Influence of the geometry of extra column volumes on band broadening in a chromatographic system. Predictions by computational fluid dynamics.

Bartłomiej Filip1, Roman Bochenek2, Krystian Baran2, Dominik Strzałka3, Dorota Antos4.   

Abstract

A computational fluid dynamics method was used for prediction of flow behavior and band profiles of small- and macro-molecule compounds eluting in extra-column volumes (ECV) of an Äkta chromatographic system. The model compounds were: acetone, bovine serum albumin and an antibody. The construction of ECV was approximated by different types of geometries, starting from the simplest two-dimensional (2D) arrangement consisting of a straight capillary tube, and ending with a three-dimensional system (3D), which accounted for the flow path curvature of individual elements of ECV, including: injection loop capillary, multi-way valve, connecting capillary and detector cell. The accuracy of the model predictions depended on the flow path length and the eluent flowrate. 2D-geometry models reproduced pretty well the shapes of band profiles recorded at the lowest eluent flowrate used, but they failed for increased flowrates. The 3D-geometry model was found to be sufficiently accurate for all conditions investigated. It was exploited to analyze band broadening in the individual ECV elements. The simulation results revealed that the flow behavior in the injection loop capillaries strongly influenced the shape of band profiles, particularly at higher eluent velocities. This was attributed to the formation of Dean vertices triggered by centrifugal forces in curved parts of the eluent flow path.
Copyright © 2021 The Authors. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Computational fluid dynamics; Extra-column volume effect; Protein chromatography

Year:  2021        PMID: 34332316     DOI: 10.1016/j.chroma.2021.462410

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Peak broadening caused by using different micro-liquid chromatography detectors.

Authors:  Tobias Werres; Torsten C Schmidt; Thorsten Teutenberg
Journal:  Anal Bioanal Chem       Date:  2022-06-15       Impact factor: 4.478

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.