, the IDs of some of the larger BCG systems such as the MERCI 9F (Stryker,
Kalamazoo, MI; ID of 0.085 inch) remain smaller than the BMX™ 96 system, raising the issue
of compatibility with large-bore catheters. With increasing advances in large-bore
aspiration catheters, the need for larger-bore BCGs may be needed.
Figure 4.
Poiseuille’s law rearrangement, highlighting the relationship between volumetric flow
rate (Q) and radius (r) to the 4th power. Small
increases in aspiration catheter inner diameter can thus translate into significant
increases in force for endovascular thrombectomy. n = fluid viscosity,
l = length of tubing, Δp = pressure difference.
Poiseuille’s law rearrangement, highlighting the relationship between volumetric flow
rate (Q) and radius (r) to the 4th power. Small
increases in aspiration catheter inner diameter can thus translate into significant
increases in force for endovascular thrombectomy. n = fluid viscosity,
l = length of tubing, Δp = pressure difference.Large-bore aspiration catheters are also a promising approach for the treatment of cerebral
venous thrombosis, which otherwise has high morbidity and mortality rates, with a proclivity
for younger patients. Small case series demonstrate the utility and safety of large-bore
aspiration catheters in cerebral venous thrombosis.
In particular, given the association of cerebral sinus thrombosis with infections,
such as coronavirus disease 2019, continued refinement of approaches for effective treatment
is urgently required. One of the additional benefits of a large-bore aspiration system, such
as the BMX™ 96, in the setting of ADAPT is the reduced cost compared to more traditional
catheters that are typically used in conjunction with stents and/or BCGs. Although
promising, further analysis is warranted to analyze the efficacy and complication risk of
larger-bore neurointerventional catheters as they gain more widespread use.Finally, management approaches for symptomatic acute carotid artery occlusions with
preserved circle of Willis flow, as demonstrated in our case report, are not well
established. The natural history of such occlusions may be more favorable due to collateral
flow via the circle of Willis, thus arguing against placing the patient at potential risk of
procedure-related complications. However, these patients remain at increased risk of
infarct, particularly in watershed regions. A recent retrospective analysis, using the MR
CLEAN registry, suggests generally improved clinical outcomes for patients who underwent EVT
for isolated intracranial ICA occlusions. However, 20% of these patients did demonstrate
distal clot migration with subsequent poor clinical outcome.
The use of an ADAPT approach with a large-bore catheter may minimize the risk of
intervention in these patients. Future studies, ideally using the most recent large-bore
catheters, will likely aid to elucidate further safe and efficacious management options for
patients with isolated acute carotid occlusions.
Conclusion
The BENCHMARK™ BMX™ 96 access system is a promising advance in neurointervention, and to
our knowledge, this is the first reported case of use in ischemic stroke. Our early
institution experience highlights the ease of use and versatility for neurointerventional
procedures such as EVT.
Authors: G Marnat; X Barreau; L Detraz; R Bourcier; B Gory; A Sgreccia; F Gariel; J Berge; P Menegon; M Kyheng; J Labreuche; A Consoli; R Blanc; B Lapergue Journal: AJNR Am J Neuroradiol Date: 2019-05-23 Impact factor: 3.825
Authors: Aquilla S Turk; Adnan Siddiqui; Johanna T Fifi; Reade A De Leacy; David J Fiorella; Eugene Gu; Elad I Levy; Kenneth V Snyder; Ricardo A Hanel; Amin Aghaebrahim; B Keith Woodward; Harry R Hixson; Mohammad I Chaudry; Alejandro M Spiotta; Ansaar T Rai; Donald Frei; Josser E Delgado Almandoz; Mike Kelly; Adam Arthur; Blaise Baxter; Joey English; Italo Linfante; Kyle M Fargen; J Mocco Journal: Lancet Date: 2019-03-09 Impact factor: 79.321
Authors: William J Powers; Alejandro A Rabinstein; Teri Ackerson; Opeolu M Adeoye; Nicholas C Bambakidis; Kyra Becker; José Biller; Michael Brown; Bart M Demaerschalk; Brian Hoh; Edward C Jauch; Chelsea S Kidwell; Thabele M Leslie-Mazwi; Bruce Ovbiagele; Phillip A Scott; Kevin N Sheth; Andrew M Southerland; Deborah V Summers; David L Tirschwell Journal: Stroke Date: 2019-10-30 Impact factor: 7.914
Authors: Federico Bolognini; Pablo A Lebedinsky; Mariano Musacchio; Mariette Delaitre; Abdoulaye M Traoré; Francis Vuillemet; François Sellal; Jean-François Cerfon; Eric Schluck; Daniela Iancu; Elena A Cora; Sébastien Richard; René Anxionnat; Benjamin Gory; Stephanos N Finitsis Journal: Interv Neuroradiol Date: 2021-04-04 Impact factor: 1.610