Lawrence Torkan1,2, Maria Theresa Servito1, Gianluigi Bisleri1.
Abstract
Entities:
Year: 2020 PMID: 34317986 PMCID: PMC8303071 DOI: 10.1016/j.xjtc.2020.08.012
Source DB: PubMed Journal: JTCVS Tech ISSN: 2666-2507
Figure 1A multiloop suture constructed from 2 polytetrafluoroethylene sutures (CV-4 Gore-Tex; WL Gore & Associates Inc, Newark, Del) tied to a third, which was passed through a felt pledget.
Figure 2A, Diagram of the sagittal view of the multiloop anchored to the posteromedial papillary muscle with neochordae at P2 and P3. B, Superior view.
Comparison of echocardiographic data from preoperation to follow-up
| Variable | Preoperation | Follow-up |
|---|---|---|
| LVEDV (mL) | 142 ± 70 | 113 ± 37 |
| LVESV (mL) | 73 ± 56 | 56 ± 53 |
| LVEDD (mm) | 60 ± 11 | 60 ± 5 |
| LVESD (mm) | 42 ± 10 | 45 ± 11 |
| LVEF% | 37 ± 10 | 49 ± 26 |
| WMSI | 1.8 ± 0.4 | 1.6 ± 0.6 |
Values are presented as mean ± standard deviation. LVEDV, Left ventricular end-diastolic volume; LVESV, left ventricular end-systolic volume; LVEDD, left ventricular end-diastolic diameter; LVESD, Left ventricular end-systolic diameter; LVEF, left ventricular ejection fraction; WMSI, wall motion score index.