Sebastian Lehner1,2, Isabel Nowak3, Mathias Zacherl3, Julia Brosch-Lenz3, Maximilian Fischer4, Harun Ilhan3, Johannes Rübenthaler5, Astrid Gosewisch3, Peter Bartenstein3, Andrei Todica3.
Abstract
BACKGROUND: To evaluate quantitative myocardial perfusion SPECT/CT datasets for routine clinical reporting and the assessment of myocardial tracer uptake in patients with severe TVCAD.Entities:
Keywords: Myocardial perfusion scintigraphy; attenuation correction; balanced ischemia; quantitative SPECT; three-vessel disease
Mesh:
Year: 2021 PMID: 34341952 PMCID: PMC9553851 DOI: 10.1007/s12350-021-02735-2
Source DB: PubMed Journal: J Nucl Cardiol ISSN: 1071-3581 Impact factor: 3.872
Figure 1Excellent correlation between the global TPDs as determined with the three different reconstruction methods. The methods showed good agreement in the Bland-Altman plots, however, some variability was present. mod, mean of differences
Comparison of SUVmax, SUVpeak, and SUVmean between Q_INT and Q_EXT groups as well as mean paired differences and SD
| Q_INT | Q_EXT | Q_INT − Q_EXT | ||
|---|---|---|---|---|
| SUVmax | 5.80 ± 1.76 | 5.89 ± 1.72 | .599 | − .089 ± .91 |
| SUVpeak | 5.21 ± 1.54 | 5.24 ± 1.50 | .849 | − .027 ± .78 |
| SUVmean | 2.47 ± .90 | 2.63 ± .81 | .031 | − .16 ± .35 |
Figure 2Excellent correlation between the global SUVs and good agreement in the Bland-Altman plots, however, some variability was present. mod, mean of differences
Figure 3Patients with TVCAD present showed significantly reduced SUVmax and SUVpeak as compared to patients with no TVCAD present
Figure 4ROC analysis for SUVpeak and SUVmax to differentiate between patients with and without TVCAD, AUC = .75 for SUVpeak and AUC=.73 for SUVmax. Cut-off values optimized for sensitivity and specificity are 4.62 for SUVpeak (sensitivity 90%, specificity 55%) and 5.04 for SUVmax (sensitivity 90% and specificity 55%)