Myocardial recovery following pulsatile biventricular assist device support in infants: Report of 2 cases.

Joseph Philip1, Desiree Machado1, Renata Shih1, Archana Shenoy2, Mohammed Ebraheem1, Kevin Sullivan1,3, Himesh Vyas1, Dipankar Gupta1, Mark Bleiweis1,4.   

Abstract

Entities:  

Year:  2020        PMID: 34318118      PMCID: PMC8299981          DOI: 10.1016/j.xjtc.2020.04.025

Source DB:  PubMed          Journal:  JTCVS Tech        ISSN: 2666-2507


× No keyword cloud information.
BiVAD, Pulsatile, Myocardial recovery, VAD wean Delayed myocardial recovery post VAD implant has been well described in adults. In most cases, complete recovery has been reported in the first 4 to 6 weeks. Herein we report 2 cases of delayed myocardial recovery with successful VAD explant after 2 months and at 6 months in a pediatric population. One is a patient with CHD and the other with dilated cardiomyopathy, both on BiVAD support. See Commentaries on pages 93 and 95. Use of pulsatile ventricular assist devices (VAD) is a well-established strategy for supporting children with end-stage heart failure (HF), as a bridge to heart transplant. In 2005, Stiller and colleagues described a cohort of 73 pediatric patients managed using Excor pumps (Berlin Heart Inc, Woodlands, Tex), of whom 12.5% were successfully weaned of support. Since then, there have been multiple reports of successful pulsatile VAD weans in children., Myocardial recovery has been usually reported within a window of 1 to 2 weeks, on extracorporeal support and longer on VAD support.3, 4, 5 Here, we describe 2 patients with HF requiring biventricular VAD (BiVAD) support, successfully weaned 2 and 6 months after VAD implantation.

Case 1

A 6-month old girl (7 kg) was transferred for severe HF status postrepair of a late diagnosis of severe coarctation of the aorta (CoA) with minimal residual gradient and requiring escalating inotropic support. Echocardiography demonstrated severely diminished biventricular systolic function and left ventricular end diastolic volume (LVEDVi) of 87.14 mL (z score, 8.02). The etiology for HF was likely multifactorial: late repair of severe CoA, ischemic insult postbypass. Due to severe biventricular dysfunction, BiVAD support was initiated using Excor 10 mL pumps based on our institutional approach. LV biopsy at implant demonstrated interstitial fibrosis with no inflammation, patchy cardiomyocyte atrophy, and hypertrophy of LV trabeculae (Figure 1, A). Surveillance echocardiograms were performed. Normalization of right ventricular function was noticed at 2.5 months. A gradual decrease in LVEDVi was noted over the next 2 months to z score of +2 with improvement in LV function to normal at 4.5 months. Biventricular function was monitored over 4 weeks, with no evidence of deterioration. A weaning trial was performed (Table 1) and good hemodynamic parameters were measured by cardiac catheterization (Table 2). VAD was explanted with good clinical function (Table 3). Patient was changed to Status 7 (inactive) once evidence of recovery was seen. This patient is currently 21 months postexplant with normal function and no recurrent CoA (LVEDVi after explant was 23.44 mL).
Figure 1

Left ventricular biopsy at implant. A, Case 1: Left ventricular biopsy at implant- hematoxylin and eosin stained section (200× magnification) that demonstrates evenly spaced cardiomyocytes with mild interstitial expansion. Rare enlarged nuclei of hypertrophic cardiomyoctes are present. B, Case 2: Hematoxylin and eosin stained section (200× magnification) that demonstrates mild cardiomyocyte fiber size variation with focal interstitial fibrosis. C, Image (400×) magnification of a trichrome stained slide that confirms the interstitial fibrous expansion (blue-purple). The eosinophilic cardiomyocytes diameter ranges from 12 to 20 microns.

Table 1

University of Florida ventricular assist device (VAD) weaning protocol (heparin bolus 75 IU/kg, 15 minutes before VAD action)

Day of weanVAD actionParameters monitored
1BiVAD rate decreased by 50% for 30 minVital signs (heart rate, respiratory rate, blood pressure) every 3 minContinuous NIRSMental statusSvo2 measurement every 10 minEchocardiogram every 10 min (left ventricular end diastolic volume, ejection fraction, and qualitative assessment of wall function)
2BiVAD rate decreased by 75% or 35 (whichever was higher) for 30 min; following this, BiVAD completely paused for 3 minImportant: Both pumps have to be manually pumped to fill and eject every 30 sSame as Day 1Svo2 measurement every 10 min and at the end of pauseEchocardiogram every 10 min and at the end of the pause (size and function)
3BiVAD rate decreased by 75% or 35 (whichever was higher) for 30 min; following this, BiVAD completely paused for 6 minImportant: Both pumps have to be manually pumped to fill and eject every 30 sSame as Day 2
4BiVAD rate decreased by 75% or 35 (whichever was higher) for 30 min; following this, BiVAD completely paused for 10 minImportant: Both pumps have to be manually pumped to fill and eject every 30 sSame as Day 2
5- Right heart catheterizationBiVAD rate decreased by 75% or 35 (whichever was higher) for 30 min; following this, BiVAD completely paused for 10 minImportant: Both pumps have to be manually pumped to fill and eject every 30 sSame as Day 2 with hemodynamic measurements in catheterization include right atrial, right ventricular end diastolic pressures, and pulmonary venous wedge pressures in addition to estimating cardiac output

VAD, Ventricular assist device; BiVAD, biventricular assist device; NIRS, near-infrared spectroscopy; Sv, mixed venous oxygen saturation.

Table 2

Hemodynamic findings during the stages of wean from ventricular assist device (VAD)

Hemodynamics at VAD weanRAP (mm Hg)PAP (mm Hg)PCWP (mm Hg)MvO2 (%)SBP (mm Hg)PVR (Wood units/m2)CI (L/min/m2)
Case 1
 BiVAD rate 65/601127/18 (21)1260121/622.73.3
 BiVAD rate 35/301228/13 (21)1459105/572.63.3
 BiVAD paused 10 min1337/17 (25)1458120/662.63.3
Case 2
 BiVAD rate 75/70835/20 (25)116185/403.63.8
 BiVAD rate 35/301132/15 (23)115782/473.23.7
 BiVAD paused 10 min1135/15 (24)106082/463.53.9

VAD, Ventricular assist device; RAP, right atrial pressure; PAP, pulmonary artery pressure; PCWP, pulmonary capillary wedge pressure; MvO2, mixed venous saturation; SBP, systolic blood pressure; PVR, pulmonary vascular resistance; CI, cardiac index; BiVAD, biventricular assist device.

Table 3

Echocardiographic findings for patients 1 and 2 both at diagnosis and wean

PatientLVEF (at diagnosis)RV function at diagnosisLVEDi (cm) at diagnosisLVEF at VAD weanRV function at VAD weanLVEDi (cm) at VAD wean
18%-10%Severely depressed8.4 (z score, +8)55%-60%Normal3 (z score, +1.5)
25%-7%Mildly depressed4.4 (z score, +9)60%-65%Normal2.34 (z score, –1)

LV, Left ventricle; EF, ejection fraction; RV, right ventricle; LVEDi, left ventricular end diastolic dimension; VAD, ventricular assist device.

Left ventricular biopsy at implant. A, Case 1: Left ventricular biopsy at implant- hematoxylin and eosin stained section (200× magnification) that demonstrates evenly spaced cardiomyocytes with mild interstitial expansion. Rare enlarged nuclei of hypertrophic cardiomyoctes are present. B, Case 2: Hematoxylin and eosin stained section (200× magnification) that demonstrates mild cardiomyocyte fiber size variation with focal interstitial fibrosis. C, Image (400×) magnification of a trichrome stained slide that confirms the interstitial fibrous expansion (blue-purple). The eosinophilic cardiomyocytes diameter ranges from 12 to 20 microns. University of Florida ventricular assist device (VAD) weaning protocol (heparin bolus 75 IU/kg, 15 minutes before VAD action) VAD, Ventricular assist device; BiVAD, biventricular assist device; NIRS, near-infrared spectroscopy; Sv, mixed venous oxygen saturation. Hemodynamic findings during the stages of wean from ventricular assist device (VAD) VAD, Ventricular assist device; RAP, right atrial pressure; PAP, pulmonary artery pressure; PCWP, pulmonary capillary wedge pressure; MvO2, mixed venous saturation; SBP, systolic blood pressure; PVR, pulmonary vascular resistance; CI, cardiac index; BiVAD, biventricular assist device. Echocardiographic findings for patients 1 and 2 both at diagnosis and wean LV, Left ventricle; EF, ejection fraction; RV, right ventricle; LVEDi, left ventricular end diastolic dimension; VAD, ventricular assist device.

Case 2

A 9-month-old girl (8 kg) was admitted with cardiogenic shock (biventricular dysfunction due to dilated cardiomyopathy), with initial LVEDVi of 87.2 mL (z score, 7.79). BiVAD support was initiated using Excor 10 mL pumps as a bridge to heart transplant. A biopsy from LV at VAD implantation showed cardiomyocyte size variation with patchy interstitial fibrosis without inflammation. (Figure 1, B). On surveillance, recovery in LV function with decrease in LVEDi was noticed at 4 weeks with recovery of right ventricle. LV function was normal at 6 weeks and was monitored for another 2 weeks. Weaning trial was performed at 8 weeks and was tolerated. The patient underwent successful a VAD explant. Follow-up echocardiogram showed LVEDVi of 24.51 mL (z score, 0.7) and currently the patient has normal biventricular function 18 months postexplant (Table 3).

Discussion

Although recovery is well described in the adult population on VAD support, there is limited data in children using pulsatile VADs, with some of the case reports of recovery being on continuous flow devices.3, 4, 5, 6, 7, 8 In 2016, Hezter and colleagues, described that 15% (n = 18) of patients had recovery over a median time frame of 1 to 2 months. Ihnat and colleagues reported a higher frequency of recovery than other cohorts—61% (8 out of 13)—with the longest duration of support before wean being 21 days. Recovery was seen more frequently in myocarditis and idiopathic HF, with mean patient age <2 years. Current evidence suggests initial myocardial injury could arise from high levels of transforming growth factor beta and imbalance of the renin-angiotensin system, which promotes interstitial fibrosis and LV dilation with myocyte hypertrophy, respectively. Reverse modeling with VAD support has been seen in adults over longer time periods, leading to reduction in myocyte size and the area of fibrosis. Our imaging surveillance protocol includes echocardiogram twice a month for the first 3 months and monthly thereafter, monitoring LV dimensions, ejection fraction, and wall motion assessment. If wall motion recovery is noted, then imaging is performed weekly. Once function is identified as low normal qualitatively, ejection fraction >45% to 55% with LVEDVi within +2 z scores, a VAD wean trial is initiated after ensuring recovery is sustained over the next 2 to 4 weeks. Recurrence of HF postexplant has been described by Irving in 3 of 11 patients, requiring VAD reimplant and heart transplant. The etiology in all 3 was dilated cardiomyopathy. We suggest an active surveillance protocol for any signs of myocardial recovery in all pediatric patients with a VAD.