The hybrid technique of projection-reconstruction echo-planar (PREP) imaging for obtaining chemical-shift images is demonstrated experimentally using a fluorine sample. The technique which is a variation on echo-planar imaging (EPI) relies on a multipass procedure. It is nevertheless quite efficient and consequently chemical-shift images may be produced in a few minutes. The method produces images in 64 chemical-shift regions, each region mapped spatially by 64 X 64 pixels. The imaging time was just over 4 min. These 64 chemical-shifted images can be straightforwardly added together to form an undistorted image of the complete object. In addition the chemical-shift spectrum can be extracted and the various chemical-shift images can be unambiguously assigned to the spectral peaks.Entities: